Evaluation of SenseGuard™- a Noninvasive Respiratory Rate Measuring Device

Evaluation of SenseGuard™ - a Noninvasive Respiratory Rate Measuring Device for Adult Patients

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03964675
Enrollment
25
Registered
2019-05-28
Start date
2019-01-13
Completion date
2019-04-08
Last updated
2019-05-28

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Condition

Brief summary

The SenseGuard™ is a new respiratory rate measuring and logging non-invasive device, and measures the respiratory rate by responding to changes in air moisture content in exhaled air. A prospective single-center, single arm study, designed to assess the bias, accuracy , precision and safety of the SenseGuard™ device in measuring of respiratory rate in adult patients. The study shall include a single invitational visit with a respiratory rate measurement duration of 30 minutes for each of 25 adult subjects hospitalized in internal wards.Subjects will be invited to participate should they meet all of the eligibility criteria. Following informed consent, subjects will be measured with both Microstream capnograph and SenseGuard™ measuring simultaneously so their results can be compared.

Detailed description

Study description: Respiratory rate (RR) is a vital sign used to monitor clinical condition of a patient. It is defined as the number of breaths per minute. Various devices using different techniques are available to measure RR. Currently this is measured at the bedside in clinical practice by counting the breaths, however the gold standard for measuring this vital sign is the capnograph. For the patient, this involves limiting the movement and talking so that the measurements can be taken. A new respiratory rate measuring and logging device, SenseGuard™, is non-invasive and measures the respiratory rate by responding to changes in air moisture content (small amounts of water condensation) in exhaled air. The study is planned as a prospective single-center, single arm study, designed to assess the bias, accuracy , precision and safety of the SenseGuard™ non-invasive device in measuring of respiratory rate in adult patients. Subjects will be invited to participate should they meet all of the eligibility criteria. Following informed consent, subjects will be measured with both Microstream capnograph and SenseGuard™ measuring simultaneously so their results can be compared. Study population: 25 hospitalized patients from Internal/General wards aged 18 and older. Objective: Evaluate the performance and safety of the SenseGuard™ non-invasive device in measuring respiratory rate in hospitalized adult patients. Inclusion Criteria: 1. Hospitalized patients from the Internal/General wards. 2. Provision of signed and dated informed consent form 3. Stated willingness to comply with all study procedures and availability for the duration of the study 4. Male or female, aged ≥18 5. Willing to adhere to the study regimen Exclusion Criteria: 1. Intubated patients (i.e. endobronchial intubation) or patients currently undergoing a BPAP/CPAP treatment. 2. Presence of implanted electronic device intended to regulate cardiac activity (e.g., pacemaker or cardioverter/defibrillator) 3. Pregnancy or lactation 4. Known allergic reactions to materials used in the components of the SenseGuard™ system, i.e. polyethylene or silicon, or gold, nickel, copper or alloys containing any of the above. 5. Participation in another study in the last 4 weeks Participant Duration: A single invitational visit with a respiratory rate measurement duration of 30 minutes Study Duration: 1 year.

Interventions

A single invitational visit with a respiratory rate measurement duration of 30 minutes.

Sponsors

NanoVation
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

1. Hospitalized patients from the Internal/General wards. 2. Provision of signed and dated informed consent form 3. Stated willingness to comply with all study procedures and availability for the duration of the study 4. Male or female, aged ≥18 5. Willing to adhere to the study regimen

Exclusion criteria

1. Intubated patients (i.e. endobronchial intubation) or patients currently undergoing a BPAP/CPAP treatment. 2. Presence of implanted electronic device intended to regulate cardiac activity (e.g., pacemaker or cardioverter/defibrillator) 3. Pregnancy or lactation 4. Known allergic reactions to materials used in the components of the SenseGuard™ system, i.e. polyethylene or silicon, or gold, nickel, copper or alloys containing any of the above. 5. Participation in another study in the last 4 weeks

Design outcomes

Primary

MeasureTime frame
Accuracy of Respiratory Rate (RR) as measured by the SenseGuard™ system versus the gold standard (capnography).4 months
Frequency and incidence of any adverse events and serious adverse events related and unrelated to the device use4 months