Comparative Study of the Topcon Tonometers CT-800 and TRK-2P and the Haag-Streit Goldmann Manual Tonometer and Topcon SP-1P

Comparative Study of the Topcon Tonometers CT-800 and TRK-2P and the Haag-Streit Goldmann Manual Tonometer (Predicate) to Demonstrate Conformance to ANSI Z80.10-2014 Ophthalmic Instruments - Tonometers, to the FDA Guidance for Industry and FDA Staff, Tonometer-Premarket Notification [510(k)] Submissions, and to the Applicable Supplemental Information Sheet and Comparison of Pachymetry Values for Topcon TRK-2P With the Topcon SP-1P (Predicate)

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT03439774
Enrollment
345
Registered
2018-02-20
Start date
2017-05-05
Completion date
2018-06-30
Last updated
2022-06-08

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Intraocular Pressure

Brief summary

The study will assess the CT-800 and TRK-2P for conformity to ANSI Z80.10-2014 Ophthalmic Instruments - Tonometers, to the FDA Guidance for Industry and FDA Staff, Tonometer-Premarket Notification \[510(k)\] Submissions, and to the applicable Supplemental Information Sheet. This clinical device study will also assess the substantial equivalence of TRK-2P to a predicate device in regard to pachymetry.

Interventions

DEVICETopcon CT-800

tonometer

DEVICETopcon TRK-2P

tonometer, pachymeter

Sponsors

Topcon Medical Systems, Inc.
Lead SponsorINDUSTRY

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

1. be at least 18 years of age of either sex and any race or ethnicity; 2. be willing and able to provide written informed consent prior to any study procedures being performed; 3. be willing and able to follow all instructions and attend all study visits;

Exclusion criteria

1. have only one functional eye; 2. have poor or eccentric fixation in either eye; 3. have corneal scarring or have had corneal surgery, including corneal laser surgery in either eye; 4. have microphthalmos in either eye; 5. have buphthalmos in either eye; 6. be a contact lens wearer, meaning having worn soft contact lenses within the past 3 months and/or rigid permeable gas lenses within the past 6 months; 7. have dry eyes, meaning having been diagnosed by a physician with dry eyes and currently using a prescribed medication or daily use of an artificial tear; 8. be a lid squeezer - blepharospasm; 9. have nystagmus in either eye; 10. have keratoconus in either eye; 11. have any other corneal or conjunctival pathology or infection in either eye; 12. have a condition or a situation, which in the Investigator's opinion, may put the subject at increased risk, confound study data, or interfere significantly with the subject's study participation.

Design outcomes

Primary

MeasureTime frameDescription
Agreement of IOP1 dayAgreement of the measured IOP between the test devices and the predicate device for the tonometer function of CT-800 and TRK-2P. Presentation of results and analyses will include descriptive statistics (mean, standard deviation, limits of agreement), Bland-Altman plot and linear regression.
Agreement of Corneal Thickness1 dayAgreement of measured corneal thickness between the test device and the predicate device for the pachymeter function of the TRK-2P. Presentation of results and analyses will include descriptive statistics (mean, standard deviation, limits of agreement), Bland-Altman plot and Deming regression.