Skip to content

Enhanced Lung Protective Ventilation With ECCO2R During ARDS

Enhanced Lung Protective Ventilation With Extracorporeal CO2 Removal During Acute Respiratory Distress Syndrome

Status
Terminated
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03525691
Acronym
PROVE
Enrollment
3
Registered
2018-05-16
Start date
2018-05-23
Completion date
2023-12-05
Last updated
2023-12-12

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

ARDS, Human, Ventilator-Induced Lung Injury

Brief summary

Acute Respiratory Distress Syndrome (ARDS) is associated with a mortality rate of 30 - 45 % and required invasive mechanical ventilation (MV) in almost 85 % of patients\[1\]. During controlled MV, driving pressure (i.e., the difference between end-inspiratory and end-expiratory airway pressure) depends of both tidal volume and respiratory system compliance. Either excessive tidal volume or reduced lung aeration may increase the driving pressure. ARDS patients receiving tidal volume of 6 ml/kg predicted body weight (PBW) and having a day-1 driving pressure ≥ 14 cmH2O have an increased risk of death in the hospital\[2\]. Seemly, in the LUNG SAFE observational cohort, ARDS patients having a day-1 driving pressure \< 11 cmH2O had the lowest risk of death in the hospital\[1\]. Hence, driving pressure acts as a major contributor of mortality in ARDS, and probably reflects excessive regional lung distension resulting in pro-inflammatory and fibrotic biological processes. Whether decreasing the driving pressure by an intervention change mortality remains an hypothesis; but one of means is to decrease the tidal volume from 6 to 4 ml/ kg predicted body weight (PBW). However, this strategy promotes hypercarbia, at constant respiratory rate, by decreasing the alveolar ventilation. In this setting, implementing an extracorporeal CO2 removal (ECCO2R) therapy prevents from hypercarbia. A number of low-flow ECCO2R devices are now available and some of those use renal replacement therapy (RRT) platform. The investigators previously reported that combining a membrane oxygenator (0.65 m²) within a hemofiltration circuit provides efficacious low flow ECCO2R and blood purification in patients presenting with both ARDS and Acute Kidney injury\[3\]. This study aims to investigate the efficacy of an original ECCO2R system combining a 0.67 m² membrane oxygenator (Lilliput 2, SORIN) inserted within a specific circuit (HP-X, BAXTER) and mounted on a RRT monitor (PrismafleX, BAXTER). Such a therapy only aims to provide decarboxylation but not blood purification and has the huge advantage to be potentially implemented in most ICUs without requiring a specific ECCO2R device. The study will consist in three periods: * The first period will address the efficacy of this original ECCO2R system at tidal volume of 6 and 4 ml/kg PBW using an off-on-off design. * The second part will investigate the effect of varying the sweep gas flow (0-2-4-6-8-10 l/min) and the mixture of the sweep gas (Air/O2) on the CO2 removal rate. * The third part will compare three ventilatory strategies applied in a crossover design: 1. Minimal distension: Tidal volume 4 ml/kg PBW and positive end-expiratory pressure (PEEP) based on the ARDSNet PEEP/FiO2 table (ARMA). 2. Maximal recruitment: 4 ml/kg PBW and PEEP adjusted to maintain a plateau pressure between 23 - 25 cmH2O. 3. Standard: Tidal volume 6 ml/kg and PEEP based on the ARDSNet PEEP/FiO2 table (ARMA).

Interventions

DEVICELow flow Extracorporeal CO2 removal

Low flow Extracorporeal CO2 removal using a 0.67 m² membrane oxygenator (Lilliput 2) and a specific circuit (HP-X) mounted on a RRT monitor (PrismafleX)

Sponsors

Hôpital Européen Marseille
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
SUPPORTIVE_CARE
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* ARDS moderate or severe (Berlin criteria) * Onset \< 48 h * Driving pressure ≥ 11 cmH2O

Exclusion criteria

* Lack of consent or social protection * Chronic respiratory failure (requiring Oxygen or NIPPV) * Severe hypoxemia: PaO2/FIO2 \< 80 with PEEP ≥ 18 cmH2O AND FIO2= 1 * Acute Renal Failure requiring RRT * DNR order or death expected within the next 72 hours * Planned surgery or transport out-of-ICU expected within the next 72 hours * Heparin allergy * Contraindication to jugular vein catheterization * Intracranial Hypertension

Design outcomes

Primary

MeasureTime frameDescription
Change in PaCO215 minutes after initiation of ECCO2R at tidal volume of 4 ml/kg PBW.20 % decrease in PaCO2 after initiation of ECCO2R at tidal volume of 4 ml/kg PBW (as compared to 4 ml/kg without ECCO2R)

Secondary

MeasureTime frameDescription
Transpulmonary pressure and work of breathingeach 15 minutes up to the third hour (Part I and II of the study). In the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.Using an oesophageal balloon catheter (NutriVent catheter) and a dedicated monitor (FluxMed, MBMed)
End-expiratory Lung Volumeeach 15 minutes up to the third hour (Part I and II of the study). In the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.Using the nitrogen wash-in wash-out method (Engstrom GE)
Plasma CytokinesOnly in the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.Using Elisa custom kit (Qiagen) from plasma samples
Pulmonary CytokinesOnly in the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.Using Elisa custom kit (Qiagen) from BAL samples
Type III ProcollagenOnly in the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.Using both RIA and Elisa methods from plasma and BAL samples
Pulmonary Inflammatory and Fibrotic pathwayOnly in the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.Using mRNA custom kit RT-PCR analysis (Qiagen) from BAL samples
PaCO2each 15 minutes up to the third hour (Part I and II of the study). In the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.Arterial blood gas analyser (RAPIDPoint 500)
CO2 removal rateeach 15 minutes up to the third hour (Part I and II of the study). In the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.Using measurements from both the blood side and the gas side (two methods)
Regional tidal ventilationeach 15 minutes up to the third hour (Part I and II of the study). In the third part, measurement at baseline and at 1 hour and at 22 hours into each arm.Using an Electrical Impedance Tomography device (BB², Swisstom)

Other

MeasureTime frameDescription
Lacticodéshydrogenase (LDH)every 24 hours, up to 72 hours.serum samples
schizocytesevery 24 hours, up to 72 hours.serum samples
Bilirubinevery 24 hours, up to 72 hours.serum samples
Plasma Free Hemoglobinevery 24 hours, up to 72 hours.serum samples
Haptoglobinevery 24 hours, up to 72 hours.serum samples

Countries

France

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026