Clinical Trial Testing the BioSerenity Pulse Oximeter (EOXY) Efficacy, Performance and Safety in Healthy Volunteer

Clinical Trial Testing the BioSerenity Pulse Oximeter (EOXY) Efficacy, Performance and Safety in Healthy Volunteer

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03614416
Acronym
EOXY
Enrollment
13
Registered
2018-08-03
Start date
2017-12-08
Completion date
2018-11-06
Last updated
2021-01-12

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Healthy

Brief summary

The purpose of the study is to test the performance and safety of the Bioserenity pulse oxymeter EOXY. For that, and in agreement with ISO60601-2-61 norm and FDA regulation, measures of heart rate, with a gold-standard, and SaO2 measures generated by blood samples will be compared with Sp02 and heart rate measures that are generated by our EOXY. Thirteen healthy subjects will participate in the study in total. Three subjects will be enrolled during a primary phase .They will participate at a secondary phase as well as ten others subjects. The purpose of the primary phase is to test the body area where the measurement can be done. The purpose of the secondary phase is to measure SpO2 and HR in desaturation conditions.

Detailed description

The primary objective is to evaluate and document the SpO2 accuracy of the pulse oximeter equipment, evaluating the performance of the device compared to invasive method blood sampling (SaO2) by CO-Oximeter in healthy subjects during invasive controlled desaturation study at different ranges (100 % to 70% SaO2). Subject safety during desaturation and saturation period, will be evaluated by reviewing vital sign: heart rate (HR) continuously and blood pressure/temperature between each desaturation plateau. There are two secondaries objectives, the measurement of the variation of the fraction of inspired oxygen (FiO2), and the validation of the algorithm for data review and analysis. For the study primary phase, the purpose of this phase is to test the body area where the SpO2 is better detected and define 3 measurement areas. For this, three healthy volunteers will be placed in normal O2 saturation, the SpO2 will be measured using the Bioserenity oximeter (noninvasive method) and will be compared to a gold-standard oximeter CE-marked. The investigator will determine the best area in the body where the SpO2 could be taken for the following subjects. During a second phase, the measurement of Sp02 and HR with EOXY, will be performed in O2 saturation conditions up to 100% and then in desaturation by plateau from 100 % to 70%. 3 EOXY pulse oximeter recorded the SpO2 at the same time in 3 different areas of the body. As required by the standard, Sp02 measures will be simultaneously compared to SaO2 values measured by blood sampling taken from an indwelling arterial catheter. The HR obtained with EOXY will be compared simultaneously to a gold standard oximeter CE-marked. As the example in the standard, it was choosen to do the test on the 70-100% range, with 5 plateaus (100%-97%, 97%-92%, 92%-85%, 84%-78%, 77%-70%) and to collect five Sp02 and Sa02 values (blood samples) and five HR measures, per plateau, for each of 13 subjects. The desaturation will be controlled thanks' the FiO2 measures, that should allow to bring subjects near target levels of desaturation plateau.

Interventions

DEVICEEOXY device

Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard

DEVICEGold standard oximeter

Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard

PROCEDURESaO2 sampling

Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard

Sponsors

NAMSA
CollaboratorOTHER
Université Catholique de Louvain
CollaboratorOTHER
BioSerenity
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Intervention model description

No randomisation. All subjects will test simultaneously the gold standard and the EOXY device

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* Healthy subject (Man/ Women) \> 18 years old * Subject signed informed consent prior to any screening procedure * Subject in good health (COHb \< 3% , MetHb \< 2%, CtHb \> 10mg/dL) * Healthy adult capable of undergoing controlled hypoxemia to the level called for in the protocol with minimal medical risk

Exclusion criteria

* Smokers or individuals exposed to high levels of carbon monoxide that result in elevation of carboxy hemoglobin levels * Individual subject to conditions that result in elevated levels of methemoglobin * Subject who would be placed at undue medical risk associated with any procedure called for in the protocol * Subject with open wounds * Pregnant woman (negative pregnancy test needed) * Subject allergic to silicon, polyamide and silver yarn * Subject with cardiorespiratory disorders likely to worsen with thoracic pressure applied by the textile band * Subject with mental or motor impairment preventing him from expressing pain * Subject with blood or skin disorder that may impacting results * Subject with behavioral disorders, too agitated or too aggressive * Subject with sensorial disorders, insensible to skin pain * Subject susceptible to tension/pressure based headaches

Design outcomes

Primary

MeasureTime frameDescription
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Forehead Position1 dayFor each desaturation plateau, SpO2 ARMS are calculated by the difference between the investigational pulse oximeter SpO2 measurement and the co-oximeter's SaO2 measurements
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Torso Position1 dayFor each desaturation plateau, SpO2 ARMS (Accuracy Root-Mean Square) are calculated by the difference between the investigational pulse oximeter SpO2 measurement and the co-oximeter's SaO2 measurements
Percentage SPO2 Accuracy Root-Mean Square (ARMS) at BACK Position1 dayFor each desaturation plateau, SpO2 ARMS are calculated by the difference between the investigational pulse oximeter SpO2 measurement and the co-oximeter's SaO2 measurements
HR Accuracy Root-Mean Square (ARMS) at Forehead Position1 dayFor each desaturation plateau, HR ARMS are calculated by the difference between the investigational pulse oximeter HR measurement and the HR gold standard measurements
Percentage of HR Accuracy Root-Mean Square (ARMS) at TORSO Position1 dayFor each desaturation plateau, HR ARMS are calculated by the difference between the investigational pulse oximeter HR measurement and the HR gold standard measurements
Percentage of HR Accuracy Root-Mean Square (ARMS) at BACK Position1 dayFor each desaturation plateau, HR ARMS are calculated by the difference between the investigational pulse oximeter HR measurement and the HR gold standard measurements

Countries

Belgium

Participant flow

Participants by arm

ArmCount
EOXY Device and Gold Standard Oximeter and SaO2 Measures
Heart rate and SPO2 measures provided from EOXY device Heart rate measures provided from gold standard oximeter. SaO2 measures provided from blood sampling There is only one arm: all subjects have simultaneously three interventions (as required in the European standard ISO 80601-2-61), to qualify a pulse oximeter EOXY device: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard Gold standard oximeter: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard SaO2 sampling: Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
10
Total10

Baseline characteristics

CharacteristicEOXY Device and Gold Standard Oximeter and SaO2 Measures
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
10 Participants
Age, Continuous23.5 years
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
Race (NIH/OMB)
Asian
0 Participants
Race (NIH/OMB)
Black or African American
2 Participants
Race (NIH/OMB)
More than one race
2 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants
Race (NIH/OMB)
White
6 Participants
Region of Enrollment
Belgium
10 participants
Sex: Female, Male
Female
3 Participants
Sex: Female, Male
Male
7 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 10
other
Total, other adverse events
0 / 10
serious
Total, serious adverse events
0 / 10

Outcome results

Primary

HR Accuracy Root-Mean Square (ARMS) at Forehead Position

For each desaturation plateau, HR ARMS are calculated by the difference between the investigational pulse oximeter HR measurement and the HR gold standard measurements

Time frame: 1 day

Population: People respecting study conditions (good irrigation and no movements)

ArmMeasureGroupValue (NUMBER)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresHR Accuracy Root-Mean Square (ARMS) at Forehead PositionOverall range 70-100% of O212.2 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresHR Accuracy Root-Mean Square (ARMS) at Forehead Position97-100% of O210.0 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresHR Accuracy Root-Mean Square (ARMS) at Forehead Position97-92% of O211.7 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresHR Accuracy Root-Mean Square (ARMS) at Forehead Position92-85% of O211.5 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresHR Accuracy Root-Mean Square (ARMS) at Forehead Position84-78% of O214.3 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresHR Accuracy Root-Mean Square (ARMS) at Forehead Position77-70% of O213.0 percentage of ARMS (accuracy)
Primary

Percentage of HR Accuracy Root-Mean Square (ARMS) at BACK Position

For each desaturation plateau, HR ARMS are calculated by the difference between the investigational pulse oximeter HR measurement and the HR gold standard measurements

Time frame: 1 day

Population: People respecting study conditions (good irrigation and no movements)

ArmMeasureGroupValue (NUMBER)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage of HR Accuracy Root-Mean Square (ARMS) at BACK PositionOverall range 70-100% of O221.1 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage of HR Accuracy Root-Mean Square (ARMS) at BACK Position97-100% of O222.3 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage of HR Accuracy Root-Mean Square (ARMS) at BACK Position97-92% of O224.6 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage of HR Accuracy Root-Mean Square (ARMS) at BACK Position92-85% of O219.2 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage of HR Accuracy Root-Mean Square (ARMS) at BACK Position84-78% of O219.1 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage of HR Accuracy Root-Mean Square (ARMS) at BACK Position77-70% of O220.6 percentage of ARMS (accuracy)
Primary

Percentage of HR Accuracy Root-Mean Square (ARMS) at TORSO Position

For each desaturation plateau, HR ARMS are calculated by the difference between the investigational pulse oximeter HR measurement and the HR gold standard measurements

Time frame: 1 day

Population: People respecting study conditions (good irrigation, no movements)

ArmMeasureGroupValue (NUMBER)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage of HR Accuracy Root-Mean Square (ARMS) at TORSO Position84-78% of O230.4 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage of HR Accuracy Root-Mean Square (ARMS) at TORSO Position77-70% of O236.5 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage of HR Accuracy Root-Mean Square (ARMS) at TORSO PositionOverall range 70-100% of O232.1 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage of HR Accuracy Root-Mean Square (ARMS) at TORSO Position97-100% of O230.8 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage of HR Accuracy Root-Mean Square (ARMS) at TORSO Position97-92% of O231.5 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage of HR Accuracy Root-Mean Square (ARMS) at TORSO Position92-85% of O231.5 percentage of ARMS (accuracy)
Primary

Percentage SPO2 Accuracy Root-Mean Square (ARMS) at BACK Position

For each desaturation plateau, SpO2 ARMS are calculated by the difference between the investigational pulse oximeter SpO2 measurement and the co-oximeter's SaO2 measurements

Time frame: 1 day

ArmMeasureGroupValue (NUMBER)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage SPO2 Accuracy Root-Mean Square (ARMS) at BACK PositionOverall range 70-100% of O26.02 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage SPO2 Accuracy Root-Mean Square (ARMS) at BACK Position97-100% of O27.09 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage SPO2 Accuracy Root-Mean Square (ARMS) at BACK Position97-92% of O26.63 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage SPO2 Accuracy Root-Mean Square (ARMS) at BACK Position92-85% of O26.57 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage SPO2 Accuracy Root-Mean Square (ARMS) at BACK Position84-78% of O25.06 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage SPO2 Accuracy Root-Mean Square (ARMS) at BACK Position77-70% of O23.96 percentage of ARMS (accuracy)
Primary

Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Forehead Position

For each desaturation plateau, SpO2 ARMS are calculated by the difference between the investigational pulse oximeter SpO2 measurement and the co-oximeter's SaO2 measurements

Time frame: 1 day

Population: Subjects respecting the study conditions (no movements, good irrigation) 3 subjects excluded from analysis(no respect of study conditions) so 10 subjects analyzed

ArmMeasureGroupValue (NUMBER)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage SPO2 Accuracy Root-Mean Square (ARMS) at Forehead PositionOverall range 70-100% of O23.91 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage SPO2 Accuracy Root-Mean Square (ARMS) at Forehead Position97-100% of O23.08 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage SPO2 Accuracy Root-Mean Square (ARMS) at Forehead Position92-85% of O22.85 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage SPO2 Accuracy Root-Mean Square (ARMS) at Forehead Position84-78% of O23.48 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage SPO2 Accuracy Root-Mean Square (ARMS) at Forehead Position77-70% of O26.21 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage SPO2 Accuracy Root-Mean Square (ARMS) at Forehead Position97-92% of O23.36 percentage of ARMS (accuracy)
Primary

Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Torso Position

For each desaturation plateau, SpO2 ARMS (Accuracy Root-Mean Square) are calculated by the difference between the investigational pulse oximeter SpO2 measurement and the co-oximeter's SaO2 measurements

Time frame: 1 day

Population: Subjects respecting the study conditions (no movements, good irrigation) 3 subjects excluded from analysis(no respect of study conditions) so 10 subjects analyzed

ArmMeasureGroupValue (NUMBER)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage SPO2 Accuracy Root-Mean Square (ARMS) at Torso Position92-85% of O29.91 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage SPO2 Accuracy Root-Mean Square (ARMS) at Torso Position84-78% of O27.31 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage SPO2 Accuracy Root-Mean Square (ARMS) at Torso PositionOverall range 70-100% of O211.14 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage SPO2 Accuracy Root-Mean Square (ARMS) at Torso Position97-100% of O212.99 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage SPO2 Accuracy Root-Mean Square (ARMS) at Torso Position97-92% of O212.51 percentage of ARMS (accuracy)
EOXY Device and Gold Standard Oximeter and SaO2 MeasuresPercentage SPO2 Accuracy Root-Mean Square (ARMS) at Torso Position77-70% of O210.76 percentage of ARMS (accuracy)