Skip to content

Clinical Accuracy of the welloStationX

welloStationX Determination of Clinical Accuracy

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT03758157
Enrollment
110
Registered
2018-11-29
Start date
2018-10-15
Completion date
2018-11-15
Last updated
2021-07-07

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Fever

Brief summary

This study evaluates the clinical accuracy of the welloStationX, an automated non-touch self-service electronic thermometer using an infrared sensor of the surface of the forehead to screen for fever.

Detailed description

Traditionally, body temperature is measured using a contact thermometer placed on the forehead, in the mouth, ear, armpit or rectum. Non-contact thermometers allow a person's temperature to be taken with no contact with the person, reducing the discomfort of having to sit still for several minutes. Additionally, there is no need for disinfection between patients to occur, allowing for easier and faster use when screening large numbers of people. This study is to evaluate the clinical accuracy of the welloStationX, a non-contact thermometer by comparing its measurements with that of the gold standard Welch Allyn SureTemp oral thermometer.

Interventions

DIAGNOSTIC_TESTwelloStationX

welloStationX is an automated electronic thermometer using an infrared sensor of the surface of the forehead to measure human body skin temperature to screen for fever.

DIAGNOSTIC_TESTSureTemp

SureTemp is an oral thermometer to measure human body skin temperature.

Sponsors

Wello Inc.
Lead SponsorINDUSTRY

Study design

Observational model
COHORT
Time perspective
OTHER

Eligibility

Sex/Gender
ALL
Age
5 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* Male or female. * 5 years and up. * Signed informed consent form.

Exclusion criteria

* Use of medications known to affect body temperature (for example, antipyretics, barbiturates, thyroid preparations, antipsychotics, etc.) within 3 hours of the test. * immunization within seven days of the test

Design outcomes

Primary

MeasureTime frameDescription
Clinical Agreement Through Clinical Bias Evaluationone dayAgreement through clinical bias evaluation was calculated as the Mean Bias ± Standard Deviation (SD) by subtracting the first output measurement recorded with the welloStationX thermometer (DUT) from the corresponding output measurement recorded with the Welch Allyn SureTemp oral thermometer for each subject, and then calculating the Mean Bias ± SD of the biases.

Secondary

MeasureTime frameDescription
Clinical Repeatabilityone dayClinical Repeatability was calculated using the pooled standard deviations of the triplicate measurements recorded by the welloStationX.

Countries

United States

Participant flow

Participants by arm

ArmCount
Temperature Measurement
welloStationX: welloStationX is an automated electronic thermometer using an infrared sensor of the surface of the forehead to measure human body skin temperature to screen for fever. SureTemp: SureTemp is an oral thermometer to measure human body skin temperature.
110
Total110

Baseline characteristics

CharacteristicTemperature Measurement
Age, Continuous44.78 years
STANDARD_DEVIATION 16.14
Ethnicity (NIH/OMB)
Hispanic or Latino
54 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
56 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants
febrility
afebrile subjects (temperature < 99.5 deg. C)
77 Participants
febrility
febrile subjects (temperature >=99.5 deg. C)
33 Participants
Region of Enrollment
United States
110 Participants
Sex: Female, Male
Female
74 Participants
Sex: Female, Male
Male
36 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 110
other
Total, other adverse events
0 / 110
serious
Total, serious adverse events
0 / 110

Outcome results

Primary

Clinical Agreement Through Clinical Bias Evaluation

Agreement through clinical bias evaluation was calculated as the Mean Bias ± Standard Deviation (SD) by subtracting the first output measurement recorded with the welloStationX thermometer (DUT) from the corresponding output measurement recorded with the Welch Allyn SureTemp oral thermometer for each subject, and then calculating the Mean Bias ± SD of the biases.

Time frame: one day

ArmMeasureValue (MEAN)Dispersion
Temperature MeasurementClinical Agreement Through Clinical Bias Evaluation0.04 degrees FarehheitStandard Deviation 0.42
Secondary

Clinical Repeatability

Clinical Repeatability was calculated using the pooled standard deviations of the triplicate measurements recorded by the welloStationX.

Time frame: one day

ArmMeasureValue (NUMBER)
Temperature MeasurementClinical Repeatability0.062 degrees Farehheit

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026