Correlation of Changes in PaO2 and ORi in Adult Patients

Correlation of Changes in PaO2 and ORi in Adult Patients

Status
Completed
Phases
Unknown
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03488238
Enrollment
31
Registered
2018-04-04
Start date
2018-08-01
Completion date
2018-09-19
Last updated
2021-09-20

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Hyperoxia

Brief summary

The Oxygen Reserve Index (ORi) is a reference that could help clinicians with their assessments of normoxic and hyperoxic states by scaling the measured absorption information between 0.00 and 1.00. An ORi of 0.00 corresponds to PaO2 values of 100mmHg and below and an ORi of 1.00 corresponds to PaO2 values of 200mmHg and above. This is a prospective, non-blinded, non-randomized study of the Oxygen Reserve Index (ORi) in a clinical setting. It is designed to evaluate the correlations with ORi and changes in PaO2 and the potential use of ORi as an early warning of impending arterial oxygen desaturation.

Interventions

DEVICEORi sensor

Noninvasive pulse oximeter sensor that is placed on the finger for measurement of ORi

Sponsors

Masimo Corporation
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Age greater than 18 years * ASA physical status III or IV * Scheduled for an elective surgical procedure requiring endotracheal intubation and the use of an arterial pressure monitoring catheter placed prior to induction of general anesthesia

Exclusion criteria

* Age less than 18 years * Adults unable to give primary consent * Pregnancy * Prisoners

Design outcomes

Primary

MeasureTime frameDescription
Sensitivity for Changes in ORi Values as Compared to Corresponding Changes in PaO2 Values.Less than 30 minsORi is an index value without a unit and ORi does not directly measure PaO2. Therefore, sensitivity is calculated to evaluate changes in ORi relative to the changes in PaO2.
Specificity for Changes in ORi Values as Compared to Corresponding Changes in PaO2 Values.Less than 30 minsORi is an index value without a unit and ORi does not directly measure PaO2. Therefore specificity is calculated to evaluate changes in ORi relative to the changes in PaO2.
Concordance for Changes in ORi Values as Compared to Corresponding Changes in PaO2 Values.Less than 30 minutesORi is an index value without a unit and ORi does not directly measure PaO2. Therefore concordance is calculated to evaluate changes in ORi relative to the changes in PaO2.

Countries

United States

Participant flow

Participants by arm

ArmCount
ORi Sensor
All subjects are enrolled in the test group and receive an ORi sensor during their scheduled, general surgery procedure ORi sensor: Noninvasive pulse oximeter sensor that is placed on the finger for measurement of ORi
29
Total29

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyAdverse Event1
Overall StudyEquipment malfunction1

Baseline characteristics

CharacteristicORi Sensor
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
18 Participants
Age, Categorical
Between 18 and 65 years
11 Participants
Race and Ethnicity Not Collected— Participants
Region of Enrollment
United States
29 participants
Sex: Female, Male
Female
12 Participants
Sex: Female, Male
Male
17 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 29
other
Total, other adverse events
1 / 29
serious
Total, serious adverse events
0 / 29

Outcome results

Primary

Concordance for Changes in ORi Values as Compared to Corresponding Changes in PaO2 Values.

ORi is an index value without a unit and ORi does not directly measure PaO2. Therefore concordance is calculated to evaluate changes in ORi relative to the changes in PaO2.

Time frame: Less than 30 minutes

Population: 1 subject was excluded as investigator reported unreliable data.

ArmMeasureValue (NUMBER)
ORi SensorConcordance for Changes in ORi Values as Compared to Corresponding Changes in PaO2 Values.95.05 percentage of accuracy
Primary

Sensitivity for Changes in ORi Values as Compared to Corresponding Changes in PaO2 Values.

ORi is an index value without a unit and ORi does not directly measure PaO2. Therefore, sensitivity is calculated to evaluate changes in ORi relative to the changes in PaO2.

Time frame: Less than 30 mins

Population: 1 subject was excluded as investigator reported unreliable data.

ArmMeasureValue (NUMBER)
ORi SensorSensitivity for Changes in ORi Values as Compared to Corresponding Changes in PaO2 Values.100 percentage of true positives
Primary

Specificity for Changes in ORi Values as Compared to Corresponding Changes in PaO2 Values.

ORi is an index value without a unit and ORi does not directly measure PaO2. Therefore specificity is calculated to evaluate changes in ORi relative to the changes in PaO2.

Time frame: Less than 30 mins

Population: 1 subject was excluded as investigator reported unreliable data.

ArmMeasureValue (NUMBER)
ORi SensorSpecificity for Changes in ORi Values as Compared to Corresponding Changes in PaO2 Values.94.38 percentage of true negatives

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026