Intraocular Pressure
Conditions
Brief summary
The study will assess the CT-800 and TRK-2P for conformity to ANSI Z80.10-2014 Ophthalmic Instruments - Tonometers, to the FDA Guidance for Industry and FDA Staff, Tonometer-Premarket Notification \[510(k)\] Submissions, and to the applicable Supplemental Information Sheet. This clinical device study will also assess the substantial equivalence of TRK-2P to a predicate device in regard to pachymetry.
Interventions
tonometer
tonometer, pachymeter
Sponsors
Study design
Eligibility
Inclusion criteria
1. be at least 18 years of age of either sex and any race or ethnicity; 2. be willing and able to provide written informed consent prior to any study procedures being performed; 3. be willing and able to follow all instructions and attend all study visits;
Exclusion criteria
1. have only one functional eye; 2. have poor or eccentric fixation in either eye; 3. have corneal scarring or have had corneal surgery, including corneal laser surgery in either eye; 4. have microphthalmos in either eye; 5. have buphthalmos in either eye; 6. be a contact lens wearer, meaning having worn soft contact lenses within the past 3 months and/or rigid permeable gas lenses within the past 6 months; 7. have dry eyes, meaning having been diagnosed by a physician with dry eyes and currently using a prescribed medication or daily use of an artificial tear; 8. be a lid squeezer - blepharospasm; 9. have nystagmus in either eye; 10. have keratoconus in either eye; 11. have any other corneal or conjunctival pathology or infection in either eye; 12. have a condition or a situation, which in the Investigator's opinion, may put the subject at increased risk, confound study data, or interfere significantly with the subject's study participation.
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Agreement of IOP | 1 day | Agreement of the measured IOP between the test devices and the predicate device for the tonometer function of CT-800 and TRK-2P. Presentation of results and analyses will include descriptive statistics (mean, standard deviation, limits of agreement), Bland-Altman plot and linear regression. |
| Agreement of Corneal Thickness | 1 day | Agreement of measured corneal thickness between the test device and the predicate device for the pachymeter function of the TRK-2P. Presentation of results and analyses will include descriptive statistics (mean, standard deviation, limits of agreement), Bland-Altman plot and Deming regression. |
Countries
United States