Effect of High Frequency Chest Wall Oscillation Vests on Spirometry Measurements

Effect of High Frequency Chest Wall Oscillation Vests on Spirometry Measurements

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03534986
Enrollment
32
Registered
2018-05-23
Start date
2018-02-20
Completion date
2018-03-16
Last updated
2018-10-03

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Bronchiectasis, Cystic Fibrosis

Brief summary

The impact of high-frequency chest wall oscillation therapy on spirometry values (Forced Expiratory Volume, Forced Vital Capacity, Peak Expiratory Flow, Forced Expiratory Flow and Tidal Volume is investigated during use of several products and comparing to baseline values.

Detailed description

The study will be broken into three (3) arms: * AffloVest® & The Vest® * AffloVest® & inCourage® * AffloVest® & SmartVest® Subjects will be randomly assigned to an arm. Within each arm, the order of products will be also randomized. Baseline spirometry (Forced Expiratory Volume (FEV1), Forced Vital Capacity (FVC), Peak Expiratory Flow (PEF), Forced Expiratory Flow (FEF25-75%) and Tidal Volume (TV)) will be taken at the beginning, middle and end of each series of measurements with each subject, without any device on the subject. A product (AffloVest or Compressor-based vest) will be placed onto subject and turned ON to the highest frequency and intensity settings. Then the subject will be given a certain period of time to adjust and spirometry measurements will then be repeated, then The product will be removed and the subject will be allowed a recovery period, then the other product will be placed on the subject, turned ON and spirometry measurements will repeated.

Interventions

High-frequency chest wall oscillation vest

DEVICEHill-Rom The Vest

High-frequency chest wall oscillation vest

DEVICERespirtech inCourage

High-frequency chest wall oscillation vest

DEVICEElectromed SmartVest

High-frequency chest wall oscillation vest

Sponsors

International Biophysics Corporation
Lead SponsorINDUSTRY

Study design

Allocation
NON_RANDOMIZED
Intervention model
PARALLEL
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 50 Years
Healthy volunteers
Yes

Inclusion criteria

* Healthy subject, ages 18 - 50

Exclusion criteria

* Non-ambulatory, diagnosed neuromuscular disorder, currently using any type of oscillation vest therapy, diagnosed co-morbid condition (i.e. lung cancer, other lung disorder or disease), currently enrolled in a medical research study, non-English speaking

Design outcomes

Primary

MeasureTime frameDescription
TV30 minutesTidal Volume, the lung volume representing the normal volume of air displaced between normal inhalation and exhalation when extra effort is not applied
PEF30 minutesPeak Expiratory Flow, a person's maximum speed of expiration
FVC30 minutesForced Vital Capacity, the amount of air which can be forcibly exhaled from the lungs after taking the deepest breath possible
FEV130 minutesForced Expiratory Volume, the volume of air that can be forced out in one second after taking a deep breath
FEF25-75%30 minutesForced Expiratory Flow (25-75%), the peak expiratory flow at 25 - 75% FVC

Countries

United States

Participant flow

Participants by arm

ArmCount
AffloVest The Vest Arm
Devices placed on highest intensity / highest frequency International Biophysics AffloVest: High-frequency chest wall oscillation vest Hill-Rom The Vest: High-frequency chest wall oscillation vest
10
AffloVest inCourage Arm
Devices placed on highest intensity / highest frequency International Biophysics AffloVest: High-frequency chest wall oscillation vest Respirtech inCourage: High-frequency chest wall oscillation vest
10
AffloVest SmartVest Arm
Devices placed on highest intensity / highest frequency International Biophysics AffloVest: High-frequency chest wall oscillation vest Electromed SmartVest: High-frequency chest wall oscillation vest
12
Total32

Baseline characteristics

CharacteristicAffloVest inCourage ArmAffloVest SmartVest ArmAffloVest The Vest ArmTotal
Age, Categorical
<=18 years
0 Participants0 Participants0 Participants0 Participants
Age, Categorical
>=65 years
0 Participants0 Participants0 Participants0 Participants
Age, Categorical
Between 18 and 65 years
10 Participants12 Participants10 Participants32 Participants
Age, Continuous30.6 years
STANDARD_DEVIATION 13.3
34.7 years
STANDARD_DEVIATION 10.2
28.4 years
STANDARD_DEVIATION 11.9
31.7 years
STANDARD_DEVIATION 11.7
Race and Ethnicity Not Collected0 Participants
Region of Enrollment
United States
10 participants12 participants10 participants32 participants
Sex: Female, Male
Female
4 Participants7 Participants2 Participants13 Participants
Sex: Female, Male
Male
6 Participants5 Participants8 Participants19 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
EG002
affected / at risk
deaths
Total, all-cause mortality
0 / 100 / 100 / 12
other
Total, other adverse events
0 / 100 / 100 / 12
serious
Total, serious adverse events
0 / 100 / 100 / 12

Outcome results

Primary

FEF25-75%

Forced Expiratory Flow (25-75%), the peak expiratory flow at 25 - 75% FVC

Time frame: 30 minutes

ArmMeasureValue (MEAN)
BaselineFEF25-75%3.71 L/s
AffloVest ArmFEF25-75%3.54 L/s
Compressor ArmFEF25-75%3.19 L/s
Primary

FEV1

Forced Expiratory Volume, the volume of air that can be forced out in one second after taking a deep breath

Time frame: 30 minutes

ArmMeasureValue (MEAN)
BaselineFEV13.51 L
AffloVest ArmFEV13.46 L
Compressor ArmFEV13.30 L
Primary

FVC

Forced Vital Capacity, the amount of air which can be forcibly exhaled from the lungs after taking the deepest breath possible

Time frame: 30 minutes

ArmMeasureValue (MEAN)
BaselineFVC4.29 L
AffloVest ArmFVC4.25 L
Compressor ArmFVC4.12 L
Primary

PEF

Peak Expiratory Flow, a person's maximum speed of expiration

Time frame: 30 minutes

ArmMeasureValue (MEAN)
BaselinePEF8.19 L/s
AffloVest ArmPEF8.28 L/s
Compressor ArmPEF8.13 L/s
Primary

TV

Tidal Volume, the lung volume representing the normal volume of air displaced between normal inhalation and exhalation when extra effort is not applied

Time frame: 30 minutes

ArmMeasureValue (MEAN)
BaselineTV0.93 L
AffloVest ArmTV1.00 L
Compressor ArmTV1.07 L

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026