Clinical Study of a Diagnostic Device for NG, TV and CT in Women

Pivotal Study of a Single-Use, Point-of-Care Molecular Diagnostic Device for the Detection of Neisseria Gonorrhoeae (NG), Trichomonas Vaginalis (TV), and Chlamydia Trachomatis (CT) in Women

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03852316
Enrollment
1585
Registered
2019-02-25
Start date
2019-02-27
Completion date
2019-12-02
Last updated
2022-01-13

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Chlamydial Infection, Gonococcal Infection, Trichomoniasis

Keywords

Chlamydia trachomatis, Device, Diagnostic, Neisseria gonorrohoeae, Trichomonas vaginalis

Brief summary

This is a multi-center study with a minimum of three sites in the United States. The study will enroll approximately 1750 female subjects, 14 years of age and older, and will have a study duration of approximately 9 months after enrollment of the first subject. Female subjects seen at the participating sites for any reason will be evaluated for enrollment in this study. All subjects will be managed per standard of care as applicable. Subjects who are enrolled in the study will perform self-collection of a vaginal swab to be tested by Click device and allow the health care provider (HCP) to collect three additional vaginal swabs to be tested by recognized FDA-cleared comparator methods. Subjects will complete the study in a single visit. The primary objective is to assess the performance of the Click device for detection of Chlamydia trachomatis (CT), Neisseria gonorrhoeae (NG), and Trichomonas vaginalis (TV) in self-collected vaginal specimens as compared to Patient Infected Status (PIS) determined by three approved comparator assays using vaginal specimens collected by a qualified HCP in support of obtaining FDA clearance. and a Clinical Laboratory Improvement Amendments (CLIA) Waiver.

Detailed description

This is a multi-center study with a minimum of three clinical sites across diverse geographical areas in the United States. Approximately one-third of the total number of sites will have a patient population of low Neisseria gonorrhoeae (NG), Chlamydia trachomatis (CT), or Trichomonas vaginalis (TV) prevalence. Low prevalence sites are defined as sites with a prevalence \< / = 2 percent for any of the three targets. Sites with a prevalence higher than 2 percent for any of the three targets will be defined as high prevalence sites. The study will enroll approximately 1750 female subjects, 14 years of age and older, and will have a study duration of approximately 9 months after enrollment of the first subject. Female subjects seen at the participating sites for any reason will be evaluated for enrollment in this study. All subjects will be managed per standard of care as applicable. Subjects who are enrolled in the study will perform self-collection of a vaginal swab to be tested by Click device and allow the health care provider (HCP) to collect three additional vaginal swabs to be tested by recognized FDA-cleared comparator methods. Subjects will complete the study in a single visit, approximately 60 minutes in length. The hypothesis of this study is that Click Diagnostics Sexual Health Test performs substantially equivalent to the NAAT (Nucleic Acid Amplification Test) predicate system, and the identification of each organism (CT, NG, and TV) in self-collected vaginal swabs by women using the Click device will agree with the Patient Infected Status (PIS) with a high sensitivity and specificity. The primary objective is to assess the performance of the Click device for detection of CT, NG, and TV in self-collected vaginal specimens as compared to PIS determined by three approved comparator assays using vaginal specimens collected by a qualified HCP in support of obtaining FDA clearance and a Clinical Laboratory Improvement Amendments (CLIA) Waiver. The secondary objectives are: 1) to assess the performance of the Click device for detection of CT, NG, and TV in self-collected vaginal specimens among symptomatic subjects as compared to PIS determined by three approved comparator assays using vaginal specimens collected by a qualified HCP; 2) to assess the performance of the Click device for detection of CT, NG, and TV in self-collected vaginal specimens among asymptomatic subjects as compared to PIS determined by three approved comparator assays using vaginal specimens collected by a qualified HCP.

Interventions

A single-use device containing a polymerase chain reaction (PCR)-based nucleic acid amplification test (NAAT) for detection of deoxyribonucleic acid (DNA) from Chlamydia trachomatis (CT), Neisseria gonorrhoeae (NG), and Trichomonas vaginalis (TV) in self-collected vaginal specimens.

Sponsors

National Institute of Allergy and Infectious Diseases (NIAID)
Lead SponsorNIH

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
FEMALE
Age
14 Years to 99 Years
Healthy volunteers
Yes

Inclusion criteria

1. Willing and able to give voluntary written informed consent (or the parent/legal guardian will provide parental permission) before any study-related procedure is performed. 2. Female at birth. (Pregnant and breastfeeding women are eligible.) 3. Age \> / = 14 years at the time of enrollment. 4. Able to read and understand the procedural information provided for the study. 5. Able and willing to follow all study procedures, including performing self-collection of one vaginal swab and permitting a licensed HCP to collect three additional vaginal swabs.

Exclusion criteria

1. Have a medical condition, serious intercurrent illness, or other circumstance that, in the Investigator's judgment, could jeopardize the subject's safety, or could interfere with study procedures. 2. Enrollment in this study previously. 3. Use of antiperspirants and deodorants or the following vaginal products: douches, washes, lubricants, vaginal wipes, vaginal moisturizers, or feminine hygiene spray in the genital area, within 48 hours prior to enrollment.

Design outcomes

Primary

MeasureTime frameDescription
The i) Percent Sensitivity and ii) Percent Specificity of Click Device for Detection of Chlamydia Trachomatis (CT) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCPDay 1The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive. FN = the number of specimens where the Click result is Negative and the PIS designation is Positive. Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative. FP = the number of specimens where the Click result is Positive and the PIS designation is Negative. Specificity = 100 \* TN / (TN + FP)
The i) Percent Sensitivity and ii) Percent Specificity of Click Device for Detection of Neisseria Gonorrhoeae (NG) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCPDay 1The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive. FN = the number of specimens where the Click result is Negative and the PIS designation is Positive. Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative. FP = the number of specimens where the Click result is Positive and the PIS designation is Negative. Specificity = 100 \* TN / (TN + FP)
The i) Percent Sensitivity and ii) Percent Specificity of Click Device for Detection of Trichomonas Vaginalis (TV) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCPDay 1The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive. FN = the number of specimens where the Click result is Negative and the PIS designation is Positive. Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative. FP = the number of specimens where the Click result is Positive and the PIS designation is Negative. Specificity = 100 \* TN / (TN + FP)

Secondary

MeasureTime frameDescription
The i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Symptomatic ParticipantsDay 1If a participant reported Yes to any of the following symptoms, the participant is classified as Symptomatic: Unusual vaginal discharge; Vaginal irritation (itching, burning, soreness); Lower abdominal/pelvic pain; Painful urination; Increased urinary frequency; Abnormal bleeding/spotting; Pain or bleeding with sex/intercourse. Otherwise, the participant is classified as Asymptomatic. The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive. FN = the number of specimens where the Click result is Negative and the PIS designation is Positive. Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative. FP = the number of specimens where the Click result is Positive and the PIS designation is Negative. Specificity = 100 \* TN / (TN + FP)
The i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Asymptomatic ParticipantsDay 1If a participant reported Yes to any of the following symptoms, the participant is classified as Symptomatic: Unusual vaginal discharge; Vaginal irritation (itching, burning, soreness); Lower abdominal/pelvic pain; Painful urination; Increased urinary frequency; Abnormal bleeding/spotting; Pain or bleeding with sex/intercourse. Otherwise, the participant is classified as Asymptomatic. The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive. FN = the number of specimens where the Click result is Negative and the PIS designation is Positive. Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative. FP = the number of specimens where the Click result is Positive and the PIS designation is Negative. Specificity = 100 \* TN / (TN + FP)
The i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Symptomatic ParticipantsDay 1If a participant reported Yes to any of the following symptoms, the participant is classified as Symptomatic: Unusual vaginal discharge; Vaginal irritation (itching, burning, soreness); Lower abdominal/pelvic pain; Painful urination; Increased urinary frequency; Abnormal bleeding/spotting; Pain or bleeding with sex/intercourse. Otherwise, the participant is classified as Asymptomatic. The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive. FN = the number of specimens where the Click result is Negative and the PIS designation is Positive. Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative. FP = the number of specimens where the Click result is Positive and the PIS designation is Negative. Specificity = 100 \* TN / (TN + FP)
The i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Symptomatic ParticipantsDay 1If a participant reported Yes to any of the following symptoms, the participant is classified as Symptomatic: Unusual vaginal discharge; Vaginal irritation (itching, burning, soreness); Lower abdominal/pelvic pain; Painful urination; Increased urinary frequency; Abnormal bleeding/spotting; Pain or bleeding with sex/intercourse. Otherwise, the participant is classified as Asymptomatic. The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive. FN = the number of specimens where the Click result is Negative and the PIS designation is Positive. Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative. FP = the number of specimens where the Click result is Positive and the PIS designation is Negative. Specificity = 100 \* TN / (TN + FP)
The i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Asymptomatic ParticipantsDay 1If a participant reported Yes to any of the following symptoms, the participant is classified as Symptomatic: Unusual vaginal discharge; Vaginal irritation (itching, burning, soreness); Lower abdominal/pelvic pain; Painful urination; Increased urinary frequency; Abnormal bleeding/spotting; Pain or bleeding with sex/intercourse. Otherwise, the participant is classified as Asymptomatic. The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive. FN = the number of specimens where the Click result is Negative and the PIS designation is Positive. Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative. FP = the number of specimens where the Click result is Positive and the PIS designation is Negative. Specificity = 100 \* TN / (TN + FP)
The i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Asymptomatic ParticipantsDay 1If a participant reported Yes to any of the following symptoms, the participant is classified as Symptomatic: Unusual vaginal discharge; Vaginal irritation (itching, burning, soreness); Lower abdominal/pelvic pain; Painful urination; Increased urinary frequency; Abnormal bleeding/spotting; Pain or bleeding with sex/intercourse. Otherwise, the participant is classified as Asymptomatic. The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive. FN = the number of specimens where the Click result is Negative and the PIS designation is Positive. Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative. FP = the number of specimens where the Click result is Positive and the PIS designation is Negative. Specificity = 100 \* TN / (TN + FP)

Countries

United States

Participant flow

Recruitment details

Participants were females 14 to 80 years of age, inclusive, recruited from 10 clinic sites across 7 states at locations including but not limited to: OB/GYN and primary care offices, as well as sexually transmitted disease, teen, public health, and family planning clinics. Participants were enrolled from 27FEB2019 to 02DEC2019.

Participants by arm

ArmCount
Click Device
Female participants \>=14 years of age who may have been symptomatic or asymptomatic for STIs. Participants who met the inclusion/exclusion criteria and were enrolled performed self-collection with a vaginal swab for the Click device, and allowed the health care professional (HCP) to collect three additional vaginal swabs for the comparator methods.
1,585
Total1,585

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyEnrolled but Device not Utilized1
Overall StudyProtocol Violation1
Overall StudyWithdrawal by Subject2

Baseline characteristics

CharacteristicClick Device
Age, Categorical
<=18 years
9 Participants
Age, Categorical
>=65 years
25 Participants
Age, Categorical
Between 18 and 65 years
1551 Participants
Age, Continuous35.1 years
STANDARD_DEVIATION 14.3
Ethnicity (NIH/OMB)
Hispanic or Latino
310 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
1168 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
107 Participants
Evaluable Results
C.trachomatis
1457 Participants
Evaluable Results
N. gonorrhoeae
1468 Participants
Evaluable Results
T. vaginalis
1449 Participants
Race (NIH/OMB)
American Indian or Alaska Native
12 Participants
Race (NIH/OMB)
Asian
86 Participants
Race (NIH/OMB)
Black or African American
876 Participants
Race (NIH/OMB)
More than one race
70 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
6 Participants
Race (NIH/OMB)
Unknown or Not Reported
121 Participants
Race (NIH/OMB)
White
414 Participants
Region of Enrollment
United States
1585 participants
Sex: Female, Male
Female
1585 Participants
Sex: Female, Male
Male
0 Participants
Symptomatic Status
C. trachomatis : Asymptomatic
722 participants
Symptomatic Status
C. trachomatis : Symptomatic
735 participants
Symptomatic Status
N. gonorrhoeae : Asymptomatic
722 participants
Symptomatic Status
N. gonorrhoeae : Symptomatic
746 participants
Symptomatic Status
T. vaginalis : Asymptomatic
715 participants
Symptomatic Status
T. vaginalis : Symptomatic
734 participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 1,585
other
Total, other adverse events
0 / 1,585
serious
Total, serious adverse events
0 / 1,585

Outcome results

Primary

The i) Percent Sensitivity and ii) Percent Specificity of Click Device for Detection of Chlamydia Trachomatis (CT) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP

The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive. FN = the number of specimens where the Click result is Negative and the PIS designation is Positive. Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative. FP = the number of specimens where the Click result is Positive and the PIS designation is Negative. Specificity = 100 \* TN / (TN + FP)

Time frame: Day 1

Population: Evaluable Population: This population included all samples which provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for CT. A sample provides an evaluable result if both the Click device and the PIS provide a valid positive or negative result.

ArmMeasureGroupValue (NUMBER)
Click DeviceThe i) Percent Sensitivity and ii) Percent Specificity of Click Device for Detection of Chlamydia Trachomatis (CT) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCPSensitivity97.6 percentage of participants
Click DeviceThe i) Percent Sensitivity and ii) Percent Specificity of Click Device for Detection of Chlamydia Trachomatis (CT) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCPSpecificity98.3 percentage of participants
Primary

The i) Percent Sensitivity and ii) Percent Specificity of Click Device for Detection of Neisseria Gonorrhoeae (NG) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP

The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive. FN = the number of specimens where the Click result is Negative and the PIS designation is Positive. Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative. FP = the number of specimens where the Click result is Positive and the PIS designation is Negative. Specificity = 100 \* TN / (TN + FP)

Time frame: Day 1

Population: Evaluable Population: This population included all samples which provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for NG. A sample provides an evaluable result if both the Click device and the PIS provide a valid positive or negative result.

ArmMeasureGroupValue (NUMBER)
Click DeviceThe i) Percent Sensitivity and ii) Percent Specificity of Click Device for Detection of Neisseria Gonorrhoeae (NG) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCPSensitivity97.4 percentage of participants
Click DeviceThe i) Percent Sensitivity and ii) Percent Specificity of Click Device for Detection of Neisseria Gonorrhoeae (NG) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCPSpecificity99.4 percentage of participants
Primary

The i) Percent Sensitivity and ii) Percent Specificity of Click Device for Detection of Trichomonas Vaginalis (TV) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP

The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive. FN = the number of specimens where the Click result is Negative and the PIS designation is Positive. Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative. FP = the number of specimens where the Click result is Positive and the PIS designation is Negative. Specificity = 100 \* TN / (TN + FP)

Time frame: Day 1

Population: Evaluable Population: This population included all samples which provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for TV. A sample provides an evaluable result if both the Click device and the PIS provide a valid positive or negative result.

ArmMeasureGroupValue (NUMBER)
Click DeviceThe i) Percent Sensitivity and ii) Percent Specificity of Click Device for Detection of Trichomonas Vaginalis (TV) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCPSensitivity99.2 percentage of participants
Click DeviceThe i) Percent Sensitivity and ii) Percent Specificity of Click Device for Detection of Trichomonas Vaginalis (TV) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCPSpecificity96.9 percentage of participants
Secondary

The i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Asymptomatic Participants

If a participant reported Yes to any of the following symptoms, the participant is classified as Symptomatic: Unusual vaginal discharge; Vaginal irritation (itching, burning, soreness); Lower abdominal/pelvic pain; Painful urination; Increased urinary frequency; Abnormal bleeding/spotting; Pain or bleeding with sex/intercourse. Otherwise, the participant is classified as Asymptomatic. The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive. FN = the number of specimens where the Click result is Negative and the PIS designation is Positive. Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative. FP = the number of specimens where the Click result is Positive and the PIS designation is Negative. Specificity = 100 \* TN / (TN + FP)

Time frame: Day 1

Population: Evaluable Asymptomatic Population: This population included all samples from asymptomatic participants which provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for CT. A sample provides an evaluable result if both the Click device and the PIS provide a valid positive or negative result.

ArmMeasureGroupValue (NUMBER)
Click DeviceThe i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Asymptomatic ParticipantsSensitivity95.8 percentage of participants
Click DeviceThe i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Asymptomatic ParticipantsSpecificity99.0 percentage of participants
Secondary

The i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Symptomatic Participants

If a participant reported Yes to any of the following symptoms, the participant is classified as Symptomatic: Unusual vaginal discharge; Vaginal irritation (itching, burning, soreness); Lower abdominal/pelvic pain; Painful urination; Increased urinary frequency; Abnormal bleeding/spotting; Pain or bleeding with sex/intercourse. Otherwise, the participant is classified as Asymptomatic. The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive. FN = the number of specimens where the Click result is Negative and the PIS designation is Positive. Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative. FP = the number of specimens where the Click result is Positive and the PIS designation is Negative. Specificity = 100 \* TN / (TN + FP)

Time frame: Day 1

Population: Evaluable Symptomatic Population: This population included all samples from symptomatic participants which provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for CT. A sample provides an evaluable result if both the Click device and the PIS provide a valid positive or negative result.

ArmMeasureGroupValue (NUMBER)
Click DeviceThe i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Symptomatic ParticipantsSpecificity97.7 percentage of participants
Click DeviceThe i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Symptomatic ParticipantsSensitivity98.7 percentage of participants
Secondary

The i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Asymptomatic Participants

If a participant reported Yes to any of the following symptoms, the participant is classified as Symptomatic: Unusual vaginal discharge; Vaginal irritation (itching, burning, soreness); Lower abdominal/pelvic pain; Painful urination; Increased urinary frequency; Abnormal bleeding/spotting; Pain or bleeding with sex/intercourse. Otherwise, the participant is classified as Asymptomatic. The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive. FN = the number of specimens where the Click result is Negative and the PIS designation is Positive. Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative. FP = the number of specimens where the Click result is Positive and the PIS designation is Negative. Specificity = 100 \* TN / (TN + FP)

Time frame: Day 1

Population: Evaluable Asymptomatic Population: This population included all samples from asymptomatic participants which provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for NG. A sample provides an evaluable result if both the Click device and the PIS provide a valid positive or negative result.

ArmMeasureGroupValue (NUMBER)
Click DeviceThe i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Asymptomatic ParticipantsSensitivity94.4 percentage of participants
Click DeviceThe i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Asymptomatic ParticipantsSpecificity99.4 percentage of participants
Secondary

The i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Symptomatic Participants

If a participant reported Yes to any of the following symptoms, the participant is classified as Symptomatic: Unusual vaginal discharge; Vaginal irritation (itching, burning, soreness); Lower abdominal/pelvic pain; Painful urination; Increased urinary frequency; Abnormal bleeding/spotting; Pain or bleeding with sex/intercourse. Otherwise, the participant is classified as Asymptomatic. The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive. FN = the number of specimens where the Click result is Negative and the PIS designation is Positive. Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative. FP = the number of specimens where the Click result is Positive and the PIS designation is Negative. Specificity = 100 \* TN / (TN + FP)

Time frame: Day 1

Population: Evaluable Symptomatic Population: This population included all samples from symptomatic participants which provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for NG. A sample provides an evaluable result if both the Click device and the PIS provide a valid positive or negative result.

ArmMeasureGroupValue (NUMBER)
Click DeviceThe i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Symptomatic ParticipantsSensitivity100 percentage of participants
Click DeviceThe i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Symptomatic ParticipantsSpecificity99.4 percentage of participants
Secondary

The i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Asymptomatic Participants

If a participant reported Yes to any of the following symptoms, the participant is classified as Symptomatic: Unusual vaginal discharge; Vaginal irritation (itching, burning, soreness); Lower abdominal/pelvic pain; Painful urination; Increased urinary frequency; Abnormal bleeding/spotting; Pain or bleeding with sex/intercourse. Otherwise, the participant is classified as Asymptomatic. The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive. FN = the number of specimens where the Click result is Negative and the PIS designation is Positive. Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative. FP = the number of specimens where the Click result is Positive and the PIS designation is Negative. Specificity = 100 \* TN / (TN + FP)

Time frame: Day 1

Population: Evaluable Asymptomatic Population: This population included all samples from asymptomatic participants which provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for TV. A sample provides an evaluable result if both the Click device and the PIS provide a valid positive or negative result.

ArmMeasureGroupValue (NUMBER)
Click DeviceThe i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Asymptomatic ParticipantsSensitivity100 percentage of participants
Click DeviceThe i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Asymptomatic ParticipantsSpecificity98.0 percentage of participants
Secondary

The i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Symptomatic Participants

If a participant reported Yes to any of the following symptoms, the participant is classified as Symptomatic: Unusual vaginal discharge; Vaginal irritation (itching, burning, soreness); Lower abdominal/pelvic pain; Painful urination; Increased urinary frequency; Abnormal bleeding/spotting; Pain or bleeding with sex/intercourse. Otherwise, the participant is classified as Asymptomatic. The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive. FN = the number of specimens where the Click result is Negative and the PIS designation is Positive. Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative. FP = the number of specimens where the Click result is Positive and the PIS designation is Negative. Specificity = 100 \* TN / (TN + FP)

Time frame: Day 1

Population: Evaluable Symptomatic Population: This population included all samples from symptomatic participants which provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for TV. A sample provides an evaluable result if both the Click device and the PIS provide a valid positive or negative result.

ArmMeasureGroupValue (NUMBER)
Click DeviceThe i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Symptomatic ParticipantsSensitivity98.6 percentage of participants
Click DeviceThe i) Percent Sensitivity and ii) Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV From Self-collected Specimens as Compared to PIS Using Specimens Collected by a Qualified HCP Among Symptomatic ParticipantsSpecificity95.8 percentage of participants