Skip to content

Alternating Pressure Overlay on Weight Bearing Tissue Tolerance in People With Spinal Cord Injury

Effect of Alternating Pressure Overlay on Weight Bearing Tissue Tolerance in People With Spinal Cord Injury

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03317288
Enrollment
15
Registered
2017-10-23
Start date
2017-05-12
Completion date
2019-03-31
Last updated
2021-06-09

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Pressure Ulcer, Spinal Cord Injuries

Keywords

spinal cord injury, pressure ulcer, laser Doppler flowmetry

Brief summary

The purpose of this study is to test if the newly-developed Dabir alternating pressure (AP) overlay could be beneficial to preventing skin damage during daily activities, such as lying in bed. Specifically, the study is being done to test if the Dabir AP overlay could be used to increase skin tolerance when lying in bed for an extended period of time (40 minutes), including reducing pressure and increasing skin blood flow (amount of blood supply to oxygen and nutrients to skin) as compared to regular operation room (OR) overlay. A total of 20 participants with spinal cord injury will be recruited. Subjects will undergo study procedures including: AP and Control protocols. Skin blood flow and interface pressure will be collected non-invasively (from outside the body) during both protocols. During AP protocols, subject will be asked to lie on side (10 min), on back on AP overlay (40 min), on side (30 min), on back on OR overlay (40 min), and on side (10 min). During Control protocol, subject will be asked to lie on side (10 min), on back on OR overlay (40 min), and on side (10 min). Subject will then proceed to vascular control measures, including: non-invasive skin blood flow and tissue oxygen measurements with mild heating and electrical stimulation. Findings from this study will help us understand the effectiveness of the AP overlay on skin blood flow response during prolonged lying in bed.

Detailed description

Tissue ischemia is one of the main etiologies of pressure injury. It is inevitable during bed-ridden and wheelchair bound conditions, such as spinal cord injury (SCI), or prolonged surgery procedure in the operating room. Active alternating pressure (AP) mattresses are covered by Medicare policy for patients with multiple/large stage 2/ stage 3 pressure injuries to relief interface pressure and prevent recurrence of pressure injuries. Previous studies demonstrated that the protective mechanisms of AP on pressure injury include redistribution of surface pressure, and allowing more blood supplies while lying on the support surface. The Dabir AP overlay is low profile yet possesses similar characteristics as other AP mattresses on the market; it is intended for clinical use, such as operation room (OR) where AP mattresses may not be an option. We expect Dabir AP overlay will be beneficial to people at risk of pressure injury by maintaining the blood supply during weight bearing condition. The objective of this study is to evaluate the protective mechanisms of Dabir AP overlay on weight bearing tissue of people with chronic spinal cord injury (SCI). Our central hypothesis is that the Dabir AP overlay will cause lower amount of interface pressure and allow more skin blood flow during weight bearing as compared to regular OR overlay. Each subject will undergo two study protocols: AP and control, and the order of the protocols will be randomized by drawing one envelope from the envelope pool. We will start with a pool of total 20 envelopes (10 envelopes indicating AP protocol first, and 10 envelopes indicating control protocol first). A 30-minute washout period will be provided in between the two protocols. Two Laser Doppler flowmetry (LDF) flat probes (moor instrument, Wilmington, DE) will be taped to the sacrum and one heel to collect the skin blood flow response throughout both AP and control protocols. A whole body pressure mapping system will be used between the subject and the overlay to collect the inter pressure data. For the AP cycle, we will use four 10-minute cycles (5 minutes of high pressure and 5 minutes of low pressure). AP protocol: The subject will first lie on his/her side quietly for 10 minutes to collect the baseline skin blood flow, followed by lying supine on the Dabir AP overlay for 40 minutes, and then side lying for 30 minutes to collect the first reactive hyperemic response. To test the effect of preconditioning, the subject will then lie supine on the OR overlay for 40 minutes followed by side lying for 30 minutes for the second reactive hyperemic response. Control protocol: The subject will first lie on his/her side quietly for 10 minutes to collect the baseline skin blood flow, followed by lying supine on the OR overlay for 40 minutes, and then side lying for 30 minutes to collect the reactive hyperemic response for control protocol. The PI, or key personnel will then collect the vascular control measures, including the heat hyperemic (HH) blood flow response, and the skin vasomotor response. A non-invasive combined laser Doppler and white light reflectance spectroscopy (LDF-WLRS) probe (CP2T-1000, Moor Instrument, Wilmington, DE) will be used for both measures. HH blood flow response will be measured at the sacrum and one heel while the subject is in side-lying position. HH will be induced with the heater at the tip of the probe head (41°C for 5 minutes). This HH measure will take approximately 7 minutes (2 min-baseline, 5 min-heating) per location. In total, the HH measures will take up to 15 minutes to complete. The skin vasomotor response will be induced using cutaneous electrical stimulation (3-10 mA, 0.2-1.0 ms pulses at 20 Hz, 1 s trains) delivered via 1 cm Ag-AgCl surface electrodes applied bilaterally (4 cm apart) to the forehead (above lesion) and the abdominal wall (below lesion) while the subject is in a semi-reclined position.12 For each site, 5-8 stimuli will be delivered randomly. CP2T-1000 probe will be used to collect SBF and tissue oxygenation data at pulp of the second finger and toe. Subject will be in a semi-reclined position (about 30° from horizontal plane) during this protocol, and it will take up to 15 minutes to complete this protocol. Pressure mapping of wheelchair: (approximately 10 minutes) The subject will be asked to transfer back to his/her wheelchair with the pressure mapping system placed between the wheelchair and the subject to collect the interface pressure data. The pressure mapping system is the same as that used between the subject and overlay to collect the interface pressure data.

Interventions

DEVICEDabir Air overlay

The Dabir Air overlay is a low profile alternating pressure overlay that is designed to place on top of mattress and operation room table.

Sponsors

Dabir Surfaces Inc
CollaboratorINDUSTRY
University of Illinois at Chicago
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
PREVENTION
Masking
NONE

Masking description

Here is no masking of the intervention.

Intervention model description

Repeated measures design. Subject serves as his or her own control.

Eligibility

Sex/Gender
ALL
Age
18 Years to 65 Years
Healthy volunteers
No

Inclusion criteria

* have had spinal cord injury for more than one year * have injury level at T10 or above * not ambulatory and use a wheelchair for mobility * none smokers or able to refrain from smoking four hours prior to and during the study

Exclusion criteria

* current pressure ulcer * history of diabetes mellitus * history of cardiovascular diseases * history of hypertension * history of pulmonary diseases * pregnant women (based on a positive pregnancy test) * Non-English speakers

Design outcomes

Primary

MeasureTime frameDescription
Heel Averaged Interface Pressure During Supine Lying40 minutes, the inflation-deflation cycle alternates every 10 minutesHeel averaged interface pressure measured using pressure mapping system during supine lying on operating room pad with and without alternating pressure overlay. This is the averaged interface pressure of the nine pressure cells located at the heel.
Sacral Skin Blood Flow During Supine Lying40 minutes, the inflation-deflation cycle alternates every 10 minutesSacral skin blood flow measured using laser Doppler flowmetry system (moor instrument) during supine lying on operating room pad with and without alternating pressure overlay
Heel Skin Blood Flow During Supine Lying40 minutes, the inflation-deflation cycle alternates every 10 minutesHeel skin blood flow measured using laser Doppler flowmetry system (moor instrument) during supine lying on operating room pad with and without alternating pressure overlay
Sacral Peak Interface Pressure During Supine Lying40 minutes, the inflation-deflation cycle alternates every 10 minutesSacral peak interface pressure measured using pressure mapping system during supine lying on operating room pad with and without alternating pressure overlay. This is the maximum interface pressure of the nine pressure cells located at the sacrum.
Sacral Averaged Interface Pressure During Supine Lying40 minutes, the inflation-deflation cycle alternates every 10 minutesSacral averaged interface pressure measured using pressure mapping system during supine lying on operating room pad with and without alternating pressure overlay. This is the averaged interface pressure of the nine pressure cells located at the sacrum.
Heel Peak Interface Pressure During Supine Lying40 minutes, the inflation-deflation cycle alternates every 10 minutesHeel peak interface pressure measured using pressure mapping system during supine lying on operating room pad with and without alternating pressure overlay. This is the maximum interface pressure of the nine pressure cells located at the heel.

Secondary

MeasureTime frameDescription
Post Alternating Pressure Sacral Skin Blood Flow40 minutesSacral skin blood flow measured using laser Doppler flowmetry system (moor instrument) during supine lying on operating room pad after 40 minutes on operating room pad with or without alternating pressure and a 30 minutes washout period.
Post Alternating Pressure Sacral Peak Interface Pressure40 minutesSacral peak interface pressure during supine lying on operating room pad after 40 minutes on operating room pad with or without alternating pressure and a 30 minutes washout period.
Post Alternating Pressure Sacral Averaged Interface Pressure40 minutesSacral averaged interface pressure during supine lying on operating room pad after 40 minutes on operating room pad with or without alternating pressure and a 30 minutes washout period.
Post Alternating Pressure Heel Skin Blood Flow40 minutesHeel skin blood flow measured using laser Doppler flowmetry system (moor instrument) during supine lying on operating room pad after 40 minutes on operating room pad with or without alternating pressure and a 30 minutes washout period.
Post Alternating Pressure Heel Peak Interface Pressure40 minutesHeel peak interface pressure during supine lying on operating room pad after 40 minutes on operating room pad with or without alternating pressure and a 30 minutes washout period.
Post Alternating Pressure Heel Averaged Interface Pressure40 minutesHeel averaged interface pressure during supine lying on operating room pad after 40 minutes on operating room pad with or without alternating pressure and a 30 minutes washout period.

Countries

United States

Participant flow

Participants by arm

ArmCount
SCI Subjects
The subject act as his or her own control. Each subject will undergo two procedures, intervention: alternating pressure overlay on top of standard operation room overlay, vs. control: operation room overlay. Dabir Air overlay: The Dabir Air overlay is a low profile alternating pressure overlay that is designed to place on top of mattress and operation room table.
15
Total15

Baseline characteristics

CharacteristicSCI Subjects
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
15 Participants
Age, Continuous41.87 years
STANDARD_DEVIATION 14.89
American Spinal Injury Association Scale score
Complete
5 Participants
American Spinal Injury Association Scale score
Incomplete
10 Participants
body mass index26.81 kg/m2
STANDARD_DEVIATION 4.13
duration of spinal cord injury17 years
STANDARD_DEVIATION 14.62
level of spinal cord injury
At T6 and above
6 Participants
level of spinal cord injury
Below T6
9 Participants
Race/Ethnicity, Customized
Ethnicity
African American
11 Participants
Race/Ethnicity, Customized
Ethnicity
Asian
1 Participants
Race/Ethnicity, Customized
Ethnicity
Caucasian
1 Participants
Race/Ethnicity, Customized
Ethnicity
Hispanic
2 Participants
Region of Enrollment
United States
15 participants
Sex: Female, Male
Female
4 Participants
Sex: Female, Male
Male
11 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 0
other
Total, other adverse events
0 / 0
serious
Total, serious adverse events
0 / 0

Outcome results

Primary

Heel Averaged Interface Pressure During Supine Lying

Heel averaged interface pressure measured using pressure mapping system during supine lying on operating room pad with and without alternating pressure overlay. This is the averaged interface pressure of the nine pressure cells located at the heel.

Time frame: 40 minutes, the inflation-deflation cycle alternates every 10 minutes

Population: All participants with spinal cord injury. Each participant serves as his/her own control.

ArmMeasureValue (MEAN)Dispersion
Alternating Pressure -- Inflation CycleHeel Averaged Interface Pressure During Supine Lying37.88 mmHgStandard Deviation 17.27
Alternating Pressure -- Deflation CycleHeel Averaged Interface Pressure During Supine Lying14.60 mmHgStandard Deviation 8.37
Control (Operating Room Pad Only)Heel Averaged Interface Pressure During Supine Lying34.83 mmHgStandard Deviation 16.62
Primary

Heel Peak Interface Pressure During Supine Lying

Heel peak interface pressure measured using pressure mapping system during supine lying on operating room pad with and without alternating pressure overlay. This is the maximum interface pressure of the nine pressure cells located at the heel.

Time frame: 40 minutes, the inflation-deflation cycle alternates every 10 minutes

Population: All participants with spinal cord injury. Each participant serves as his/her own control.

ArmMeasureValue (MEAN)Dispersion
Alternating Pressure -- Inflation CycleHeel Peak Interface Pressure During Supine Lying52.49 mmHgStandard Deviation 21.21
Alternating Pressure -- Deflation CycleHeel Peak Interface Pressure During Supine Lying26.79 mmHgStandard Deviation 12.91
Control (Operating Room Pad Only)Heel Peak Interface Pressure During Supine Lying53.05 mmHgStandard Deviation 18.22
Primary

Heel Skin Blood Flow During Supine Lying

Heel skin blood flow measured using laser Doppler flowmetry system (moor instrument) during supine lying on operating room pad with and without alternating pressure overlay

Time frame: 40 minutes, the inflation-deflation cycle alternates every 10 minutes

Population: All participants with spinal cord injury. Each participant serves as his/her own control.

ArmMeasureValue (MEAN)Dispersion
Alternating Pressure -- Inflation CycleHeel Skin Blood Flow During Supine Lying15.10 arbitrary unitStandard Deviation 20.78
Alternating Pressure -- Deflation CycleHeel Skin Blood Flow During Supine Lying27.92 arbitrary unitStandard Deviation 32.15
Control (Operating Room Pad Only)Heel Skin Blood Flow During Supine Lying10.43 arbitrary unitStandard Deviation 11.16
Primary

Sacral Averaged Interface Pressure During Supine Lying

Sacral averaged interface pressure measured using pressure mapping system during supine lying on operating room pad with and without alternating pressure overlay. This is the averaged interface pressure of the nine pressure cells located at the sacrum.

Time frame: 40 minutes, the inflation-deflation cycle alternates every 10 minutes

Population: All participants with spinal cord injury. Each participant serves as his/her own control.

ArmMeasureValue (MEAN)Dispersion
Alternating Pressure -- Inflation CycleSacral Averaged Interface Pressure During Supine Lying57.21 mmHgStandard Deviation 28.54
Alternating Pressure -- Deflation CycleSacral Averaged Interface Pressure During Supine Lying36.16 mmHgStandard Deviation 18.47
Control (Operating Room Pad Only)Sacral Averaged Interface Pressure During Supine Lying81.50 mmHgStandard Deviation 46.39
Primary

Sacral Peak Interface Pressure During Supine Lying

Sacral peak interface pressure measured using pressure mapping system during supine lying on operating room pad with and without alternating pressure overlay. This is the maximum interface pressure of the nine pressure cells located at the sacrum.

Time frame: 40 minutes, the inflation-deflation cycle alternates every 10 minutes

Population: All participants with spinal cord injury. Each participant serves as his/her own control.

ArmMeasureValue (MEAN)Dispersion
Alternating Pressure -- Inflation CycleSacral Peak Interface Pressure During Supine Lying89.27 mmHgStandard Deviation 53.92
Alternating Pressure -- Deflation CycleSacral Peak Interface Pressure During Supine Lying51.47 mmHgStandard Deviation 30.18
Control (Operating Room Pad Only)Sacral Peak Interface Pressure During Supine Lying114.13 mmHgStandard Deviation 60.97
Primary

Sacral Skin Blood Flow During Supine Lying

Sacral skin blood flow measured using laser Doppler flowmetry system (moor instrument) during supine lying on operating room pad with and without alternating pressure overlay

Time frame: 40 minutes, the inflation-deflation cycle alternates every 10 minutes

Population: All participants with spinal cord injury. Each participant serves as his/her own control.

ArmMeasureValue (MEAN)Dispersion
Alternating Pressure -- Inflation CycleSacral Skin Blood Flow During Supine Lying12.65 arbitrary unitStandard Deviation 12.45
Alternating Pressure -- Deflation CycleSacral Skin Blood Flow During Supine Lying15.54 arbitrary unitStandard Deviation 15.33
Control (Operating Room Pad Only)Sacral Skin Blood Flow During Supine Lying11.96 arbitrary unitStandard Deviation 10.26
Secondary

Post Alternating Pressure Heel Averaged Interface Pressure

Heel averaged interface pressure during supine lying on operating room pad after 40 minutes on operating room pad with or without alternating pressure and a 30 minutes washout period.

Time frame: 40 minutes

Population: All participants with spinal cord injury. Each participant serves as his/her own control.

ArmMeasureValue (MEAN)Dispersion
Alternating Pressure -- Inflation CyclePost Alternating Pressure Heel Averaged Interface Pressure30.35 mmHgStandard Deviation 17.59
Alternating Pressure -- Deflation CyclePost Alternating Pressure Heel Averaged Interface Pressure34.83 mmHgStandard Deviation 16.62
Secondary

Post Alternating Pressure Heel Peak Interface Pressure

Heel peak interface pressure during supine lying on operating room pad after 40 minutes on operating room pad with or without alternating pressure and a 30 minutes washout period.

Time frame: 40 minutes

Population: All participants with spinal cord injury. Each participant serves as his/her own control.

ArmMeasureValue (MEAN)Dispersion
Alternating Pressure -- Inflation CyclePost Alternating Pressure Heel Peak Interface Pressure47.69 mmHgStandard Deviation 16.21
Alternating Pressure -- Deflation CyclePost Alternating Pressure Heel Peak Interface Pressure53.05 mmHgStandard Deviation 18.22
Secondary

Post Alternating Pressure Heel Skin Blood Flow

Heel skin blood flow measured using laser Doppler flowmetry system (moor instrument) during supine lying on operating room pad after 40 minutes on operating room pad with or without alternating pressure and a 30 minutes washout period.

Time frame: 40 minutes

Population: All participants with spinal cord injury. Each participant serves as his/her own control.

ArmMeasureValue (MEAN)Dispersion
Alternating Pressure -- Inflation CyclePost Alternating Pressure Heel Skin Blood Flow16.31 arbitrary unitStandard Deviation 29.18
Alternating Pressure -- Deflation CyclePost Alternating Pressure Heel Skin Blood Flow10.43 arbitrary unitStandard Deviation 11.16
Secondary

Post Alternating Pressure Sacral Averaged Interface Pressure

Sacral averaged interface pressure during supine lying on operating room pad after 40 minutes on operating room pad with or without alternating pressure and a 30 minutes washout period.

Time frame: 40 minutes

Population: All participants with spinal cord injury. Each participant serves as his/her own control.

ArmMeasureValue (MEAN)Dispersion
Alternating Pressure -- Inflation CyclePost Alternating Pressure Sacral Averaged Interface Pressure81.07 mmHgStandard Deviation 48.62
Alternating Pressure -- Deflation CyclePost Alternating Pressure Sacral Averaged Interface Pressure81.50 mmHgStandard Deviation 46.39
Secondary

Post Alternating Pressure Sacral Peak Interface Pressure

Sacral peak interface pressure during supine lying on operating room pad after 40 minutes on operating room pad with or without alternating pressure and a 30 minutes washout period.

Time frame: 40 minutes

Population: All participants with spinal cord injury. Each participant serves as his/her own control.

ArmMeasureValue (MEAN)Dispersion
Alternating Pressure -- Inflation CyclePost Alternating Pressure Sacral Peak Interface Pressure104.62 mmHgStandard Deviation 58.17
Alternating Pressure -- Deflation CyclePost Alternating Pressure Sacral Peak Interface Pressure114.13 mmHgStandard Deviation 60.97
Secondary

Post Alternating Pressure Sacral Skin Blood Flow

Sacral skin blood flow measured using laser Doppler flowmetry system (moor instrument) during supine lying on operating room pad after 40 minutes on operating room pad with or without alternating pressure and a 30 minutes washout period.

Time frame: 40 minutes

Population: All participants with spinal cord injury. Each participant serves as his/her own control.

ArmMeasureValue (MEAN)Dispersion
Alternating Pressure -- Inflation CyclePost Alternating Pressure Sacral Skin Blood Flow15.78 arbitrary unitStandard Deviation 15.82
Alternating Pressure -- Deflation CyclePost Alternating Pressure Sacral Skin Blood Flow11.96 arbitrary unitStandard Deviation 10.26

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026