Capillary Blood Volume and Pain Perception Obtained in a Process of Puncturing With Different Safety Lancets

A Randomized, Single Blind, Uni-centre Study in Healthy Volunteers, to Determine Capillary Blood Volume and Pain Perception Obtained in a Process of Puncturing With Different Safety Lancets

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03752229
Enrollment
300
Registered
2018-11-23
Start date
2017-12-08
Completion date
2018-01-30
Last updated
2018-11-26

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Healthy

Keywords

safety lancets, lancet, capillary blood, capillary blood sampling, blood volume, pain

Brief summary

The study was a randomized, single blind, uni-center pivotal study in male and female healthy volunteers, to determine capillary blood volume and pain perception obtained in a process of puncturing with different safety lancets. A group of three hundred (300) Caucasian male and female Subjects was randomized (aged \> 18 to \< 65 years old) who were in good physical and mental condition. The Subjects had 4 (four) fingers (two fingers from each hand) lanced each with a different type of safety lancet. Overall, 4 (four) different safety lancets were investigated per each Subject. Overall, 48 (forty eight) different safety lancets (4 different safety lancets per one Subject) were investigated, the capillary blood volume and pain were measured.

Interventions

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEHaemolance Micro Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEHaemolance Low Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEHaemolnace Normal Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEHaemolance High Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEHaemolance Max Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEHaemolance Pediatric

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEergoLance Micro Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEergoLance Normal Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEergoLance High Flow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEMedlance (light blue)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEMedlance (blue)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEMedlance (dark blue)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEMedlance (yellow)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEMedlance (orange)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEMedlance (red)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEUnistik 3 Extra

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEUnistik 3 Comfort

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEUnistik Touch

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEReadyLance (blue)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEReadyLance (orange)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEAssureLance

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEAssureLance MicroFlow

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICESurgilance

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICESurgilance Lite (grey)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICESurgilance Lite (purple)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICESteriLance Lite2

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICESensiLance (pink)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICESteriLance Press

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICESensiLance (orange)

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEmylife

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEVeriFine

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

DEVICEMicrodot

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Sponsors

HTL-Strefa S.A.
Lead SponsorINDUSTRY

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
OTHER
Masking
SINGLE (Subject)

Eligibility

Sex/Gender
ALL
Age
18 Years to 65 Years
Healthy volunteers
Yes

Inclusion criteria

* Willing to sign the informed consent. * Male and female volunteers * Age between 18 and 65 inclusive * Able to communicate well with the Investigator and comply with the requirements of the study * Volunteers who are in good physical and mental health

Exclusion criteria

* Confirmed or suspected malignant cancer * Pregnancy or breastfeeding (only for women with childbearing potential based on pregnancy test conducted at Site) * History of poor blood circulation. * Any skin condition on his or her fingers that prevents blood sampling * Anxiety with needles or finger pricks * Clotting disorders (including bleeding) in medical history * Neuropathy or other condition affecting sensation in the hands. * History of blood borne infection (e.g., HIV, hepatitis B or C, syphilis, malaria, babesiosis, brucellosis, leptospirosis, arboviral infections, relapsing fever, T lymphotropic virus Type 1, Creutzfeldt-Jakob disease) * Currently participating in another study * History of drug or alcohol abuse within the 12 months prior to screening or evidence of such abuse * Intake of alcohol within 48 hours prior to the start of the study (self-reporting) * Donation or loss of 400 mL or more of blood within 4 weeks prior to the start of the study * Any other condition that in the Investigator opinion may negatively influence Subject's participation in the study * Intake of medicines that affect blood coagulability (including anticoagulants such as vitamin K, antivirals and anticoagulants such as heparin, aspirin, thrombin inhibitors, vitamin K antagonists).

Design outcomes

Primary

MeasureTime frameDescription
Volume (microliters) of blood in the capillary tubeThe blood volume was noted after 20 secondsThe amount of capillary blood volume collected after a single lancing of the fingertip.

Secondary

MeasureTime frameDescription
Pain perceptiondirectly after lancingThe intensity of pain perceived by the patient was assessed with the use of the numeric Visual Analog Scale (VAS). The Subject marked on the scale the number from 0 (no pain) to 10 (worst imaginable pain) that represented their perception of pain. A higher score indicated greater pain intensity.

Countries

Poland

Outcome results

None listed