Ametropia
Conditions
Brief summary
The aim of this prospective study is to evaluate the clinical performance of habitual wearers of methafilcon A toric lenses after a refit with fanfilcon A toric lenses for 1-month of daily wear.
Interventions
Bilateral daily wear of methafilcon A toric contact lenses
Bilateral daily wear of fanfilcon A toric contact lenses
Sponsors
Study design
Masking description
Study lenses will be transferred, by an assistant, out of their packaging to unmarked new contact lens cases filled with unpreserved sterile saline just prior to dispensing to maintain subject masking of the study lenses
Eligibility
Inclusion criteria
* Is between 18 and 40 years of age (inclusive) * Has had a self-reported visual exam in the last two years * Is an adapted soft contact lens wearer * Has a contact lens spherical prescription between +6.00 to - 8.00 (inclusive)-Have no less than -0.75D of astigmatism and no more than -2.25 D in both eyes * Can achieve best corrected spectacle distance visual acuity of 20/25 (0.10 logMAR) or better in each eye. * Can achieve a distance visual acuity of 20/30 (0.18 logMAR) or better in each eye with the study contact lenses. * Has clear corneas and no active ocular disease * Has read, understood and signed the information consent letter * Patient contact lens refraction should fit within the available parameters of the study lenses * Is willing to comply with the wear schedule (at least 5 days per week, \>8 hours/day assuming there are no contraindications for doing so) * Is willing to comply with the visit schedule
Exclusion criteria
* Has a Contact Lens (CL) prescription outside the range of the available parameters of the study lenses * Has a spectacle cylinder less than -0.75D or more than -2.50 D of cylinder in either eye * Has a history of not achieving comfortable Contact Lens (CL) wear (5 days per week; \> 8 hours/day) * Has contact lens best corrected distance vision worse than 20/25 (0.10 logMAR) in either eye * Presence of clinically significant (grade 2-4) anterior segment abnormalities * Presence of ocular or systemic disease or need of medications which might interfere with contact lens wear * Slit lamp findings that would contraindicate contact lens wear such as: * Pathological dry eye or associated findings * Pterygium, pinguecula, or corneal scars within the visual axis * Neovascularization \> 0.75 mm in from of the limbus * Giant papillary conjunctivitis (GCP) worse than grade 1 * Anterior uveitis or iritis (past or present) * Seborrheic eczema, Seborrheic conjunctivitis * History of corneal ulcers or fungal infections * Poor personal hygiene * Has a known history of corneal hypoesthesia (reduced corneal sensitivity) * Has aphakia, keratoconus or a highly irregular cornea * Has Presbyopia or has dependence on spectacles for near work over the contact lenses * Has undergone corneal refractive surgery * Is participating in any other type of eye related clinical or research study.
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Overall Lens Fit Acceptance by the Investigator - Fanfilcon A Toric Contact Lenses | Baseline | Overall lens fit acceptance assessed using grade (0-4) 0=Should not be worn, 1=Borderline but unacceptable, 2=Minimal acceptable, early review, 3=Not perfect but okay to dispense, 4=perfect. |
| Rotational Recovery Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | Baseline | Lens rotational recovery was assessed in 5-degree steps 1 minute after the lens was manually rotated 45degrees temporally from the primary gaze position |
| Overall Stability Assessment by the Investigator - Methafilcon A Toric Contact Lenses | Baseline | Overall lens stability was assessed using a 0 - 4 scale (where 0 = very poor stability, and 4 = excellent stability). |
| Overall Stability Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | Baseline | Overall lens stability was assessed using a 0 - 4 scale (where 0 = very poor stability, and 4 = excellent stability). |
| Overall Lens Fit Acceptance by the Investigator - Methafilcon A Toric Contact Lenses | Baseline | Overall lens fit acceptance assessed using grade (0-4) 0=Should not be worn, 1=Borderline but unacceptable, 2=Minimal acceptable, early review, 3=Not perfect but okay to dispense, 4=perfect. |
| Number of Participants With Lens Centration Assessment by the Investigator - Methafilcon A Toric Contact Lenses | Baseline | Lens centration was assessed on a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable). |
| Number of Participants With Lens Centration Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | Baseline | Lens centration was assessed on a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable). |
| Percentage of Participants With Corneal Coverage Assessment by the Investigator - Methafilcon A Toric Contact Lenses | Baseline | Corneal coverage was collected on a YES/NO response. Yes - lens covers the cornea, No - lens does not cover the cornea |
| Percentage of Participants With Corneal Coverage Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | Baseline | Corneal coverage was collected on a YES/NO response. Yes - lens covers the cornea, No - lens does not cover the cornea |
| Post-blink Movement Assessment by the Investigator - Methafilcon A Toric Contact Lenses | Baseline | Post-blink movement was assessed on a 0-4 scale (where: 0=Insufficient, unacceptable movement,1=Minimal, but acceptable movement, 2=Optimal movement, 3=Moderate, but acceptable movement, 4=Excessive, unacceptable movement). |
| Post-blink Movement Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | Baseline | Post-blink movement was assessed on a 0-4 scale (where: 0=Insufficient, unacceptable movement,1=Minimal, but acceptable movement, 2=Optimal movement, 3=Moderate, but acceptable movement, 4=Excessive, unacceptable movement). |
| Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Methalfilcon A Toric Contact Lenses | Baseline | Lens orientation was assessed in 5-degree steps with the slit lamp by measuring the amount of mislocation of the axis mark on the lens relative to the 6 o'clock position. |
| Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | Baseline | Lens orientation was assessed in 5-degree steps with the slit lamp by measuring the amount of mislocation of the axis mark on the lens relative to the 6 o'clock position. |
| Rotational Recovery Assessment by the Investigator - Methafilcon A Toric Contact Lenses | Baseline | Lens rotational recovery was assessed in 5-degree steps 1 minute after the lens was manually rotated 45degrees temporally from the primary gaze position |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Average Daily Wearing Time - Fanfilcon A Toric Contact Lenses | 2 weeks | Average daily wearing time response by the subject |
| Average Comfortable Wearing Time - Methafilcon A Toric Contact Lenses | 4 weeks | Average comfortable wearing time - response by the subject |
| Average Comfortable Wearing Time - Fanfilcon A Toric Contact Lenses | 2 weeks | Average comfortable wearing time - response by the subject |
| Average Daily Wearing Time - Methafilcon A Toric Contact Lenses | 4 weeks | Average daily wearing time response by the subject |
Countries
Mexico
Participant flow
Participants by arm
| Arm | Count |
|---|---|
| Overall Study Total Participants | 43 |
| Total | 43 |
Withdrawals & dropouts
| Period | Reason | FG000 |
|---|---|---|
| Methafilcon A | Withdrawal by Subject | 1 |
Baseline characteristics
| Characteristic | Overall Study | — |
|---|---|---|
| Age, Categorical <=18 years | 0 Participants | — |
| Age, Categorical >=65 years | 0 Participants | — |
| Age, Categorical Between 18 and 65 years | 43 Participants | — |
| Age, Continuous | 24.2 years STANDARD_DEVIATION 6.9 | — |
| Race and Ethnicity Not Collected | — | — Participants |
| Region of Enrollment Mexico | 43 participants | — |
| Sex: Female, Male Female | 26 Participants | — |
| Sex: Female, Male Male | 17 Participants | — |
Adverse events
| Event type | EG000 affected / at risk | EG001 affected / at risk |
|---|---|---|
| deaths Total, all-cause mortality | 0 / 43 | 0 / 42 |
| other Total, other adverse events | 0 / 43 | 0 / 42 |
| serious Total, serious adverse events | 0 / 43 | 0 / 42 |
Outcome results
Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Fanfilcon A Toric Contact Lenses
Lens orientation was assessed in 5-degree steps with the slit lamp by measuring the amount of mislocation of the axis mark on the lens relative to the 6 o'clock position.
Time frame: 4 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | 0.47 Degrees | Standard Deviation 1.9 |
Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Fanfilcon A Toric Contact Lenses
Lens orientation was assessed in 5-degree steps with the slit lamp by measuring the amount of mislocation of the axis mark on the lens relative to the 6 o'clock position.
Time frame: Baseline
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | 0.71 Degrees | Standard Deviation 1.8 |
Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Fanfilcon A Toric Contact Lenses
Lens orientation was assessed in 5-degree steps with the slit lamp by measuring the amount of mislocation of the axis mark on the lens relative to the 6 o'clock position.
Time frame: 2 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | 0.71 Degrees | Standard Deviation 1.8 |
Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Methalfilcon A Toric Contact Lenses
Lens orientation was assessed in 5-degree steps with the slit lamp by measuring the amount of mislocation of the axis mark on the lens relative to the 6 o'clock position.
Time frame: Baseline
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Methalfilcon A Toric Contact Lenses | 2.21 Degrees | Standard Deviation 4 |
Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Methalfilcon A Toric Contact Lenses
Lens orientation was assessed in 5-degree steps with the slit lamp by measuring the amount of mislocation of the axis mark on the lens relative to the 6 o'clock position.
Time frame: 4 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Methalfilcon A Toric Contact Lenses | 1.55 Degrees | Standard Deviation 3.2 |
Number of Participants With Lens Centration Assessment by the Investigator - Fanfilcon A Toric Contact Lenses
Lens centration was assessed on a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable).
Time frame: Baseline
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Methafilcon A | Number of Participants With Lens Centration Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | Optimal centration | 42 Participants |
| Methafilcon A | Number of Participants With Lens Centration Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | Decentration acceptable | 0 Participants |
| Methafilcon A | Number of Participants With Lens Centration Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | Decentration unacceptable | 0 Participants |
Number of Participants With Lens Centration Assessment by the Investigator - Fanfilcon A Toric Contact Lenses
Lens centration was assessed on a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable).
Time frame: 4 weeks
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Methafilcon A | Number of Participants With Lens Centration Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | Optimal centration | 42 Participants |
| Methafilcon A | Number of Participants With Lens Centration Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | Decentration acceptable | 0 Participants |
| Methafilcon A | Number of Participants With Lens Centration Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | Decentration unacceptable | 0 Participants |
Number of Participants With Lens Centration Assessment by the Investigator - Fanfilcon A Toric Contact Lenses
Lens centration was assessed on a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable).
Time frame: 2 weeks
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Methafilcon A | Number of Participants With Lens Centration Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | Optimal centration | 42 Participants |
| Methafilcon A | Number of Participants With Lens Centration Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | Decentration acceptable | 0 Participants |
| Methafilcon A | Number of Participants With Lens Centration Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | Decentration unacceptable | 0 Participants |
Number of Participants With Lens Centration Assessment by the Investigator - Methafilcon A Toric Contact Lenses
Lens centration was assessed on a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable).
Time frame: Baseline
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Methafilcon A | Number of Participants With Lens Centration Assessment by the Investigator - Methafilcon A Toric Contact Lenses | Optimal Centration | 40 Participants |
| Methafilcon A | Number of Participants With Lens Centration Assessment by the Investigator - Methafilcon A Toric Contact Lenses | Decentration acceptable | 3 Participants |
| Methafilcon A | Number of Participants With Lens Centration Assessment by the Investigator - Methafilcon A Toric Contact Lenses | Decentration unacceptable | 0 Participants |
Number of Participants With Lens Centration Assessment by the Investigator - Methafilcon A Toric Contact Lenses
Lens centration was assessed on a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable).
Time frame: 4 weeks
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Methafilcon A | Number of Participants With Lens Centration Assessment by the Investigator - Methafilcon A Toric Contact Lenses | Optimal centration | 41 Participants |
| Methafilcon A | Number of Participants With Lens Centration Assessment by the Investigator - Methafilcon A Toric Contact Lenses | Decentration acceptable | 1 Participants |
| Methafilcon A | Number of Participants With Lens Centration Assessment by the Investigator - Methafilcon A Toric Contact Lenses | Decentration unacceptable | 0 Participants |
Overall Lens Fit Acceptance by the Investigator - Fanfilcon A Toric Contact Lenses
Overall lens fit acceptance assessed using grade (0-4) 0=Should not be worn, 1=Borderline but unacceptable, 2=Minimal acceptable, early review, 3=Not perfect but okay to dispense, 4=perfect.
Time frame: 2 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Overall Lens Fit Acceptance by the Investigator - Fanfilcon A Toric Contact Lenses | 4.0 units on a scale | Standard Deviation 0.2 |
Overall Lens Fit Acceptance by the Investigator - Fanfilcon A Toric Contact Lenses
Overall lens fit acceptance assessed using grade (0-4) 0=Should not be worn, 1=Borderline but unacceptable, 2=Minimal acceptable, early review, 3=Not perfect but okay to dispense, 4=perfect.
Time frame: Baseline
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Overall Lens Fit Acceptance by the Investigator - Fanfilcon A Toric Contact Lenses | 4.0 units on a scale | Standard Deviation 0.2 |
Overall Lens Fit Acceptance by the Investigator - Fanfilcon A Toric Contact Lenses
Overall lens fit acceptance assessed using grade (0-4) 0=Should not be worn, 1=Borderline but unacceptable, 2=Minimal acceptable, early review, 3=Not perfect but okay to dispense, 4=perfect.
Time frame: 4 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Overall Lens Fit Acceptance by the Investigator - Fanfilcon A Toric Contact Lenses | 4.0 units on a scale | Standard Deviation 0.2 |
Overall Lens Fit Acceptance by the Investigator - Methafilcon A Toric Contact Lenses
Overall lens fit acceptance assessed using grade (0-4) 0=Should not be worn, 1=Borderline but unacceptable, 2=Minimal acceptable, early review, 3=Not perfect but okay to dispense, 4=perfect.
Time frame: 4 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Overall Lens Fit Acceptance by the Investigator - Methafilcon A Toric Contact Lenses | 3.8 units on a scale | Standard Deviation 0.4 |
Overall Lens Fit Acceptance by the Investigator - Methafilcon A Toric Contact Lenses
Overall lens fit acceptance assessed using grade (0-4) 0=Should not be worn, 1=Borderline but unacceptable, 2=Minimal acceptable, early review, 3=Not perfect but okay to dispense, 4=perfect.
Time frame: Baseline
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Overall Lens Fit Acceptance by the Investigator - Methafilcon A Toric Contact Lenses | 3.8 units on a scale | Standard Deviation 0.4 |
Overall Stability Assessment by the Investigator - Fanfilcon A Toric Contact Lenses
Overall lens stability was assessed using a 0 - 4 scale (where 0 = very poor stability, and 4 = excellent stability).
Time frame: Baseline
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Overall Stability Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | 4.0 units on a scale | Standard Deviation 0.2 |
Overall Stability Assessment by the Investigator - Fanfilcon A Toric Contact Lenses
Overall lens stability was assessed using a 0 - 4 scale (where 0 = very poor stability, and 4 = excellent stability).
Time frame: 2 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Overall Stability Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | 4.0 units on a scale | Standard Deviation 0.2 |
Overall Stability Assessment by the Investigator - Fanfilcon A Toric Contact Lenses
Overall lens stability was assessed using a 0 - 4 scale (where 0 = very poor stability, and 4 = excellent stability).
Time frame: 4 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Overall Stability Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | 4.0 units on a scale | Standard Deviation 0.2 |
Overall Stability Assessment by the Investigator - Methafilcon A Toric Contact Lenses
Overall lens stability was assessed using a 0 - 4 scale (where 0 = very poor stability, and 4 = excellent stability).
Time frame: 4 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Overall Stability Assessment by the Investigator - Methafilcon A Toric Contact Lenses | 3.8 units on a scale | Standard Deviation 0.4 |
Overall Stability Assessment by the Investigator - Methafilcon A Toric Contact Lenses
Overall lens stability was assessed using a 0 - 4 scale (where 0 = very poor stability, and 4 = excellent stability).
Time frame: Baseline
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Overall Stability Assessment by the Investigator - Methafilcon A Toric Contact Lenses | 3.8 units on a scale | Standard Deviation 0.4 |
Percentage of Participants With Corneal Coverage Assessment by the Investigator - Fanfilcon A Toric Contact Lenses
Corneal coverage was collected on a YES/NO response. Yes - lens covers the cornea, No - lens does not cover the cornea
Time frame: Baseline
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Methafilcon A | Percentage of Participants With Corneal Coverage Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | Yes | 100 percentage of participants |
| Methafilcon A | Percentage of Participants With Corneal Coverage Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | No | 0 percentage of participants |
Percentage of Participants With Corneal Coverage Assessment by the Investigator - Fanfilcon A Toric Contact Lenses
Corneal coverage was collected on a YES/NO response. Yes - lens covers the cornea, No - lens does not cover the cornea
Time frame: 2 weeks
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Methafilcon A | Percentage of Participants With Corneal Coverage Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | Yes | 100 percentage of participants |
| Methafilcon A | Percentage of Participants With Corneal Coverage Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | No | 0 percentage of participants |
Percentage of Participants With Corneal Coverage Assessment by the Investigator - Fanfilcon A Toric Contact Lenses
Corneal coverage was collected on a YES/NO response. Yes - lens covers the cornea, No - lens does not cover the cornea
Time frame: 4 weeks
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Methafilcon A | Percentage of Participants With Corneal Coverage Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | Yes | 100 percentage of participants |
| Methafilcon A | Percentage of Participants With Corneal Coverage Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | No | 0 percentage of participants |
Percentage of Participants With Corneal Coverage Assessment by the Investigator - Methafilcon A Toric Contact Lenses
Corneal coverage was collected on a YES/NO response. Yes - lens covers the cornea, No - lens does not cover the cornea
Time frame: Baseline
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Methafilcon A | Percentage of Participants With Corneal Coverage Assessment by the Investigator - Methafilcon A Toric Contact Lenses | Yes | 100 percentage of participants |
| Methafilcon A | Percentage of Participants With Corneal Coverage Assessment by the Investigator - Methafilcon A Toric Contact Lenses | No | 0 percentage of participants |
Percentage of Participants With Corneal Coverage Assessment by the Investigator - Methafilcon A Toric Contact Lenses
Corneal coverage was collected on a YES/NO response. Yes - lens covers the cornea, No - lens does not cover the cornea
Time frame: 4 weeks
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Methafilcon A | Percentage of Participants With Corneal Coverage Assessment by the Investigator - Methafilcon A Toric Contact Lenses | Yes | 100 percentage of participants |
| Methafilcon A | Percentage of Participants With Corneal Coverage Assessment by the Investigator - Methafilcon A Toric Contact Lenses | No | 0 percentage of participants |
Post-blink Movement Assessment by the Investigator - Fanfilcon A Toric Contact Lenses
Post-blink movement was assessed on a 0-4 scale (where: 0=Insufficient, unacceptable movement,1=Minimal, but acceptable movement, 2=Optimal movement, 3=Moderate, but acceptable movement, 4=Excessive, unacceptable movement).
Time frame: Baseline
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Post-blink Movement Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | 2.0 units on a scale | Standard Deviation 0 |
Post-blink Movement Assessment by the Investigator - Fanfilcon A Toric Contact Lenses
Post-blink movement was assessed on a 0-4 scale (where: 0=Insufficient, unacceptable movement,1=Minimal, but acceptable movement, 2=Optimal movement, 3=Moderate, but acceptable movement, 4=Excessive, unacceptable movement).
Time frame: 4 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Post-blink Movement Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | 1.97 units on a scale | Standard Deviation 0.15 |
Post-blink Movement Assessment by the Investigator - Fanfilcon A Toric Contact Lenses
Post-blink movement was assessed on a 0-4 scale (where: 0=Insufficient, unacceptable movement,1=Minimal, but acceptable movement, 2=Optimal movement, 3=Moderate, but acceptable movement, 4=Excessive, unacceptable movement).
Time frame: 2 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Post-blink Movement Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | 1.97 units on a scale | Standard Deviation 0.15 |
Post-blink Movement Assessment by the Investigator - Methafilcon A Toric Contact Lenses
Post-blink movement was assessed on a 0-4 scale (where: 0=Insufficient, unacceptable movement,1=Minimal, but acceptable movement, 2=Optimal movement, 3=Moderate, but acceptable movement, 4=Excessive, unacceptable movement).
Time frame: Baseline
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Post-blink Movement Assessment by the Investigator - Methafilcon A Toric Contact Lenses | 2.0 units on a scale | Standard Deviation 0 |
Post-blink Movement Assessment by the Investigator - Methafilcon A Toric Contact Lenses
Post-blink movement was assessed on a 0-4 scale (where: 0=Insufficient, unacceptable movement,1=Minimal, but acceptable movement, 2=Optimal movement, 3=Moderate, but acceptable movement, 4=Excessive, unacceptable movement).
Time frame: 4 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Post-blink Movement Assessment by the Investigator - Methafilcon A Toric Contact Lenses | 1.97 units on a scale | Standard Deviation 0.15 |
Rotational Recovery Assessment by the Investigator - Fanfilcon A Toric Contact Lenses
Lens rotational recovery was assessed in 5-degree steps 1 minute after the lens was manually rotated 45degrees temporally from the primary gaze position
Time frame: Baseline
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Rotational Recovery Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | 0.83 Degrees | Standard Deviation 2.2 |
Rotational Recovery Assessment by the Investigator - Fanfilcon A Toric Contact Lenses
Lens rotational recovery was assessed in 5-degree steps 1 minute after the lens was manually rotated 45degrees temporally from the primary gaze position
Time frame: 2 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Rotational Recovery Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | 0.95 Degrees | Standard Deviation 2.3 |
Rotational Recovery Assessment by the Investigator - Fanfilcon A Toric Contact Lenses
Lens rotational recovery was assessed in 5-degree steps 1 minute after the lens was manually rotated 45degrees temporally from the primary gaze position
Time frame: 4 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Rotational Recovery Assessment by the Investigator - Fanfilcon A Toric Contact Lenses | 0.48 Degrees | Standard Deviation 1.5 |
Rotational Recovery Assessment by the Investigator - Methafilcon A Toric Contact Lenses
Lens rotational recovery was assessed in 5-degree steps 1 minute after the lens was manually rotated 45degrees temporally from the primary gaze position
Time frame: 4 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Rotational Recovery Assessment by the Investigator - Methafilcon A Toric Contact Lenses | 2.02 Degrees | Standard Deviation 3.5 |
Rotational Recovery Assessment by the Investigator - Methafilcon A Toric Contact Lenses
Lens rotational recovery was assessed in 5-degree steps 1 minute after the lens was manually rotated 45degrees temporally from the primary gaze position
Time frame: Baseline
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Rotational Recovery Assessment by the Investigator - Methafilcon A Toric Contact Lenses | 2.79 Degrees | Standard Deviation 4.8 |
Average Comfortable Wearing Time - Fanfilcon A Toric Contact Lenses
Average comfortable wearing time - response by the subject
Time frame: 2 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Average Comfortable Wearing Time - Fanfilcon A Toric Contact Lenses | 8.1 hours/day | Standard Deviation 2.9 |
Average Comfortable Wearing Time - Fanfilcon A Toric Contact Lenses
Average comfortable wearing time - response by the subject
Time frame: 4 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Average Comfortable Wearing Time - Fanfilcon A Toric Contact Lenses | 8.6 hours/day | Standard Deviation 3.1 |
Average Comfortable Wearing Time - Methafilcon A Toric Contact Lenses
Average comfortable wearing time - response by the subject
Time frame: 4 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Average Comfortable Wearing Time - Methafilcon A Toric Contact Lenses | 7.6 hours/day | Standard Deviation 2.8 |
Average Daily Wearing Time - Fanfilcon A Toric Contact Lenses
Average daily wearing time response by the subject
Time frame: 2 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Average Daily Wearing Time - Fanfilcon A Toric Contact Lenses | 10.6 hours/day | Standard Deviation 2.8 |
Average Daily Wearing Time - Fanfilcon A Toric Contact Lenses
Average daily wearing time response by the subject
Time frame: 4 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Average Daily Wearing Time - Fanfilcon A Toric Contact Lenses | 11.0 hours/day | Standard Deviation 2.8 |
Average Daily Wearing Time - Methafilcon A Toric Contact Lenses
Average daily wearing time response by the subject
Time frame: 4 weeks
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Methafilcon A | Average Daily Wearing Time - Methafilcon A Toric Contact Lenses | 10.4 hours / day | Standard Deviation 2.8 |