Comparison Between NICOM-Cheetah and Flotrac-Vigileo for Cardiac Output Monitoring.

An Observational Study Comparing NICOM-Cheetah Versus Flotrac-Vigileo for Cardiac Output Monitoring in the Surgical Intensive Care Unit.

Status
Terminated
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT03394768
Enrollment
26
Registered
2018-01-09
Start date
2018-08-10
Completion date
2021-01-05
Last updated
2021-05-11

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Cardiac Output, Low

Brief summary

Cardiac output (CO) monitoring is an important tool in critically ill patients with haemodynamic instability. Traditionally, this has been accomplished using the pulmonary artery catheter (PAC). However, its use has been associated with various complications such as pneumothorax, arrhythmia, infection, pulmonary valve injury and embolism. This has led to the development of various minimally invasive cardiac output monitoring devices such as the oesophageal Doppler, PiCCO®, NiCCO® and FloTrac®. These devices however have their inherent limitations and though minimally invasive, are still not without risks to the patient. More recently a completely non-invasive CO monitoring device, NICOM Cheetah® has been introduced. Its use is based on thoracic bioreactance and it involves placement of 4 electrodes on either side of the thorax and it analyses changes in the phase of electrical voltage signal to the current applied across the thorax. Initial validation studies on the NICOM Cheetah® device revealed promising results. This study intends to compare cardiac output monitoring capabilities of the NICOM Cheetah® device using the FloTrac® CO monitor as a reference.

Detailed description

This will be a single centre prospective observational study. 50 consecutive patients admitted to the surgical intensive care unit (SICU) who are deemed to require cardiac output monitoring by the treating intensivist will be studied. Data collection is planned to be carried out over 2 consecutive years. Inclusion criteria are: 1) Patients who are deemed to require cardiac output monitoring. 2) Patients 21 years old and above. 3) American Society of Anesthesiologist (ASA) physical status 1 to 4. Exclusion criteria are: 1) Patients who are not on mechanical ventilation. 2) Patients who are not in sinus rhythm. 3) Patients with allergy to adhesive tapes. Patients will be treated as per department protocols and no additional intervention will be performed. Each patient will have an arterial catheter inserted as per our usual practice. All patients will have the FloTrac® connected to the arterial catheter (standard of care in CGH SICU) and the NICOM Cheetah® electrodes placed on the skin across the anterior thoracic wall. The FloTrac CO monitor is the current standard of care for cardiac output monitoring in the SICU of CGH. All patients deemed to require cardiac output monitoring will receive the FloTrac (as per departmental practice). The NICOM CO monitor involves the application of non-invasive sensor strips. In this study, it will be applied to patients receiving the FloTrac (standard of care), on top of the standard care of monitoring with Flotrac. For each patient, standard demographic and SICU data (age, gender, comorbidities, ASA score, admission diagnosis, type of surgery, cause of haemodynamic instability, mean arterial pressure (MAP) and central venous pressure (CVP) at the time of instituting cardiac output (CO) monitoring, use of inotropic support, presence of renal replacement therapy, presence of mechanical ventilation and APACHE II score) will be collected. Cardiac Index (CI) values simultaneously recorded by the NICOM Cheetah® and FloTrac® on a 15-minutely basis for the first 10-hour period will also be collected. These data will be automatically recorded by the FloTrac® Vigilance monitor and the Cheetah NICOM monitor and transcribed to a data entering/analysis software by the principal and co-principal investigators for data analysis.

Interventions

Patients will be treated as per department protocols and no additional intervention will be performed

Sponsors

Changi General Hospital
Lead SponsorOTHER

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
21 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Patients who are deemed to require cardiac output monitoring. * Patients 21 years old and above * American Society of Anesthesiologist (ASA) physical status 1 to 4

Exclusion criteria

* Patients who are not on mechanical ventilation. * Patients who are not in sinus rhythm. * Patients with allergy to adhesive tapes

Design outcomes

Primary

MeasureTime frameDescription
Degree of Disagreement, %10 hours of data collected for each participantComparison of cardiac output monitoring capabilities of the NICOM Cheetah® versus the FloTrac®. Cardiac output defined as the amount of blood the heart pumps through the circulatory system in a minute. Bland-Altman analysis used to compare the level of disagreement between the 2 devices, %.

Countries

Singapore

Participant flow

Recruitment details

Participants recruited from the surgical intensive care unit of Changi General Hospital, Singapore, from 2018 until 2021

Participants by arm

ArmCount
All Participants
Patients will be treated as per department protocols and no additional intervention will be performed. Each patient will have an arterial catheter inserted as per our usual practice. All patients will have the FloTrac® connected to the arterial catheter (standard of care in CGH SICU) and the NICOM Cheetah® electrodes placed on the skin across the anterior thoracic wall. The NICOM CO monitor involves the application of non-invasive sensor strips. In this study, it will be applied to patients receiving the FloTrac (standard of care), on top of the standard care of monitoring with Flotrac. No additional intervention: Patients will be treated as per department protocols and no additional intervention will be performed
26
Total26

Baseline characteristics

CharacteristicAll Participants
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
13 Participants
Age, Categorical
Between 18 and 65 years
13 Participants
Age, Continuous68.58 years
STANDARD_DEVIATION 12.4
Ethnicity (NIH/OMB)
Hispanic or Latino
0 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
0 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
26 Participants
Region of Enrollment
Singapore
26 participants
Sex: Female, Male
Female
11 Participants
Sex: Female, Male
Male
15 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 26
other
Total, other adverse events
0 / 26
serious
Total, serious adverse events
0 / 26

Outcome results

Primary

Degree of Disagreement, %

Comparison of cardiac output monitoring capabilities of the NICOM Cheetah® versus the FloTrac®. Cardiac output defined as the amount of blood the heart pumps through the circulatory system in a minute. Bland-Altman analysis used to compare the level of disagreement between the 2 devices, %.

Time frame: 10 hours of data collected for each participant

ArmMeasureValue (NUMBER)
All ParticipantsDegree of Disagreement, %69.2 percentage of disagreement