Skip to content

Evaluation of Precision and Accuracy of INR Measurements in a Point of Care Device

Evaluation of the Precision and Accuracy of International Normalised Ratio (INR) Measurements as Assessed by the LumiraDx Instrument in a Professional Point of Care Setting With Multiple Operators

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT03682419
Acronym
OPTIMAL
Enrollment
420
Registered
2018-09-24
Start date
2018-09-17
Completion date
2018-11-23
Last updated
2026-06-11

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Antiphospholipid Syndrome, Artificial Heart Device User, Atrial Fibrillation, Deep Vein Thrombosis, Post-Myocardial Infarction Syndrome, Pulmonary Embolism, Thrombophilia

Brief summary

Comparison of capillary whole blood INR determined by LumiraDx Instrument to venous plasma INR determined by laboratory reference method (IL ACL ELITE PRO) for method comparison and assessment of accuracy and bias by regression analysis and other analytical methods.

Detailed description

A performance evaluation study designed to assess the precision and accuracy of the LumiraDx Instrument, a point of care in vitro diagnostic system employing a portable diagnostic meter in conjunction with single use analyte test strips, for the quantitative determination of Prothrombin Time expressed as International Normalised Ratio (INR), as sampled from individuals evaluated for INR at the point of care (POC). The accuracy of the LumiraDx Instrument INR results will be assessed by, in the first instance, comparison to the INR results obtained from the same individuals as analysed by trained laboratory professionals using the paired venous plasma samples in the reference device IL ACL ELITE PRO Coagulation Analyzer. The accuracy of INR measurements of capillary whole blood across the range of haematocrit (25% - 55%) will be assessed using paired haematocrit results obtained from the HemoCue Hb 201+ on venous whole blood. The results of these evaluations are intended to be used for regulatory filings for the LumiraDx Instrument and LumiraDx INR test as an in vitro diagnostic test in the European Union (EU) and other relevant geographies. For information only, the accuracy of the LumiraDx Instrument INR Test will also be assessed by comparison to INR results from an alternative point of care instrument, the Roche Coaguchek PRO II. Adult males and females currently in receipt of vitamin K antagonist therapy will be included on the study. Approximately 500 subjects will be recruited to ensure that sufficient suitable subjects are enrolled and to allow for a small level of exclusions. The power calculation as well as recommendations of the FDA CLIA Waiver Guidance document (1) requires a minimum of 360 valid subjects in total (inclusive of non-therapy subset). A subsection of the subject population (\ 50 subjects) will consist of adult males and females who fulfil all other study inclusion criteria but are not in receipt of vitamin K antagonist therapy. This will be done in order to test the lower measurement limit of the LumiraDx Instrument INR test. At the end of the study the Healthcare Professional will be asked to complete a User Acceptability Questionnaire which is intended to assess the acceptability of the LumiraDx INR system in terms of ease of use.

Interventions

PROCEDUREVKA Patients

Venepuncture and capillary blood collection from Patients receiving Warfarin Therapy

PROCEDURENon-VKA Patients

Venepuncture and capillary blood collection from Patients not receiving Warfarin Therapy

Sponsors

LumiraDx UK Limited
Lead SponsorINDUSTRY

Study design

Observational model
OTHER
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

- VKA Therapy Subjects: * Persons \>18 years of age * Willing and able to provide written informed consent and comply with study procedures * Currently prescribed vitamin K antagonist therapy * Deemed medically appropriate for study participation by the Investigator

Exclusion criteria

- VKA Therapy Subjects: * Persons \<18 years of age * Subject has previously participated in this study * Subject is within 4 weeks of first prescription of vitamin K antagonist therapy * Confirmed or suspected pregnancy * Unwilling or unable to provide written informed consent and comply with study procedures * Vulnerable populations deemed inappropriate for study by the Investigator * Deemed medically inappropriate for study by the Investigator (i.e. patients with a known inherited \[e.g. haemophilia or von Willebrand's disease\] or acquired \[e.g. liver cirrhosis\] condition likely to be associated with a coagulopathy; or patients receiving non-VKA anticoagulant medications) Inclusion Criteria - Non-VKA Therapy Subjects: * Persons \>18 years of age * Willing and able to provide written informed consent and comply with study procedures * Deemed medically appropriate for study participation by the Investigator

Design outcomes

Primary

MeasureTime frameDescription
To determine the clinical performance of the LumiraDx instrument8 weeksMeasurement of INR using Capillary \& Venous blood samples

Secondary

MeasureTime frameDescription
To determine the accuracy of the LumiraDx instrument8 weeksMeasurement of INR \& Haematocrit using Capillary \& Venous blood samples

Countries

United Kingdom

Contacts

PRINCIPAL_INVESTIGATORRobert C Tait, BSc (Hons)

Consultant Haematologist

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Jun 12, 2026