Evaluating Refractive and Visual Outcomes With the Alcon Toric ReSTOR +2.5 and +3.0 Models

Evaluating Refractive and Visual Outcomes With the Alcon Toric ReSTOR +2.5 and +3.0 Models

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03856944
Enrollment
29
Registered
2019-02-27
Start date
2018-10-05
Completion date
2019-11-18
Last updated
2021-08-06

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Cataract

Keywords

Astigmatism, intraocular lens, multifocal

Brief summary

The study will evaluate visual acuity and visual function with a 'blended' ReSTOR Toric intraocular lens (IOL) implantation after cataract surgery.

Detailed description

The objective is to provide a normative standard for refractive and visual outcomes in eyes with significant corneal astigmatism undergoing cataract surgery with the Alcon Toric ReSTOR lens. Patients will be bilaterally implanted, with the Toric ReSTOR +2.5D model implanted in the dominant eye and the Toric ReSTOR +3.0D model implanted in the non-dominant eye.

Interventions

DEVICEReSTOR Toric

ReSTOR Toric bilateral IOL implantation

Sponsors

SiV Consulting
CollaboratorOTHER
Gainesville Eye Associates
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Intervention model description

This study is a prospective, single center, open clinical trial to collect normative performance data.

Eligibility

Sex/Gender
ALL
Age
30 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* Subjects are eligible for the study if they meet the following criteria: Note: Ocular criteria must be met in the eligible eye. * Bilateral visually-significant cataracts * Willing and able to provide written informed consent for participation in the study. * Willing and able to comply with scheduled visits and other study procedures. * have good ocular health, with no pathology that compromises visual acuity (outside of residual refractive error and cataract) * Regular corneal astigmatism of 1.00D to 2.50D in both eyes * Potential postoperative acuity of 20/25 or better

Exclusion criteria

* If any of the following

Design outcomes

Primary

MeasureTime frameDescription
Residual Refractive Cylinder3 monthsManifest postoperative refractive cylinder (measured in diopters, with a phoropter or trial lens set) after cataract surgery and IOL implantation

Secondary

MeasureTime frameDescription
Monocular Uncorrected Distance Visual Acuity3 monthsMonocular uncorrected distance visual acuity measured in logMAR notation
Monocular Uncorrected Intermediate Visual Acuity3 monthsMonocular Uncorrected intermediate visual acuity measured in logMAR notation (log of the minimum angle of resolution)
Monocular Best Corrected Distance Visual Acuity3 monthsMonocular best corrected distance visual acuity measured in logMAR notation
Monocular Best Distance-corrected Intermediate Visual Acuity3 monthsMonocular best distance-corrected intermediate visual acuity in logMAR notation
Monocular Best Distance-corrected Near Visual Acuity3 monthsMonocular best distance-corrected near visual acuity measured in logMAR notation
Binocular Uncorrected Near Visual Acuity3 monthsBinocular uncorrected near (40cm) visual acuity measured in logMAR notation
Binocular Uncorrected Distance Visual Acuity3 monthsBinocular uncorrected distance visual acuity measured in logMAR notation
Monocular Uncorrected Near Visual Acuity3 monthsMonocular uncorrected near visual acuity measuring in logMAR
Binocular Best-corrected Distance Visual Acuity3 monthsBinocular best-corrected distance visual acuity measured in logMAR notation
Binocular Best Distance-corrected Intermediate Visual Acuity3 monthsBinocular best distance-corrected intermediate (60cm) visual acuity
Binocular Best Distance-corrected Near Visual Acuity3 monthsBinocular best distance-corrected near (40cm) visual acuity
IOL Orientation3 monthsPostoperative orientation changes with the Toric ReSTOR IOL up to 3 months after surgery will be measured by determining the angle of orientation of the IOL in degrees (0-180) using a slit lamp.
Binocular Best Distance-corrected Defocus Curve3 monthsBinocular best distance-corrected defocus curve.
Binocular Uncorrected Defocus Curve.3 monthsBinocular uncorrected defocus curve, from +1.0D to -4.00D in 0.50D increments
Subjective Visual Quality3 monthsQuality of vision, measured with a subjective questionnaire (Q of V). Frequency, Severity and Degree of Bother aggregate statistics. Scored on a Rasch scale (0-100) where lower is better.
Binocular Uncorrected Intermediate Visual Acuity3 monthsBinocular uncorrected intermediate (60cm) visual acuity

Countries

United States

Participant flow

Participants by arm

ArmCount
ReSTOR Toric
Patients will be bilaterally implanted, with the Toric ReSTOR +2.5D model implanted in the dominant eye and the Toric ReSTOR +3.0D model implanted in the non-dominant eye. Patients will self-select for multifocal implantation. ReSTOR Toric: ReSTOR Toric bilateral IOL implantation
29
Total29

Baseline characteristics

CharacteristicReSTOR Toric
Age, Continuous69 years
STANDARD_DEVIATION 9
Race and Ethnicity Not Collected— Participants
Region of Enrollment
United States
29 Participants
Sex: Female, Male
Female
18 Participants
Sex: Female, Male
Male
11 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 29
other
Total, other adverse events
0 / 29
serious
Total, serious adverse events
0 / 29

Outcome results

Primary

Residual Refractive Cylinder

Manifest postoperative refractive cylinder (measured in diopters, with a phoropter or trial lens set) after cataract surgery and IOL implantation

Time frame: 3 months

ArmMeasureValue (MEAN)Dispersion
ReSTOR ToricResidual Refractive Cylinder0.02 dioptersStandard Deviation 0.08
Secondary

Binocular Best-corrected Distance Visual Acuity

Binocular best-corrected distance visual acuity measured in logMAR notation

Time frame: 3 months

ArmMeasureValue (MEAN)Dispersion
ReSTOR ToricBinocular Best-corrected Distance Visual Acuity-0.03 logMARStandard Deviation 0.05
Secondary

Binocular Best Distance-corrected Defocus Curve

Binocular best distance-corrected defocus curve.

Time frame: 3 months

ArmMeasureGroupValue (MEAN)Dispersion
ReSTOR ToricBinocular Best Distance-corrected Defocus CurveDefocus +1.00.24 logMAR AcuityStandard Deviation 0.2
ReSTOR ToricBinocular Best Distance-corrected Defocus CurveDefocus +0.50.07 logMAR AcuityStandard Deviation 0.11
ReSTOR ToricBinocular Best Distance-corrected Defocus CurveDefocus 0.00-.01 logMAR AcuityStandard Deviation 0.07
ReSTOR ToricBinocular Best Distance-corrected Defocus CurveDefocus -4.00.55 logMAR AcuityStandard Deviation 0.19
ReSTOR ToricBinocular Best Distance-corrected Defocus CurveDefocus -3.50.40 logMAR AcuityStandard Deviation 0.15
ReSTOR ToricBinocular Best Distance-corrected Defocus CurveDefocus -3.00.28 logMAR AcuityStandard Deviation 0.12
ReSTOR ToricBinocular Best Distance-corrected Defocus CurveDefocus -2.50.19 logMAR AcuityStandard Deviation 0.11
ReSTOR ToricBinocular Best Distance-corrected Defocus CurveDefocus -2.00.16 logMAR AcuityStandard Deviation 0.11
ReSTOR ToricBinocular Best Distance-corrected Defocus CurveDefocus -1.50.21 logMAR AcuityStandard Deviation 0.11
ReSTOR ToricBinocular Best Distance-corrected Defocus CurveDefocus -1.00.19 logMAR AcuityStandard Deviation 0.08
ReSTOR ToricBinocular Best Distance-corrected Defocus CurveDefocus -0.500.08 logMAR AcuityStandard Deviation 0.1
Secondary

Binocular Best Distance-corrected Intermediate Visual Acuity

Binocular best distance-corrected intermediate (60cm) visual acuity

Time frame: 3 months

ArmMeasureValue (MEAN)Dispersion
ReSTOR ToricBinocular Best Distance-corrected Intermediate Visual Acuity-0.06 logMARStandard Deviation 0.03
Secondary

Binocular Best Distance-corrected Near Visual Acuity

Binocular best distance-corrected near (40cm) visual acuity

Time frame: 3 months

ArmMeasureValue (MEAN)Dispersion
ReSTOR ToricBinocular Best Distance-corrected Near Visual Acuity0.01 logMARStandard Deviation 0.01
Secondary

Binocular Uncorrected Defocus Curve.

Binocular uncorrected defocus curve, from +1.0D to -4.00D in 0.50D increments

Time frame: 3 months

ArmMeasureGroupValue (MEAN)Dispersion
ReSTOR ToricBinocular Uncorrected Defocus Curve.Defocus -4.00.55 logMAR AcuityStandard Deviation 0.19
ReSTOR ToricBinocular Uncorrected Defocus Curve.Defocus -3.50.41 logMAR AcuityStandard Deviation 0.16
ReSTOR ToricBinocular Uncorrected Defocus Curve.Defocus -3.00.29 logMAR AcuityStandard Deviation 0.13
ReSTOR ToricBinocular Uncorrected Defocus Curve.Defocus -2.50.18 logMAR AcuityStandard Deviation 0.11
ReSTOR ToricBinocular Uncorrected Defocus Curve.Defocus 1.00.22 logMAR AcuityStandard Deviation 0.2
ReSTOR ToricBinocular Uncorrected Defocus Curve.Defocus 0.50.07 logMAR AcuityStandard Deviation 0.11
ReSTOR ToricBinocular Uncorrected Defocus Curve.Defocus 0.00-.01 logMAR AcuityStandard Deviation 0.08
ReSTOR ToricBinocular Uncorrected Defocus Curve.Defocus -2.00.16 logMAR AcuityStandard Deviation 0.11
ReSTOR ToricBinocular Uncorrected Defocus Curve.Defocus -1.50.20 logMAR AcuityStandard Deviation 0.11
ReSTOR ToricBinocular Uncorrected Defocus Curve.Defocus -1.00.19 logMAR AcuityStandard Deviation 0.08
ReSTOR ToricBinocular Uncorrected Defocus Curve.Defocus -0.500.09 logMAR AcuityStandard Deviation 0.1
Secondary

Binocular Uncorrected Distance Visual Acuity

Binocular uncorrected distance visual acuity measured in logMAR notation

Time frame: 3 months

ArmMeasureValue (MEAN)Dispersion
ReSTOR ToricBinocular Uncorrected Distance Visual Acuity-0.01 logMARStandard Deviation 0.07
Secondary

Binocular Uncorrected Intermediate Visual Acuity

Binocular uncorrected intermediate (60cm) visual acuity

Time frame: 3 months

ArmMeasureValue (MEAN)Dispersion
ReSTOR ToricBinocular Uncorrected Intermediate Visual Acuity-0.02 logMARStandard Deviation 0.07
Secondary

Binocular Uncorrected Near Visual Acuity

Binocular uncorrected near (40cm) visual acuity measured in logMAR notation

Time frame: 3 months

ArmMeasureValue (MEAN)Dispersion
ReSTOR ToricBinocular Uncorrected Near Visual Acuity0.01 logMARStandard Deviation 0.08
Secondary

IOL Orientation

Postoperative orientation changes with the Toric ReSTOR IOL up to 3 months after surgery will be measured by determining the angle of orientation of the IOL in degrees (0-180) using a slit lamp.

Time frame: 3 months

Population: Subjects are enrolled, eyes (2x subjects) were evaluated for some tests.

ArmMeasureValue (NUMBER)
ReSTOR ToricIOL Orientation56 Eyes with < 5 degrees rot'n from 1-3 mos
Secondary

Monocular Best Corrected Distance Visual Acuity

Monocular best corrected distance visual acuity measured in logMAR notation

Time frame: 3 months

ArmMeasureValue (MEAN)Dispersion
ReSTOR ToricMonocular Best Corrected Distance Visual Acuity.04 logMARStandard Deviation 0.09
Secondary

Monocular Best Distance-corrected Intermediate Visual Acuity

Monocular best distance-corrected intermediate visual acuity in logMAR notation

Time frame: 3 months

ArmMeasureValue (MEAN)Dispersion
ReSTOR ToricMonocular Best Distance-corrected Intermediate Visual Acuity.06 logMARStandard Deviation 0.11
Secondary

Monocular Best Distance-corrected Near Visual Acuity

Monocular best distance-corrected near visual acuity measured in logMAR notation

Time frame: 3 months

ArmMeasureValue (MEAN)Dispersion
ReSTOR ToricMonocular Best Distance-corrected Near Visual Acuity.11 logMARStandard Deviation 0.12
Secondary

Monocular Uncorrected Distance Visual Acuity

Monocular uncorrected distance visual acuity measured in logMAR notation

Time frame: 3 months

ArmMeasureValue (MEAN)Dispersion
ReSTOR ToricMonocular Uncorrected Distance Visual Acuity0.04 logMARStandard Deviation 0.09
Secondary

Monocular Uncorrected Intermediate Visual Acuity

Monocular Uncorrected intermediate visual acuity measured in logMAR notation (log of the minimum angle of resolution)

Time frame: 3 months

ArmMeasureValue (MEAN)Dispersion
ReSTOR ToricMonocular Uncorrected Intermediate Visual Acuity0.06 logMARStandard Deviation 0.11
Secondary

Monocular Uncorrected Near Visual Acuity

Monocular uncorrected near visual acuity measuring in logMAR

Time frame: 3 months

ArmMeasureValue (MEAN)Dispersion
ReSTOR ToricMonocular Uncorrected Near Visual Acuity0.12 logMARStandard Deviation 0.11
Secondary

Subjective Visual Quality

Quality of vision, measured with a subjective questionnaire (Q of V). Frequency, Severity and Degree of Bother aggregate statistics. Scored on a Rasch scale (0-100) where lower is better.

Time frame: 3 months

ArmMeasureGroupValue (MEDIAN)
ReSTOR ToricSubjective Visual QualityFrequency32 score on a Rasch scale (0-100)
ReSTOR ToricSubjective Visual QualitySeverity22 score on a Rasch scale (0-100)
ReSTOR ToricSubjective Visual QualityDegree of Bother0 score on a Rasch scale (0-100)