Refinement of the OBMedical LaborView System

Refinement of LaborView Electronic Fetal Monitor

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT03244865
Enrollment
35
Registered
2017-08-10
Start date
2016-09-29
Completion date
2018-04-11
Last updated
2019-09-20

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Fetal Monitoring

Brief summary

Despite its limitations, fetal heart rate (FHR) tracing analysis is the best monitor of fetal well-being during labor or preparation for cesarean delivery. Current monitoring methods: transabdominal ultrasound (CTG: cardiotocograph) and fetal scalp electrodes (FSE) have limitations (tracing loss during fetal movement, potential to confuse maternal for fetal heart rate, inability to monitor during cesarean section or abdominal surgery) and, in the latter case, risks (infection, hematoma). A reliable, non-invasive monitor of the fetal heart rate that can be used during labor, non-obstetric and obstetric surgery during pregnancy is needed. Recent developments in use of maternal abdominal electrodes for extraction of the FHR and contraction information inform this study. LaborView®, developed by researchers at University of Florida (UF) College of Medicine and OBMedical, is an FDA-approved non-invasive labor monitoring device. Refinement of the extraction algorithms continues, and that work is most effectively performed at UF, where the primary researchers and OBMedical are located. LaborView monitoring will be recorded from up to150 subjects admitted to Labor and Delivery at UFHealth. Data will be simultaneously obtained from the standard labor monitor. These signals will then be analyzed off-line for comparison and refinement of algorithms.

Detailed description

Despite its limitations, fetal heart rate (FHR) tracing analysis is the best monitor of fetal well being during labor. Typically, continuous electronic fetal monitoring is via transabdominal ultrasound (CTG: cardiotocograph). A transducer is strapped to the maternal abdomen and positioned in such a way to detect fetal rather than maternal heart rate. There are occasional periods of signal loss as the patient or her fetus moves, as well as times when the monitor incorrectly records the maternal signal. Maternal obesity presents particular challenges to non-invasive monitoring. When there is particular concern regarding the fetal heart rate tracing the obstetrician may opt for a fetal scalp electrode (FSE). This is screwed into the fetus' presenting part through the opening in the cervix (requires ruptured membranes and adequate dilation). This signal is usually more reliable, but entails risks of chorioamnionitis, fetal scalp infection, scalp hematoma, and potential vertical transmission of maternal viruses including HIV and hepatitis. Recently, devices have come to market that extract the FHR and uterine electromyograph (EMG) from maternal abdominal electrodes. The proposed study will provide data for refinement of the LaborView device produced by OBMedical. Specific Aims: The goal of this research is to evaluate the LaborView device under various settings and fetal heart rate accuracy as compared to a traditional CTG monitor. Research Plan: All patients admitted to the Labor & Delivery Suite at UF Health, will be eligible for inclusion in the study except minors and those unable to consent for themselves. Following written informed consent, the external monitors (tocodynamometer and ultrasound) may not be removed. Per LaborView directions, electrode application sites will be prepared to reduce skin impedance, electrodes will be applied to the maternal abdomen and data acquired for comparison with CTG data obtained through the traditional monitoring system. FHR, ECG (electrocardiography), and uterine activity data from the existing fetal monitors will be collected via a laptop PC connected to the systems for comparison. The comparison will be performed off-line. Evaluation of the data obtained will include (1) comparison of the calculated FHR with that of the CTG and/or fetal scalp electrode. Nothing is required of the subjects. They will be non-invasively monitored for one to several hours. The information obtained will not be used for medical decision-making and there is no significant risk to the patient or her fetus. No longitudinal follow-up is planned or indicated.

Interventions

DEVICELaborView Electronic Fetal Monitor

Per LaborView directions, electrode application sites will be prepared to reduce skin impedance, electrodes will be applied to the maternal abdomen and FHR data acquired for comparison with CTG data obtained through the traditional monitoring system. The comparison will be performed off-line. Nothing is required of the subjects. They will be non-invasively monitored for one to several hours. The information obtained will not be used for medical decision-making and there is no significant risk to the patient or her fetus.

Sponsors

University of Florida
CollaboratorOTHER
OBMedical Company
Lead SponsorOTHER

Study design

Observational model
OTHER
Time perspective
OTHER

Eligibility

Sex/Gender
FEMALE
Age
18 Years to 100 Years
Healthy volunteers
Yes

Inclusion criteria

* Admitted to Labor and Delivery at Shands UF Health * Are able to consent for themselves

Exclusion criteria

• Under the age of 18

Design outcomes

Primary

MeasureTime frameDescription
Fetal Heart Rate With the Laborview System Versus the CTG Monitor - Pilot Comparison Study1 - 8 hoursUsing standard of care CTG (Cardiotocography) already in place and the LaborView electrode-based FHR monitoring simultaneously, we will collect FHR data from both monitors simultaneously. We will compare the two FHR calculations using root mean squared error (RMSE).

Countries

United States

Participant flow

Participants by arm

ArmCount
Pregnant Females
All patients admitted to the Labor & Delivery Suite at UF Health, will be eligible for inclusion in the study except minors and those unable to consent for themselves. Nothing is required of the subjects. They will be non-invasively monitored for one to several hours. The information obtained will not be used for medical decision-making and there is no significant risk to the patient or her fetus. No longitudinal follow-up is planned or indicated. LaborView Electronic Fetal Monitor: Per LaborView directions, electrode application sites will be prepared to reduce skin impedance, electrodes will be applied to the maternal abdomen and data acquired for comparison with CTG data obtained through the traditional monitoring system. FHR, ECG, and uterine activity data from the existing fetal monitors will be collected via a laptop PC connected to the systems for comparison. The comparison will be performed off-line.
35
Total35

Baseline characteristics

CharacteristicPregnant Females
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
35 Participants
Age, Continuous27 years
STANDARD_DEVIATION 30.69
Number of Subjects in active labor35 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
Race (NIH/OMB)
Asian
0 Participants
Race (NIH/OMB)
Black or African American
9 Participants
Race (NIH/OMB)
More than one race
2 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
Race (NIH/OMB)
Unknown or Not Reported
3 Participants
Race (NIH/OMB)
White
21 Participants
Region of Enrollment
United States
35 participants
Sex: Female, Male
Female
35 Participants
Sex: Female, Male
Male
0 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 35
other
Total, other adverse events
0 / 35
serious
Total, serious adverse events
0 / 35

Outcome results

Primary

Fetal Heart Rate With the Laborview System Versus the CTG Monitor - Pilot Comparison Study

Using standard of care CTG (Cardiotocography) already in place and the LaborView electrode-based FHR monitoring simultaneously, we will collect FHR data from both monitors simultaneously. We will compare the two FHR calculations using root mean squared error (RMSE).

Time frame: 1 - 8 hours

ArmMeasureValue (MEAN)
Pregnant FemalesFetal Heart Rate With the Laborview System Versus the CTG Monitor - Pilot Comparison Study5.282 beats per minute
Comparison: There is only one ARM in the study. Standard monitoring and LaborView monitoring will be collected simultaneously and analyzed.p-value: <0.1t-test, 2 sided