Skip to content

PMCF Study on the Safety and Performance of the Alloclassic Variall Cup Ceramic Bearing System in Total Hip Arthroplasty

Alloclassic® Variall® Cup Ceramic Bearing System in Total Hip Arthroplasty A Multi-center, Prospective, Non-controlled Post Market Surveillance Study

Status
Terminated
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT03672370
Enrollment
100
Registered
2018-09-14
Start date
2015-03-24
Completion date
2020-05-01
Last updated
2021-01-14

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Avascular Necrosis, Inflammatory Arthritis, Osteoarthritis, Posttraumatic Arthritis

Keywords

Total hip arthroplasty, Medical device, Performance, Safety, Hip prosthesis

Brief summary

The study is a multicenter, prospective, non-randomized, non-controlled consecutive cohort post-market clinical follow-up study.

Detailed description

The objectives of this study are to obtain survival and outcomes data on the Alloclassic® Variall® Cup in combination with the BIOLOX® delta Taper Liner and a BIOLOX® delta Femoral Head used in primary total hip arthroplasty. This will be done by analysis of standard scoring systems, radiographs and adverse event records. Two sites are involved in this study, enrolling a total of 100 patients for the study. All subjects were required to participate in the Informed Consent Process.

Interventions

None listed

Sponsors

Zimmer Biomet
Lead SponsorINDUSTRY

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to 75 Years
Healthy volunteers
No

Inclusion criteria

* Patient is skeletally mature. * Patient qualifies for primary unilateral or bilateral total hip arthroplasty (THA) based on physical exam and medical history including the following: * Avascular necrosis (AVN) * Osteoarthritis (OA) * Inflammatory arthritis * Post-traumatic arthritis * Patient has no history of previous prosthetic replacement device (any type, including surface replacement arthroplasty, endoprosthesis, etc.) of the affected hip joint(s). * Patient has a Harris Hip Score \<70 in the affected hip * Patient is willing and able to provide written informed consent. * Patient is willing and able to cooperate in the required post-operative therapy. * Patient is willing and able to complete scheduled follow-up evaluations as described in the Informed Consent. * Patient has participated in the Informed Consent process and has signed the Ethics Committee approved informed consent.

Exclusion criteria

* The patient is: * A prisoner * Mentally incompetent or unable to understand what participation in the study entails * A known alcohol or drug abuser * Anticipated to be non-compliant. * The patient has a neuromuscular disorder, vascular disorder or other conditions that could contribute to prosthesis instability, prosthesis fixation failure, or complications in postoperative care. * The patient has a neurologic condition in the ipsilateral or contralateral limb which affects lower limb function. * The patient has a known history of systemic disease that could affect his/her safety or the study outcome. * The patient is known to be pregnant. * The patient is unwilling or unable to give informed consent, or to comply with the follow-up program. * The patient is not having any other investigational drug or device simultaneously. * The patient has an active or latent infection in or about the affected hip joint or an infection distant from the hip joint that may spread to the hip hematogenously. * The patient has insufficient bone stock to fix the component. Insufficient bone stock exists in the presence of metabolic bone disease (i.e. clinical relevant osteoporosis), cancer, and radiation. Note: Dual Energy X-ray Absorptiometry (DEXA) may be used to assess the presence of adequate bone stock. * The patient has osteoradionecrosis in the operative hip joint * The patient has a known sensitivity or allergic reaction to one or more of the implanted materials. * The patient has known local bone tumors in the operative hip. * The patient is Grade III obese with a Body Mass Inldex (BMI) \> 40.

Design outcomes

Primary

MeasureTime frameDescription
Clinical Success Will be Based on the Harris Hip Scoring System3 yearsThe Harris Hip Score (HHS) was developed for the assessment of the results of hip surgery, and is intended to evaluate various hip disabilities and methods of treatment in an adult population. The domains covered are pain, function, absence of deformity, and range of motion. The score has a minimum of 0 points and a maximum of 100 points. Clinical success will be defined as a modified Harris Hip score of \> 80 that included a rating of 'mild', or 'no pain'; a failure will be defined as a modified Harris Hip score \< 80.
Percentage of Participants Who Survived According to the Survival Analysis With the Kaplan Meier (K-M) Estimate4 yearsEndpoint revision of any investigational device for any reason attributable to any investigational device

Secondary

MeasureTime frameDescription
Oxford Hip Score3 yearsThe purpose of the Oxford Hip Score (OHS) is to assess the outcome after total hip arthroplasty by measuring cases' perceptions in adjunction to surgery. The OHS assesses pain (6 items) and function (6 items) of the hip in relation to daily activities such as walking, dressing, sleeping, etc. The scoring is the following: 0-4 (worst to best) with overall scores ranging from 0-48 where 48 represents the best score
EQ-5D3 yearsThe EQ-5D questionnaire comprises two parts. For the first part, respondents are asked to tick boxes to indicate the level of problem on each of the five dimensions (mobility, self-care, usual activities, pain/discomfort, and anxiety/depression). In total, there are three levels per dimension (no problems/no pain/not worried, some problems/some pain/a bit worried, a lot of problems/a lot of pain/very worried). Responses are coded as single-digit numbers expressing the severity level selected in each dimension. The digits for the five dimensions are combined in a 5-digit code. The second part of the questionnaire is the EQ Visual Analogue Scale. This captures the respondent's overall assessment of their health from 0 (worst health imaginable) to 100 (best health imaginable). The results from both parts are indexed, weightened and pulled together. The final score lies between 1 for full health and 0 for death. Values less than 0 are possible considered to be worse than death.

Countries

Austria

Participant flow

Participants by arm

ArmCount
Alloclassic Variall Cup
Subjects who received the Alloclassic Variall Cup Ceramic Bearing System
89
Total89

Baseline characteristics

CharacteristicAlloclassic Variall Cup
Age, Continuous63.9 years
STANDARD_DEVIATION 9
BMI29.0 kg/m^2
STANDARD_DEVIATION 1.2
Race and Ethnicity Not Collected— Participants
Region of Enrollment
Austria
89 participants
Sex: Female, Male
Female
51 Participants
Sex: Female, Male
Male
38 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
1 / 89
other
Total, other adverse events
2 / 89
serious
Total, serious adverse events
7 / 89

Outcome results

Primary

Clinical Success Will be Based on the Harris Hip Scoring System

The Harris Hip Score (HHS) was developed for the assessment of the results of hip surgery, and is intended to evaluate various hip disabilities and methods of treatment in an adult population. The domains covered are pain, function, absence of deformity, and range of motion. The score has a minimum of 0 points and a maximum of 100 points. Clinical success will be defined as a modified Harris Hip score of \> 80 that included a rating of 'mild', or 'no pain'; a failure will be defined as a modified Harris Hip score \< 80.

Time frame: 3 years

ArmMeasureValue (MEAN)Dispersion
Alloclassic Variall CupClinical Success Will be Based on the Harris Hip Scoring System92.6 score on a scaleStandard Deviation 8.9
Primary

Percentage of Participants Who Survived According to the Survival Analysis With the Kaplan Meier (K-M) Estimate

Endpoint revision of any investigational device for any reason attributable to any investigational device

Time frame: 4 years

ArmMeasureValue (NUMBER)
Alloclassic Variall CupPercentage of Participants Who Survived According to the Survival Analysis With the Kaplan Meier (K-M) Estimate98.63 percentage of participants
Secondary

EQ-5D

The EQ-5D questionnaire comprises two parts. For the first part, respondents are asked to tick boxes to indicate the level of problem on each of the five dimensions (mobility, self-care, usual activities, pain/discomfort, and anxiety/depression). In total, there are three levels per dimension (no problems/no pain/not worried, some problems/some pain/a bit worried, a lot of problems/a lot of pain/very worried). Responses are coded as single-digit numbers expressing the severity level selected in each dimension. The digits for the five dimensions are combined in a 5-digit code. The second part of the questionnaire is the EQ Visual Analogue Scale. This captures the respondent's overall assessment of their health from 0 (worst health imaginable) to 100 (best health imaginable). The results from both parts are indexed, weightened and pulled together. The final score lies between 1 for full health and 0 for death. Values less than 0 are possible considered to be worse than death.

Time frame: 3 years

ArmMeasureValue (MEAN)Dispersion
Alloclassic Variall CupEQ-5D0.9 score on a scaleStandard Deviation 0.2
Secondary

Oxford Hip Score

The purpose of the Oxford Hip Score (OHS) is to assess the outcome after total hip arthroplasty by measuring cases' perceptions in adjunction to surgery. The OHS assesses pain (6 items) and function (6 items) of the hip in relation to daily activities such as walking, dressing, sleeping, etc. The scoring is the following: 0-4 (worst to best) with overall scores ranging from 0-48 where 48 represents the best score

Time frame: 3 years

ArmMeasureValue (MEAN)Dispersion
Alloclassic Variall CupOxford Hip Score44.7 score on a scaleStandard Deviation 7.9