Skip to content

Hemodilution Validation of INVSENSOR00026

Hemodilution Validation of INVSENSOR00026

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03677375
Enrollment
173
Registered
2018-09-19
Start date
2018-09-04
Completion date
2019-02-04
Last updated
2019-05-20

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Healthy

Brief summary

In this study, the concentration of hemoglobin within the subject's blood will be reduced in a controlled manner by administering fluids intravenously. The accuracy of a noninvasive hemoglobin sensor will be assessed by comparison to hemoglobin measurements from a laboratory analyzer.

Interventions

DEVICEINVSENSOR00026

Noninvasive pulse oximeter sensor

Sponsors

Masimo Corporation
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 50 Years
Healthy volunteers
Yes

Inclusion criteria

* Subject weighs a minimum of 110 lbs and no more than 250 lbs unless subject is over 6 feet tall. * Hemoglobin value is greater than or equal to 11 g/dL at time of screening * Baseline heart rate ≥ 45 bpm and ≤ 85 bpm. * CO value ≤ 2.0% FCOHb * Subject has a physical status of ASA I or II (American Society of Anesthesiology Class I; Healthy subjects without any systemic disease at all. American Society of Anesthesiology Class II; subjects with mild systemic disease) as it applies to the systemic disease portion of the classification. * Systolic Blood Pressure ≤ 140 mmHg and Diastolic Blood Pressure ≤ 90 mmHg. * Subject is able to read and communicate in English and understands the study and risks involved.

Exclusion criteria

* Subject is pregnant * Subject smokes (including e-cigarette use) * Subject has a BMI \> 35 and has been classified as morbidly obese or at an increased risk for participation by a medical professional. * Subject has a history of fainting (vasovagal), blacking out or losing consciousness during or after a blood draw. * Subject has open wounds, inflamed tattoos or piercings, or any visible healing wounds. * Subject has known drug or alcohol abuse or uses recreational drugs. * Subject has experienced a head injury with loss of consciousness within the last year. * Subject has taken anticoagulant medication within the last 30 days. * Subject has any chronic bleeding disorders (i.e. hemophilia). * Subject has any history of a stroke, myocardial infarction, heart attack, or seizures. * Subject has any cancer or history of cancer (not including skin cancer). * Subject has a chronic neurological disease (i.e. multiple sclerosis, Huntington's disease). * Subject has any cardiac dysrhythmias (i.e. atrial fibrillation) and has not received clearance from their physician. * Subject has known neurological and/or psychiatric disorder (i.e. schizophrenia, bipolar disorder) that interferes with the subject's level of consciousness. * Subject has Wolff-Parkinson-White Syndrome or Stokes-Adams Syndrome * Subject has any medical condition which in the judgment of the investigator and/or medical staff, renders them ineligible for participation in this study (Discretion of investigator). * Subject has taken opioid pain medication within 24 hours of start of study. * Subject has any type of infectious disease (i.e. Hepatitis, HIV, Tuberculosis, Flu, Malaria, Measles, etc.). * Subject is taking medications known to treat any type of infectious disease. * Subject has either signs or history of peripheral ischemia or carpal tunnel. * Subject has had invasive surgery within the past year- including but not limited to major dental surgery, gallbladder, heart, appendix, major fracture repairs (involving plates/ screws), jaw surgery, Urinary tract surgery, plastic surgery, major ENT surgery, joint replacement or gynecological surgeries, heart surgery or thoracic surgery. * Subject has symptoms of congestion, head colds, flu, or other illnesses * Subject has donated blood within the last 2 weeks. * Subject is claustrophobic or has generalized anxiety disorder. * Subject has been in severe car accident(s) or a similar type of accident(s) requiring hospitalization within the last 12 months. * Subject has had a concussion within the last 12 months. * Subject has chronic unresolved asthma, lung disease or respiratory disease. * Subject is allergic to lidocaine, latex, adhesives, or plastic. * Subject has heart conditions, insulin-dependent diabetes or uncontrolled hypertension. * Subject has delivered vaginally, has had a pregnancy terminated, a miscarriage with hospitalization, or had a C-section within the last 6 months. * Subject intends on participating in any heavy lifting, repetitive movement of their wrist (including riding a motorcycle) or exercise (working out, riding a bike, riding a skate board etc.), or any activity that will put additional stress on the wrist within 24 hours following a study that involves an arterial line. * Discretion of investigator/study staff

Design outcomes

Primary

MeasureTime frameDescription
Accuracy of Noninvasive Hemoglobin Sensor by Arms Calculation1-5 hoursAccuracy will be determined by comparing the noninvasive hemoglobin measurement of the pulse oximeter to the hemoglobin value obtained from a blood sample and calculating the arithmetic root mean square (Arms) error value. In order to obtain the Arms value, the blood sample hemoglobin value is subtracted from the pulse oximeter hemoglobin value for a number of samples, the average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the Arms Error value.

Countries

United States

Participant flow

Pre-assignment details

Protocol enrollment number includes screen failures and subjects that did not proceed due to time constraints

Participants by arm

ArmCount
Test Subject
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive INVSENSOR00026. INVSENSOR00026: Noninvasive pulse oximeter sensor
40
Total40

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyAdverse Event3
Overall StudyIncomplete procedure8
Overall StudyPhysician Decision1

Baseline characteristics

CharacteristicTest Subject
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
40 Participants
Race/Ethnicity, Customized
Asian or Pacific Islander
4 Participants
Race/Ethnicity, Customized
Black or African American
12 Participants
Race/Ethnicity, Customized
Caucasian
15 Participants
Race/Ethnicity, Customized
Hispanic
9 Participants
Sex: Female, Male
Female
31 Participants
Sex: Female, Male
Male
9 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 93
other
Total, other adverse events
8 / 93
serious
Total, serious adverse events
0 / 93

Outcome results

Primary

Accuracy of Noninvasive Hemoglobin Sensor by Arms Calculation

Accuracy will be determined by comparing the noninvasive hemoglobin measurement of the pulse oximeter to the hemoglobin value obtained from a blood sample and calculating the arithmetic root mean square (Arms) error value. In order to obtain the Arms value, the blood sample hemoglobin value is subtracted from the pulse oximeter hemoglobin value for a number of samples, the average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the Arms Error value.

Time frame: 1-5 hours

ArmMeasureValue (NUMBER)
Test SubjectAccuracy of Noninvasive Hemoglobin Sensor by Arms Calculation1.13 g/dL