Repeatability and Reproducibility of the IOLMaster 700 Vers 1.70 and Agreement With IOLMaster 700 Vers. 1.50 and Pentacam

Repeatability and Reproducibility of the IOLMaster 700 Vers 1.70 and Agreement With IOLMaster 700 Vers. 1.50 and Pentacam

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT03518775
Enrollment
98
Registered
2018-05-08
Start date
2018-03-16
Completion date
2018-04-20
Last updated
2018-05-31

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Cornea

Brief summary

The objectives of this study are to determine the agreement of measurements between the IOLMaster 700 SW Vers. 1.70 and IOLMaster 700 SW Vers. 1.50 and Pentacam, as well as to evaluate the repeatability and reproducibility of the IOLMaster 700 SW Vers 1.70. The following parameters will be evaluated for Total Keratometry, (TK) and Posterior Corneal Surface (PCS): * cylinder (CYL), Axis (A) * spherical equivalent (SE) Specific Objectives: 1. To compare the agreement of SE between IOLMaster 700 SW Vers. 1.70 and IOLMaster 700 SW Vers. 1.50 (Carl Zeiss Meditec AG, Jena, Germany) 2. To compare the agreement of Cylinder magnitude and axis between IOLMaster 700 SW Version 1.70 and Pentacam 3. To evaluate the repeatability and reproducibility of the IOLMaster 700 SW Vers. 1.70 for all measurement parameters listed above

Interventions

DEVICEBiometry

Measurement of Cornea

Sponsors

ClinReg Consulting Services, Inc.
CollaboratorOTHER
Sierra Clinical Services, LLC
CollaboratorUNKNOWN
Carl Zeiss Meditec, Inc.
Lead SponsorINDUSTRY

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
22 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

1. Male or Female 22 years of age and older. 2. No history of rigid contact lens wear for at least 2 weeks. Soft lenses, if worn, should be removed at least one hour prior to the measurement. 3. Able and willing to make the required study visits. 4. Able and willing to give consent and follow study instructions. 5. Group 1: normal eyes with best-corrected visual acuity of 20/20 or better, and lens opacities of 1.0 or less in the study eye using the LOCS III system, and no prior Laser Vision Correction. 6. Group 2: cataract in the study eye greater than Grade 1 using the LOCS III system for one or more: nuclear opacity, nuclear color, cortical opacity, or PSC. 7. Group 3: history of Laser Vision Correction (LVC).

Exclusion criteria

1. Previous intraocular surgery other than Laser Vision Correction in the study eye. 2. Previous corneal surgery other than Laser Vision Correction in the study eye. 3. Presence of any corneal or retinal abnormality on slit lamp examination in the study eye. 4. Corneal Cylinder \< 0.75D measured with a IOLMaster. 5. Active ocular infection or inflammation. 6. Poor tear film/insufficient corneal reflex. 7. Best corrected visual acuity worse than 20/200 in the study eye. 8. Physical inability to be positioned at the slit-lamp biomicroscope or study devices (e.g. torticollis, head tremor, etc.). 9. Insufficient eyelid opening or eyelashes which partially conceal alignment or keratometry reflections and prevent capture of all measurements. 10. Inability to fixate (e.g. due to nystagmus, amblyopia or ocular disease, such as macular degeneration). 11. Rigid contact lens wear during past two weeks. 12. Principal Investigator's determination that the subject is not suitable due to other medical condition (reasons to be listed on eCRFs).

Design outcomes

Primary

MeasureTime frameDescription
Measurement Agreement between IOL Master Software Versions 1.70 and 1.50.1 hourIOLMaster 700 Vers. 1.70 SE\_TK (Spherical Equivalent\_Total Keratometry) Measurements will be compared to IOLMaster 700 SW Vers. 1.50 SE\_K (Spherical Equivalent\_Keratometry) Measurements for Agreement.
Measurement Agreement between IOLMaster Software Version 1.70 and Pentacam.1 hourIOLMaster 700 Vers. 1.70 A\_TK (Axis\_Total Keratometry), Cyl\_TK (Cylinder\_Total Keratometry), A\_PCS (Axis\_Posterior Corneal Surface), SE\_PCS (Spherical Equivalent\_Posterior Corneal Surface), and Cyl\_PCS (Cylinder\_Posterior Corneal Surface) Measurements will be compared to Pentacam Measurements for Agreement.
Repeatability and Reproducibility of IOLMaster 700 Software Version 700 Measurements.1 hour3\. Evaluate the Repeatability and Reproducibility of the IOLMaster 700 SW Version 1.70 for A\_TK, A\_PCS, Cyl\_TK, Cyl\_PCS, SE\_TK and SE\_PCS.