Skip to content

Hemospec Device for the Diagnosis of Sepsis

Integration of Clinical and Laboratory Information to Generate Technological Advance for the Diagnosis of Sepsis

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03306186
Acronym
INTELLIGENCE-1
Enrollment
183
Registered
2017-10-10
Start date
2017-10-09
Completion date
2018-03-31
Last updated
2018-05-15

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Sepsis

Keywords

diagnosis, prognosis

Brief summary

A diagnostic devise, namely HemoSpec, had been developed that integrates clinical information, along with information on circulating protein biomarkers and the morphology of white blood cells to achieve early diagnosis of sepsis. The current study is aiming to validate and improve performance of HemoSpec for the rapid assessment of the critically ill patient.

Detailed description

Sepsis is a life-threating organ dysfunction resulting from the dysregulated response of the host to an infection. It is estimated that 1.5 million people present with sepsis annually in Northern America and another 1.5 million people in Europe; 30 to 50% of them die making sepsis the leading cause of death. The key-point in the management of sepsis is the early resuscitation with broad-spectrum antimicrobials and intravenous fluids, if possible within the first hour. The great mortality of sepsis indicates that this goal is not easy to be achieved for two main reasons: the first is the delay in recognition of the septic patients and the second is the resistance of the implicated pathogen to broad-spectrum antimicrobials. In an attempt to improve the failure of physicians for early sepsis recognition, several markers have been developed. Some of them rely on clinical signs of the host and others on the measurements of circulating biomarkers. Recently, qSOFA (quick SOFA score) has been introduced to help the early recognition of sepsis in patients who present with infection outside the Intensive care Unit (ICU) i.e. either in the community or during hospitalization in the general ward1. However, there are concerns of the sensitivity of qSOFA and many introduce the need to measure biomarkers in serum. These biomarkers are usually protein molecules that are over-produced in the host as a result of the interaction with an infective insult. However, these protein molecules are produced by white blood cells. What is currently known is that although most of patients present with a similar phenotype, their pathophysiology is diverse. More precisely, although the majority of patients with sepsis present with high concentrations of protein molecules like interleukin (IL)-6, C-reactive protein (CRP) and procalcitonin (PCT) in their blood, in some patients circulating white blood cells remain over-active and in other patients they are significantly anergic, a situation often known as sepsis-induced immunoparalysis. Another molecule, called soluble urokinase plasminogen activator receptor (suPAR), is the shed uPAR receptor on neutrophils and is released in the circulation as a result of neutrophil activation; concentrations greater than 12 ng/ml can trace with negative predictive value almost 95% the patient at great chance of unfavorable outcome. As such, the robust diagnosis of sepsis may rely on a combination of clinical assessment, measurement of protein biomarkers and validation of the activity of circulating white blood cells. One FrameWork 7-funded initiative from seven European countries aims to develop a rapid score that can integrate all clinical and laboratory information and provide early diagnosis whether a patient has sepsis or not. The vision of this initiative is to build a device that is called HemoSpec. With this approach, whole blood coming from patients will be in parallel analyzed into three aspects: a) absolute white blood cell counting; b) information on the fluidity and activity of the white blood cells using Raman spectroscopy; and c) measurement of serum levels of IL-6, CRP, PCT and suPAR. The end result is building a diagnostic algorithm where clinical information is also taken into consideration. The project was started in November 2013 and the HemoSpec device is anticipated to be ready by February 2017. The diagnostic performance of HemoSpec is currently based on preliminary data coming from 60 patients (20 controls, 20 with systemic inflammatory response syndrome and 20 with sepsis) hospitalized in Jena University Hospital. The current study is aiming to validate and improve performance of HemoSpec for the rapid assessment of the critically ill patient in a larger phase II diagnostic study.

Interventions

DEVICEHemoSpec

Blood sampling for analysis

Sponsors

Ippokration General Hospital
CollaboratorOTHER
Tzanion General Hospital of Piraeus
CollaboratorUNKNOWN
Aghia Olga Konstantopouleion General Hospital
CollaboratorUNKNOWN
University of Athens
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Age above or equal to 18 years old * Both genders * Written consent provided from patients or their first-degree relatives for patients unable to consent * Patients with acute pancreatitis or post-operative or with clinical signs of infection * Considerable risk of death as as indicated by the presence of at least one of the following: i) sudden alteration of mental status; ii) systolic blood pressure less than 100 mmHg; and iii) high respiratory rate defined as more than or equal to 22 breaths per minute.

Exclusion criteria

* Known infection by the human immunodeficiency virus-1; * Neutropenia defined as an absolute neutrophil count lower than 1000 neutrophils/mm3 due to reasons other than an infection. * Single trauma or multiple injuries

Design outcomes

Primary

MeasureTime frameDescription
Sensitivity of HemoSpec for the diagnosis of sepsis3 daysThe sensitivity of HemoSpec output to diagnose the presence of sepsis compared to the absence of sepsis. HemoSpec output will be considered to provide a satisfactory diagnosis of sepsis if sensitivity for the diagnosis is greater than 85%.

Secondary

MeasureTime frameDescription
Prognostic performance for organ dysfunction28 daysThe prognostic performance of HemoSpec output to predict progression into organ dysfunction. The prognostic performance is composed by the aggregation of sensitivity, specificity, positive predictive value and negative predictive value.
Diagnostic performance over qSOFA3 daysThe aggregation of sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) of HemoSpec output to indicate patients with infection among those scoring positive for qSOFA.
Diagnostic performance for sepsis3 daysThe diagnostic performance of HemoSpec output to diagnose the presence of sepsis compared to the absence of sepsis. The diagnostic performance is composed by the aggregation of specificity, positive predictive value and negative predictive value.
Prognostics performance for sepsis28 daysThe prognostic performance of HemoSpec output to predict unfavorable outcome compared to survivors. The prognostic performance is composed by the aggregation of sensitivity, specificity, positive predictive value and negative predictive value.
Diagnostic performance and microbiology3 daysThe diagnostic performance of HemoSpec output to diagnose the presence of sepsis between patients with microbiologically-proven infection and patients without microbiologically-proven infection. The diagnostic performance is composed by the aggregation of sensitivity, specificity, positive predictive value and negative predictive value.

Countries

Greece

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026