Clinical Study of a Single-Use, Point-of-Care Molecular Diagnostic Device for the Detection of Neisseria Gonorrhoeae (NG), Trichomonas Vaginalis (TV), and Chlamydia Trachomatis (CT) Utilizing Vaginal Swabs

Clinical Study of a Single-Use, Point-of-Care Molecular Diagnostic Device for the Detection of Neisseria Gonorrhoeae (NG), Trichomonas Vaginalis (TV), and Chlamydia Trachomatis (CT) Utilizing Vaginal Swabs

Status
Terminated
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03596151
Enrollment
295
Registered
2018-07-23
Start date
2018-07-09
Completion date
2018-08-20
Last updated
2021-08-24

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Chlamydial Infection, Gonorrhoea, Trichomoniasis

Keywords

Chlamydia trachomatis, device, diagnostic, molecular, Neisseria gonorrhoeae, Trichomonas vaginalis

Brief summary

This is a multi-center study with a minimum of three sites in the United States. The study will enroll approximately 1750 female subjects and will have a study duration of approximately 9 months after enrollment of the first subject. Female subjects seen at the participating sites for any reason will be evaluated for enrollment in this study. All subjects will be managed per standard of care as applicable. Subjects who are enrolled in the study will perform self-collection of a vaginal swab to be tested by Click device, and allow the health care provider (HCP) to collect three additional vaginal swabs to be tested by recognized FDA-cleared comparator methods. Subjects will complete the study in a single visit. The primary objective is to assess the performance of the Click device for detection of Chlamydia trachomatis (CT), Neisseria gonorrhoeae (NG), and Trichomonas vaginalis (TV) in self-collected vaginal specimens as compared to Patient Infected Status (PIS) determined by three approved comparator assays using vaginal specimens collected by a qualified HCP in support of obtaining FDA clearance.

Detailed description

This study is a multi-center study with a minimum of three Clinical Laboratory Improvement Amendments-waived (CLIA-waived) intended operator sites in the United States; approximately one-third of the total number of sites will have a patient population of low Neisseria gonorrhoeae (NG), Chlamydia trachomatis (CT), or Trichomonas vaginalis (TV) prevalence. Low prevalence sites are defined as sites with a prevalence \< / = 2% for any of the three targets. Sites with a prevalence higher than 2% for any of the three targets will be defined as high prevalence sites. The study will enroll approximately 1750 female subjects and will have a study duration of approximately 9 months after enrollment of the first subject. Approximately one-third of the total subjects shall be recruited from low prevalence sites for the infections of interest. Female subjects seen at the participating sites for any reason will be evaluated for enrollment in this study. All subjects will be managed per standard of care as applicable. Subjects who are enrolled in the study will perform self-collection of a vaginal swab to be tested by Click device, and allow the health care provider (HCP) to collect three additional vaginal swabs to be tested by recognized FDA-cleared comparator methods. Subjects will complete the study in a single visit. The hypothesis of this study is that the identification of each organism (CT, NG, and TV) in self-collected vaginal swabs by women using the Click device will agree with the Patient Infected Status (PIS) with a high sensitivity and specificity. The primary objective is to assess the performance of the Click device for detection of CT, NG, and TV in self-collected vaginal specimens as compared to PIS determined by three approved comparator assays using vaginal specimens collected by a qualified HCP in support of obtaining FDA clearance. The secondary objectives are: 1) to assess the performance of the Click device for detection of CT, NG, and TV in self-collected vaginal specimens among symptomatic subjects as compared to PIS determined by three approved comparator assays using vaginal specimens collected by a qualified HCP; 2) to assess the performance of the Click device for detection of CT, NG, and TV in self-collected vaginal specimens among asymptomatic subjects as compared to PIS determined by three approved comparator assays using vaginal specimens collected by a qualified HCP.

Interventions

A single-use device containing a polymerase chain reaction (PCR)-based nucleic acid amplification test (NAAT) for detection of deoxyribonucleic acid (DNA) from Chlamydia trachomatis (CT), Neisseria gonorrhoeae (NG), and Trichomonas vaginalis (TV) in self-collected vaginal specimens.

Sponsors

National Institute of Allergy and Infectious Diseases (NIAID)
Lead SponsorNIH

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
FEMALE
Age
14 Years to 99 Years
Healthy volunteers
Yes

Inclusion criteria

1. Willing and able to give voluntary written informed consent (or the parent/legal guardian will provide parental permission) before any study-related procedure is performed. 2. Female at birth. (Pregnant and breastfeeding women are eligible) 3. Age \> / = 14 years at the time of enrollment. 4. Able to read and understand the procedural information provided for the study. 5. Able and willing to follow all study procedures, including performing self-collection of one vaginal swab and permitting a licensed health care provider to collect three additional vaginal swabs.

Exclusion criteria

1. Have a medical condition, serious intercurrent illness, or other circumstance that, in the Investigator's judgment, could jeopardize the subject's safety, or could interfere with study procedures. 2. Enrollment in this study previously.

Design outcomes

Primary

MeasureTime frameDescription
The Percent Sensitivity and Percent Specificity of Click Device for Detection of Chlamydia Trachomatis (CT) in Self-collected Vaginal Specimens as Compared to Patient Infection Status (PIS) Using Vaginal Specimens Collected by a Qualified HCPDay 1The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
The Percent Sensitivity and Percent Specificity of Click Device for Detection of Neisseria Gonorrhoeae (NG) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCPDay 1The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
The Percent Sensitivity and Percent Specificity of Click Device for Detection of Trichomonas Vaginalis (TV) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCPDay 1The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic ParticipantsDay 1The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic ParticipantsDay 1The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic ParticipantsDay 1The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic ParticipantsDay 1The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants.Day 1The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants.Day 1The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)

Countries

United States

Participant flow

Participants by arm

ArmCount
Click Device
The study population was comprised of female participants \>=14 years of age who visited any of the participating clinics for any reason, who performed self-collection with a vaginal swab for the Click device, and allowed the health care professional (HCP) to collect three additional vaginal swabs for the comparator methods.
295
Total295

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyWithdrawal by Subject1

Baseline characteristics

CharacteristicClick Device
Age, Categorical
<=18 years
7 Participants
Age, Categorical
>=65 years
2 Participants
Age, Categorical
Between 18 and 65 years
286 Participants
Age, Continuous36.2 years
STANDARD_DEVIATION 12.7
Ethnicity (NIH/OMB)
Hispanic or Latino
24 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
194 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
77 Participants
Evaluable for CT
Asymptomatic
107 Participants
Evaluable for CT
Symptomatic
149 Participants
Evaluable for NG
Asymptomatic
106 Participants
Evaluable for NG
Symptomatic
149 Participants
Evaluable for TV
Asymptomatic
107 Participants
Evaluable for TV
Symptomatic
148 Participants
Race (NIH/OMB)
American Indian or Alaska Native
1 Participants
Race (NIH/OMB)
Asian
13 Participants
Race (NIH/OMB)
Black or African American
151 Participants
Race (NIH/OMB)
More than one race
23 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
2 Participants
Race (NIH/OMB)
Unknown or Not Reported
25 Participants
Race (NIH/OMB)
White
80 Participants
Region of Enrollment
United States
295 participants
Sex: Female, Male
Female
295 Participants
Sex: Female, Male
Male
0 Participants
Symptomatic Status
Asymptomatic
127 Participants
Symptomatic Status
Symptomatic
168 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 294
other
Total, other adverse events
0 / 294
serious
Total, serious adverse events
0 / 294

Outcome results

Primary

The Percent Sensitivity and Percent Specificity of Click Device for Detection of Chlamydia Trachomatis (CT) in Self-collected Vaginal Specimens as Compared to Patient Infection Status (PIS) Using Vaginal Specimens Collected by a Qualified HCP

The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)

Time frame: Day 1

Population: This population includes all participants with samples that provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for CT. An evaluable result is indicated if both the Click device and the PIS provide a valid positive or negative result.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Click DeviceThe Percent Sensitivity and Percent Specificity of Click Device for Detection of Chlamydia Trachomatis (CT) in Self-collected Vaginal Specimens as Compared to Patient Infection Status (PIS) Using Vaginal Specimens Collected by a Qualified HCPSensitivity10 Participants
Click DeviceThe Percent Sensitivity and Percent Specificity of Click Device for Detection of Chlamydia Trachomatis (CT) in Self-collected Vaginal Specimens as Compared to Patient Infection Status (PIS) Using Vaginal Specimens Collected by a Qualified HCPSpecificity241 Participants
Primary

The Percent Sensitivity and Percent Specificity of Click Device for Detection of Neisseria Gonorrhoeae (NG) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP

The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)

Time frame: Day 1

Population: This population includes all participants with samples that provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for NG. An evaluable result is indicated if both the Click device and the PIS provide a valid positive or negative result.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Click DeviceThe Percent Sensitivity and Percent Specificity of Click Device for Detection of Neisseria Gonorrhoeae (NG) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCPSensitivity0 Participants
Click DeviceThe Percent Sensitivity and Percent Specificity of Click Device for Detection of Neisseria Gonorrhoeae (NG) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCPSpecificity249 Participants
Primary

The Percent Sensitivity and Percent Specificity of Click Device for Detection of Trichomonas Vaginalis (TV) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP

The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)

Time frame: Day 1

Population: This population includes all participants with samples that provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for TV. An evaluable result is indicated if both the Click device and the PIS provide a valid positive or negative result.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Click DeviceThe Percent Sensitivity and Percent Specificity of Click Device for Detection of Trichomonas Vaginalis (TV) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCPSensitivity29 Participants
Click DeviceThe Percent Sensitivity and Percent Specificity of Click Device for Detection of Trichomonas Vaginalis (TV) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCPSpecificity212 Participants
Primary

The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants

The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)

Time frame: Day 1

Population: This population includes all asymptomatic participants with samples that provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for CT. An evaluable result is indicated if both the Click device and the PIS provide a valid positive or negative result. If a participant reported Yes to any of the symptoms on the Clinical Symptoms Assessment form, the participant is classified as Symptomatic. Otherwise, the participant is classified as Asymptomatic.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Click DeviceThe Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic ParticipantsSensitivity4 Participants
Click DeviceThe Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic ParticipantsSpecificity102 Participants
Primary

The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic Participants

The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)

Time frame: Day 1

Population: This population includes all symptomatic participants with samples that provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for CT. An evaluable result is indicated if both the Click device and the PIS provide a valid positive or negative result. If a participant reported Yes to any of the symptoms on the Clinical Symptoms Assessment form, the participant is classified as Symptomatic. Otherwise, the participant is classified as Asymptomatic.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Click DeviceThe Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic ParticipantsSensitivity6 Participants
Click DeviceThe Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic ParticipantsSpecificity139 Participants
Primary

The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants.

The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)

Time frame: Day 1

Population: This population includes all asymptomatic participants with samples that provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for NG. An evaluable result is indicated if both the Click device and the PIS provide a valid positive or negative result. If a participant reported Yes to any of the symptoms on the Clinical Symptoms Assessment form, the participant is classified as Symptomatic. Otherwise, the participant is classified as Asymptomatic.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Click DeviceThe Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants.Sensitivity0 Participants
Click DeviceThe Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants.Specificity104 Participants
Primary

The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic Participants

The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)

Time frame: Day 1

Population: This population includes all symptomatic participants with samples that provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for NG. An evaluable result is indicated if both the Click device and the PIS provide a valid positive or negative result. If a participant reported Yes to any of the symptoms on the Clinical Symptoms Assessment form, the participant is classified as Symptomatic. Otherwise, the participant is classified as Asymptomatic.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Click DeviceThe Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic ParticipantsSensitivity0 Participants
Click DeviceThe Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic ParticipantsSpecificity145 Participants
Primary

The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants.

The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)

Time frame: Day 1

Population: This population includes all asymptomatic participants with samples that provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for TV. An evaluable result is indicated if both the Click device and the PIS provide a valid positive or negative result. If a participant reported Yes to any of the symptoms on the Clinical Symptoms Assessment form, the participant is classified as Symptomatic. Otherwise, the participant is classified as Asymptomatic.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Click DeviceThe Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants.Sensitivity12 Participants
Click DeviceThe Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants.Specificity88 Participants
Primary

The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic Participants

The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)

Time frame: Day 1

Population: This population includes all symptomatic participants with samples that provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for TV. An evaluable result is indicated if both the Click device and the PIS provide a valid positive or negative result. If a participant reported Yes to any of the symptoms on the Clinical Symptoms Assessment form, the participant is classified as Symptomatic. Otherwise, the participant is classified as Asymptomatic.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Click DeviceThe Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic ParticipantsSensitivity17 Participants
Click DeviceThe Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic ParticipantsSpecificity124 Participants