Skip to content

Clinical Investigation of ENO/TEO/OTO Pacing System Under MRI Environment

Clinical Investigation of ENO/TEO/OTO Pacing System Under MRI Environment

Status
UNKNOWN
Phases
Phase 4
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03811691
Acronym
CAPRI
Enrollment
270
Registered
2019-01-22
Start date
2019-07-18
Completion date
2023-08-31
Last updated
2021-06-09

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Bradycardia

Brief summary

The interest of the CAPRI study is to confirm the safety and the efficacy of the ENO/TEO/OTO pacing system when used under 1.5 or 3 Tesla MRI environment and in accordance with the MRI solutions guideline.

Detailed description

The purpose of CAPRI study is to confirm the safety and the efficacy of the CE marked MR conditional pacing system composed of the ENO, TEO or OTO pacemaker with VEGA pacing lead(s). Conditions to undergo an MRI scan are provided in the MRI solutions guideline. The primary objective is to confirm the clinical safety of the ENO/TEO/OTO pacing system when used under 1.5 and 3 Tesla specific MRI conditions without scan exclusion zone. The study secondary main objectives are aiming to assess the performance of the MR conditional pacing system in the right atrium and ventricle at 1 month following the MRI scan: * Stability of the Pacing Capture Threshold * Stability of the lead sensed amplitude

Interventions

DIAGNOSTIC_TESTMRI exam

Pacing system safety and stability under MRI scan environment

Sponsors

MicroPort CRM
Lead SponsorINDUSTRY

Study design

Allocation
NON_RANDOMIZED
Intervention model
PARALLEL
Primary purpose
OTHER
Masking
NONE

Intervention model description

This study is an interventional, prospective, multicenter, open label, two arms parallel non-comparative non-randomized study.

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

Subjects who meet all the following criteria at the time of inclusion visit may be included: * Already implanted in the left or right pectoral region for at least 6 weeks with: * ENO, TEO or OTO single chamber rate response (SR) pacemaker with a VEGA pacing lead, or * ENO, TEO or OTO dual chamber rate response (DR) pacemaker with two VEGA pacing leads. * Implanted pacing system must fulfill the following parameters: * Battery impedance is \< 5 Kilo Ohm (kΩ) * Pacing capture threshold value is ≤ 2 Volts (V) at 0.35 millisecond (ms) * Lead impedance value between 200Ω and 3000Ω * No diaphragmatic or pectoral stimulation at 5V/1ms * P-wave minimum sensed amplitude ≥ 1mV for patients with sinus rhythm * R-wave minimum sensed amplitude ≥ 4mV for patients with spontaneous conduction rhythm * Must agree to undergo a non clinically-indicated MRI scan without intravenous injection and without sedation; * Have reviewed, signed and dated informed consent. Subjects who meet any of the following criteria are not eligible to be included in the study: * Included in another clinical study that could confound the results of this study such as studies involving intra-cardiac device; * Have other active or abandoned cardiac implants already implanted; * Have other active or passive non MR conditional devices implanted such as metallic foreign body; * Have a history of brain aneurysm with ferromagnetic clipping; * Have a planned cardiac surgery within the 3 months of inclusion; * Have a medical MRI examination prescription planned within the 3 months of inclusion; * Age less than 18 years old or under guardianship or kept in detention; * Known pregnancy, women breastfeeding or in childbearing age without an adequate contraceptive method ; * Be unavailable for the scheduled follow-up associated with this clinical study or refusal to cooperate.

Design outcomes

Primary

MeasureTime frameDescription
Freedom from MRI-related complications following the MRI scan of the ENO/TEO/OTO pacing system1 month post MRIAn MRI-related complication is defined as a Serious Adverse Event (SAE) that led to death or to an invasive intervention or a Device Deficiency that causes the termination of any significant device function per Clinical Event Committee (CEC) adjudication. The primary endpoint will be assessed independently for each arm of the study (1.5 and 3 Tesla MRI scan).

Secondary

MeasureTime frameDescription
Ventricular Pacing Capture Threshold (PCT) stability in the right atrium and ventricle following the MRI scan1 month post MRISubject's ventricular pacing capture threshold will be measured in Volt (V) at 0.5ms at baseline (at the MRI visit - just before the MRI scan) and at 1 month Post-MRI visit. A success is defined as a PCT increase less than or equal to 0.5V at 0.5ms between the two visits. The secondary endpoints will be assessed independently for each arm of the study (1.5 and 3 Tesla MRI scan).
Atrial Pacing Capture Threshold stability in the right atrium and ventricle following the MRI scan1 month post MRISubject's atrial pacing capture threshold will be measured in Volt (V) at 0.5ms at Baseline (at the MRI visit - just before the MRI scan) and at 1 month Post-MRI visit. A success is defined as a PCT increase less than or equal to 0.5V at 0.5ms between the two visits. The secondary endpoints will be assessed independently for each arm of the study (1.5 and 3 Tesla MRI scan).
Ventricular sensed amplitude stability in the right atrium and ventricle following the MRI scan1 month post MRISubject's ventricular minimum sensed amplitude will be measured in mV at Baseline (at the MRI visit - just before the MRI scan) and at 1 month Post-MRI visit. A success is defined as 50% or less decrease between the two visits. If subject's ventricular minimum sensed amplitude is less or equal to 4mV at time of baseline, then this subject will not count for this endpoint. The secondary endpoints will be assessed independently for each arm of the study (1.5 and 3 Tesla MRI scan).
Atrial sensed amplitude stability in the right atrium and ventricle following the MRI scan1 month post MRISubject's atrial minimum sensed amplitude will be measured in mV at baseline. A success is defined as 50% or less decrease between the two visits. If subject's atrial minimum sensed amplitude is less or equal to 1mV at time of Baseline (at the MRI visit - just before the MRI scan) and at 1 month Post-MRI visit, then this subject will not count for this endpoint.The secondary endpoints will be assessed independently for each arm of the study (1.5 and 3 Tesla MRI scan).

Countries

France

Contacts

Primary ContactAnne RousseauPlasse
Anne.RousseauPlasse@crm.microport.com01 46 01 30 82
Backup ContactHanan Fawaz
hanan.fawaz@crm.microport.com01 46 01 89 01

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026