Automatic Oxygen Control (SPOC) in Preterm Infants

Automatic Oxygen Control (SPOC) in Preterm Infants - Evaluation of a Revised Algorithm and Effect of Averaging Time of Pulse Oximetry Signal

Status
UNKNOWN
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03785899
Acronym
optimalSPOC
Enrollment
24
Registered
2018-12-24
Start date
2018-08-03
Completion date
2020-08-01
Last updated
2019-01-04

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Infantile Respiratory Distress Syndrome, Ventilator Lung; Newborn

Brief summary

Single-center, randomised controlled, cross-over clinical trial in preterm infants born at gestational age below 34+1/7 weeks receiving supplemental oxygen and respiratory support (continous positive airway pressure (CPAP) or non-invasive ventilation (NIV) or invasive ventilation (IV)). Routine manual control (RMC) of the fraction of inspired oxygen (FiO2) will be tested against RMC supported by automatic control (SPOC) with old-algorithm and RMC supported by CLAC with new-algorithm. The first primary hypothesis is, that the use of the new algorithm results in more time within arterial oxygen saturation (SpO2) target range compared to RMC only. The a-priori subordinate hypothesis is, that the new algorithm results in more time within SpO2 target range compared to SPOCold. The second primary hypothesis is, that the use of 2 seconds averaging time of the SpO2 Signal results in more time within arterial oxygen saturation (SpO2) target range compared to the use of 8 seconds averaging interval of the SpO2 signal.

Detailed description

BACKGROUND AND OBJECTIVE In preterm infants receiving supplemental oxygen, routine manual control (RMC) of the fraction of inspired oxygen (FiO2) is often difficult and time consuming. The investigators developed a system for closed-loop automatic control (SPOC) of the FiO2. The objective of this study is to test a revised, new algorithm with 3 adaptions against the former old algorithm and against RMC. The 3 adaptions are: 1. Faster re-adjustment to baseline-FiO2 (baseline FiO2: mean FiO2 during the previous 5min) 2. Delayed reduction of FiO2 below baseline FiO2 3. Maximum FiO2 adjustable by user The first primary hypothesis is, that the application of SPOCnew in addition to RMC results in more time within arterial oxygen saturation (SpO2) target range compared to RMC only. The a-priori subordinate hypothesis is, that the revised algorithm is more effective as the old algorithm to maintain the SpO2 in the target range. The second primary hypothesis is, that the shortening of averaging time used for the SpO2 Signal from 8 seconds to 2 seconds results in more time within SpO2 target range for both, SPOCnew and SPOCold. Further hypotheses for exploratory testing are, that the SPOC new algorithm will achieve a lower proportion of time with SpO2 above and below the target range, hyper- and hypoxia and an improved stability of cerebral oxygenation (measured as rcStO2 and rcFtO2E determined by Near-infrared spectroscopy) compared with SPOCold and RMC. Reduction of staff workload (estimated by number of manual adjustments per hour) by SPOC. Validation of a clinical scoring tool to monitor severity of apnea of prematurity. STUDY DESIGN The Study is designed as a single-center, randomized controlled, cross-over clinical trial in preterm infants receiving mechanical ventilation or nasal continuous positive airway pressure or non-invasive ventilation and supplemental oxygen (FiO2 above 0.21). Within a 30-hour period the investigators will compare 6 hours of RMC with 12-hour periods of RMC supported by SPOCnew algorithm or SPOCold algorithm, respectively. During intervals with SPOC control the SpO2 Signal averaging time will be 2 second or 8seconds , respectively, for 6 hours each.

Interventions

DEVICESPOCnew

SPOC is an automated, algorithm based adjustment of the fraction of inspired oxygen in relation to arterial saturation (SPO2). The revised new algorithm is turned on.

DEVICE8s SpO2 averaging

The arterial saturation (SPO2) will be averaged over 8s.

DEVICESPOCold

SPOC is an automated, algorithm based adjustment of the fraction of inspired oxygen in relation to arterial saturation (SPO2). The former old algorithm is turned on.

DEVICE2s SpO2 averaging

The arterial saturation (SPO2) will be averaged over 2s.

Sponsors

Fritz Stephan GmbH
CollaboratorUNKNOWN
University Hospital Tuebingen
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
No minimum to 34 Weeks
Healthy volunteers
No

Inclusion criteria

* gestational age at birth \<34+1/7weeks and * invasive mechanical ventilation OR noninvasive ventilation OR continous positive airway pressure support and * Fraction of inspired oxygen above 0.21 before inclusion and * more than 2 hypoxaemic events (arterial oxygen saturation below 80%) within 8 hours before inclusion and * parental written informed consent

Exclusion criteria

(any of the following): * congenital pulmonary anomalies * congenital heart defects influencing SpO2 (i.e. cyanotic heart defects) * right-to -left shunt through a PDA * Severe neonatal complications during study period (sepsis, necrotising enterocolitis) * diaphragmatic hernia or other diaphragmatic disorders

Design outcomes

Primary

MeasureTime frameDescription
Proportion of time with SpO2 within target range30 hoursComparison of proportion of time with SpO2 within target range and time above target range if no supplemental oxygen was administered at that time and within the preceding 30sec between the five treatment modalities

Secondary

MeasureTime frameDescription
Severe hypoxemic episodes30 hoursRate of episodes with SpO2 \<80% for at least 60seconds
Proportion of Time with Hyperoxia30 hoursComparison of proportion of time with SpO2 above 97% if supplemental oxygen was administered at that time or at anytime during the previous 30 seconds.
Stability of cerebral oxygenation30 hoursArea under the curve of cerebral tissue saturation or fraction of tissue oxygen extraction outside of the infants Median +- 5% or outside of the safe interval of 55-80% rcStO2.
Proportion of Time with SpO2 above target range30 hoursComparison of proportion of time with SpO2 above target range if supplemental oxygen was administered at that time or within the preceding 30sec .
Proportion of Time with SpO2 below target range30 hoursComparison of proportion of time with SpO2 below target range.
Proportion of Time with Hypoxia30 hoursComparison of proportion of time with SpO2 below 80%.

Other

MeasureTime frameDescription
Validation of clinical Apnea Score30 hoursValidation of a modified Apnea Score monitored by clinical staff by correlation between Score and other secondary outcomes. The modified Apnoea Score aims to quantify the burden from apnoea-bradycardia-syndrome by assigning 1-2-4 or 8 points according to event severity. Points are summed up during each treatment period in this study.
Staff workload30 hoursnumber of manual adjustments of inspired oxygen per time