Comparison of Repeatability Between SS and SD-OCT for Measuring Retinal Thickness in Various Retinal Diseases

Comparison of Repeatability Between Swept-source and Spectral-domain Optical Coherence Tomography for Measuring Inner Retinal Thickness in Various Retinal Diseases

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT03438903
Enrollment
114
Registered
2018-02-20
Start date
2017-10-01
Completion date
2018-01-31
Last updated
2018-02-20

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Macula Abnormality

Brief summary

Comparison of intradevice repeatability of thickness measurements for macular area in the eyes with normal and various retinal diseases using SS-OCT and SD-OCT

Detailed description

The investigators studied the intradevice repeatability of thickness measurements for total retina, retinal nerve fiber layer (RNFL) and ganglion cell-inner plexiform (GC-IPL) layers of macular area in the eyes with normal and various retinal diseases using SS-OCT and SD-OCT and compared the repeatability between two devices.

Interventions

DEVICERepeat exams of OCT device

All 2 consecutive OCT examinations were performed by 1 experienced technician, in the order of SD-OCT and SS-OCT. The interval was 10 minutes between examinations by each OCT devices

Sponsors

Hanyang University
Lead SponsorOTHER

Study design

Observational model
CASE_CONTROL
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
19 Years to 80 Years
Healthy volunteers
Yes

Inclusion criteria

* Age-matched normal group was composed of subjects with age over at least 19 years, * a best-corrected visual acuity of 20/25 or higher, * normal fundus and intraocular pressure (IOP) \< 21mm Hg.

Exclusion criteria

* an spherical equivalent (SE) larger than ± 6 D, * cataract graded as more severe than grade 3 (Lens Opacities Classification System III), * a history of glaucoma and optic nerve diseases, or a history of vitrectomy.

Design outcomes

Primary

MeasureTime frameDescription
To compare repeatability between swept-source optical coherence tomography (SS-OCT) and spectral-domain OCT (SD-OCT) devices for the measurement of retinal thickness in various retinal diseases.Each device was examined twice consecutively within one minutes. After one OCT device exam, another OCT exam was performed immediately. The interval between test devices did not exceed 5 minutes.Comparison of Intraclass correlation coefficient (ICC) and coefficients of variability (CV) between devices

Secondary

MeasureTime frameDescription
To compare repeatability according to the degree of macular thicknessEach device was examined twice consecutively within one minutes. After one OCT device exam, another OCT exam was performed immediately. The interval between test devices did not exceed 5 minutes.To compare the repeatability of retinal thickness according to central macular thickness in eyes with retinal diseases

Countries

South Korea

Outcome results

None listed