Evaluation of Imaging of Peripheral Arteries by Optical Coherence Tomography and Intravascular Ultrasound Imaging

A Post-market, Multi-vessel Evaluation of the Imaging of Peripheral Arteries for Diagnostic Purposes Comparing Optical Coherence Tomography and Intravascular Ultrasound Imaging

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03480685
Acronym
SCAN
Enrollment
12
Registered
2018-03-29
Start date
2018-05-24
Completion date
2018-10-01
Last updated
2021-06-14

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Peripheral Artery Disease

Brief summary

A non inferiority trial to determine the capability of optical coherence tomography (OCT) imaging to show vessel morphology in comparison to imaging provided by intravascular ultrasound imaging (IVUS).

Detailed description

Patients requiring diagnostic imaging of peripheral vessels to determine vessel condition and planning treatment strategy will have vessel segments imaged by both OCT and IVUS imaging catheters. The images will be collected into a library and independent readers will rank each image on the quality of vessel morphology and disease.

Interventions

DEVICEIVUS and OCT Imaging

Use of an imaging catheter

Sponsors

Avinger, Inc.
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 90 Years
Healthy volunteers
No

Inclusion criteria

* adult * suspected vascular disease that might be a candidate for IVUS * reference vessel can accommodate imaging catheters * successful diagnostic imaging and removal of IVUS catheter with no adverse events

Exclusion criteria

* if female, pregnant or breast-feeding * unwilling to give informed consent

Design outcomes

Primary

MeasureTime frameDescription
Freedom From Adverse EventsOn day of diagnostic imaging procedures, up to 1 hourPrimary safety endpoint is freedom from diagnostic imaging procedure-related and device-associated adverse events.
Visualization of Vessel Morphology: Stent StructureDuring the diagnostic imaging procedures, up to 5 minutesThe primary endpoint was met when OCT imaging had the equivalent or lower mean score than IVUS imaging when comparing stent structure. Images were scored on a scale from 1 to 3. A lower score means better image quality and a higher score means worse image quality. Scoring Scale for image quality of Stent Structure: 1. Excellent image; 2. Acceptable image: 3. Unacceptably poor image.
Visualization of Vessel Morphology: ArtifactsDuring the diagnostic imaging procedures, up to 5 minutesThe primary endpoint was met when OCT imaging had the equivalent or lower mean score than IVUS imaging when comparing imaging artifacts. Images were scored on a scale from 1 to 3. A lower score means better image quality and a higher score means worse image quality. Scoring Scale for image quality of artifacts: 1. None; 2. Tolerable/not limiting; 3. Is intense and limits image quality.
Visualization of Vessel Morphology: Layered StructureDuring the diagnostic imaging procedures, up to 5 minutesThe primary endpoint was met when OCT imaging had the equivalent or lower mean score than IVUS imaging when comparing layered structures. Images were scored on a scale from 1 to 4. A lower score means better image quality and a higher score means worse image quality. Scoring Scale for image quality of layered structure: 1. clear differentiation of vessel wall layers; 2. differentiation of 3 wall layers; 3. differentiation of 2 wall layers; 4. no differentiation visible
Visualization of Vessel Morphology: Non-layered StructureDuring the diagnostic imaging procedures, up to 5 minutesThe primary endpoint was met when OCT imaging had the equivalent or lower mean score than IVUS imaging when comparing non-layered structures. Images were scored on a scale from 1 to 5. A lower score means better image quality and a higher score means worse image quality. Scoring scale for image quality of non-layered structure: 1-Excellent histology-like image quality to 5-unacceptably poor image quality
Visualization of Vessel Morphology: CalcificationDuring the diagnostic imaging procedures, up to 5 minutesThe primary endpoint was met when OCT imaging had the equivalent or lower mean score than IVUS imaging when comparing images for calcification. Images were scored on a scale from 1 to 5. A lower score means better image quality and a higher score means worse image quality. Scoring Scale for image quality for calcification: 1- Excellent histology-like image quality to 5- Unacceptably poor image quality.

Secondary

MeasureTime frameDescription
Vessel Measurement: Total Luminal AreaDuring the diagnostic imaging procedures, up to 5 minutesTotal luminal area is measured from the longest diameter at proximal, middle and distal locations.
Vessel Measurement: Mean DiameterDuring the diagnostic imaging procedures, up to 5 minutesMean diameter measurements include longest and shortest diameter at proximal, middle, and distal locations.

Countries

United States

Participant flow

Participants by arm

ArmCount
IVUS Imaging vs. OCT Imaging
A vessel segment will be imaged with intravascular ultrasound (IVUS). The same vessel segment will be imaged with optical coherence tomography (OCT).
12
IVUS Imaging vs. OCT Imaging
A vessel segment will be imaged with intravascular ultrasound (IVUS). The same vessel segment will be imaged with optical coherence tomography (OCT).
240
Total252

Baseline characteristics

CharacteristicIVUS Imaging vs. OCT Imaging
Age, Continuous68 years
Height176.8 cm
Leg of vessel accessed
Left Leg
9 Participants
Leg of vessel accessed
Right Leg
3 Participants
Mean length of target segment of vessel24 cm
Number of images containing a specific vessel characteristic to be scored
Artifact
120 Images
Number of images containing a specific vessel characteristic to be scored
Calcification
40 Images
Number of images containing a specific vessel characteristic to be scored
Layered Structure
24 Images
Number of images containing a specific vessel characteristic to be scored
Nonlayered Structure
104 Images
Number of images containing a specific vessel characteristic to be scored
Stent Structure
21 Images
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
Race (NIH/OMB)
Asian
0 Participants
Race (NIH/OMB)
Black or African American
1 Participants
Race (NIH/OMB)
More than one race
0 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants
Race (NIH/OMB)
White
11 Participants
Sex: Female, Male
Female
2 Participants
Sex: Female, Male
Male
10 Participants
Vessel imaged
Popliteal
36 Percentage of Participants
Vessel imaged
SFA
64 Percentage of Participants
Weight82 kg

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 120 / 12
other
Total, other adverse events
0 / 120 / 12
serious
Total, serious adverse events
0 / 120 / 12

Outcome results

Primary

Freedom From Adverse Events

Primary safety endpoint is freedom from diagnostic imaging procedure-related and device-associated adverse events.

Time frame: On day of diagnostic imaging procedures, up to 1 hour

ArmMeasureValue (NUMBER)
IVUS ImagingFreedom From Adverse Events0 Adverse Events
OCT ImagingFreedom From Adverse Events0 Adverse Events
Primary

Visualization of Vessel Morphology: Artifacts

The primary endpoint was met when OCT imaging had the equivalent or lower mean score than IVUS imaging when comparing imaging artifacts. Images were scored on a scale from 1 to 3. A lower score means better image quality and a higher score means worse image quality. Scoring Scale for image quality of artifacts: 1. None; 2. Tolerable/not limiting; 3. Is intense and limits image quality.

Time frame: During the diagnostic imaging procedures, up to 5 minutes

Population: One hundred and twenty imaging pairs (240 images total) were analyzed for artifacts.

ArmMeasureValue (MEAN)
IVUS ImagingVisualization of Vessel Morphology: Artifacts1.87 score on a scale
OCT ImagingVisualization of Vessel Morphology: Artifacts1.79 score on a scale
Primary

Visualization of Vessel Morphology: Calcification

The primary endpoint was met when OCT imaging had the equivalent or lower mean score than IVUS imaging when comparing images for calcification. Images were scored on a scale from 1 to 5. A lower score means better image quality and a higher score means worse image quality. Scoring Scale for image quality for calcification: 1- Excellent histology-like image quality to 5- Unacceptably poor image quality.

Time frame: During the diagnostic imaging procedures, up to 5 minutes

Population: Forty imaging pairs of the 120 imaging pairs (240 images total) were analyzed for calcification.

ArmMeasureValue (MEAN)
IVUS ImagingVisualization of Vessel Morphology: Calcification2.45 score on a scale
OCT ImagingVisualization of Vessel Morphology: Calcification2.11 score on a scale
Primary

Visualization of Vessel Morphology: Layered Structure

The primary endpoint was met when OCT imaging had the equivalent or lower mean score than IVUS imaging when comparing layered structures. Images were scored on a scale from 1 to 4. A lower score means better image quality and a higher score means worse image quality. Scoring Scale for image quality of layered structure: 1. clear differentiation of vessel wall layers; 2. differentiation of 3 wall layers; 3. differentiation of 2 wall layers; 4. no differentiation visible

Time frame: During the diagnostic imaging procedures, up to 5 minutes

Population: Twenty-four imaging pairs of the 120 imaging pairs (240 images total) were analyzed for layered structures.

ArmMeasureValue (MEAN)
IVUS ImagingVisualization of Vessel Morphology: Layered Structure1.61 score on a scale
OCT ImagingVisualization of Vessel Morphology: Layered Structure1.49 score on a scale
Primary

Visualization of Vessel Morphology: Non-layered Structure

The primary endpoint was met when OCT imaging had the equivalent or lower mean score than IVUS imaging when comparing non-layered structures. Images were scored on a scale from 1 to 5. A lower score means better image quality and a higher score means worse image quality. Scoring scale for image quality of non-layered structure: 1-Excellent histology-like image quality to 5-unacceptably poor image quality

Time frame: During the diagnostic imaging procedures, up to 5 minutes

Population: One hundred and four imaging pairs of the 120 imaging pairs (240 images total) were analyzed for non-layered structures.

ArmMeasureValue (MEAN)
IVUS ImagingVisualization of Vessel Morphology: Non-layered Structure2.70 score on a scale
OCT ImagingVisualization of Vessel Morphology: Non-layered Structure1.82 score on a scale
Primary

Visualization of Vessel Morphology: Stent Structure

The primary endpoint was met when OCT imaging had the equivalent or lower mean score than IVUS imaging when comparing stent structure. Images were scored on a scale from 1 to 3. A lower score means better image quality and a higher score means worse image quality. Scoring Scale for image quality of Stent Structure: 1. Excellent image; 2. Acceptable image: 3. Unacceptably poor image.

Time frame: During the diagnostic imaging procedures, up to 5 minutes

Population: Twenty-one imaging pairs of the 120 imaging pairs (240 images total) were analyzed for stent structure.

ArmMeasureValue (MEAN)
IVUS ImagingVisualization of Vessel Morphology: Stent Structure1.79 score on a scale
OCT ImagingVisualization of Vessel Morphology: Stent Structure1.43 score on a scale
Secondary

Vessel Measurement: Mean Diameter

Mean diameter measurements include longest and shortest diameter at proximal, middle, and distal locations.

Time frame: During the diagnostic imaging procedures, up to 5 minutes

ArmMeasureGroupValue (MEAN)Dispersion
IVUS ImagingVessel Measurement: Mean DiameterProximal Location: Longest Diameter4.7 mmStandard Deviation 1.2
IVUS ImagingVessel Measurement: Mean DiameterProximal Location: Shortest Diameter4.0 mmStandard Deviation 1.1
IVUS ImagingVessel Measurement: Mean DiameterMiddle Location: Longest Diameter4.3 mmStandard Deviation 1
IVUS ImagingVessel Measurement: Mean DiameterMiddle Location: Shortest Diameter3.7 mmStandard Deviation 1
IVUS ImagingVessel Measurement: Mean DiameterDistal Location: Longest Diameter3.5 mmStandard Deviation 0.9
IVUS ImagingVessel Measurement: Mean DiameterDistal Location: Shortest Diameter3.0 mmStandard Deviation 0.9
OCT ImagingVessel Measurement: Mean DiameterDistal Location: Longest Diameter3.5 mmStandard Deviation 0.8
OCT ImagingVessel Measurement: Mean DiameterProximal Location: Longest Diameter4.7 mmStandard Deviation 1
OCT ImagingVessel Measurement: Mean DiameterMiddle Location: Shortest Diameter3.5 mmStandard Deviation 0.9
OCT ImagingVessel Measurement: Mean DiameterProximal Location: Shortest Diameter3.8 mmStandard Deviation 0.9
OCT ImagingVessel Measurement: Mean DiameterDistal Location: Shortest Diameter3.1 mmStandard Deviation 0.8
OCT ImagingVessel Measurement: Mean DiameterMiddle Location: Longest Diameter4.3 mmStandard Deviation 0.9
Secondary

Vessel Measurement: Total Luminal Area

Total luminal area is measured from the longest diameter at proximal, middle and distal locations.

Time frame: During the diagnostic imaging procedures, up to 5 minutes

ArmMeasureGroupValue (MEAN)Dispersion
IVUS ImagingVessel Measurement: Total Luminal AreaProximal Location: Luminal area from longest diameter18.55 square mmStandard Deviation 8.8
IVUS ImagingVessel Measurement: Total Luminal AreaMiddle Location: Luminal area from longest diameter15.36 square mmStandard Deviation 6.6
IVUS ImagingVessel Measurement: Total Luminal AreaDistal Location: Luminal area from longest diameter10.06 square mmStandard Deviation 6.2
OCT ImagingVessel Measurement: Total Luminal AreaProximal Location: Luminal area from longest diameter17.95 square mmStandard Deviation 7.3
OCT ImagingVessel Measurement: Total Luminal AreaMiddle Location: Luminal area from longest diameter14.83 square mmStandard Deviation 6.3
OCT ImagingVessel Measurement: Total Luminal AreaDistal Location: Luminal area from longest diameter10.26 square mmStandard Deviation 5.4