Non-invasive Tidal Volume Monitoring Using the Linshom Respiratory Monitoring Device

Non-invasive Tidal Volume Monitoring Using the Linshom Respiratory Monitoring Device

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03279458
Enrollment
40
Registered
2017-09-12
Start date
2017-09-08
Completion date
2017-09-08
Last updated
2018-07-05

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Capnography, Pulmonary Ventilation

Keywords

Capnography

Brief summary

Many post-operative complications arise from patients who breathe inadequately. Inadequate respiration, whether the result of surgery or the anesthesia, causes a decrease in blood oxygen saturation and an increase in carbon dioxide partial pressure. Both of these surrogate measurements of respiration may pose a challenge to measure. Some administer exogenous oxygen to all patients as they leave the operating room in order to maintain the blood oxygen saturation. This renders the oximeter a less sensitive metric of depressed respiration. In the face of decreased respiration, the carbon dioxide levels continue to increase slowly and often go undetected unless blood gases are measured. Indeed carbon dioxide blood levels are the only metric to detect inadequate ventilation using this surrogate index. Monitoring ventilation is a serious challenge outside of critical care settings. In fact, there are no monitors available that can measure tidal volume or relative tidal volume outside of these settings. Linshom is a novel instrument that tracks relative respiration by measuring the excursions of the temperature swings between inspiration and expiration and normalizing them to the patient's breathing. This monitor may be the first non-invasive monitor to measure relative tidal volume in non-critical care settings. The purpose of this study is to determine whether a non-invasive, temperature-based respiratory instrument can track tidal volume (Vt) in patients. The investigators hypothesize that the Linshom device can accurately and consistently track tidal volume as measured by closed loop mechanical ventilator.

Detailed description

Forty volunteers will be enrolled in this open label, prospective study to determine the correlation between the tidal volumes measured with the Linshom detector and those measured with the mechanical ventilator. after Institutional Review Board (IRB) approval and written informed consent from the volunteers. The Linshom detector, which is comprised of two rapid responding medical-grade thermistors in close proximity to the mouth/nose (sensor) and a thermistor remote to the airway, will be mounted in the continuous positive airway pressure (CPAP) face mask to measure the temperature during breathing. The CPAP will be connected to a Servo-I ventilator (Maquet) with a circuit and disposable filter. The volunteers will be instructed to breathe normal through the CPAP mask on room air. The excursions of the thermistor tracings (from valley to peak) will be recorded by the Linshom device and displayed continuously on a laptop monitor in a waveform. The tidal volume will also be measured by the ventilator and the data downloaded in a Compact Flash card. The temperature profiles from the sensors and the relative tidal volume will be correlated with the tidal volumes measured by the ventilator.

Interventions

DEVICELinshom Respiratory Monitoring Device

Volunteers will breathe through a continuous positive airway pressure (CPAP) face mask fitted wth the Linshom device. The volunteers will be instructed to breathe normal through the CPAP mask on room air. The excursions of the thermistor tracings (from valley to peak) will be recorded by the Linshom device and displayed continuously on a laptop monitor in a waveform. The tidal volume will also be measured by the ventilator and the data downloaded in a Compact Flash card.

Sponsors

University of Mississippi Medical Center
Lead SponsorOTHER

Study design

Allocation
NON_RANDOMIZED
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* healthy

Exclusion criteria

* pregnant * suffer from claustrophobia * had recent respiratory illness * had recent gastrointestinal illness * unable to provide informed consent

Design outcomes

Primary

MeasureTime frameDescription
Tidal Volume of Each Breath Was Measured Using the Same Breathing Mask CPAP Apparatus Combined With Respiratory Monitoring Device Allowing for Simultaneous Recording.5 minutesTidal volume measured by CPAP mask with Lishom respiratory monitoring device. Tidal volume determined from the ventilator will be compared to tidal volume measured by Linshom (calculated indirectly by temperature changes by computer algorithm)

Countries

United States

Participant flow

Participants by arm

ArmCount
Linshom
Linshom Respiratory Monitoring Device: Volunteers will breathe through a continuous positive airway pressure (CPAP) face mask fitted wth the Linshom device. The volunteers will be instructed to breathe normal through the CPAP mask on room air. The excursions of the thermistor tracings (from valley to peak) will be recorded by the Linshom device and displayed continuously on a laptop monitor in a waveform. The tidal volume will also be measured by the ventilator and the data downloaded in a Compact Flash card.
40
Total40

Baseline characteristics

CharacteristicLinshom
Age, Customized
>= 18 years
40 Participants
Number of participants40 Participants
Race and Ethnicity Not Collected— Participants
Region of Enrollment
United States
40 participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 400 / 40
other
Total, other adverse events
0 / 400 / 40
serious
Total, serious adverse events
0 / 400 / 40

Outcome results

Primary

Tidal Volume of Each Breath Was Measured Using the Same Breathing Mask CPAP Apparatus Combined With Respiratory Monitoring Device Allowing for Simultaneous Recording.

Tidal volume measured by CPAP mask with Lishom respiratory monitoring device. Tidal volume determined from the ventilator will be compared to tidal volume measured by Linshom (calculated indirectly by temperature changes by computer algorithm)

Time frame: 5 minutes

Population: Thirty participants successfully completed the experiment, which involved breathing through a CPAP mask in order to have the Linshom device record excursions of the thermistor tracings while the ventilator concurrently measured tidal volume.

ArmMeasureValue (LEAST_SQUARES_MEAN)
LinshomTidal Volume of Each Breath Was Measured Using the Same Breathing Mask CPAP Apparatus Combined With Respiratory Monitoring Device Allowing for Simultaneous Recording.0.94 ml
Maquet VentilatorTidal Volume of Each Breath Was Measured Using the Same Breathing Mask CPAP Apparatus Combined With Respiratory Monitoring Device Allowing for Simultaneous Recording.0.94 ml
p-value: <0.0595% CI: [0.88, 0.97]Regression, Linear