Icare Tonometry Effects on Keratometry Readings, Topography Readings, and Corneal Staining

Icare Tonometry Effects on Keratometry Readings, Topography Readings, and Corneal Staining

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03275142
Enrollment
60
Registered
2017-09-07
Start date
2017-06-19
Completion date
2018-07-31
Last updated
2019-01-30

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Evaluation of Corneal Stability Post-applanation

Brief summary

To demonstrate that the Icare tonometer does not alter Keratometry readings, topography, or corneal staining. This would therefore allow intraocular pressure (IOP) testing to be done at any time during an exam, without affecting other testing. This may improve office-flow and spare patients from returning for another exam for additional testing.

Detailed description

Four sites will enroll fifteen subjects each. Individual sites will identify their patients using numbers 1-15. The link to their identity will be stored in their records but will not be forwarded to other investigators. Only the originating site will have the identity log linking the participants to the study records. All patients over age 18 in the patient clinic are eligible for inclusion. Staff members and employees may also participate. Following informed consent, each patient will undergo visual acuity testing, keratometry readings by IOL Master, topography testing by Pentacam, and corneal staining evaluation with fluorescein using the Oxford scoring scale on both eyes. Once all three tests are completed, the Icare tonometer will be used to check the intraocular pressure (IOP) on one eye only. The investigator will be masked as to which eye will undergo applanation. Immediately following the IOP check, corneal staining, topography and keratometry readings will be repeated. The unchecked eye will serve as a control as there may be mild variations in topography or IOL Master even without applanation. Following the repeated testing, the control eye will also undergo applanation to document IOP.

Interventions

Icare tonometry effects of keratometry readings, topography, and corneal staining to demonstrate that the Icare tonometer does not alter subsequent evaluation or testing. This would allow IOP testing to be performed by technical staff at any time during the examination process without affecting subsequent tests. A total of fifteen patients at each of four sites will be evaluated using Icare tonometers provided by Icare USA. The investigator will be masked as to which eye will undergo Icare tonometry. Both the study eye and the control eye will undergo pre-tonometry and post-tonometry testing; Keratometry readings, IOL master, Topography via Pentacam, and corneal staining with fluorescein using the Oxford scoring scale.

Sponsors

Icare USA
CollaboratorUNKNOWN
Comprehensive EyeCare of Central Ohio
Lead SponsorOTHER

Study design

Allocation
NON_RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
BASIC_SCIENCE
Masking
DOUBLE (Investigator, Outcomes Assessor)

Masking description

Only the Investigator will be interpreting outcomes. The technician/care provider will perform all testing and will mask the investigator to which eye tonometry was performed.

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* 18 years of age

Exclusion criteria

* Under 18 years of age

Design outcomes

Primary

MeasureTime frameDescription
Demonstration of corneal stability post-applanation with the Icare tonometer.30 minutesDemonstration that the Icare tonometer does not alter keratometry readings performed by IOL Master, topography performed by Pentacam HR, or corneal staining as evaluated at slit-lamp with fluorescence as compared to the Oxford Grading Scale

Secondary

MeasureTime frameDescription
Average keratometric measurement differences pre and post- tonometry with ICare Tonometer.10 minutesEvaluation of average keratometric difference pre and post-tonometry with ICare via Pentacam HR and IOL Master.
Total cylinder differences pre and post-tonometry with ICare tonometer.10 minutesTotal cylinder differences pre and post-tonometry with ICare Tonometer via Pentacam HR and IOL Master.
Difference in corneal staining pre and post-tonometry with ICare Tonometer.10 minutesDifference in corneal staining pre and post-tonometry with ICare Tonometer as compared to the Oxford Scoring Scale.

Countries

United States

Outcome results

None listed