Effects of Different Intra-arterial Catheter Size on Frequency of Optimal Dynamic Response in Radial Arterial Waveform

Effects of Different Intra-arterial Catheter Size on the Frequency of Optimal Dynamic Response in Radial Arterial Waveform

Status
UNKNOWN
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03642756
Enrollment
366
Registered
2018-08-22
Start date
2018-10-01
Completion date
2019-12-01
Last updated
2018-08-22

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Catheters

Brief summary

The aim of this study is to compare the effects of different intra-arterial catheter size on the frequency of optimal dynamic response in radial arterial waveform.

Interventions

DEVICEdifferent intra-arterial catheter size

in placebo group, patients were cannulated with 20 gauge intra-arterial catheter. In experiment groups, patients were cannulated with 22 gauge or 24 gauge intra-arterial catheter.

Sponsors

Seoul National University Hospital
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
DIAGNOSTIC
Masking
DOUBLE (Subject, Outcomes Assessor)

Eligibility

Sex/Gender
ALL
Age
20 Years to 79 Years
Healthy volunteers
No

Inclusion criteria

* patients who were scheduled for elective neurosurgery under general anesthesia * patients with age of 20-79 and ASA class I-III * patients who are needed to establish invasive arterial pressure monitoring

Exclusion criteria

* patients who refuted to be involved in the study * patients with age of \<20 or \>80 * patients who had both arm saved * patients who had a previous history of surgery on radial artery

Design outcomes

Primary

MeasureTime frameDescription
The frequency of adequately damped arterial waveform1 minute after cannulationThe accuracy of arterial blood pressure waveform is depedent on damping coefficient (DC) and natural frequency (NF). The extent to which a monitoring system is damped is expressed by the DC, while the frequency of oscillation that a monitoring system resonates at is known as the NF. The NF and DC are required to be within certain rages for accurate blood pressure monitoring, and they can be affected by different arterial catheter size. Adequately damped arterial waveform can be defined as NF and DC being within certain ranges. For example, waveforms with DC of 0.5 at a NF of 15Hz can be regarded as adequately damped arterial waveform. However, if waveforms have DC of 0.1 at a NF of 15Hz, these can be regarded as not having optimal dynamic response. (There is a reference figure explaining these factors in detail in a previous article. Gardner RM: Direct blood pressure measurement--dynamic response requirements. Anesthesiology 1981; 54:227-3)

Secondary

MeasureTime frameDescription
The difference of mean blood pressure between invasive blood pressure and non-invasive blood pressure1 minute after cannulationAt 1 minute after cannulating radial artery, we will obtain information about the mean blood pressure measured in both invasive blood pressure and non-invasive blood pressure. Invasive blood pressure will be measured from radial artery cannulation, and non-invasive blood pressure will be measured by blood pressure cuff at Rt. upper arm. Thereafter, the difference of mean blood pressure between two values will be used for data analysis.

Countries

South Korea

Outcome results

None listed