INVSENSOR00013 Respiratory Rate Clinical Performance Study

INVSENSOR00013 Respiratory Rate Clinical Performance Study

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03440619
Enrollment
28
Registered
2018-02-22
Start date
2018-02-06
Completion date
2018-02-14
Last updated
2019-04-23

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Healthy

Brief summary

The objective of this prospective study is to compare the noninvasive respiration from pleth (RRp) performance of the Masimo INVSENSOR00013 device against the respiratory rate measured by capnography (RRref), Masimo's RAM technology and the manual annotation.

Interventions

DEVICEINVSENSOR00013

Investigational pulse oximeter device that will be placed on the subject's finger. This will be used as a spotcheck device

Sponsors

Masimo Corporation
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 70 Years
Healthy volunteers
Yes

Inclusion criteria

* 18 to 70 years old * Physical status of ASA I or II * Must be able to read and communicate in English * Has signed all necessary related documents, e.g. written informed consent, confidentiality agreement * Passed health assessment screening * Negative pregnancy test for female subjects of child bearing potential

Exclusion criteria

* Physical status of ASA III, IV, or V * Subject has any medical condition which in the judgement of the investigator, renders them inappropriate for participation in the study * Inability to tolerate sitting still or minimal movement for at least 30 minutes * Positive pregnancy test for female subjects * Refusal to take pregnancy test for women of child bearing potential * Nursing female subjects * Refusal of male subjects to agree to shave hair off areas where the sensor will be applied (neck) when deemed necessary * Subjects wearing acrylic nails or subjects refusing to remove nail polish * Subjects who have a nail deformity on the measurement finger * Subjects who do not have adequate skin integrity on the measurement finger * Excluded at the Principal Investigator's discretion

Design outcomes

Primary

MeasureTime frameDescription
RRp Arms of Sensor Accuracy1-5 hoursAccuracy will be determined by comparing the noninvasive respiratory rate by pleth (RRp) measurement of the INVSENSOR00013 to the respiratory rate reference (RRef) and calculating the arithmetic root mean square (Arms) value. In order to obtain the Arms value, the RR measurement from the reference is subtracted from the INVSENSOR00013 RRp measurement for a number of samples. The average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the Arms value.

Countries

United States

Participant flow

Participants by arm

ArmCount
INVSENSOR00013 Sensor
All subjects are enrolled into the test group and all subjects received the INVSENSOR00013 sensor.
28
Total28

Baseline characteristics

CharacteristicINVSENSOR00013 Sensor
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
28 Participants
Race/Ethnicity, Customized
African American
9 Participants
Race/Ethnicity, Customized
Asian or Pacific Islander
2 Participants
Race/Ethnicity, Customized
Caucasian
9 Participants
Race/Ethnicity, Customized
Hispanic
7 Participants
Race/Ethnicity, Customized
Other
1 Participants
Region of Enrollment
United States
28 Participants
Sex: Female, Male
Female
9 Participants
Sex: Female, Male
Male
19 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 28
other
Total, other adverse events
0 / 28
serious
Total, serious adverse events
0 / 28

Outcome results

Primary

RRp Arms of Sensor Accuracy

Accuracy will be determined by comparing the noninvasive respiratory rate by pleth (RRp) measurement of the INVSENSOR00013 to the respiratory rate reference (RRef) and calculating the arithmetic root mean square (Arms) value. In order to obtain the Arms value, the RR measurement from the reference is subtracted from the INVSENSOR00013 RRp measurement for a number of samples. The average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the Arms value.

Time frame: 1-5 hours

ArmMeasureValue (NUMBER)
INVSENSOR00013 SensorRRp Arms of Sensor Accuracy1.2 respiration per minute (RPM)

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026