Post-Market Clinical Investigation of the Clareon® IOL

Post-Market Clinical Investigation of the Clareon® IOL

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03316885
Enrollment
245
Registered
2017-10-20
Start date
2018-03-14
Completion date
2021-09-23
Last updated
2023-08-21

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Aphakia, Cataract

Brief summary

The primary purpose of this study is to evaluate the long-term (3 years) visual acuity and adverse event outcomes for the Clareon® Intraocular Lens (IOL). A comparison to historical safety and performance endpoint (SPE) rates as reported in the European Standard International Organization for Standardization (EN ISO) 11979-7:2014 will be conducted at one year. The secondary purpose of this study is to evaluate the visual acuity outcomes with the Clareon IOL at Years 2 and 3.

Detailed description

Subjects will attend a total of 12 study visits over a period of approximately 36 months: One preoperative screening visit, two operative visits, and nine post-implantation visits. Both eyes will be implanted, with the second eye implantation to occur 2-21 days after the first. Primary endpoint data will be collected at the Year 1 visit (Day 330-420 post-implantation from second eye surgery).

Interventions

DEVICEClareon® IOL

Foldable, single-piece posterior chamber IOL intended for long-term use over the lifetime of the cataract patient. This device is CE-marked (European Conformity) in the countries with participating Investigators.

PROCEDURECataract Surgery

Routine small incision cataract surgery with unilateral IOL implantation

Sponsors

Alcon Research
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
22 Years to No maximum
Healthy volunteers
No

Inclusion criteria

Key Inclusion Criteria: * Diagnosed with bilateral cataracts * Planned small incision cataract removal surgery * Able to comprehend and willing to sign a statement of informed consent and complete all required post-implantation visits Key

Exclusion criteria

* Subjects who may reasonably be expected to require ocular surgical treatment or refractive surgical procedures at any time during the study * Certain eye conditions, as specified in the protocol, including but not limited to glaucoma, diabetic retinopathy, diabetic macular edema, and macular degeneration * Pregnant or lactating, current or planned during the course of the study Other protocol-defined inclusion/

Design outcomes

Primary

MeasureTime frameDescription
Percentage of First Implanted Eyes Achieving Best-Corrected Distance Visual Acuity (BCDVA) of 0.3 Logarithm Minimum Angle of Resolution (logMAR) or Better at 1 Year Post-ImplantationYear 1 post-implantation from second eye surgeryVisual acuity (VA) was tested for the first implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic Early Treatment Diabetic Retinopathy Study (ETDRS) charts at a distance of 4 meters (m) from the eye. VA was measured in logarithm of the minimum angle of resolution (logMAR), with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. VA was compared to the historical safety and performance endpoint (SPE) of 92.5% \[as reported in the European Standard International Organization for Standardization (EN ISO) 11979-7:2014\].
Percentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 1 Year Post-ImplantationYear 1 post-implantation from second eye surgeryVA was tested for the second implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. VA was compared to the historical SPE of 92.5% (EN ISO 11979-7:2014).

Secondary

MeasureTime frameDescription
Percentage of First Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 3 Years Post-ImplantationYear 3 post-implantation from second eye surgeryVA was tested for the first implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.
Percentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 3 Years Post-ImplantationYear 3 post-implantation from second eye surgeryVA was tested for the second implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.
Uncorrected Distance Visual Acuity (UCDVA) at 1 Year Post-Implantation - First Implanted EyeYear 1 post-implantation from second eye surgeryVA was tested for the first implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 diopter (D) spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.
UCDVA at 1 Year Post-Implantation - Second Implanted EyeYear 1 post-implantation from second eye surgeryVA was tested for the second implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.
Percentage of First Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 2 Years Post-ImplantationYear 2 post-implantation from second eye surgeryVA was tested for the first implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.
UCDVA at 2 Years Post-Implantation - Second Implanted EyeYear 2 post-implantation from second eye surgeryVA was tested for the second implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.
UCDVA at 3 Years Post-Implantation - First Implanted EyeYear 3 post-implantation from second eye surgeryVA was tested for the first implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.
UCDVA at 3 Years Post-Implantation - Second Implanted EyeYear 3 post-implantation from second eye surgeryVA was tested for the second implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.
UCDVA at 2 Years Post-Implantation - First Implanted EyeYear 2 post-implantation from second eye surgeryVA was tested for the first implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.
Percentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 2 Years Post-ImplantationYear 2 post-implantation from second eye surgeryVA was tested for the second implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.

Countries

Australia, France, Germany, Italy, Netherlands, Spain, United Kingdom

Participant flow

Recruitment details

Study participants were recruited from 19 study centers located in Australia (n = 3), France (n = 2), Germany (n = 2), Italy (n = 2), the Netherlands (n = 2), Spain (n = 5), and the United Kingdom (n = 3).

Pre-assignment details

Of the 245 enrolled, 30 participants were discontinued from the study prior to attempted implantation with the IOL. This reporting group includes all participants with attempted implantation (215).

Participants by arm

ArmCount
Clareon IOL
Clareon® aspheric hydrophobic acrylic intraocular lens implanted with the Alcon Monarch III-D delivery system as a replacement of the human crystalline lens during cataract surgery
215
Total215

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyAdverse Event1
Overall StudyCOVID related8
Overall StudyDeath10
Overall StudyLost to Follow-up4
Overall StudyWithdrawal by Subject9

Baseline characteristics

CharacteristicClareon IOL
Age, Continuous72.1 years
STANDARD_DEVIATION 8.07
Ethnicity (NIH/OMB)
Hispanic or Latino
17 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
195 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
3 Participants
Race/Ethnicity, Customized
American Indian or Alaska Native
0 Participants
Race/Ethnicity, Customized
Asian
6 Participants
Race/Ethnicity, Customized
Black or African American
3 Participants
Race/Ethnicity, Customized
More than one race
0 Participants
Race/Ethnicity, Customized
Native Hawaiian or Other Pacific Islander
1 Participants
Race/Ethnicity, Customized
Other
1 Participants
Race/Ethnicity, Customized
White
204 Participants
Sex: Female, Male
Female
126 Participants
Sex: Female, Male
Male
89 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
EG002
affected / at risk
EG003
affected / at risk
deaths
Total, all-cause mortality
0 / 2150 / 2150 / 20910 / 215
other
Total, other adverse events
0 / 21534 / 21539 / 2090 / 215
serious
Total, serious adverse events
0 / 2156 / 2154 / 20931 / 215

Outcome results

Primary

Percentage of First Implanted Eyes Achieving Best-Corrected Distance Visual Acuity (BCDVA) of 0.3 Logarithm Minimum Angle of Resolution (logMAR) or Better at 1 Year Post-Implantation

Visual acuity (VA) was tested for the first implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic Early Treatment Diabetic Retinopathy Study (ETDRS) charts at a distance of 4 meters (m) from the eye. VA was measured in logarithm of the minimum angle of resolution (logMAR), with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. VA was compared to the historical safety and performance endpoint (SPE) of 92.5% \[as reported in the European Standard International Organization for Standardization (EN ISO) 11979-7:2014\].

Time frame: Year 1 post-implantation from second eye surgery

Population: All Implanted Analysis Set with data

ArmMeasureValue (NUMBER)
Clareon IOLPercentage of First Implanted Eyes Achieving Best-Corrected Distance Visual Acuity (BCDVA) of 0.3 Logarithm Minimum Angle of Resolution (logMAR) or Better at 1 Year Post-Implantation99.5 percentage of first implanted eyes
Primary

Percentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 1 Year Post-Implantation

VA was tested for the second implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. VA was compared to the historical SPE of 92.5% (EN ISO 11979-7:2014).

Time frame: Year 1 post-implantation from second eye surgery

Population: All Implanted Analysis Set with data

ArmMeasureValue (NUMBER)
Clareon IOLPercentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 1 Year Post-Implantation99.5 percentage of second implanted eyes
Secondary

Percentage of First Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 2 Years Post-Implantation

VA was tested for the first implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.

Time frame: Year 2 post-implantation from second eye surgery

Population: All Implanted Analysis Set with data

ArmMeasureValue (NUMBER)
Clareon IOLPercentage of First Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 2 Years Post-Implantation99.5 percentage of second implanted eyes
Secondary

Percentage of First Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 3 Years Post-Implantation

VA was tested for the first implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.

Time frame: Year 3 post-implantation from second eye surgery

Population: All Implanted Analysis Set with data

ArmMeasureValue (NUMBER)
Clareon IOLPercentage of First Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 3 Years Post-Implantation99.5 percentage of first implanted eyes
Secondary

Percentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 2 Years Post-Implantation

VA was tested for the second implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.

Time frame: Year 2 post-implantation from second eye surgery

Population: All Implanted Analysis Set with data

ArmMeasureValue (NUMBER)
Clareon IOLPercentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 2 Years Post-Implantation98.9 percentage of second implanted eyes
Secondary

Percentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 3 Years Post-Implantation

VA was tested for the second implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.

Time frame: Year 3 post-implantation from second eye surgery

Population: All Implanted Analysis Set

ArmMeasureValue (NUMBER)
Clareon IOLPercentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 3 Years Post-Implantation98.9 percentage of second implanted eyes
Secondary

UCDVA at 1 Year Post-Implantation - Second Implanted Eye

VA was tested for the second implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.

Time frame: Year 1 post-implantation from second eye surgery

Population: All Implanted Analysis Set with data

ArmMeasureValue (MEAN)Dispersion
Clareon IOLUCDVA at 1 Year Post-Implantation - Second Implanted Eye0.058 logMARStandard Deviation 0.1364
Secondary

UCDVA at 2 Years Post-Implantation - First Implanted Eye

VA was tested for the first implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.

Time frame: Year 2 post-implantation from second eye surgery

Population: All Implanted Analysis Set with data

ArmMeasureValue (MEAN)Dispersion
Clareon IOLUCDVA at 2 Years Post-Implantation - First Implanted Eye0.088 logMARStandard Deviation 0.1533
Secondary

UCDVA at 2 Years Post-Implantation - Second Implanted Eye

VA was tested for the second implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.

Time frame: Year 2 post-implantation from second eye surgery

Population: All Implanted Analysis Set with data

ArmMeasureValue (MEAN)Dispersion
Clareon IOLUCDVA at 2 Years Post-Implantation - Second Implanted Eye0.083 logMARStandard Deviation 0.1624
Secondary

UCDVA at 3 Years Post-Implantation - First Implanted Eye

VA was tested for the first implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.

Time frame: Year 3 post-implantation from second eye surgery

Population: All Implanted Analysis Set

ArmMeasureValue (MEAN)Dispersion
Clareon IOLUCDVA at 3 Years Post-Implantation - First Implanted Eye0.091 logMARStandard Deviation 0.1614
Secondary

UCDVA at 3 Years Post-Implantation - Second Implanted Eye

VA was tested for the second implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.

Time frame: Year 3 post-implantation from second eye surgery

Population: All Implanted Analysis Set with data

ArmMeasureValue (MEAN)Dispersion
Clareon IOLUCDVA at 3 Years Post-Implantation - Second Implanted Eye0.086 logMARStandard Deviation 0.1706
Secondary

Uncorrected Distance Visual Acuity (UCDVA) at 1 Year Post-Implantation - First Implanted Eye

VA was tested for the first implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 diopter (D) spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.

Time frame: Year 1 post-implantation from second eye surgery

Population: All Implanted Analysis Set with data

ArmMeasureValue (MEAN)Dispersion
Clareon IOLUncorrected Distance Visual Acuity (UCDVA) at 1 Year Post-Implantation - First Implanted Eye0.079 logMARStandard Deviation 0.1499

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026