Assessment of Portable Oxygen Concentrators in Infants Undergoing Hypoxic Challenge Testing.

Assessment of Portable Oxygen Concentrators in Infants Undergoing Hypoxic Challenge Testing. A Randomized Crossover Study.

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03976986
Enrollment
26
Registered
2019-06-06
Start date
2016-02-11
Completion date
2019-02-04
Last updated
2019-06-06

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Hypoxia, Altitude

Keywords

Infants, Hypoxia Altitude Simulation Test, Oxygen Delivery Device, Portable Oxygen Concentrator, Hypoxic Challenge Testing

Brief summary

Hypoxic Challenge Testing (HCT) is the recommended method for inflight hypoxia risk assessment. Onboard oxygen administration remains controversial. The Federal Aviation Administration approved portable oxygen concentrators (POCs) for onboard oxygen supply but there is lack of evidence about the use, especially in children. The aim of our study is to establish the effectiveness and safety of POCs in infants undergoing HCT.

Detailed description

According to a forecast by International Air Transport Association the number of people travelling on commercial aircrafts is predicted to rise up to 8.2 billion passengers in 2037. Therefore evidence-based flight recommendations will gain importance in the future, especially for patients suffering from chronic diseases. Hypoxic Challenge Testing (HCT) is the recommended method for inflight hypoxia risk assessment where nitrogen is introduced in a plethysmograph reducing FiO2 (fraction of inspired oxygen) to 0.15. Oxygen supply is recommended if PaO2 (partial pressure of oxygen in arterial blood) drops \<50-55 mmHg (in adults) or Sat.O2 (oxygen saturation) ≤ 85% (in infants) where non-invasive pulse oximetry is the recommended method for hypoxia assessment. Onboard oxygen administration remains controversial. The effectiveness of pulsed-dosed systems remains unclear in small patients due to shallow breathing patterns. The aim of our study is to establish the effectiveness and safety of POCs in infants undergoing HCT.

Interventions

DEVICEpulsed-flow portable oxygen concentrator (pPOC)

InogenOne G3® (Inogen,Goleta,CA) on pulsed-flow mode: setting 2 (flow rate 420 ml/min, 16.8 ml +/- 3ml per bolus at 25 rpm).

DEVICEcontinuous-flow portable oxygen concentrator (cPOC)

SeQal Eclipse 3® on continuous-flow (SeQual,Ball Ground,GA): flow rate 2 litre per minute (lpm).

Sponsors

Hospital Son Espases
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
No minimum to 16 Years
Healthy volunteers
No

Inclusion criteria

Patients with baseline Sat.O2 \>94% AND * \<1 year with neonatal respiratory disease * required oxygen supply in the last 6 months * chronic respiratory disease (eg cystic fibrosis, obstructive or restrictive pulmonary disease) with FEV 1 (forced expiratory volume in 1 second) or FVC (forced vital capacity) \<50%.

Exclusion criteria

* acute respiratory infection

Design outcomes

Primary

MeasureTime frameDescription
Assessment of portable oxygen concentrators (POCs) to change HCT induced hypoxia (Sat.O2 drop ≤ 85%).20 minutesHypoxia (Sat.O2 drop ≤ 85%) measured by Masimo SET Radical-7 Electron® pulse oximeter is induced performing HCT. Thereafter oxygen is administered through cPOC or pPOC until baseline Sat.O2 (Sat.O2 \>93%) is achieved.

Secondary

MeasureTime frameDescription
POC comparison to change HCT induced hypoxia (Sat.O2 drop ≤ 85%).20 minutesCompare the capacity of different POCs: continuous flow (cPOC) versus pulsed flow (pPOC) to change HCT induced hypoxia (Sat.O2 ≤ 85%) until baseline Sat.O2 (Sat.O2 \>93%) is achieved.
Relate patient age (months) to POCs capacity to change HCT induced hypoxia (Sat.O2 drop ≤ 85%).20 minutesRelationship between patient age (months) and POCs capacity to change hypoxic state (≤ 85%) to baseline Sat.O2 (\>93%) measured by Masimo SET Radical-7 Electron®
Relate patient weight (kilograms) to POCs capacity to change HCT induced hypoxia (Sat.O2 drop ≤ 85%).20 minutesRelationship between patient weight (kilograms) and POCs capacity to change hypoxic state (≤ 85%) to baseline Sat.O2 (\>93%) measured by Masimo SET Radical-7 Electron®