A Comparison of Capnography Sampling Lines

A Comparison of Capnography Sampling Lines

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03554629
Acronym
FRS
Enrollment
30
Registered
2018-06-13
Start date
2018-05-01
Completion date
2020-09-15
Last updated
2021-01-06

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Capnography

Brief summary

This study evaluates the performance of eight (8) cleared capnography carbon dioxide (CO2) sampling filterlines during patient simulated scripted activities when connected to a Medtronic (MDT) dual parameter monitor.

Detailed description

The eight (8) Capnography CO2 Sampling Filterlines (CCSF) have different designs for obtaining a gas sample for CO2 measurement and also for delivering supplemental oxygen (O2). In this study awake, healthy volunteers will evaluate the 8 CCSF (Capnography CO2 Sampling Filterline) to determine their performance variability in providing a representative exhaled CO2 sample for measurement on the Capnostream 35 dual parameter monitor during scripted activities of closed mouth/open mouth breathing during both high and low respiration rate with supplemental oxygen delivery at 5lpm.

Interventions

DEVICECapnography CO2 cannula sampling Flexiterline (CCSF) designs with supplemental O2

Use of 8 post-market nasal CO2 cannula sampling filterlines; 4 nasal only sampling ports manufactured by Medtronic, Salter, Hudson/Teleflex and FlexiCare) and 4 post-market oral/nasal sampling ports manufactured by Medtronic, Salter, Dispo-Med and Westmed during scripted activity with supplemental O2.

Sponsors

Medtronic - MITG
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

1. Non-hospitalized adults ≥ 18 years old. 2. Willing and able to give informed consent.

Exclusion criteria

1. Lack of an informed consent. 2. Subjects not able to accommodate the proper application of the cannula. 3. Subject not willing or able to comply fully with the study procedures. 4. Subjects with sensitivity to nasal cannula in both nares. 5. Subject with skin allergies to medical adhesives. 6. Subjects with runny nose the day of the study participation. 7. Subject, who in the opinion of the Principal Investigator, should not be enrolled.

Design outcomes

Primary

MeasureTime frameDescription
Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityThree hoursComparison of measured end-tidal CO2 (EtCO2), Rebreathing as measured by fractional inspiratory CO2 (FiCO2) as a function the activity for 8 different CCSF designs tested on 23 subjects with supplemental O2 at 5lmp for four activities. Device data collected a 20Hz (20 data points/second) for the 3 hours of enrollment. Participants lateral lying with head on a pillow with supplemental O2 during each nasal only or open mouth breathing at respiration rate of 6 and 24.

Secondary

MeasureTime frameDescription
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Three hoursGas sample provided by CCSF to the Capnostream 35 for measurement of the partial pressure of CO2 as a function of time. The Capnostream 35 Respiration rate algorithm counts the breaths per minute based upon the frequency of a CO2 validated breath waveform.
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Three hoursGas sample provided by CCSF to the Capnostream 35 for measurement of the partial pressure of CO2 as a function of time. The Capnostream 35 Respiration rate algorithm counts the breaths per minute based upon the frequency of a CO2 validated breath waveform.

Countries

United States

Participant flow

Recruitment details

Execution of CIP v 1 enrollment at 6; data not included in results since data inadequate. Execution of CIP V2 enrollment was 23 One subject withdrawn before participation due to failure to meet failure to meet criteria of enrollment on the study day (runny nose)

Participants by arm

ArmCount
Capnography CO2 Sampling Filterline
Capnography CO2 Sampling Filterline: Use of up to 8 Capnography CO2 Sampling FIlterlines by blinded code name during scripted activities
23
Total23

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyInadequate data6

Baseline characteristics

CharacteristicCapnography CO2 Sampling Filterline
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
5 Participants
Age, Categorical
Between 18 and 65 years
18 Participants
Age, Continuous48.5 years
STANDARD_DEVIATION 17.1
Body Mass Index (BMI)27.5 kg/m^2
STANDARD_DEVIATION 6.6
Ethnicity (NIH/OMB)
Hispanic or Latino
2 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
21 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants
Race (NIH/OMB)
American Indian or Alaska Native
1 Participants
Race (NIH/OMB)
Asian
0 Participants
Race (NIH/OMB)
Black or African American
3 Participants
Race (NIH/OMB)
More than one race
0 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants
Race (NIH/OMB)
White
19 Participants
Region of Enrollment
United States
23 Participants
Sex: Female, Male
Female
14 Participants
Sex: Female, Male
Male
9 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
EG002
affected / at risk
EG003
affected / at risk
EG004
affected / at risk
EG005
affected / at risk
EG006
affected / at risk
EG007
affected / at risk
deaths
Total, all-cause mortality
0 / 290 / 290 / 290 / 290 / 290 / 290 / 290 / 29
other
Total, other adverse events
2 / 291 / 290 / 291 / 292 / 290 / 290 / 291 / 29
serious
Total, serious adverse events
0 / 290 / 290 / 290 / 290 / 290 / 290 / 290 / 29

Outcome results

Primary

Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity

Comparison of measured end-tidal CO2 (EtCO2), Rebreathing as measured by fractional inspiratory CO2 (FiCO2) as a function the activity for 8 different CCSF designs tested on 23 subjects with supplemental O2 at 5lmp for four activities. Device data collected a 20Hz (20 data points/second) for the 3 hours of enrollment. Participants lateral lying with head on a pillow with supplemental O2 during each nasal only or open mouth breathing at respiration rate of 6 and 24.

Time frame: Three hours

Population: Performance of 8 CO2 cannula sampling filterlines (CCSF): 4 filterlines provided nasal only sampling ports and 4 filterlines provided both nasal and oral sampling ports.~Each of the 8 CCSF was assigned a letter by the investigation site to maintain blinding in data analysis.~Side-stream capnography measurement via the Capnostream 35

ArmMeasureGroupValue (MEAN)Dispersion
EtCO2 During Mouth Closed Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivitySalter Lab (Nasal)30.1 mmHgStandard Deviation 5.5
EtCO2 During Mouth Closed Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityHudson RCI /TeleFlex(Nasal)13.5 mmHgStandard Deviation 8.4
EtCO2 During Mouth Closed Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityFlexiCare (Nasal)30.6 mmHgStandard Deviation 7.5
EtCO2 During Mouth Closed Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityMedtronic Filterline (Nasal)32.6 mmHgStandard Deviation 5.4
EtCO2 During Mouth Closed Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityWestMed (Oral/Nasal)33.9 mmHgStandard Deviation 4.8
EtCO2 During Mouth Closed Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityDispo-Med (Oral/Nasal)29.9 mmHgStandard Deviation 7.5
EtCO2 During Mouth Closed Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivitySalter Lab Oral-Trac (Oral/Nasal)29.3 mmHgStandard Deviation 8.7
EtCO2 During Mouth Closed Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityMedtronic Smart Capnoline+(Oral/Nasal)32.6 mmHgStandard Deviation 5.4
FiCO2 During Mouth Closed Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityWestMed (Oral/Nasal)0.5 mmHgStandard Deviation 0.4
FiCO2 During Mouth Closed Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivitySalter Lab (Nasal)0.4 mmHgStandard Deviation 0.8
FiCO2 During Mouth Closed Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityDispo-Med (Oral/Nasal)0.1 mmHgStandard Deviation 0.3
FiCO2 During Mouth Closed Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivitySalter Lab Oral-Trac (Oral/Nasal)0.3 mmHgStandard Deviation 0.5
FiCO2 During Mouth Closed Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityFlexiCare (Nasal)0.0 mmHgStandard Deviation 0.2
FiCO2 During Mouth Closed Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityMedtronic Filterline (Nasal)0.0 mmHgStandard Deviation 0
FiCO2 During Mouth Closed Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityHudson RCI /TeleFlex(Nasal)0.0 mmHgStandard Deviation 0
FiCO2 During Mouth Closed Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityMedtronic Smart Capnoline+(Oral/Nasal)0.0 mmHgStandard Deviation 0
EtCO2 During Mouth Closed Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivitySalter Lab Oral-Trac (Oral/Nasal)25.0 mmHgStandard Deviation 6.9
EtCO2 During Mouth Closed Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityWestMed (Oral/Nasal)29.2 mmHgStandard Deviation 5.8
EtCO2 During Mouth Closed Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityMedtronic Smart Capnoline+(Oral/Nasal)28.8 mmHgStandard Deviation 4.8
EtCO2 During Mouth Closed Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivitySalter Lab (Nasal)24.1 mmHgStandard Deviation 5.9
EtCO2 During Mouth Closed Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityDispo-Med (Oral/Nasal)23.8 mmHgStandard Deviation 8.2
EtCO2 During Mouth Closed Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityMedtronic Filterline (Nasal)28.1 mmHgStandard Deviation 5.2
EtCO2 During Mouth Closed Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityFlexiCare (Nasal)26.6 mmHgStandard Deviation 6.2
EtCO2 During Mouth Closed Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityHudson RCI /TeleFlex(Nasal)12.3 mmHgStandard Deviation 6.9
FiCO2 During Mouth Closed Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityMedtronic Smart Capnoline+(Oral/Nasal)0.0 mmHgStandard Deviation 0.1
FiCO2 During Mouth Closed Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityHudson RCI /TeleFlex(Nasal)0.0 mmHgStandard Deviation 0.1
FiCO2 During Mouth Closed Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityFlexiCare (Nasal)0.2 mmHgStandard Deviation 0.8
FiCO2 During Mouth Closed Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityMedtronic Filterline (Nasal)0.0 mmHgStandard Deviation 0.2
FiCO2 During Mouth Closed Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivitySalter Lab (Nasal)1.4 mmHgStandard Deviation 1.2
FiCO2 During Mouth Closed Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityWestMed (Oral/Nasal)1.4 mmHgStandard Deviation 1.5
FiCO2 During Mouth Closed Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityDispo-Med (Oral/Nasal)1.4 mmHgStandard Deviation 0.9
FiCO2 During Mouth Closed Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivitySalter Lab Oral-Trac (Oral/Nasal)0.6 mmHgStandard Deviation 0.9
EtCO2 During Mouth Open Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityMedtronic Filterline (Nasal)24.8 mmHgStandard Deviation 8.3
EtCO2 During Mouth Open Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivitySalter Lab Oral-Trac (Oral/Nasal)26.8 mmHgStandard Deviation 5.2
EtCO2 During Mouth Open Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivitySalter Lab (Nasal)22.3 mmHgStandard Deviation 9.2
EtCO2 During Mouth Open Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityFlexiCare (Nasal)20.1 mmHgStandard Deviation 12.5
EtCO2 During Mouth Open Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityMedtronic Smart Capnoline+(Oral/Nasal)26.8 mmHgStandard Deviation 7.4
EtCO2 During Mouth Open Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityWestMed (Oral/Nasal)28.0 mmHgStandard Deviation 8.9
EtCO2 During Mouth Open Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityDispo-Med (Oral/Nasal)23.6 mmHgStandard Deviation 6.9
EtCO2 During Mouth Open Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityHudson RCI /TeleFlex(Nasal)7.6 mmHgStandard Deviation 6.6
FiCO2 During Mouth Open Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityMedtronic Smart Capnoline+(Oral/Nasal)0.0 mmHgStandard Deviation 0
FiCO2 During Mouth Open Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityHudson RCI /TeleFlex(Nasal)0.0 mmHgStandard Deviation 0
FiCO2 During Mouth Open Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivitySalter Lab (Nasal)0.2 mmHgStandard Deviation 0.3
FiCO2 During Mouth Open Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityFlexiCare (Nasal)0.1 mmHgStandard Deviation 0.2
FiCO2 During Mouth Open Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivitySalter Lab Oral-Trac (Oral/Nasal)0.2 mmHgStandard Deviation 0.3
FiCO2 During Mouth Open Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityDispo-Med (Oral/Nasal)0.1 mmHgStandard Deviation 0.2
FiCO2 During Mouth Open Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityWestMed (Oral/Nasal)0.5 mmHgStandard Deviation 0.6
FiCO2 During Mouth Open Breathing Rate of 6Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityMedtronic Filterline (Nasal)0.0 mmHgStandard Deviation 0
EtCO2 Mouth Open Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivitySalter Lab (Nasal)19.9 mmHgStandard Deviation 7.6
EtCO2 Mouth Open Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityHudson RCI /TeleFlex(Nasal)7.8 mmHgStandard Deviation 7.2
EtCO2 Mouth Open Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityMedtronic Filterline (Nasal)22.7 mmHgStandard Deviation 7.6
EtCO2 Mouth Open Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityWestMed (Oral/Nasal)24.1 mmHgStandard Deviation 8.1
EtCO2 Mouth Open Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityDispo-Med (Oral/Nasal)19.3 mmHgStandard Deviation 7
EtCO2 Mouth Open Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityMedtronic Smart Capnoline+(Oral/Nasal)24.9 mmHgStandard Deviation 6.2
EtCO2 Mouth Open Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivitySalter Lab Oral-Trac (Oral/Nasal)23.4 mmHgStandard Deviation 4.7
EtCO2 Mouth Open Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityFlexiCare (Nasal)19.1 mmHgStandard Deviation 10.2
FiCO2 Mouth Open Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityMedtronic Smart Capnoline+(Oral/Nasal)0.0 mmHgStandard Deviation 0.1
FiCO2 Mouth Open Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityDispo-Med (Oral/Nasal)1.4 mmHgStandard Deviation 1.3
FiCO2 Mouth Open Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityWestMed (Oral/Nasal)1.2 mmHgStandard Deviation 1.4
FiCO2 Mouth Open Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityMedtronic Filterline (Nasal)0.0 mmHgStandard Deviation 0
FiCO2 Mouth Open Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivitySalter Lab (Nasal)1.0 mmHgStandard Deviation 1.1
FiCO2 Mouth Open Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityHudson RCI /TeleFlex(Nasal)0.0 mmHgStandard Deviation 0
FiCO2 Mouth Open Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivityFlexiCare (Nasal)0.2 mmHgStandard Deviation 0.6
FiCO2 Mouth Open Breathing Rate of 24Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by ActivitySalter Lab Oral-Trac (Oral/Nasal)0.9 mmHgStandard Deviation 1.1
Secondary

Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24

Gas sample provided by CCSF to the Capnostream 35 for measurement of the partial pressure of CO2 as a function of time. The Capnostream 35 Respiration rate algorithm counts the breaths per minute based upon the frequency of a CO2 validated breath waveform.

Time frame: Three hours

Population: Participant actual Respiration Rate was coached using the Pace Breathing Android App: Version 2.1.~RR = 6 : Inspiratory Time was 5.0 seconds and Expiratory time was 5.0 seconds No breath alarm is deployed when both RR and EtCO2 are equal to zero. RR = 24: Inspiratory Time was 1.2 seconds and Expiratory time was 1.3 seconds

ArmMeasureGroupValue (MEAN)Dispersion
EtCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Salter Labs (Nasal)0.02 AlarmsStandard Deviation 0.15
EtCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Hudson/TeleFlex (Nasal)0.20 AlarmsStandard Deviation 0.4
EtCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24FlexiCare (Nasal)0.17 AlarmsStandard Deviation 0.53
EtCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Medtronic CO2 (Nasal)0.02 AlarmsStandard Deviation 0.15
EtCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24WestMed (Oral/Nasal)0.04 AlarmsStandard Deviation 0.29
EtCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24DispoMed (Oral/Nasal)0.04 AlarmsStandard Deviation 0.21
EtCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Salter Oral-Trac (Oral/Nasal)0.02 AlarmsStandard Deviation 0.15
EtCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Medtronic Smart CapnoLine Plus (Oral/Nasal)0.00 AlarmsStandard Deviation 0
FiCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Medtronic CO2 (Nasal)0.02 AlarmsStandard Deviation 0.15
FiCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Medtronic Smart CapnoLine Plus (Oral/Nasal)0.00 AlarmsStandard Deviation 0
FiCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24FlexiCare (Nasal)0.13 AlarmsStandard Deviation 0.4
FiCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Salter Oral-Trac (Oral/Nasal)0.02 AlarmsStandard Deviation 0.15
FiCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24DispoMed (Oral/Nasal)0.04 AlarmsStandard Deviation 0.21
FiCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Salter Labs (Nasal)0.02 AlarmsStandard Deviation 0.15
FiCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Hudson/TeleFlex (Nasal)0.13 AlarmsStandard Deviation 0.34
FiCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24WestMed (Oral/Nasal)0.04 AlarmsStandard Deviation 0.29
EtCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Hudson/TeleFlex (Nasal)0.74 AlarmsStandard Deviation 0.53
EtCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Salter Labs (Nasal)0.37 AlarmsStandard Deviation 0.61
EtCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Medtronic CO2 (Nasal)0.17 AlarmsStandard Deviation 0.38
EtCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Medtronic Smart CapnoLine Plus (Oral/Nasal)0.09 AlarmsStandard Deviation 0.28
EtCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24WestMed (Oral/Nasal)0.13 AlarmsStandard Deviation 0.34
EtCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24DispoMed (Oral/Nasal)0.37 AlarmsStandard Deviation 0.53
EtCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Salter Oral-Trac (Oral/Nasal)0.22 AlarmsStandard Deviation 0.47
EtCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24FlexiCare (Nasal)0.35 AlarmsStandard Deviation 0.57
FiCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Salter Oral-Trac (Oral/Nasal)0.11 AlarmsStandard Deviation 0.31
FiCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24DispoMed (Oral/Nasal)0.28 AlarmsStandard Deviation 0.46
FiCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24WestMed (Oral/Nasal)0.11 AlarmsStandard Deviation 0.31
FiCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24FlexiCare (Nasal)0.30 AlarmsStandard Deviation 0.51
FiCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Salter Labs (Nasal)0.22 AlarmsStandard Deviation 0.47
FiCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Medtronic CO2 (Nasal)0.11 AlarmsStandard Deviation 0.31
FiCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Medtronic Smart CapnoLine Plus (Oral/Nasal)0.09 AlarmsStandard Deviation 0.28
FiCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Hudson/TeleFlex (Nasal)0.63 AlarmsStandard Deviation 0.57
EtCO2 During Mouth Open Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Medtronic Smart CapnoLine Plus (Oral/Nasal)0.04 AlarmsStandard Deviation 0.21
EtCO2 During Mouth Open Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24FlexiCare (Nasal)0.17 AlarmsStandard Deviation 0.38
EtCO2 During Mouth Open Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24WestMed (Oral/Nasal)0.09 AlarmsStandard Deviation 0.28
EtCO2 During Mouth Open Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24DispoMed (Oral/Nasal)0.20 AlarmsStandard Deviation 0.45
EtCO2 During Mouth Open Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Salter Oral-Trac (Oral/Nasal)0.13 AlarmsStandard Deviation 0.4
EtCO2 During Mouth Open Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Hudson/TeleFlex (Nasal)0.52 AlarmsStandard Deviation 0.51
EtCO2 During Mouth Open Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Salter Labs (Nasal)0.13 AlarmsStandard Deviation 0.34
EtCO2 During Mouth Open Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Medtronic CO2 (Nasal)0.13 AlarmsStandard Deviation 0.4
FiCO2 During Mouth Open Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Medtronic Smart CapnoLine Plus (Oral/Nasal)0.04 AlarmsStandard Deviation 0.21
FiCO2 During Mouth Open Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24WestMed (Oral/Nasal)0.09 AlarmsStandard Deviation 0.28
FiCO2 During Mouth Open Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Salter Labs (Nasal)0.04 AlarmsStandard Deviation 0.21
FiCO2 During Mouth Open Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Medtronic CO2 (Nasal)0.07 AlarmsStandard Deviation 0.25
FiCO2 During Mouth Open Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24DispoMed (Oral/Nasal)0.15 AlarmsStandard Deviation 0.36
FiCO2 During Mouth Open Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Hudson/TeleFlex (Nasal)0.50 AlarmsStandard Deviation 0.51
FiCO2 During Mouth Open Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24Salter Oral-Trac (Oral/Nasal)0.07 AlarmsStandard Deviation 0.25
FiCO2 During Mouth Open Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24FlexiCare (Nasal)0.15 AlarmsStandard Deviation 0.36
Secondary

Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6

Gas sample provided by CCSF to the Capnostream 35 for measurement of the partial pressure of CO2 as a function of time. The Capnostream 35 Respiration rate algorithm counts the breaths per minute based upon the frequency of a CO2 validated breath waveform.

Time frame: Three hours

Population: Participant actual Respiration Rate was coached using the Pace Breathing Android App: Version 2.1.~RR = 6 : Inspiratory Time was 5.0 seconds and Expiratory time was 5.0 seconds No breath alarm is deployed when both RR and EtCO2 are equal to zero. RR = 24: Inspiratory Time was 1.2 seconds and Expiratory time was 1.3 seconds

ArmMeasureGroupValue (MEAN)Dispersion
EtCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6FlexiCare (Nasal)0.13 AlarmsStandard Deviation 0.34
EtCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Hudson/TeleFlex (Nasal)0.30 AlarmsStandard Deviation 0.55
EtCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Salter Labs (Nasal)0.11 AlarmsStandard Deviation 0.31
EtCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Medtronic CO2 (Nasal)0.07 AlarmsStandard Deviation 0.25
EtCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6WestMed (Oral/Nasal)0.02 AlarmsStandard Deviation 0.15
EtCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6DispoMed (Oral/Nasal)0.04 AlarmsStandard Deviation 0.21
EtCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Salter Oral-Trac (Oral/Nasal)0.04 AlarmsStandard Deviation 0.21
EtCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Medtronic Smart CapnoLine Plus (Oral/Nasal)0.0 AlarmsStandard Deviation 0
FiCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6DispoMed (Oral/Nasal)0.02 AlarmsStandard Deviation 0.15
FiCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6WestMed (Oral/Nasal)0.02 AlarmsStandard Deviation 0.15
FiCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Hudson/TeleFlex (Nasal)0.24 AlarmsStandard Deviation 0.48
FiCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Medtronic Smart CapnoLine Plus (Oral/Nasal)0.0 AlarmsStandard Deviation 0
FiCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Salter Oral-Trac (Oral/Nasal)0.04 AlarmsStandard Deviation 0.21
FiCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Medtronic CO2 (Nasal)0.07 AlarmsStandard Deviation 0.25
FiCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Salter Labs (Nasal)0.07 AlarmsStandard Deviation 0.25
FiCO2 During Mouth Closed Breathing Rate of 6Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6FlexiCare (Nasal)0.09 AlarmsStandard Deviation 0.28
EtCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Salter Oral-Trac (Oral/Nasal)0.11 AlarmsStandard Deviation 0.31
EtCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Salter Labs (Nasal)0.11 AlarmsStandard Deviation 0.31
EtCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Medtronic CO2 (Nasal)0.13 AlarmsStandard Deviation 0.34
EtCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6WestMed (Oral/Nasal)0.07 AlarmsStandard Deviation 0.33
EtCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6DispoMed (Oral/Nasal)0.11 AlarmsStandard Deviation 0.31
EtCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Medtronic Smart CapnoLine Plus (Oral/Nasal)0.07 AlarmsStandard Deviation 0.25
EtCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6FlexiCare (Nasal)0.24 AlarmsStandard Deviation 0.43
EtCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Hudson/TeleFlex (Nasal)0.54 AlarmsStandard Deviation 0.55
FiCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Salter Labs (Nasal)0.09 AlarmsStandard Deviation 0.28
FiCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Medtronic CO2 (Nasal)0.09 AlarmsStandard Deviation 0.28
FiCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Hudson/TeleFlex (Nasal)0.43 AlarmsStandard Deviation 0.5
FiCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6FlexiCare (Nasal)0.22 AlarmsStandard Deviation 0.42
FiCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6WestMed (Oral/Nasal)0.04 AlarmsStandard Deviation 0.21
FiCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Medtronic Smart CapnoLine Plus (Oral/Nasal)0.07 AlarmsStandard Deviation 0.25
FiCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6Salter Oral-Trac (Oral/Nasal)0.07 AlarmsStandard Deviation 0.25
FiCO2 During Mouth Closed Breathing Rate of 24Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6DispoMed (Oral/Nasal)0.07 AlarmsStandard Deviation 0.25

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026