The Effectiveness of SensAwake™ for Continue Positive Airway Pressure Treatment of Obstructive Sleep Apnea Patients

Effects of Pressure Control Device (SensAwake™) on Obstructive Sleep Apnea (OSA) Patients Who Remove the Mask for Unknown Reasons During Automatic Continuous Positive Airway Pressure (Auto-CPAP) Therapy

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03294629
Enrollment
35
Registered
2017-09-27
Start date
2015-03-31
Completion date
2016-03-31
Last updated
2018-06-12

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Obstructive Sleep Apnea

Keywords

cPAP, SenseAwake

Brief summary

This study will investigate Obstructive Sleep Apnea (OSA) and an automatically Adjusting Positive Airway Pressure (APAP) device with new technology called SensAwake™. This requires experimental confirmation in a randomised controlled trial with crossover design, comparing compliance on standard APAP with compliance using APAP modified by the addition of the SensAwake™ modification on consecutive nights in participants with moderate-to-severe OSA.

Interventions

DEVICESensAwake™ modification

SensAwake™ is a new design based on the research of Doctor Ayappa 5 years ago. The SensAwake™ modification to the Fisher and Paykel automatically titrating positive airway pressure (APAP) device aims to sense whether the patient is awake via respiratory patterns that differentiate between sleep and wake. Upon sensing that the patient is awake the device is able to reduce positive airway pressure PAP aiming to improve patient comfort which should result in more consolidated sleep.

Sponsors

Fisher and Paykel Healthcare
CollaboratorINDUSTRY
Chang Gung Memorial Hospital
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
TREATMENT
Masking
DOUBLE (Subject, Investigator)

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

1\. Participants will be recruited from OSA patients who will remove CPAP himself without any notable reasons.

Exclusion criteria

1. Unstable cardiovascular disease (untreated or resistant hypertension acceptable). 2. Inability to tolerate CPAP due to nasal obstruction or claustrophobia as determined by the study investigator. 3. Any known factor or disease that might interfere with treatment compliance, study conduct or interpretation of the results such as psychiatric disease, history of non compliance to medical regimens, or unwillingness to comply with study requirements. 4. Any known factor that result in mask taken of such as skin discomfort, nasal obstruction, water dripping inside the tubing system etc.

Design outcomes

Primary

MeasureTime frameDescription
Compliance of Auto-CPAP Therapy2 weeksObjective data recorded in the auto-CPAP machine, Percentage of nights of using auto-CPAP for at least 4 hours during therapy.

Secondary

MeasureTime frameDescription
Average Pressures of Auto-CPAP Machine2 weeksAverage Pressures of Auto-CPAP Machine (Objective data recorded in the auto-CPAP machine)
90th Percentile Pressures of Auto-CPAP Machine2 weeks90th percentile pressures of auto-CPAP machine (Objective data recorded in the auto-CPAP machine)
Average Leaks of Auto-CPAP Machine2 weeksAverage leaks of auto-CPAP machine (Objective data recorded in the auto-CPAP machine)
90th Percentile Leaks of Auto-CPAP Machine2 weeksAverage leaks of auto-CPAP machine (Objective data recorded in the auto-CPAP machine)
Residual Apnea-Hypopnea Index (AHI)2 weeksResidual Apnea-Hypopnea index (Objective data recorded in the auto-CPAP machine) means events of apnea and hypopnea per hour during CPAP therapy.
Average SA Detections2 weeksAverage detects of arousal per hours. (Objective data recorded in the auto-CPAP machine)
Percentage of Day Used of Auto-CPAP Machine2 weeksPercentage of day used of auto-CPAP machine (Objective data recorded in the auto-CPAP machine)
Pittsburgh Sleep Quality Index (PSQI)2 weeksThe Pittsburgh Sleep Quality Index (PSQI) is a self-report questionnaire that assesses sleep quality over a 1-month time interval. Overall score is ranging from 0 to 21, where lower scores denote a healthier sleep quality.
Epworth Sleepiness Score (ESS)2 weeksThe ESS is a self-administered questionnaire with 8 questions. The ESS score can range from 0 to 24, where higher scores denote a worse daytime sleepiness.
Nasal Obstruction Symptom Evaluation (NOSE)2 weeksThe Nasal Obstruction Symptom Evaluation (NOSE) is a self-administered questionnaire with 11 questions. The NOSE score can range from 0 to 44, where higher scores denote a worse nasal obstruction.
Time Used of Auto-CPAP Machine2 weeksAverage Time used of auto-CPAP machine (Objective data recorded in the auto-CPAP machine)

Participant flow

Recruitment details

March 2015 to Mar 2016, at medical clinic.

Participants by arm

ArmCount
CPAP w/SensAwake First, Then CPAP w/o SensAwake
Patients will receive CPAP treatment with SensAwake™ activation for two weeks, then crossed-over to CPAP treatment without SensAwake™ activation for additional two weeks. No washout period.
12
CPAP w/o SensAwake First, Then CPAP w/SensAwake
Patients will receive CPAP treatment without SensAwake™ activation for two weeks, then crossed-over to CPAP treatment with SensAwake™ activation for additional two weeks. No washout period.
13
Total25

Baseline characteristics

CharacteristicCPAP w/o SensAwake First, Then CPAP w/SensAwakeCPAP w/SensAwake First, Then CPAP w/o SensAwakeTotal
Age, Continuous41.5 years
STANDARD_DEVIATION 12
47.6 years
STANDARD_DEVIATION 10.9
44.0 years
STANDARD_DEVIATION 11.7
Apnea-hypopnea index (AHI)54.0 events per hour
STANDARD_DEVIATION 35.3
55.3 events per hour
STANDARD_DEVIATION 30.8
54.6 events per hour
STANDARD_DEVIATION 32.5
Average used time of CPAP227.9 minutes
STANDARD_DEVIATION 104.1
187.4 minutes
STANDARD_DEVIATION 104.6
208.5 minutes
STANDARD_DEVIATION 104.2
Body Mass Index (BMI)32.9 kg/m2
STANDARD_DEVIATION 6.9
31.8 kg/m2
STANDARD_DEVIATION 4.3
32.4 kg/m2
STANDARD_DEVIATION 5.7
Compliance of CPAP46.2 %
STANDARD_DEVIATION 30.6
31.9 %
STANDARD_DEVIATION 31.9
39.3 %
STANDARD_DEVIATION 31.4
Deep sleep - Stage 34.4 % (percentage of minutes of stage 3)
STANDARD_DEVIATION 5.6
6.1 % (percentage of minutes of stage 3)
STANDARD_DEVIATION 10.4
5.2 % (percentage of minutes of stage 3)
STANDARD_DEVIATION 8
Epworth Sleepiness Scale (ESS)13.1 units on a scale
STANDARD_DEVIATION 6.7
13.3 units on a scale
STANDARD_DEVIATION 5.5
13.2 units on a scale
STANDARD_DEVIATION 6
Nasal Obstruction Symptom Evaluation (NOSE)15.6 units on a scale
STANDARD_DEVIATION 8.4
13.5 units on a scale
STANDARD_DEVIATION 6.8
14.6 units on a scale
STANDARD_DEVIATION 7.6
Neck Circumference (NC)41.0 cm
STANDARD_DEVIATION 2.2
42.3 cm
STANDARD_DEVIATION 2.1
41.6 cm
STANDARD_DEVIATION 2.2
Pittsburgh Sleep Quality Index (PSQI)10.1 units on a scale
STANDARD_DEVIATION 3.8
7.4 units on a scale
STANDARD_DEVIATION 2.3
8.8 units on a scale
STANDARD_DEVIATION 3.4
Residual AHI under CPAP Tx4.4 events per hour
STANDARD_DEVIATION 2.8
2.9 events per hour
STANDARD_DEVIATION 1.3
3.7 events per hour
STANDARD_DEVIATION 2.3
Sex: Female, Male
Female
3 Participants0 Participants3 Participants
Sex: Female, Male
Male
10 Participants12 Participants22 Participants
Sleep efficiency73.0 % (percentage of minutes slept)
STANDARD_DEVIATION 21.3
82.0 % (percentage of minutes slept)
STANDARD_DEVIATION 8.9
77.1 % (percentage of minutes slept)
STANDARD_DEVIATION 17.1
Total sleep time276.2 minutes
STANDARD_DEVIATION 79.2
301.0 minutes
STANDARD_DEVIATION 34.2
287.6 minutes
STANDARD_DEVIATION 62.8

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 250 / 25
other
Total, other adverse events
0 / 250 / 25
serious
Total, serious adverse events
0 / 250 / 25

Outcome results

Primary

Compliance of Auto-CPAP Therapy

Objective data recorded in the auto-CPAP machine, Percentage of nights of using auto-CPAP for at least 4 hours during therapy.

Time frame: 2 weeks

ArmMeasureValue (MEAN)Dispersion
SA ONCompliance of Auto-CPAP Therapy41.8 percentage of nightsStandard Deviation 35
SA OFFCompliance of Auto-CPAP Therapy39.7 percentage of nightsStandard Deviation 32.6
Secondary

90th Percentile Leaks of Auto-CPAP Machine

Average leaks of auto-CPAP machine (Objective data recorded in the auto-CPAP machine)

Time frame: 2 weeks

ArmMeasureValue (MEAN)Dispersion
SA ON90th Percentile Leaks of Auto-CPAP Machine57.6 liters per minuteStandard Deviation 24.7
SA OFF90th Percentile Leaks of Auto-CPAP Machine58.9 liters per minuteStandard Deviation 26
Secondary

90th Percentile Pressures of Auto-CPAP Machine

90th percentile pressures of auto-CPAP machine (Objective data recorded in the auto-CPAP machine)

Time frame: 2 weeks

ArmMeasureValue (MEAN)Dispersion
SA ON90th Percentile Pressures of Auto-CPAP Machine8.6 cmH2OStandard Deviation 3
SA OFF90th Percentile Pressures of Auto-CPAP Machine9.2 cmH2OStandard Deviation 2.9
Secondary

Average Leaks of Auto-CPAP Machine

Average leaks of auto-CPAP machine (Objective data recorded in the auto-CPAP machine)

Time frame: 2 weeks

ArmMeasureValue (MEAN)Dispersion
SA ONAverage Leaks of Auto-CPAP Machine41.3 liters per minuteStandard Deviation 20.1
SA OFFAverage Leaks of Auto-CPAP Machine44.6 liters per minuteStandard Deviation 22.5
Secondary

Average Pressures of Auto-CPAP Machine

Average Pressures of Auto-CPAP Machine (Objective data recorded in the auto-CPAP machine)

Time frame: 2 weeks

ArmMeasureValue (MEAN)Dispersion
SA ONAverage Pressures of Auto-CPAP Machine6.9 cmH2OStandard Deviation 2.7
SA OFFAverage Pressures of Auto-CPAP Machine7.3 cmH2OStandard Deviation 2.6
Secondary

Average SA Detections

Average detects of arousal per hours. (Objective data recorded in the auto-CPAP machine)

Time frame: 2 weeks

ArmMeasureValue (MEAN)Dispersion
SA ONAverage SA Detections2.1 times/hourStandard Deviation 0.9
SA OFFAverage SA Detections0 times/hourStandard Deviation 0
Secondary

Epworth Sleepiness Score (ESS)

The ESS is a self-administered questionnaire with 8 questions. The ESS score can range from 0 to 24, where higher scores denote a worse daytime sleepiness.

Time frame: 2 weeks

ArmMeasureValue (MEAN)Dispersion
SA ONEpworth Sleepiness Score (ESS)11.1 units on a scaleStandard Deviation 6.1
SA OFFEpworth Sleepiness Score (ESS)12.1 units on a scaleStandard Deviation 7.2
Secondary

Nasal Obstruction Symptom Evaluation (NOSE)

The Nasal Obstruction Symptom Evaluation (NOSE) is a self-administered questionnaire with 11 questions. The NOSE score can range from 0 to 44, where higher scores denote a worse nasal obstruction.

Time frame: 2 weeks

ArmMeasureValue (MEAN)Dispersion
SA ONNasal Obstruction Symptom Evaluation (NOSE)14.6 units on a scaleStandard Deviation 8.1
SA OFFNasal Obstruction Symptom Evaluation (NOSE)13.3 units on a scaleStandard Deviation 6.7
Secondary

Percentage of Day Used of Auto-CPAP Machine

Percentage of day used of auto-CPAP machine (Objective data recorded in the auto-CPAP machine)

Time frame: 2 weeks

ArmMeasureValue (MEAN)Dispersion
SA ONPercentage of Day Used of Auto-CPAP Machine87.1 Percentage of daysStandard Deviation 13.8
SA OFFPercentage of Day Used of Auto-CPAP Machine88.1 Percentage of daysStandard Deviation 11.2
Secondary

Pittsburgh Sleep Quality Index (PSQI)

The Pittsburgh Sleep Quality Index (PSQI) is a self-report questionnaire that assesses sleep quality over a 1-month time interval. Overall score is ranging from 0 to 21, where lower scores denote a healthier sleep quality.

Time frame: 2 weeks

ArmMeasureValue (MEAN)Dispersion
SA ONPittsburgh Sleep Quality Index (PSQI)8.0 units on a scaleStandard Deviation 2.6
SA OFFPittsburgh Sleep Quality Index (PSQI)8.0 units on a scaleStandard Deviation 2.8
Secondary

Residual Apnea-Hypopnea Index (AHI)

Residual Apnea-Hypopnea index (Objective data recorded in the auto-CPAP machine) means events of apnea and hypopnea per hour during CPAP therapy.

Time frame: 2 weeks

ArmMeasureValue (MEAN)Dispersion
SA ONResidual Apnea-Hypopnea Index (AHI)6.3 events per hourStandard Deviation 6.7
SA OFFResidual Apnea-Hypopnea Index (AHI)5.0 events per hourStandard Deviation 4.6
Secondary

Time Used of Auto-CPAP Machine

Average Time used of auto-CPAP machine (Objective data recorded in the auto-CPAP machine)

Time frame: 2 weeks

ArmMeasureValue (MEAN)Dispersion
SA ONTime Used of Auto-CPAP Machine213.1 minutesStandard Deviation 99.4
SA OFFTime Used of Auto-CPAP Machine206.8 minutesStandard Deviation 87.6