Skip to content

Non-inferiority of Portable Versus Desktop Spirometry

Non-inferiority of Portable Versus Desktop Spirometry

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03894475
Enrollment
62
Registered
2019-03-28
Start date
2017-12-18
Completion date
2018-06-21
Last updated
2019-03-28

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Asthma, COPD

Keywords

Asthma, COPD, Mobile Spirometry

Brief summary

The investigators compared the ventilation parameters for volume and flow obtained from standard spirometry procedures from patients presently monitored and treated for asthma or chronic obstructive pulmonary disease (COPD) using AioCare (HealthUp Sp. z o.o., Serial Number: MS082017005412, software version: MySpiroo app 1.1.14) as the tested device and Spirometer USB CPFS/D (MGC Diagnostics) as the reference, which required calibration prior to each session. Spirometry measurements were performed on sixty-two patients (forty-four females (58±17 years old) and eighteen males (52±19 years old)) at the Institute of Tuberculosis and Lung Disease in Warsaw, Poland. Participants were asked to perform correct spirometry examinations (which means at least three technically correct exhales and meeting repeatability criteria for FEV1 and FVC) on both measuring devices with a five-minute break between devices to prevent respiratory muscle fatigue. The highest value from all acceptable spirometry results was then used for analysis. All spirometry examinations followed ERS/ATS standards.

Interventions

DEVICEAioCare

AioCare is an ultraportable, handheld hardware module that contains the MEMS-based flow sensor and electronics, with a dedicated mobile application that works on iOS and Android operating systems. The unit is used with a disposable mouthpiece fitted to the tip of the flow tube, and a nose clip. AioCare is connected to its dedicated mobile application, which contains software that will show flow-volume graphs and results in real time. The device encompasses all of the widely used spirometry parameters, including: peak expiratory flow (PEF), forced expiratory volume in 1 second (FEV1), forced vital capacity (FVC), FEV1/FVC ratio, forced expiratory flow at 25% of expiration (FEF25), forced expiratory flow at 50% of expiration (FEF50), forced expiratory flow at 75% of expiration (FEF75), and forced expiratory volume in 6 second (FEV6).

DEVICESpirometer USB CPFS/D (MGC Diagnostics)

Spirometer USB CPFS/D (MGC Diagnostics) is a diagnostic desktop spirometer used in hospitals/clinics. CPFS/D USB spirometer is compatible with desktop and laptop computers.

Sponsors

Healthup Sp. z o.o.
CollaboratorINDUSTRY
National Institute for Tuberculosis and Lung Diseases, Poland
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

1. Age\>18. 2. Diagnosed asthma or COPD 3. Signed consent to participation in the research experiment. 4. Ability to comply with the spirometry protocol.

Exclusion criteria

1. Pregnant females (based on declaration; no pregnancy tests are planned before the test). 2. Recent myocardial infarction (\<30 days). 3. Known thoracic, aortic or cerebral aneurysm. 4. Recent stroke, eye surgery, thoracic/abdominal surgery. 5. Haemoptysis. 6. Recent pneumothorax. 7. Uncontrolled hypertension. 8. Pulmonary Embolism. 9. Angina. 10. Chest or abdominal pain of any etiology. 11. Oral or facial pain exacerbated by a mouthpiece. 12. Stress incontinence. 13. Dementia or state of confusion. 14, Acute Diarrhea

Design outcomes

Primary

MeasureTime frameDescription
Non-inferiority of PEF between AioCare and CPFS/D USB Spirometer6 monthsComparison of Peak Expiratory Flow obtained during a standard dynamic spirometry measured with AioCare (HealthUp) and reference spirometer (MGC)
Non-inferiority of FEV1 between AioCare and CPFS/D USB Spirometer6 monthsComparison of Forced Expiratory Volume in the first second (FEV1) obtained during a standard dynamic spirometry measured with AioCare (HealthUp) and reference spirometer (MGC)
Non-inferiority of FVC between AioCare and CPFS/D USB Spirometer6 monthsComparison of Forced Vital Capacity (FVC) obtained during a standard dynamic spirometry measured with AioCare (HealthUp) and reference spirometer (MGC)

Countries

Poland

Outcome results

None listed