Skip to content

Carotid Artery Implant for Trapping Upstream Emboli for Preventing Stroke in Atrial Fibrillation Patients

Carotid Artery Implant for Trapping Upstream Emboli for Preventing Stroke in Atrial Fibrillation Patients

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03571789
Acronym
CAPTURE
Enrollment
25
Registered
2018-06-28
Start date
2017-09-12
Completion date
2023-11-28
Last updated
2025-02-20

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Atrial Fibrillation

Keywords

AF

Brief summary

Vine™ is a permanent carotid filter designed to provide protection against embolic stroke in people with atrial fibrillation. It is implanted bilaterally in the common carotid arteries from a thin needle under ultrasound guidance. The procedure is performed without general anesthesia and takes minutes. The safety, feasibility and tolerability of Vine™ will be evaluated. Patients who are eligible will receive Vine™ and will be followed-up for a year after device implantation.

Detailed description

CAPTURE is a prospective, single-arm, multicenter trial, designed to evaluate the safety, feasibility and tolerability of the Vine™ permanent carotid filter. The trial will enroll up to 30 patients, and will be conducted in up to five sites worldwide. All enrolled patients will be invited for non-invasive follow-up tests and clinical evaluations according to the study protocol.

Interventions

DEVICEVine™

The Vine™ is made of a super-elastic nitinol wire comprises a helix that resides within the CCA lumen and a linear stem that traverses the CCA wall. The helix includes three segments: supporting coils, filter, and leading coils. The device is available in sizes of 6.0mm - 10.0mm (in 0.5mm intervals) to accommodate individual CCA diameters

Sponsors

Javelin Medical
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
PREVENTION
Masking
NONE

Intervention model description

Multicenter, prospective, non-randomized, open-label, first-in-human (FIH) study

Eligibility

Sex/Gender
ALL
Age
50 Years to No maximum
Healthy volunteers
No

Inclusion criteria

1. Atrial fibrillation (AF): documented persistent or permanent 2. CHA2DS2-VASc score ≥ 4 3. Age \> 50 4. Unsuitable for oral anticoagulation therapy (OAC), defined as contraindicated for OAC, patient refusing OAC, or physician is reluctant to prescribe OAC 5. Maximal (systolic) CCA diameter range: ≥ 5.3mm and ≤ 9.8mm 6. CCA accessibility: up to 60mm from skin to CCA center, safe approach 7. Patient is willing to provide informed consent 8. Patient is willing to complete all scheduled follow-up

Exclusion criteria

1. Evidence of carotid stenosis \> 30% \[CCA, internal carotid artery (ICA), or external carotid artery (ECA) 2. Evidence of any atherosclerotic disease in CCA above the clavicles 3. Evidence of carotid dissection 4. Pre-existing stent(s) in CCA 5. Contraindicated or allergic to antiplatelet therapy, or any medication required during the study 6. Recent stroke, TIA, or myocardial infarction (MI) within two months prior to index procedure 7. Female who is pregnant or who is planning to become pregnant during the course of the study 8. Life expectancy of less than 1 year 9. Active systemic infection 10. Known sensitivity to nickel or titanium metals, or their alloys 11. Known hereditary or acquired coagulation disorders 12. Any planned surgical or endovascular procedure within 14 days prior to or 30 days after the index procedure 13. A co-morbid disease or condition that could confound the neurological and functional evaluations or compromise survival or ability to complete follow-up assessments 14. Current use or a recent history of illicit drug(s) use or alcohol abuse (defined as regular or daily consumption of more than four alcoholic drinks per day) 15. Active participation in another investigational drug or device treatment study 16. Any other condition that in the opinion of the investigator may adversely affect the safety of the subject or would limit the subject's ability to complete the study

Design outcomes

Primary

MeasureTime frameDescription
Number of patients with no incidence of device and/or procedure related Major Adverse Events (MAEs)30 days from implantation procedureMajor Adverse Events (MAEs) are defined as: * Death * Major and minor strokes * Major bleeding * Common carotid artery (CCA) stenosis \> 70% * Vine™ migration * CCA thrombus * Any complications in the CCA requiring endovascular treatment or surgery
Number of patients with Procedure Success30 days from implantation procedureProcedure Success is defined as Proper Vine™ Position in each CCA Proper Vine™ Position is defined as: * Supporting coil in contact with artery walls * No migration * No fracture * No Vine™ coils or portions thereof visible outside the arterial lumen * No entangled or overlapping coils

Secondary

MeasureTime frameDescription
Number of patients with no incidence of device and/or procedure related Major Adverse Events (MAEs)within 3, 6, months,1, 2, 3, 4 and 5 years of implantation procedureDeath * Major and minor strokes * Major bleeding * Common carotid artery (CCA) stenosis \> 70% * Vine™ migration * CCA thrombus * Any complications in the CCA requiring endovascular treatment or surgery
Number of Successful Delivery and Deployment Attemptswithin 4 hours of implantation procedureSuccessful Delivery and Deployment Attempts is defined as Proper Vine™ Position
Number of patients with properly positioned Vine™ in each CCAwithin 3, 6, months,1, 2, 3, 4 and 5 years of implantation procedureProper implant position

Countries

Belgium, Czechia, Netherlands

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 21, 2026