Chlamydial Infection, Gonorrhoea, Trichomoniasis
Conditions
Keywords
Chlamydia trachomatis, device, diagnostic, molecular, Neisseria gonorrhoeae, Trichomonas vaginalis
Brief summary
This is a multi-center study with a minimum of three sites in the United States. The study will enroll approximately 1750 female subjects and will have a study duration of approximately 9 months after enrollment of the first subject. Female subjects seen at the participating sites for any reason will be evaluated for enrollment in this study. All subjects will be managed per standard of care as applicable. Subjects who are enrolled in the study will perform self-collection of a vaginal swab to be tested by Click device, and allow the health care provider (HCP) to collect three additional vaginal swabs to be tested by recognized FDA-cleared comparator methods. Subjects will complete the study in a single visit. The primary objective is to assess the performance of the Click device for detection of Chlamydia trachomatis (CT), Neisseria gonorrhoeae (NG), and Trichomonas vaginalis (TV) in self-collected vaginal specimens as compared to Patient Infected Status (PIS) determined by three approved comparator assays using vaginal specimens collected by a qualified HCP in support of obtaining FDA clearance.
Detailed description
This study is a multi-center study with a minimum of three Clinical Laboratory Improvement Amendments-waived (CLIA-waived) intended operator sites in the United States; approximately one-third of the total number of sites will have a patient population of low Neisseria gonorrhoeae (NG), Chlamydia trachomatis (CT), or Trichomonas vaginalis (TV) prevalence. Low prevalence sites are defined as sites with a prevalence \< / = 2% for any of the three targets. Sites with a prevalence higher than 2% for any of the three targets will be defined as high prevalence sites. The study will enroll approximately 1750 female subjects and will have a study duration of approximately 9 months after enrollment of the first subject. Approximately one-third of the total subjects shall be recruited from low prevalence sites for the infections of interest. Female subjects seen at the participating sites for any reason will be evaluated for enrollment in this study. All subjects will be managed per standard of care as applicable. Subjects who are enrolled in the study will perform self-collection of a vaginal swab to be tested by Click device, and allow the health care provider (HCP) to collect three additional vaginal swabs to be tested by recognized FDA-cleared comparator methods. Subjects will complete the study in a single visit. The hypothesis of this study is that the identification of each organism (CT, NG, and TV) in self-collected vaginal swabs by women using the Click device will agree with the Patient Infected Status (PIS) with a high sensitivity and specificity. The primary objective is to assess the performance of the Click device for detection of CT, NG, and TV in self-collected vaginal specimens as compared to PIS determined by three approved comparator assays using vaginal specimens collected by a qualified HCP in support of obtaining FDA clearance. The secondary objectives are: 1) to assess the performance of the Click device for detection of CT, NG, and TV in self-collected vaginal specimens among symptomatic subjects as compared to PIS determined by three approved comparator assays using vaginal specimens collected by a qualified HCP; 2) to assess the performance of the Click device for detection of CT, NG, and TV in self-collected vaginal specimens among asymptomatic subjects as compared to PIS determined by three approved comparator assays using vaginal specimens collected by a qualified HCP.
Interventions
A single-use device containing a polymerase chain reaction (PCR)-based nucleic acid amplification test (NAAT) for detection of deoxyribonucleic acid (DNA) from Chlamydia trachomatis (CT), Neisseria gonorrhoeae (NG), and Trichomonas vaginalis (TV) in self-collected vaginal specimens.
Sponsors
Study design
Eligibility
Inclusion criteria
1. Willing and able to give voluntary written informed consent (or the parent/legal guardian will provide parental permission) before any study-related procedure is performed. 2. Female at birth. (Pregnant and breastfeeding women are eligible) 3. Age \> / = 14 years at the time of enrollment. 4. Able to read and understand the procedural information provided for the study. 5. Able and willing to follow all study procedures, including performing self-collection of one vaginal swab and permitting a licensed health care provider to collect three additional vaginal swabs.
Exclusion criteria
1. Have a medical condition, serious intercurrent illness, or other circumstance that, in the Investigator's judgment, could jeopardize the subject's safety, or could interfere with study procedures. 2. Enrollment in this study previously.
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| The Percent Sensitivity and Percent Specificity of Click Device for Detection of Chlamydia Trachomatis (CT) in Self-collected Vaginal Specimens as Compared to Patient Infection Status (PIS) Using Vaginal Specimens Collected by a Qualified HCP | Day 1 | The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP) |
| The Percent Sensitivity and Percent Specificity of Click Device for Detection of Neisseria Gonorrhoeae (NG) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP | Day 1 | The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP) |
| The Percent Sensitivity and Percent Specificity of Click Device for Detection of Trichomonas Vaginalis (TV) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP | Day 1 | The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP) |
| The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic Participants | Day 1 | The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP) |
| The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic Participants | Day 1 | The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP) |
| The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic Participants | Day 1 | The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP) |
| The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants | Day 1 | The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP) |
| The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants. | Day 1 | The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP) |
| The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants. | Day 1 | The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP) |
Countries
United States
Participant flow
Participants by arm
| Arm | Count |
|---|---|
| Click Device The study population was comprised of female participants \>=14 years of age who visited any of the participating clinics for any reason, who performed self-collection with a vaginal swab for the Click device, and allowed the health care professional (HCP) to collect three additional vaginal swabs for the comparator methods. | 295 |
| Total | 295 |
Withdrawals & dropouts
| Period | Reason | FG000 |
|---|---|---|
| Overall Study | Withdrawal by Subject | 1 |
Baseline characteristics
| Characteristic | Click Device |
|---|---|
| Age, Categorical <=18 years | 7 Participants |
| Age, Categorical >=65 years | 2 Participants |
| Age, Categorical Between 18 and 65 years | 286 Participants |
| Age, Continuous | 36.2 years STANDARD_DEVIATION 12.7 |
| Ethnicity (NIH/OMB) Hispanic or Latino | 24 Participants |
| Ethnicity (NIH/OMB) Not Hispanic or Latino | 194 Participants |
| Ethnicity (NIH/OMB) Unknown or Not Reported | 77 Participants |
| Evaluable for CT Asymptomatic | 107 Participants |
| Evaluable for CT Symptomatic | 149 Participants |
| Evaluable for NG Asymptomatic | 106 Participants |
| Evaluable for NG Symptomatic | 149 Participants |
| Evaluable for TV Asymptomatic | 107 Participants |
| Evaluable for TV Symptomatic | 148 Participants |
| Race (NIH/OMB) American Indian or Alaska Native | 1 Participants |
| Race (NIH/OMB) Asian | 13 Participants |
| Race (NIH/OMB) Black or African American | 151 Participants |
| Race (NIH/OMB) More than one race | 23 Participants |
| Race (NIH/OMB) Native Hawaiian or Other Pacific Islander | 2 Participants |
| Race (NIH/OMB) Unknown or Not Reported | 25 Participants |
| Race (NIH/OMB) White | 80 Participants |
| Region of Enrollment United States | 295 participants |
| Sex: Female, Male Female | 295 Participants |
| Sex: Female, Male Male | 0 Participants |
| Symptomatic Status Asymptomatic | 127 Participants |
| Symptomatic Status Symptomatic | 168 Participants |
Adverse events
| Event type | EG000 affected / at risk |
|---|---|
| deaths Total, all-cause mortality | 0 / 294 |
| other Total, other adverse events | 0 / 294 |
| serious Total, serious adverse events | 0 / 294 |
Outcome results
The Percent Sensitivity and Percent Specificity of Click Device for Detection of Chlamydia Trachomatis (CT) in Self-collected Vaginal Specimens as Compared to Patient Infection Status (PIS) Using Vaginal Specimens Collected by a Qualified HCP
The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
Time frame: Day 1
Population: This population includes all participants with samples that provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for CT. An evaluable result is indicated if both the Click device and the PIS provide a valid positive or negative result.
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Click Device | The Percent Sensitivity and Percent Specificity of Click Device for Detection of Chlamydia Trachomatis (CT) in Self-collected Vaginal Specimens as Compared to Patient Infection Status (PIS) Using Vaginal Specimens Collected by a Qualified HCP | Sensitivity | 10 Participants |
| Click Device | The Percent Sensitivity and Percent Specificity of Click Device for Detection of Chlamydia Trachomatis (CT) in Self-collected Vaginal Specimens as Compared to Patient Infection Status (PIS) Using Vaginal Specimens Collected by a Qualified HCP | Specificity | 241 Participants |
The Percent Sensitivity and Percent Specificity of Click Device for Detection of Neisseria Gonorrhoeae (NG) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP
The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
Time frame: Day 1
Population: This population includes all participants with samples that provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for NG. An evaluable result is indicated if both the Click device and the PIS provide a valid positive or negative result.
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Click Device | The Percent Sensitivity and Percent Specificity of Click Device for Detection of Neisseria Gonorrhoeae (NG) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP | Sensitivity | 0 Participants |
| Click Device | The Percent Sensitivity and Percent Specificity of Click Device for Detection of Neisseria Gonorrhoeae (NG) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP | Specificity | 249 Participants |
The Percent Sensitivity and Percent Specificity of Click Device for Detection of Trichomonas Vaginalis (TV) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP
The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
Time frame: Day 1
Population: This population includes all participants with samples that provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for TV. An evaluable result is indicated if both the Click device and the PIS provide a valid positive or negative result.
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Click Device | The Percent Sensitivity and Percent Specificity of Click Device for Detection of Trichomonas Vaginalis (TV) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP | Sensitivity | 29 Participants |
| Click Device | The Percent Sensitivity and Percent Specificity of Click Device for Detection of Trichomonas Vaginalis (TV) in Self-collected Vaginal Specimens as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP | Specificity | 212 Participants |
The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants
The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
Time frame: Day 1
Population: This population includes all asymptomatic participants with samples that provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for CT. An evaluable result is indicated if both the Click device and the PIS provide a valid positive or negative result. If a participant reported Yes to any of the symptoms on the Clinical Symptoms Assessment form, the participant is classified as Symptomatic. Otherwise, the participant is classified as Asymptomatic.
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Click Device | The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants | Sensitivity | 4 Participants |
| Click Device | The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants | Specificity | 102 Participants |
The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic Participants
The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
Time frame: Day 1
Population: This population includes all symptomatic participants with samples that provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for CT. An evaluable result is indicated if both the Click device and the PIS provide a valid positive or negative result. If a participant reported Yes to any of the symptoms on the Clinical Symptoms Assessment form, the participant is classified as Symptomatic. Otherwise, the participant is classified as Asymptomatic.
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Click Device | The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic Participants | Sensitivity | 6 Participants |
| Click Device | The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of CT as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic Participants | Specificity | 139 Participants |
The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants.
The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
Time frame: Day 1
Population: This population includes all asymptomatic participants with samples that provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for NG. An evaluable result is indicated if both the Click device and the PIS provide a valid positive or negative result. If a participant reported Yes to any of the symptoms on the Clinical Symptoms Assessment form, the participant is classified as Symptomatic. Otherwise, the participant is classified as Asymptomatic.
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Click Device | The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants. | Sensitivity | 0 Participants |
| Click Device | The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants. | Specificity | 104 Participants |
The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic Participants
The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
Time frame: Day 1
Population: This population includes all symptomatic participants with samples that provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for NG. An evaluable result is indicated if both the Click device and the PIS provide a valid positive or negative result. If a participant reported Yes to any of the symptoms on the Clinical Symptoms Assessment form, the participant is classified as Symptomatic. Otherwise, the participant is classified as Asymptomatic.
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Click Device | The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic Participants | Sensitivity | 0 Participants |
| Click Device | The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of NG as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic Participants | Specificity | 145 Participants |
The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants.
The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
Time frame: Day 1
Population: This population includes all asymptomatic participants with samples that provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for TV. An evaluable result is indicated if both the Click device and the PIS provide a valid positive or negative result. If a participant reported Yes to any of the symptoms on the Clinical Symptoms Assessment form, the participant is classified as Symptomatic. Otherwise, the participant is classified as Asymptomatic.
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Click Device | The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants. | Sensitivity | 12 Participants |
| Click Device | The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Asymptomatic Participants. | Specificity | 88 Participants |
The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic Participants
The sensitivity of the Click device is defined as follows: TP = the number of specimens where both the Click result and the PIS designation are Positive; FN = the number of specimens where the Click result is Negative and the PIS designation is Positive; Sensitivity = 100 \* TP / (TP + FN) The specificity of the Click device is defined as follows: TN = the number of specimens where both the Click result and the PIS designation are Negative; FP = the number of specimens where the Click result is Positive and the PIS designation is Negative; Specificity = 100 \* TN / (TN + FP)
Time frame: Day 1
Population: This population includes all symptomatic participants with samples that provide an evaluable Click Diagnostics Sexual Health Test result and an evaluable PIS result for TV. An evaluable result is indicated if both the Click device and the PIS provide a valid positive or negative result. If a participant reported Yes to any of the symptoms on the Clinical Symptoms Assessment form, the participant is classified as Symptomatic. Otherwise, the participant is classified as Asymptomatic.
| Arm | Measure | Group | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| Click Device | The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic Participants | Sensitivity | 17 Participants |
| Click Device | The Percent Sensitivity and Percent Specificity of Click Device Using Self-collected Vaginal Swabs for Detection of TV as Compared to PIS Using Vaginal Specimens Collected by a Qualified HCP Among Symptomatic Participants | Specificity | 124 Participants |