Dexcom G6 Observational Study

Inpatient and Post-hospital Discharge Assessment of Glycemic Control by Capillary Point-of-care Glucose Testing and by Continuous Glucose Monitoring in Insulin-treated Patients With Type 1 and Type 2 Diabetes: Dexcom G6 Observational Study

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03832907
Enrollment
101
Registered
2019-02-06
Start date
2019-02-21
Completion date
2020-03-31
Last updated
2021-08-05

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes

Keywords

Glucose monitoring, POC BG, DexcomG6 CGM

Brief summary

This is an exploratory prospective study primary aiming to evaluate differences between Point-of-care (POC) testing (standard of care) and Dexcom G6 CGM (Continues Glucose Monitoring sensor system) in 1) glycemic control measured by mean daily blood glucose and 2) identifying hypoglycemic and hyperglycemic events - in patients with type 1 and type 2 diabetes treated with basal bolus insulin regimen.

Detailed description

Several studies have shown that the use of a device that can measure blood glucose levels every 5 minutes is more effective in identifying low blood glucose (hypoglycemia) and high blood glucose (hyperglycemia) when compared to regular measurements with fingersticks \[point of care (POC)\]. This is an exploratory prospective study primary aiming to evaluate differences between Point-of-care (POC) testing (standard of care) and Dexcom G6 CGM (Continues Glucose Monitoring sensor system) in 1) glycemic control measured by mean daily blood glucose and 2) identifying hypoglycemic and hyperglycemic events - in patients with type 1 and type 2 diabetes treated with basal bolus insulin regimen.

Interventions

A blinded factory-calibrated continues glucose monitoring sensor system Dexcom G6 will be placed shortly after admission. Two CGM devices will be inserted in all patients - one in the abdomen and one in the arm to also assess differences in blood glucose readings between upper extremity and abdominal insertion sites. Information on CGM readings will be collected daily during the hospital stay and after hospital discharge for 10 days using the Dexcom Studio software to download the Dexcom receiver data.

Standard of care - bedside point-of-care (POC) capillary blood glucose (BG) monitoring will be done before meals and bedtime daily during the hospital stay and after hospital discharge for 10 days.

Sponsors

Emory University
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

1. \> 18 years admitted to a general medicine or surgical service. 2. Known history of T1D or T2D receiving insulin therapy 3. Subjects must have a randomization BG between 140 mg and 400 mg/dL without laboratory evidence of diabetic ketoacidosis (bicarbonate \< 18 mEq/L, pH \< 7.30, or positive serum or urinary ketones). 4. Patients with expected hospital length-of-stay of 3 or more days

Exclusion criteria

1. Patients with acute illness admitted to the ICU or expected to require admission to the ICU. 2. Patients expected to require MRI procedures during hospitalization. 3. Mental condition rendering the subject unable to understand the nature, scope, and possible consequences of the study. 4. Female subjects who are pregnant or breast-feeding at time of enrollment into the study.

Design outcomes

Primary

MeasureTime frameDescription
Mean Daily Blood Glucose by POC Reading сompared to CGM ReadingDuring hospital stay (3 days - up to 30 days)Mean daily blood glucose will be calculated
Number of Clinical Significant Hypoglycemia (<54 mg/dl) Events by POC Reading Compared to CGM ReadingDuring hospital stay (3 days - up to 30 days)Number of clinical significant hypoglycemia (\<54 mg/dl) events will be calculated
Number of Severe Hyperglycemia (>240 mg/dl) Events by POC Reading Compared to CGM ReadingDuring hospital stay (3 days - up to 30 days)Number of severe hyperglycemia (\>240 mg/dl) events will be recorded

Secondary

MeasureTime frameDescription
Number of Nocturnal Hypoglycemic Events < 70 mg/dl (Between 22:00 and 06:00) by POC Reading Compared to CGM ReadingDuring hospital stay (3 days - up to 30 days)Number of nocturnal hypoglycemic events \< 70 mg/dl (between 22:00 and 06:00) will be recorded
Number of of Hypoglycemic Events < 70mg/dl in POC Group by POC Reading Compared to CGM ReadingDuring hospital stay (3 days - up to 30 days)Number of of hypoglycemic events \< 70mg/dl will be recorded
Number of Hypoglycemic Events < 70mg/dl by POC Reading Compared to CGM Reading10 days post dischargeNumber of of hypoglycemic events \< 70mg/dl will be recorded
Time in Hypoglycemia (Minutes) During the Day and Night by CGM ReadingDuring hospital stay (3 days - up to 30 days)Time in hypoglycemia (minutes) during the day and night will be recorded. Only measured by CGM, Discrete POC values are not used to assess amount of time spent within a specific glucose range. .
Time in Hyperglycemia (Minutes) During the Day and Night> 240 mg/dl by CGM ReadingDuring hospital stay (3 days - up to 30 days)Time in hyperglycemia (minutes) during the day and night\> 240 mg/dl will be recorded. Only measured by CGM, Discrete POC values are not used to assess amount of time spent within a specific glucose range. .
Number of Hyperglycemic Events > 240 mg/dl by POC Reading Compared to CGM ReadingDuring hospital stay (3 days - up to 30 days)Number of hyperglycemic events \> 240 mg/dl will be recorded
Percentage of BG Readings Within Target BG of 70 and 180 mg/dl by POC Reading Compared to CGM ReadingDuring hospital stay (3 days - up to 30 days)Percentage of BG readings within target BG of 70 and 180 mg/dl will be calculated
Percentage of Patients With Greater Than 5%-Time Below Glucose Target <70mg/dl by CGM ReadingDuring hospital stay (3 days - up to 30 days)Percentage of patients with greater than 5%-time below glucose target \<70mg/dl will be calculated. Only measured by CGM, Discrete POC values are not used to assess amount of time spent within a specific glucose range.
Percentage of Patients With Greater Than 5%-Time Below Glucose Target <70 mg/dl by CGM Reading10 days post dischargePercentage of patients with greater than 5%-time below glucose target \<70 mg/dl will be calculated
Percentage of Patients With Greater Than 5%-Time Above Glucose Target >180mg/dl by CGM ReadingDuring hospital stay (3 days - up to 30 days)Percentage of patients with greater than 5%-time above glucose target \>180 mg/dl will be calculated
Percentage of Patients With Greater Than 5%-Time Above Glucose Target >180 mg/dl by CGM Reading10 days post dischargePercentage of patients with greater than 5%-time above glucose target \>180 will be calculated
Glycemic Variability Calculated by Standard Deviation by CGM ReadingDuring hospital stay (3 days - up to 30 days)Glycemic variability will be calculated using Standard Deviation
Glycemic Variability Calculated by MAGE by CGM Reading10 days post dischargeMAGE: average amplitude of glycemic excursions with values exceeding a certain threshold.
Number of Sensor Removal/Failures/DislodgmentsDuring hospital stay (3 days - up to 30 days)Number of sensor removal/failures/dislodgments will be recorded
Number of of Hypoglycemic Events < 54 in POC Group by POC Reading Compared to CGM ReadingDuring hospital stay (3 days - up to 30 days)Number of of hypoglycemic events \< 54 will be recorded
Number of Hypoglycemic Events < 54mg/dl by POC Reading Compared to CGM Reading10 days post dischargeNumber of of hypoglycemic events \< 54mg/dl will be recorded
Mean Absolute Relative Difference (MARD) of CGM Compared to POCDuring hospital stay (3 days - up to 30 days)Accuracy of Dexcom G6 CGM, as defined by overall MARD, compared to POC
Percentage of Dexcom G6 CGM Sensor Values Within 30/30% for Values < 70 mg/d Compared to the Standard-of-care POC BGDuring hospital stay (3 days - up to 30 days)Percentage of CGM glucose values in the hypoglycemic range (\<70 mg/dl) meeting %30/30 accuracy criteria.
Percentage of Values Within Zone A, B, C, D of the Clarke Error Grid Measured in AbdomenDuring hospital stay (3 days - up to 30 days)Percentage of CGM glucose values within defined performance zones (A-E) indicating potential for clinical error. The Clarke error grid analysis uses a Cartesian diagram, in which the estimated/predicted values are displayed on the y-axis, whereas the reference values are presented on the x-axis. This diagram is subdivided into 5 zones: A, B, C, D, and E. The points that fall within zones A and B represent, respectively, sufficiently accurate and acceptable glucose results (better outcome), points in zone C may prompt unnecessary corrections, points in zones D and E represent erroneous and incorrect treatment.
Percentage of Values Within Zone A, B, C of the Clarke Error Grid Measured in ArmDuring hospital stay (3 days - up to 30 days) measured in armPercentage of CGM glucose values within defined performance zones (A-E) indicating potential for clinical error. The Clarke error grid analysis uses a Cartesian diagram, in which the estimated/predicted values are displayed on the y-axis, whereas the reference values are presented on the x-axis. This diagram is subdivided into 5 zones: A, B, C, D, and E. The points that fall within zones A and B represent, respectively, sufficiently accurate and acceptable glucose results (better outcome), points in zone C may prompt unnecessary corrections, points in zones D and E represent erroneous and incorrect treatment.
Number of Nocturnal Hypoglycemic Events < 54 mg/dl (Between 22:00 and 06:00) by POC Reading Compared to CGM ReadingDuring hospital stay (3 days - up to 30 days)Number of nocturnal hypoglycemic events \< 54 mg/dl (between 22:00 and 06:00) will be recorded
Percentage of Patients With Greater Than 5%-Time Below Glucose Target <54mg/dl by CGM ReadingDuring hospital stay (3 days - up to 30 days)Percentage of patients with greater than 5%-time below glucose target \<54mg/dl will be calculated. Only measured by CGM, Discrete POC values are not used to assess amount of time spent within a specific glucose range.
Percentage of Patients With Greater Than 5%-Time Above Glucose Target >240 mg/dl by CGM Reading10 days post dischargePercentage of patients with greater than 5%-time above glucose target \>240 will be calculated
Percentage of Patients With Greater Than 5%-Time Below Glucose Target <54 mg/dl by CGM Reading10 days post dischargePercentage of patients with greater than 5%-time below glucose target \<54 mg/dl will be calculated
Percentage of Dexcom G6 CGM Sensor Values Within 30/30% for Values 70-140 mg/dl Compared to the Standard-of-care POC BGDuring hospital stay (3 days - up to 30 days)Percentage of CGM glucose values within the 70-140 mg/dl range that meet %30/30 accuracy criteria.
Percentage of Dexcom G6 CGM Sensor Values Within 30/30% for Values 140-180 mg/dl , Compared to the Standard-of-care POC BGDuring hospital stay (3 days - up to 30 days)Percentage of CGM glucose values within the 140-180 mg/dl range that meet %30/30 accuracy criteria.
Percentage of Dexcom G6 CGM Sensor Values Within 30/30%for Values > 180 mg/dl, Compared to the Standard-of-care POC BGDuring hospital stay (3 days - up to 30 days)Percentage of CGM glucose values in the hyperglycemic range (\>180 mg/dl) that meet %30/30 accuracy criteria.
Percentage of Dexcom G6 CGM Sensor Values Within 20/20% for Values < 70 mg/dl Compared to the Standard-of-care POC BGDuring hospital stay (3 days - up to 30 days)Percentage of CGM glucose values in the hypoglycemic range (\<70 mg/dl) that meet %20/20 accuracy criteria.
Percentage of Dexcom G6 CGM Sensor Values Within 20/20% for Values 70-140 mg/dl Compared to the Standard-of-care POC BGDuring hospital stay (3 days - up to 30 days)Percentage of CGM glucose values within the 70-140 mg/dl range that meet %20/20 accuracy criteria.
Percentage of Dexcom G6 CGM Sensor Values Within 20/20%for Values > 180 mg/dl, Compared to the Standard-of-care POC BGDuring hospital stay (3 days - up to 30 days)Percentage of CGM glucose values in the hyperglycemic range (\>180 mg/dl) that meet %20/20 accuracy criteria..
Percentage of Dexcom G6 CGM Sensor Values Within 20/20% for Values 140-180 mg/dl , Compared to the Standard-of-care POC BGDuring hospital stay (3 days - up to 30 days)Percentage of CGM glucose values within the 140-180 mg/dl range that meet %20/20 accuracy criteria.
Percentage of Dexcom G6 CGM Sensor Values Within 15/15% for Values < 70 mg/dl Compared to the Standard-of-care POC BGDuring hospital stay (3 days - up to 30 days)Percentage of CGM glucose values in the hypoglycemic range (\<70 mg/dl) that meet %15/15 accuracy criteria..
Mean Daily Blood Glucose by POC Reading Compared to CGM Reading10 days post dischargeMean daily blood glucose will be calculated
Percentage of Dexcom G6 CGM Sensor Values Within 15/15% for Values 140-180 mg/dl , Compared to the Standard-of-care POC BGDuring hospital stay (3 days - up to 30 days)Percentage of CGM glucose values within the 140-180 mg/dl range that meet %15/15 accuracy criteria.
Percentage of Dexcom G6 CGM Sensor Values Within 15/15%for Values > 180 mg/dl, Compared to the Standard-of-care POC BGDuring hospital stay (3 days - up to 30 days)Percentage of CGM glucose values in the hyperglycemic range (\>180 mg/dl) that meet %15/15 accuracy criteria.
Percentage of Dexcom G6 CGM Sensor Values Within 15/15% for Values 70-140 mg/dl Compared to the Standard-of-care POC BGDuring hospital stay (3 days - up to 30 days)Percentage of CGM glucose values within the 70-140 mg/dl range that meet %15/15 accuracy criteria.
Number of Clinical Significant Hypoglycemia (<54 mg/dl) Events by POC Reading Compared to CGM Reading10 days post dischargeNumber of clinical significant hypoglycemia (\<54 mg/dl) events will be calculated
Number of Clinical Significant Severe Hyperglycemia (>240 mg/dl) Events by POC Reading Compared to CGM Reading10 days post dischargeNumber of severe hyperglycemia (\>240 mg/dl) events will be recorded

Countries

United States

Participant flow

Participants by arm

ArmCount
Dexcom G6 CGM - Continues Glucose Monitoring Sensor System
Dexcom G6 CGM - Continues Glucose Monitoring sensor system: A blinded factory-calibrated continues glucose monitoring sensor system Dexcom G6 will be placed shortly after admission. Two CGM devices will be inserted in all patients - one in the abdomen and one in the arm to also assess differences in blood glucose readings between upper extremity and abdominal insertion sites. Information on CGM readings will be collected daily during the hospital stay and after hospital discharge for 10 days using the Dexcom Studio software to download the Dexcom receiver data.
101
Total101

Baseline characteristics

CharacteristicDexcom G6 CGM - Continues Glucose Monitoring Sensor System
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
26 Participants
Age, Categorical
Between 18 and 65 years
75 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
Race (NIH/OMB)
Asian
1 Participants
Race (NIH/OMB)
Black or African American
85 Participants
Race (NIH/OMB)
More than one race
0 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
Race (NIH/OMB)
Unknown or Not Reported
2 Participants
Race (NIH/OMB)
White
13 Participants
Region of Enrollment
United States
101 participants
Sex: Female, Male
Female
47 Participants
Sex: Female, Male
Male
54 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 101
other
Total, other adverse events
4 / 101
serious
Total, serious adverse events
0 / 101

Outcome results

Primary

Mean Daily Blood Glucose by POC Reading сompared to CGM Reading

Mean daily blood glucose will be calculated

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (MEAN)Dispersion
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemMean Daily Blood Glucose by POC Reading сompared to CGM ReadingPOC175.55 mg/dLStandard Deviation 42.37
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemMean Daily Blood Glucose by POC Reading сompared to CGM ReadingCGM175.84 mg/dLStandard Deviation 42.35
Primary

Number of Clinical Significant Hypoglycemia (<54 mg/dl) Events by POC Reading Compared to CGM Reading

Number of clinical significant hypoglycemia (\<54 mg/dl) events will be calculated

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Clinical Significant Hypoglycemia (<54 mg/dl) Events by POC Reading Compared to CGM ReadingPOC9 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Clinical Significant Hypoglycemia (<54 mg/dl) Events by POC Reading Compared to CGM ReadingCGM60 number of events
Primary

Number of Severe Hyperglycemia (>240 mg/dl) Events by POC Reading Compared to CGM Reading

Number of severe hyperglycemia (\>240 mg/dl) events will be recorded

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Severe Hyperglycemia (>240 mg/dl) Events by POC Reading Compared to CGM ReadingPOC305 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Severe Hyperglycemia (>240 mg/dl) Events by POC Reading Compared to CGM ReadingCGM518 number of events
Secondary

Glycemic Variability Calculated by MAGE by CGM Reading

The average amplitude of glycemic excursions with values exceeding a certain threshold.

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (MEAN)Dispersion
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemGlycemic Variability Calculated by MAGE by CGM ReadingCGM abdomen58.73 mg/dlStandard Deviation 27.87
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemGlycemic Variability Calculated by MAGE by CGM ReadingCGM arm56.68 mg/dlStandard Deviation 30.87
Secondary

Glycemic Variability Calculated by MAGE by CGM Reading

MAGE: average amplitude of glycemic excursions with values exceeding a certain threshold.

Time frame: 10 days post discharge

ArmMeasureGroupValue (MEAN)Dispersion
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemGlycemic Variability Calculated by MAGE by CGM ReadingCGM abdomen58.33 mg/dlStandard Deviation 24.43
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemGlycemic Variability Calculated by MAGE by CGM ReadingCGM arm55 mg/dlStandard Deviation 23.76
Secondary

Glycemic Variability Calculated by Standard Deviation by CGM Reading

Glycemic variability will be calculated using Standard Deviation

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (MEAN)Dispersion
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemGlycemic Variability Calculated by Standard Deviation by CGM ReadingCGM abdomen46.37 mg/dlStandard Deviation 15.45
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemGlycemic Variability Calculated by Standard Deviation by CGM ReadingCGM arm43.37 mg/dlStandard Deviation 15.31
Secondary

Glycemic Variability Calculated by Standard Deviation by CGM Reading

Glycemic variability will be calculated using Standard Deviation

Time frame: 10 days post discharge

ArmMeasureGroupValue (MEAN)Dispersion
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemGlycemic Variability Calculated by Standard Deviation by CGM ReadingCGM abdomen46.37 mg/dlStandard Deviation 18.17
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemGlycemic Variability Calculated by Standard Deviation by CGM ReadingCGM arm43.87 mg/dlStandard Deviation 20.01
Secondary

Mean Absolute Relative Difference (MARD) of CGM Compared to POC

Accuracy of Dexcom G6 CGM, as defined by overall MARD, compared to POC

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (MEAN)Dispersion
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemMean Absolute Relative Difference (MARD) of CGM Compared to POCCGM abdomen12.54 percent differenceStandard Deviation 15.24
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemMean Absolute Relative Difference (MARD) of CGM Compared to POCCGM arm12.22 percent differenceStandard Deviation 15.11
Secondary

Mean Daily Blood Glucose by POC Reading Compared to CGM Reading

Mean daily blood glucose will be calculated

Time frame: 10 days post discharge

ArmMeasureGroupValue (MEAN)Dispersion
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemMean Daily Blood Glucose by POC Reading Compared to CGM ReadingPOC175.55 mg/dLStandard Deviation 42.37
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemMean Daily Blood Glucose by POC Reading Compared to CGM ReadingCGM173.13 mg/dLStandard Deviation 41.77
Secondary

Number of Clinical Significant Hypoglycemia (<54 mg/dl) Events by POC Reading Compared to CGM Reading

Number of clinical significant hypoglycemia (\<54 mg/dl) events will be calculated

Time frame: 10 days post discharge

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Clinical Significant Hypoglycemia (<54 mg/dl) Events by POC Reading Compared to CGM ReadingPOC9 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Clinical Significant Hypoglycemia (<54 mg/dl) Events by POC Reading Compared to CGM ReadingCGM (arm)29 number of events
Secondary

Number of Clinical Significant Severe Hyperglycemia (>240 mg/dl) Events by POC Reading Compared to CGM Reading

Number of severe hyperglycemia (\>240 mg/dl) events will be recorded

Time frame: 10 days post discharge

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Clinical Significant Severe Hyperglycemia (>240 mg/dl) Events by POC Reading Compared to CGM ReadingPOC305 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Clinical Significant Severe Hyperglycemia (>240 mg/dl) Events by POC Reading Compared to CGM ReadingCGM399 number of events
Secondary

Number of Hyperglycemic Events > 240 mg/dl by POC Reading Compared to CGM Reading

Number of hyperglycemic events \> 240 mg/dl will be recorded

Time frame: 10 days post discharge

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Hyperglycemic Events > 240 mg/dl by POC Reading Compared to CGM ReadingPOC153 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Hyperglycemic Events > 240 mg/dl by POC Reading Compared to CGM ReadingCGM abdomen554 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Hyperglycemic Events > 240 mg/dl by POC Reading Compared to CGM ReadingCGM arm316 number of events
Secondary

Number of Hyperglycemic Events > 240 mg/dl by POC Reading Compared to CGM Reading

Number of hyperglycemic events \> 240 mg/dl will be recorded

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Hyperglycemic Events > 240 mg/dl by POC Reading Compared to CGM ReadingPOC305 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Hyperglycemic Events > 240 mg/dl by POC Reading Compared to CGM ReadingCGM abdomen518 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Hyperglycemic Events > 240 mg/dl by POC Reading Compared to CGM ReadingCGM arm399 number of events
Secondary

Number of Hypoglycemic Events < 54mg/dl by POC Reading Compared to CGM Reading

Number of of hypoglycemic events \< 54mg/dl will be recorded

Time frame: 10 days post discharge

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Hypoglycemic Events < 54mg/dl by POC Reading Compared to CGM ReadingPOC3 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Hypoglycemic Events < 54mg/dl by POC Reading Compared to CGM ReadingCGM abdomen48 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Hypoglycemic Events < 54mg/dl by POC Reading Compared to CGM ReadingCGM arm16 number of events
Secondary

Number of Hypoglycemic Events < 70mg/dl by POC Reading Compared to CGM Reading

Number of of hypoglycemic events \< 70mg/dl will be recorded

Time frame: 10 days post discharge

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Hypoglycemic Events < 70mg/dl by POC Reading Compared to CGM ReadingPOC10 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Hypoglycemic Events < 70mg/dl by POC Reading Compared to CGM ReadingCGM abdomen100 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Hypoglycemic Events < 70mg/dl by POC Reading Compared to CGM ReadingCGM arm48 number of events
Secondary

Number of Nocturnal Hypoglycemic Events < 54 mg/dl (Between 22:00 and 06:00) by POC Reading Compared to CGM Reading

Number of nocturnal hypoglycemic events \< 54 mg/dl (between 22:00 and 06:00) will be recorded

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Nocturnal Hypoglycemic Events < 54 mg/dl (Between 22:00 and 06:00) by POC Reading Compared to CGM ReadingPOC1 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Nocturnal Hypoglycemic Events < 54 mg/dl (Between 22:00 and 06:00) by POC Reading Compared to CGM ReadingCGM abdomen12 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Nocturnal Hypoglycemic Events < 54 mg/dl (Between 22:00 and 06:00) by POC Reading Compared to CGM ReadingCGM arm4 number of events
Secondary

Number of Nocturnal Hypoglycemic Events < 54 mg/dl (Between 22:00 and 06:00) by POC Reading Compared to CGM Reading

Number of nocturnal hypoglycemic events \< 54 mg/dl(between 22:00 and 06:00) will be recorded

Time frame: 10 days post discharge

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Nocturnal Hypoglycemic Events < 54 mg/dl (Between 22:00 and 06:00) by POC Reading Compared to CGM ReadingPOC0 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Nocturnal Hypoglycemic Events < 54 mg/dl (Between 22:00 and 06:00) by POC Reading Compared to CGM ReadingCGM abdomen8 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Nocturnal Hypoglycemic Events < 54 mg/dl (Between 22:00 and 06:00) by POC Reading Compared to CGM ReadingCGM arm3 number of events
Secondary

Number of Nocturnal Hypoglycemic Events < 70 mg/dl (Between 22:00 and 06:00) by POC Reading Compared to CGM Reading

Number of nocturnal hypoglycemic events \< 70 mg/dl (between 22:00 and 06:00) will be recorded

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Nocturnal Hypoglycemic Events < 70 mg/dl (Between 22:00 and 06:00) by POC Reading Compared to CGM ReadingPOC1 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Nocturnal Hypoglycemic Events < 70 mg/dl (Between 22:00 and 06:00) by POC Reading Compared to CGM ReadingCGM abdomen48 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Nocturnal Hypoglycemic Events < 70 mg/dl (Between 22:00 and 06:00) by POC Reading Compared to CGM ReadingCGM arm34 number of events
Secondary

Number of Nocturnal Hypoglycemic Events < 70 mg/dl (Between 22:00 and 06:00) by POC Reading Compared to CGM Reading

Number of nocturnal hypoglycemic events \< 70 mg/dl(between 22:00 and 06:00) will be recorded

Time frame: 10 days post discharge

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Nocturnal Hypoglycemic Events < 70 mg/dl (Between 22:00 and 06:00) by POC Reading Compared to CGM ReadingPOC1 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Nocturnal Hypoglycemic Events < 70 mg/dl (Between 22:00 and 06:00) by POC Reading Compared to CGM ReadingCGM abdomen22 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Nocturnal Hypoglycemic Events < 70 mg/dl (Between 22:00 and 06:00) by POC Reading Compared to CGM ReadingCGM arm9 number of events
Secondary

Number of of Hypoglycemic Events < 54 in POC Group by POC Reading Compared to CGM Reading

Number of of hypoglycemic events \< 54 will be recorded

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of of Hypoglycemic Events < 54 in POC Group by POC Reading Compared to CGM ReadingPOC9 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of of Hypoglycemic Events < 54 in POC Group by POC Reading Compared to CGM ReadingCGM abdomen60 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of of Hypoglycemic Events < 54 in POC Group by POC Reading Compared to CGM ReadingCGM arm29 number of events
Secondary

Number of of Hypoglycemic Events < 70mg/dl in POC Group by POC Reading Compared to CGM Reading

Number of of hypoglycemic events \< 70mg/dl will be recorded

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of of Hypoglycemic Events < 70mg/dl in POC Group by POC Reading Compared to CGM ReadingPOC36 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of of Hypoglycemic Events < 70mg/dl in POC Group by POC Reading Compared to CGM ReadingCGM abdomen144 number of events
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of of Hypoglycemic Events < 70mg/dl in POC Group by POC Reading Compared to CGM ReadingCGM arm115 number of events
Secondary

Number of Sensor Removal/Failures/Dislodgments

Number of sensor removal/failures/dislodgments will be recorded

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemNumber of Sensor Removal/Failures/Dislodgments31 number of events
Secondary

Percentage of BG Readings Within Target BG of 70 and 180 mg/dl by POC Reading Compared to CGM Reading

Percentage of BG readings within target BG of 70 and 180 mg/dl will be calculated

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (MEAN)Dispersion
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of BG Readings Within Target BG of 70 and 180 mg/dl by POC Reading Compared to CGM ReadingPOC59.70 percentage of readingsStandard Deviation 24.73
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of BG Readings Within Target BG of 70 and 180 mg/dl by POC Reading Compared to CGM ReadingCGM (abdomen)58.91 percentage of readingsStandard Deviation 26.83
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of BG Readings Within Target BG of 70 and 180 mg/dl by POC Reading Compared to CGM ReadingCGM arm60.97 percentage of readingsStandard Deviation 28.56
Secondary

Percentage of BG Readings Within Target BG of 70 and 180 mg/dl by POC Reading Compared to CGM Reading

Percentage of BG readings within target BG of 70 and 180 mg/dl will be calculated

Time frame: 10 days post discharge

ArmMeasureGroupValue (MEAN)Dispersion
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of BG Readings Within Target BG of 70 and 180 mg/dl by POC Reading Compared to CGM ReadingPOC59.56 percentage of readingsStandard Deviation 29.31
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of BG Readings Within Target BG of 70 and 180 mg/dl by POC Reading Compared to CGM ReadingCGM abdomen52.62 percentage of readingsStandard Deviation 30.64
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of BG Readings Within Target BG of 70 and 180 mg/dl by POC Reading Compared to CGM ReadingCGM arm51.60 percentage of readingsStandard Deviation 31.94
Secondary

Percentage of Dexcom G6 CGM Sensor Values Within 15/15% for Values 140-180 mg/dl , Compared to the Standard-of-care POC BG

Percentage of CGM glucose values within the 140-180 mg/dl range that meet %15/15 accuracy criteria.

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 15/15% for Values 140-180 mg/dl , Compared to the Standard-of-care POC BGCGM abdomen69.6 percentage of values
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 15/15% for Values 140-180 mg/dl , Compared to the Standard-of-care POC BGCGM arm73.8 percentage of values
Secondary

Percentage of Dexcom G6 CGM Sensor Values Within 15/15%for Values > 180 mg/dl, Compared to the Standard-of-care POC BG

Percentage of CGM glucose values in the hyperglycemic range (\>180 mg/dl) that meet %15/15 accuracy criteria.

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 15/15%for Values > 180 mg/dl, Compared to the Standard-of-care POC BGCGM abdomen71.0 percentage of values
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 15/15%for Values > 180 mg/dl, Compared to the Standard-of-care POC BGCGM arm71.5 percentage of values
Secondary

Percentage of Dexcom G6 CGM Sensor Values Within 15/15% for Values 70-140 mg/dl Compared to the Standard-of-care POC BG

Percentage of CGM glucose values within the 70-140 mg/dl range that meet %15/15 accuracy criteria.

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 15/15% for Values 70-140 mg/dl Compared to the Standard-of-care POC BGCGM arm70.5 percentage of values
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 15/15% for Values 70-140 mg/dl Compared to the Standard-of-care POC BGCGM abdomen69.4 percentage of values
Secondary

Percentage of Dexcom G6 CGM Sensor Values Within 15/15% for Values < 70 mg/dl Compared to the Standard-of-care POC BG

Percentage of CGM glucose values in the hypoglycemic range (\<70 mg/dl) that meet %15/15 accuracy criteria..

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 15/15% for Values < 70 mg/dl Compared to the Standard-of-care POC BGCGM abdomen75.0 percentage of values
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 15/15% for Values < 70 mg/dl Compared to the Standard-of-care POC BGCGM arm66.7 percentage of values
Secondary

Percentage of Dexcom G6 CGM Sensor Values Within 20/20% for Values 140-180 mg/dl , Compared to the Standard-of-care POC BG

Percentage of CGM glucose values within the 140-180 mg/dl range that meet %20/20 accuracy criteria.

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 20/20% for Values 140-180 mg/dl , Compared to the Standard-of-care POC BGCGM abdomen83.2 percentage of values
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 20/20% for Values 140-180 mg/dl , Compared to the Standard-of-care POC BGCGM arm83.6 percentage of values
Secondary

Percentage of Dexcom G6 CGM Sensor Values Within 20/20%for Values > 180 mg/dl, Compared to the Standard-of-care POC BG

Percentage of CGM glucose values in the hyperglycemic range (\>180 mg/dl) that meet %20/20 accuracy criteria..

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 20/20%for Values > 180 mg/dl, Compared to the Standard-of-care POC BGCGM abdomen84.7 percentage of values
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 20/20%for Values > 180 mg/dl, Compared to the Standard-of-care POC BGCGM arm81.1 percentage of values
Secondary

Percentage of Dexcom G6 CGM Sensor Values Within 20/20% for Values 70-140 mg/dl Compared to the Standard-of-care POC BG

Percentage of CGM glucose values within the 70-140 mg/dl range that meet %20/20 accuracy criteria.

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 20/20% for Values 70-140 mg/dl Compared to the Standard-of-care POC BGCGM abdomen82.0 percentage of values
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 20/20% for Values 70-140 mg/dl Compared to the Standard-of-care POC BGCGM arm82.0 percentage of values
Secondary

Percentage of Dexcom G6 CGM Sensor Values Within 20/20% for Values < 70 mg/dl Compared to the Standard-of-care POC BG

Percentage of CGM glucose values in the hypoglycemic range (\<70 mg/dl) that meet %20/20 accuracy criteria.

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 20/20% for Values < 70 mg/dl Compared to the Standard-of-care POC BGCGM abdomen78.6 percentage of values
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 20/20% for Values < 70 mg/dl Compared to the Standard-of-care POC BGCGM arm74.1 percentage of values
Secondary

Percentage of Dexcom G6 CGM Sensor Values Within 30/30% for Values 140-180 mg/dl , Compared to the Standard-of-care POC BG

Percentage of CGM glucose values within the 140-180 mg/dl range that meet %30/30 accuracy criteria.

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 30/30% for Values 140-180 mg/dl , Compared to the Standard-of-care POC BGCGM abdomen93.3 percentage of values
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 30/30% for Values 140-180 mg/dl , Compared to the Standard-of-care POC BGCGM arm94.9 percentage of values
Secondary

Percentage of Dexcom G6 CGM Sensor Values Within 30/30%for Values > 180 mg/dl, Compared to the Standard-of-care POC BG

Percentage of CGM glucose values in the hyperglycemic range (\>180 mg/dl) that meet %30/30 accuracy criteria.

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 30/30%for Values > 180 mg/dl, Compared to the Standard-of-care POC BGCGM abdomen95.5 percentage of values
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 30/30%for Values > 180 mg/dl, Compared to the Standard-of-care POC BGCGM arm94 percentage of values
Secondary

Percentage of Dexcom G6 CGM Sensor Values Within 30/30% for Values 70-140 mg/dl Compared to the Standard-of-care POC BG

Percentage of CGM glucose values within the 70-140 mg/dl range that meet %30/30 accuracy criteria.

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 30/30% for Values 70-140 mg/dl Compared to the Standard-of-care POC BGCGM abdomen93.1 percentage of values
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 30/30% for Values 70-140 mg/dl Compared to the Standard-of-care POC BGCGM arm92.9 percentage of values
Secondary

Percentage of Dexcom G6 CGM Sensor Values Within 30/30% for Values < 70 mg/d Compared to the Standard-of-care POC BG

Percentage of CGM glucose values in the hypoglycemic range (\<70 mg/dl) meeting %30/30 accuracy criteria.

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 30/30% for Values < 70 mg/d Compared to the Standard-of-care POC BGCGM abdomen89.3 percentage of values
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Dexcom G6 CGM Sensor Values Within 30/30% for Values < 70 mg/d Compared to the Standard-of-care POC BGCGM arm85.2 percentage of values
Secondary

Percentage of Patients With Greater Than 5%-Time Above Glucose Target >180 mg/dl by CGM Reading

Percentage of patients with greater than 5%-time above glucose target \>180 will be calculated

Time frame: 10 days post discharge

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Patients With Greater Than 5%-Time Above Glucose Target >180 mg/dl by CGM ReadingCGM abdomen89 percentage of patients
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Patients With Greater Than 5%-Time Above Glucose Target >180 mg/dl by CGM ReadingCGM arm94 percentage of patients
Secondary

Percentage of Patients With Greater Than 5%-Time Above Glucose Target >180mg/dl by CGM Reading

Percentage of patients with greater than 5%-time above glucose target \>180 mg/dl will be calculated

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Patients With Greater Than 5%-Time Above Glucose Target >180mg/dl by CGM ReadingCGM abdomen91 percentage of patients
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Patients With Greater Than 5%-Time Above Glucose Target >180mg/dl by CGM ReadingCGM arm89 percentage of patients
Secondary

Percentage of Patients With Greater Than 5%-Time Above Glucose Target >240 mg/dl by CGM Reading

Percentage of patients with greater than 5%-time above glucose target \>240 will be calculated

Time frame: 10 days post discharge

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Patients With Greater Than 5%-Time Above Glucose Target >240 mg/dl by CGM ReadingCGM abdomen56 percentage of patients
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Patients With Greater Than 5%-Time Above Glucose Target >240 mg/dl by CGM ReadingCGM arm60 percentage of patients
Secondary

Percentage of Patients With Greater Than 5%-Time Above Glucose Target >240 mg/dl by CGM Reading

Percentage of patients with greater than 5%-time above glucose target \>240 mg/dl will be calculated

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Patients With Greater Than 5%-Time Above Glucose Target >240 mg/dl by CGM ReadingCGM abdomen53 percentage of patients
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Patients With Greater Than 5%-Time Above Glucose Target >240 mg/dl by CGM ReadingCGM arm43 percentage of patients
Secondary

Percentage of Patients With Greater Than 5%-Time Below Glucose Target <54 mg/dl by CGM Reading

Percentage of patients with greater than 5%-time below glucose target \<54 mg/dl will be calculated

Time frame: 10 days post discharge

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Patients With Greater Than 5%-Time Below Glucose Target <54 mg/dl by CGM ReadingCGM abdomen1.8 percentage of patients
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Patients With Greater Than 5%-Time Below Glucose Target <54 mg/dl by CGM ReadingCGM arm2.9 percentage of patients
Secondary

Percentage of Patients With Greater Than 5%-Time Below Glucose Target <54mg/dl by CGM Reading

Percentage of patients with greater than 5%-time below glucose target \<54mg/dl will be calculated. Only measured by CGM, Discrete POC values are not used to assess amount of time spent within a specific glucose range.

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Patients With Greater Than 5%-Time Below Glucose Target <54mg/dl by CGM ReadingCGM <70 abdomen3.4 percentage of patients
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Patients With Greater Than 5%-Time Below Glucose Target <54mg/dl by CGM ReadingCGM <70 arm0 percentage of patients
Secondary

Percentage of Patients With Greater Than 5%-Time Below Glucose Target <70 mg/dl by CGM Reading

Percentage of patients with greater than 5%-time below glucose target \<70 mg/dl will be calculated

Time frame: 10 days post discharge

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Patients With Greater Than 5%-Time Below Glucose Target <70 mg/dl by CGM ReadingCGM abdomen5.3 percentage of patients
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Patients With Greater Than 5%-Time Below Glucose Target <70 mg/dl by CGM ReadingCGM arm2.9 percentage of patients
Secondary

Percentage of Patients With Greater Than 5%-Time Below Glucose Target <70mg/dl by CGM Reading

Percentage of patients with greater than 5%-time below glucose target \<70mg/dl will be calculated. Only measured by CGM, Discrete POC values are not used to assess amount of time spent within a specific glucose range.

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Patients With Greater Than 5%-Time Below Glucose Target <70mg/dl by CGM ReadingCGM <70 abdomen10 percentage of readings
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Patients With Greater Than 5%-Time Below Glucose Target <70mg/dl by CGM ReadingCGM <70 arm4.6 percentage of readings
Secondary

Percentage of Values Within Zone A, B, C, D of the Clarke Error Grid Measured in Abdomen

Percentage of CGM glucose values within defined performance zones (A-E) indicating potential for clinical error. The Clarke error grid analysis uses a Cartesian diagram, in which the estimated/predicted values are displayed on the y-axis, whereas the reference values are presented on the x-axis. This diagram is subdivided into 5 zones: A, B, C, D, and E. The points that fall within zones A and B represent, respectively, sufficiently accurate and acceptable glucose results (better outcome), points in zone C may prompt unnecessary corrections, points in zones D and E represent erroneous and incorrect treatment.

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Values Within Zone A, B, C, D of the Clarke Error Grid Measured in AbdomenA82.4 percentage of values
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Values Within Zone A, B, C, D of the Clarke Error Grid Measured in AbdomenB16.5 percentage of values
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Values Within Zone A, B, C, D of the Clarke Error Grid Measured in AbdomenC0.1 percentage of values
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Values Within Zone A, B, C, D of the Clarke Error Grid Measured in AbdomenD1 percentage of values
Secondary

Percentage of Values Within Zone A, B, C of the Clarke Error Grid Measured in Arm

Percentage of CGM glucose values within defined performance zones (A-E) indicating potential for clinical error. The Clarke error grid analysis uses a Cartesian diagram, in which the estimated/predicted values are displayed on the y-axis, whereas the reference values are presented on the x-axis. This diagram is subdivided into 5 zones: A, B, C, D, and E. The points that fall within zones A and B represent, respectively, sufficiently accurate and acceptable glucose results (better outcome), points in zone C may prompt unnecessary corrections, points in zones D and E represent erroneous and incorrect treatment.

Time frame: During hospital stay (3 days - up to 30 days) measured in arm

ArmMeasureGroupValue (NUMBER)
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Values Within Zone A, B, C of the Clarke Error Grid Measured in ArmA81.4 percentage of values
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Values Within Zone A, B, C of the Clarke Error Grid Measured in ArmB17.6 percentage of values
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Values Within Zone A, B, C of the Clarke Error Grid Measured in ArmC0.1 percentage of values
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemPercentage of Values Within Zone A, B, C of the Clarke Error Grid Measured in ArmD0.9 percentage of values
Secondary

Time in Hyperglycemia (Minutes) During the Day and Night> 240 mg/dl by CGM Reading

Time in hyperglycemia (minutes) during the day and night\> 240 mg/dl will be recorded. Only measured by CGM, Discrete POC values are not used to assess amount of time spent within a specific glucose range.

Time frame: 10 days post discharge

ArmMeasureGroupValue (MEAN)Dispersion
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemTime in Hyperglycemia (Minutes) During the Day and Night> 240 mg/dl by CGM ReadingCGM abdomen4624.3 minStandard Deviation 13554
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemTime in Hyperglycemia (Minutes) During the Day and Night> 240 mg/dl by CGM ReadingCGM arm1706.1 minStandard Deviation 2255.6
Secondary

Time in Hyperglycemia (Minutes) During the Day and Night> 240 mg/dl by CGM Reading

Time in hyperglycemia (minutes) during the day and night\> 240 mg/dl will be recorded. Only measured by CGM, Discrete POC values are not used to assess amount of time spent within a specific glucose range. .

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (MEAN)Dispersion
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemTime in Hyperglycemia (Minutes) During the Day and Night> 240 mg/dl by CGM ReadingCGM abdomen972.30 minStandard Deviation 1214.5
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemTime in Hyperglycemia (Minutes) During the Day and Night> 240 mg/dl by CGM ReadingCGM arm723.23 minStandard Deviation 1061.4
Secondary

Time in Hypoglycemia (Minutes) During the Day and Night by CGM Reading

Time in hypoglycemia (minutes) during the day and night will be recorded. Only measured by CGM, Discrete POC values are not used to assess amount of time spent within a specific glucose range. .

Time frame: 10 days post discharge

ArmMeasureGroupValue (MEAN)Dispersion
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemTime in Hypoglycemia (Minutes) During the Day and Night by CGM ReadingCGM abdomen118.70 minStandard Deviation 217.83
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemTime in Hypoglycemia (Minutes) During the Day and Night by CGM ReadingCGM arm89.14 minStandard Deviation 140.16
Secondary

Time in Hypoglycemia (Minutes) During the Day and Night by CGM Reading

Time in hypoglycemia (minutes) during the day and night will be recorded. Only measured by CGM, Discrete POC values are not used to assess amount of time spent within a specific glucose range. .

Time frame: During hospital stay (3 days - up to 30 days)

ArmMeasureGroupValue (MEAN)Dispersion
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemTime in Hypoglycemia (Minutes) During the Day and Night by CGM ReadingCGM abdomen)119.08 minStandard Deviation 216.56
Dexcom G6 CGM - Continues Glucose Monitoring Sensor SystemTime in Hypoglycemia (Minutes) During the Day and Night by CGM ReadingCGM arm78.63 minStandard Deviation 135.33

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026