Brief summary

The primary objective is to compare the performance of Masimo's SedLine and the comparator's device systems during surgery. Relative accuracy of the individual depth of sedation indices will be compared.

Interventions

DEVICESedLine Vs Comparator Test Group

Subjects will receive electrodes that compare EEG signals between Masimo's SedLine system and the comparator's system.

Sponsors

Masimo Corporation
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Patients 18 years old and older at the time of consent. * ASA status I, II, or III. * English-speaking subjects. * Scheduled for surgical and non-surgical procedures scheduled under general anesthesia (Common procedures include but are not restricted to tonsillectomy, adenoidectomy, urological procedures, dental rehabilitation, orthopedic procedures, biopsies, audiogram, nuclear scans, gastroscopy, colonoscopy, etc.).

Exclusion criteria

* Any deformities or devices that may prevent application of EEG Sensor to forehead with a proper fit. * Subjects who are developmentally delayed. * Subjects deemed not suitable for study at the discretion of the investigator.

Design outcomes

Primary

MeasureTime frameDescription
Correlation of Sedation Indices Performance Across Anesthesia RangeDuration of surgeryThe correlation was computed using 1 min average values between the PSi and the comparator's sedation index.

Countries

United States

Participant flow

Participants by arm

ArmCount
SedLine Vs Comparator Test Group
SedLine Vs Comparator Test Group: Subjects will receive electrodes that compare EEG signals between Masimo's SedLine system and the comparator's system.
100
Total100

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyEquipment failure4
Overall StudyProcedure Cancelled4

Baseline characteristics

CharacteristicSedLine Vs Comparator Test Group
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
37 Participants
Age, Categorical
Between 18 and 65 years
63 Participants
Race/Ethnicity, Customized
African American
5 participants
Race/Ethnicity, Customized
American Indian
1 participants
Race/Ethnicity, Customized
Asian
1 participants
Race/Ethnicity, Customized
Not reported
14 participants
Race/Ethnicity, Customized
Other
5 participants
Race/Ethnicity, Customized
White
74 participants
Region of Enrollment
United States
100 Participants
Sex: Female, Male
Female
38 Participants
Sex: Female, Male
Male
62 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 100
other
Total, other adverse events
1 / 100
serious
Total, serious adverse events
0 / 100

Outcome results

Primary

Correlation of Sedation Indices Performance Across Anesthesia Range

The correlation was computed using 1 min average values between the PSi and the comparator's sedation index.

Time frame: Duration of surgery

Population: Due to the complexity of data collection, complete data for analysis was available from 61 subjects.

ArmMeasureValue (NUMBER)
SedLine Vs Comparator Test GroupCorrelation of Sedation Indices Performance Across Anesthesia Range0.8314 Pearson r

Source: ClinicalTrials.gov · Data processed: Feb 15, 2026