Comparison of Leakage With Nasal Prongs and Nasal Mask Interface in Newborns Receiving CPAP Treatment

Trial of NCPAP Interface Leakage (ToNIL): A Randomised Crossover Comparison of Leakage With Nasal Prongs and Nasal Mask Interface in Newborn Infants Treated With CPAP

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03586856
Acronym
ToNIL
Enrollment
50
Registered
2018-07-16
Start date
2018-08-01
Completion date
2019-10-26
Last updated
2019-11-13

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Infant,Premature, Respiration; Insufficient or Poor, Newborn

Keywords

Continuous positive airway pressure, Air Leakage, Work of breathing

Brief summary

The study is a two-armed randomized cross-over comparison of leakage with nasal prongs and nasal mask interface in newborn infants treated with CPAP, born after 28 weeks of gestational age. For infants with an interface leakage, the trial also includes an observational part evaluating simple measures to reduce leakage. The study will be carried out in the Karolinska University Hospital Stockholm and in the Östersund Hospital.

Detailed description

The study will recruit a total of 50 infants receiving CPAP for respiratory support at the Karolinska University Hospital and the Östersund Hospital. The infant should be stable and not in major respiratory distress. Consent from the parents will be obtained before enrolment. The included infants should have stable spontaneous breathing and be older than 28 weeks (corrected age) at the time of enrolment. The reason for respiratory failure and CPAP treatment is not relevant for study purposes. If an infant cannot participate in the study at a given time, he or she can be enrolled later. A child can only participate once. The primary outcome is leakage. This is recorded using equipment for measuring air flow during CPAP treatment. Measurement of leakage is first performed for one interface and then the next (cross-over). The interface order is randomized. The leakage will be measured after a few minutes of stabilization (part 1). If there is leakage, a few simple measures to reduce the leakage will be evaluated (part 2). The second part will approximately add 10-20 minutes and includes simple measures such as adjusting the nasal mask, closing the mouth, changing the position or changing the size of nasal mask/prongs. Part 1 and part 2 are then repeated for the next interface. If the child shows signs of distress or agitation the measurement will be paused or stopped. Part 1: The CPAP system with the randomized interface (nasal mask or nasal prongs) is applied and adjusted by experienced NICU staff. The staff is blinded for the outcome variable, leakage. The choice of size and other adjustments are guided by clinical experience and not by protocol. The NICU staff is not allowed to participate in the explorative part where measures to reduce leakage are evaluated. This precaution is to avoid staff learning what adjustments affect leakage and influence care of infants enrolled at a later stage. Part 2: In this part the investigators can evaluate the effect of simple measures to minimize the leakage. The level of leakage is displayed on a screen. Examples of measures are; closing the mouth of the infant, adjusting straps, adjusting position and changing the size of the interface. Not every child will undergo all measures to reduce leakage, since the aim of this explorative part is to explore ways to reduce leakage. It is estimated that these tests and measures will take 15-60 minutes per child. No other measurements or follow-up is planned. No blood samples will be collected or analysed. The study will collect information on the infant and the birth. This includes birth weight, gestational age at birth, gender, previous respiratory support and medical history. The main outcome variable is absolute leakage (via mouth or via CPAP interface). Other collected variables are oxygen saturation, respiratory rate, inspired oxygen concentration, apneas, heart rate, if the infant is awake/asleep or in distress. Safety variables include skin irritation or damage of the nose caused by the interface, instability or deterioration of respiration or circulation, problems with the equipment or other adverse events. The investigation can be stopped immediately in the case of distress or instability. The study is, as far as the investigators know, the first one of its kind. The air flow meters do not affect the CPAP system and use a technique that measures the total leakage in L/min. These improvements have been emphasized in previous studies. The study follows GCP (good clinical practice) standards and The Helsinki declaration.

Interventions

DEVICENasal mask interface

Nasal masks is a standard interface used for CPAP treatment of newborn infants. The CPAP device allows for easily switching between the nasal prongs or nasal mask interface as well as adjustment of size.

DEVICENasal prongs interface

Nasal masks is a standard interface used for CPAP treatment of newborn infants. The CPAP device allows for easily switching between the nasal prongs or nasal mask interface as well as adjustment of size.

Sponsors

Baldvin Jonsson
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
TREATMENT
Masking
NONE

Intervention model description

Main study design: Two armed randomized cross-over comparison of leakage for two interfaces. The trial also investigates simple measures to reduce leakage (observational)

Eligibility

Sex/Gender
ALL
Age
No minimum to 44 Weeks
Healthy volunteers
No

Inclusion criteria

* Gestational age over 28 weeks (corrected age) at the time of inclusion * Patient receiving respiratory support with CPAP

Exclusion criteria

* Airway, pulmonary, cardiac or gastro-intestinal malformations * Neuromuscular disease * Circulatory unstable patient * FiO2 higher than 0.5 * Nasal damage or major skin irritation. * Extubation less than 24 h before investigation * Surgery less than 5 days before investigation

Design outcomes

Primary

MeasureTime frameDescription
Leakage in the CPAP system0-60 minMeasurement of absolute leakage (at interface or mouth) using flowmeters connected to the CPAP device in flow through position (L/min, 30 seconds recording)

Secondary

MeasureTime frameDescription
Inspired oxygen level (FiO2)0-60 minFiO2 level and adjustments needed to maintain pre-intervention baseline peripheral oxygen saturation (SpO2 target range according to local guidelines)
Level of CPAP support (cmH2O)0-60 minCPAP level and adjustments needed to maintain pre-intervention baseline peripheral oxygen saturation (SpO2 target range according to local guidelines)
Peripheral oxygen saturation (SpO2)0-60 minDeviations from baseline and changes requiring adjustment of respiratory support (Outcome 2 and 3)
Measurement completion possible (yes/no)0-60 minThe number of infants (ratio of total included patients) where measurement of leakage could be determined (on first or more attempts)
Effect of simple measures to reduce leakage (L/min)0-60 minLowest level of recorded level of leakage obtainable using simple measures to reduce leakage, guided by leakage monitoring equipment (L/min and measures used to achieve reduction)

Other

MeasureTime frameDescription
Safety variable: Adverse events0 to competition of measurement (maximum 3 hours)Problems related to trial or equipment used, classified as adverse events according to GCP (Good Clinical Practice)
Safety variable: Desaturation during measurement0 to competition of measurement (maximum 3 hours)Desaturation (SpO2) from baseline requiring treatment or adjustment, during manipulation or leakage measurement.
Safety variable: Problems with equipment0 to competition of measurement (maximum 3 hours)Problems related to study equipment
Safety variable: Adherence to protocol0 to competition of measurement (maximum 3 hours)Problems related to following study protocol
Safety variable: Apneas during measurement0 to competition of measurement (maximum 3 hours)Apneas requiring treatment or adjustment, during manipulation or leakage measurement.
Safety variable: FiO2 adjustments during measurement0 to competition of measurement (maximum 3 hours)FiO2 adjustments required, during manipulation or leakage measurement.
Safety variable: Circulatory instability0 to competition of measurement (maximum 3 hours)Circulatory instability (bradycardia or hypotension) requiring treatment or adjustment, during manipulation or leakage measurement.
Safety variable: Nasal irritation0 to competition of measurement (maximum 3 hours)Nasal irritation requiring treatment or attention, during or after manipulation and leakage measurement.
Safety variable: Pneumothorax0 to competition of measurement (maximum 3 hours)Confirmed pneumothorax or air-leak, during manipulation or leakage measurement.