Skip to content

Validation of the Aktiia SA PulseWatch OBPM Device at the Wrist Against Invasive Blood Pressure Measurements

Single-center Intervention Study to Validate the Performance of the Aktiia SA PulseWatch Optical Blood Pressure Monitoring (OBPM) Device at the Wrist Against Invasive Blood Pressure Measurements With Arterial Line

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03837769
Acronym
OBPM_ICU2018
Enrollment
31
Registered
2019-02-12
Start date
2019-02-12
Completion date
2019-05-31
Last updated
2019-06-03

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Blood Pressure

Keywords

Blood Pressure, OBPM, Cuffless

Brief summary

Single-centre intervention study to validate the performance of the Aktiia SA optical blood pressure monitoring (OBPM) device at the wrist against invasive blood pressure measurements with arterial line.

Detailed description

Aktiia SA PulseWatch is an investigational device for the optical measurement of blood pressure that uses only optical sensors at the wrist. HNE is interested in exploring the potential of the device and technology (OBPM), and aims at benchmarking its performances against gold standard measurements of blood pressure. The goal of this study is thus to assess whether i) PulseWatch allows to capture blood pressure variations, and ii) the measured blood pressure values remain stable in time. The proposed study plans to include patients that are already scheduled for arterial catheterization at HNE. For these patients, the study will only require the placement of the safe optical device on the wrist to non-invasively record hemodynamic fluctuations. Therefore, no additional risk for the patient is to be foreseen. By demonstrating that the Aktiia SA OBPM technology is reliable, HNE aims to make one step further in improving blood pressure monitoring in general, and in particular in advancing the deployment of technologies that have the potential to be used in the ambulatory setting. The diagnosis and treatment of hypertension are expected to largely benefit from these advancements.

Interventions

DEVICEAktiia OBPM PulseWatch wrist device

The optical signals at the wrist are recorded non-invasively. The BP measurements are further determined from these optical signals and are compared to the reference BP readings.

Sponsors

Aktiia SA
CollaboratorINDUSTRY
Hopital Neuchatelois
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
BASIC_SCIENCE
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Age ≥ 18 years * No significant non-invasive systolic blood pressure difference between left arm and right arm (difference \<20mmHg in systolic arterial pressure) * Patient requires continuous monitoring of diastolic and systolic blood pressure continuously via a radial arterial line; * Patient information and informed consent is obtained following the procedures described in CIP

Exclusion criteria

* Arrythmias: tachycardia (resting heart rate \> 120/min) at time of study inclusion * Atrial fibrillation * Psychomotor agitation * Significant bruises or trauma on forearm * Intravenous canula around the wrist with impossibility to place the device (Aktiia OBPM PulseWatch) around the wrist

Design outcomes

Primary

MeasureTime frameDescription
Cuffless Blood Pressure performance4 MonthsThe evaluation of the accuracy (mean error and standard deviation of the error) of the non-invasive BP measurements at wrist with the PulseWatch against the reference gold-standard invasive blood pressure obtained by arterial line with respect to ISO 81060-2 (14) for two aspects: i) capturing the variations of the BP in short time and ii) maintaining the accurate stable measurements over extended periods of time.

Countries

Switzerland

Outcome results

None listed