Evaluation of Approved and Investigational Contact Lenses

Evaluation of Approved and Investigational Contact Lenses

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03228212
Enrollment
133
Registered
2017-07-24
Start date
2017-07-14
Completion date
2017-09-18
Last updated
2018-12-13

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Visual Acuity

Brief summary

This study is a randomized, 5-visit, partial subject masked, 2x3 bilateral crossover, dispensing clinical trial. During the study, each test article will be worn in a daily wear modality for at least 6 hours per day and 5 days per week for approximately 2 weeks each. The subject will wear either the Test or Control article twice and the other study article once. The primary objective of this study is to demonstrate that the Test lens works as well, if not better compared to the Control lens with respect to CLUE (Contact Lens User Experience) comfort, Slit Lamp findings, and Distance Monocular logMAR Visual Acuity. This study will also aim to show acceptable lens fit for subjects wearing the Test lens.

Interventions

DEVICECONTROL

ACUVUE OASYS

DEVICETEST

senofilcon A with new UV blocker

Sponsors

Johnson & Johnson Vision Care, Inc.
Lead SponsorINDUSTRY

Study design

Allocation
RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
TREATMENT
Masking
SINGLE (Subject)

Eligibility

Sex/Gender
ALL
Age
18 Years to 49 Years
Healthy volunteers
Yes

Inclusion criteria

* Potential subjects must satisfy all of the following inclusion criteria to be enrolled in the study: 1. The subject must read and sign the Informed Consent form. 2. The subject must appear able and willing to adhere to the instructions set forth in this clinical protocol. 3. Healthy adult males or females age ≥18 and ≤49 years of age with signed informed consent. Eligible presbyopes will be those that wear full distance contact lenses in both eyes, then wear reading glasses over them. 4. The subject's optimal vertexed spherical equivalent distance correction must be between -1.00 and - 6.00 Diopter (D). 5. The subject's refractive cylinder must be ≤ 1.00D in each eye. 6. The subject must have visual acuity best correctable to 20/25+3 or better for each eye. 7. Subjects must own a wearable pair of spectacles. 8. The subject is a current spherical soft contact lens wearer (defined as a minimum of 6 hours of Daily Wear per day, at least 5 days per week, for a minimum of 1 month prior to the study) and willing to wear the study lenses on a similar basis. 9. Subjects must be able and willing to wear the study lenses at least 6 hours a day, a minimum of 5 days per week 10. The subject must have normal eyes (i.e., no ocular medications or infections of any type).

Exclusion criteria

* Potential subjects who meet any of the following criteria will be excluded from participating in the study: 1. Currently pregnant or lactating (subjects who become pregnant during the study will be discontinued). 2. Any systemic disease, autoimmune disease, or use of medication, which may interfere with contact lens wear. This may include, but not be limited to, diabetes, hyperthyroidism, recurrent herpes simplex/zoster, Sjögren's syndrome, xerophthalmia, acne rosacea, Stevens-Johnson syndrome, and immunosuppressive diseases or any infectious diseases (e.g. hepatitis, tuberculosis). 3. Use of any of the following medications within 1 week prior to enrollment: oral retinoid isotretinoin (e.g. Accutane), oral tetracyclines, topical scopolamine, oral (e.g., Chlor-Trimeton, and Benadryl) and ophthalmic antihistamines, oral phenothiazines (e.g., Haldol, Mellaril, Thorazine, Elavil, Pamelor, Compazine), oral and ophthalmic Beta-adrenergic blockers (e.g., Propranolol, Timolol, and Practolol), systemic steroids, and any prescribed or over the counter (OTC) ocular medication. 4. Entropion, ectropion, extrusions, chalazia, recurrent styes, glaucoma, history of recurrent corneal erosions, aphakia, or moderate or above corneal distortion. 5. Any previous, or planned, ocular or intraocular surgery (e.g., radial keratotomy, photorefractive keratectomy (PRK), LASIK, etc.). 6. Any Grade 3 or greater slit lamp findings (e.g.., edema, corneal neovascularization, corneal staining, tarsal abnormalities, conjunctival injection) on the FDA classification scale, any previous history or signs of a contact lens-related corneal inflammatory event (e.g., past peripheral ulcer or round peripheral scar), or any other ocular abnormality that may contraindicate contact lens wear. 7. Any known hypersensitivity or allergic reaction to Optifree®PureMoist® multi-purpose care solution or Eye-Cept® rewetting drop solution 8. Any ocular infection, allergy or clinically significant ocular disease (e.g. corneal edema, uveitis, severe keratoconjunctivitis sicca, ocular hypertension), or ocular conditions (e.g. strabismus), which might interfere with the study. 9. Any corneal distortion resulting from previous hard or rigid gas permeable contact lens wear. 10. Toric, extended wear, monovision or multi-focal contact lens correction. 11. Participation in any contact lens or lens care product clinical trial within 30 days prior to study enrollment. 12. Participation in clinical trials involving the Test lens within 3 months prior to study enrollment. 13. History of binocular vision abnormality or strabismus. 14. Employee, relative or friends of employees of any ophthalmic company, or investigational clinic (e.g., Investigator, Coordinator, Technician).

Design outcomes

Primary

MeasureTime frameDescription
Overall Comfort2-Week Follow-upOverall comfort was assessed using the Contact Lens User Experience™ (CLUE) questionnaire. CLUE is a validated patient-reported outcomes (PRO) questionnaire to assess patient-experience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65. Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120. The average comfort score from the 3 study periods was reported.
Distance Monocular LogMAR Visual Acuity2-Week Follow-upDistance Monocular LogMAR visual acuity was assessed at 4 meters using an ETDRS chart at the 2-week follow-up for each subject eye during each of the three study periods. The average visual acuity for each lens type from the 3 study periods was reported.
Contact Lens Fitting Acceptance RateUp to 2-Week Follow-upContact lens fitting acceptance was assessed for each subject eye using a biomicroscope at post lens insertion and the 2-week follow-up. Lens fit was a binary variable where acceptable lens fit=1 and unacceptable lens fit=0. The proportion of eyes with acceptable lens fit was reported for each lens. The lens fit acceptance rate for both post lens fitting at the 2-week follow-up was reported for each lens type.
Number of Grade 3 or Higher Slit Lamp FindingsUp to 2-Week Follow-upSlit Lamp Findings (SLF) were assessed using a biomicroscope and was graded using the FDA grading scale (Grade: 0, 1,2, 3 and 4) with grade 0 represents the absence of findings and 1 to 4 representing successively worse findings (i.e. Grade 1 = trace, Grade 2 = Mild, Grade 3 = moderate and Grade 4 = severe). This was performed on each subject eye at every study visit (baseline, unscheduled visits and 2-week follow-up). The data was then dichotomized into two groups. Those with grade 3 or higher and those with grade 2 or lower. The number of SLF with grade 3 or higher by lens was reported.

Secondary

MeasureTime frameDescription
Overall Quality of Vision2-Week Follow-upOverall quality of vision was assessed using the Contact Lens User Experience™ (CLUE) questionnaire. CLUE is a validated patient-reported outcomes (PRO) questionnaire to assess patient-experience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65. Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120. The average vision score from the 3 study periods was reported.
Overall Handling2-Week Follow-upOverall Handling was assessed using the Contact Lens User Experience™ (CLUE) questionnaire. CLUE is a validated patient-reported outcomes (PRO) questionnaire to assess patient-experience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65. Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120. The average handling score from the 3 study periods was reported.

Countries

United States

Participant flow

Recruitment details

A total of 133 subjects were enrolled into this study. Of the enrolled subjects, 130 were dispensed at least one study lens and 3 subjects failed to meet all eligibility criteria. Of the dispensed subjects 127 completed the study while 3 subjects were discontinued.

Participants by arm

ArmCount
Dispensed Subjects
All subjects that were dispensed at least one study lens.
130
Total130

Withdrawals & dropouts

PeriodReasonFG000FG001
Period 1Adverse Event10
Period 1no longer meets eligibility criteria01
Period 2Lost to Follow-up10

Baseline characteristics

CharacteristicDispensed Subjects
Age, Continuous31.6 Years
STANDARD_DEVIATION 7.69
Race/Ethnicity, Customized
American Indian or Alaska Native
2 Participants
Race/Ethnicity, Customized
Asian
4 Participants
Race/Ethnicity, Customized
Black or African American
12 Participants
Race/Ethnicity, Customized
Other
1 Participants
Race/Ethnicity, Customized
White
111 Participants
Sex: Female, Male
Female
88 Participants
Sex: Female, Male
Male
42 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 1300 / 130
other
Total, other adverse events
0 / 1300 / 130
serious
Total, serious adverse events
0 / 1300 / 130

Outcome results

Primary

Contact Lens Fitting Acceptance Rate

Contact lens fitting acceptance was assessed for each subject eye using a biomicroscope at post lens insertion and the 2-week follow-up. Lens fit was a binary variable where acceptable lens fit=1 and unacceptable lens fit=0. The proportion of eyes with acceptable lens fit was reported for each lens. The lens fit acceptance rate for both post lens fitting at the 2-week follow-up was reported for each lens type.

Time frame: Up to 2-Week Follow-up

Population: Subjects that completed all study visits.

ArmMeasureValue (NUMBER)
TestContact Lens Fitting Acceptance Rate1 proportion of eyes
ControlContact Lens Fitting Acceptance Rate1 proportion of eyes
Comparison: It was calculated that 70 participants randomized in a 1:1 fashion between the two sequences would have at least 80% power to detect a the 2-week follow-up. Sample size was determined using simulations methods for repeated measures. Analysis was only performed on eyes wearing the Tesl lens.95% CI: [0.96, 1]Bayesian Methods
Primary

Distance Monocular LogMAR Visual Acuity

Distance Monocular LogMAR visual acuity was assessed at 4 meters using an ETDRS chart at the 2-week follow-up for each subject eye during each of the three study periods. The average visual acuity for each lens type from the 3 study periods was reported.

Time frame: 2-Week Follow-up

Population: All subjects that completed all study visits without a major protocol deviation.

ArmMeasureValue (MEAN)Dispersion
TestDistance Monocular LogMAR Visual Acuity-0.095 logMARStandard Deviation 0.0831
ControlDistance Monocular LogMAR Visual Acuity-0.075 logMARStandard Deviation 0.0806
Comparison: It was calculated that 40 participants randomized in a 1:1 fashion between the two sequences would have at least 80% power to detect a 0.05 difference in LogMAR visual acuity at the 2-week follow-up. Sample size was determined using simulations methods for repeated measures.95% CI: [0.01, 0.03]Bayesian multivariate normal random-effe
Primary

Number of Grade 3 or Higher Slit Lamp Findings

Slit Lamp Findings (SLF) were assessed using a biomicroscope and was graded using the FDA grading scale (Grade: 0, 1,2, 3 and 4) with grade 0 represents the absence of findings and 1 to 4 representing successively worse findings (i.e. Grade 1 = trace, Grade 2 = Mild, Grade 3 = moderate and Grade 4 = severe). This was performed on each subject eye at every study visit (baseline, unscheduled visits and 2-week follow-up). The data was then dichotomized into two groups. Those with grade 3 or higher and those with grade 2 or lower. The number of SLF with grade 3 or higher by lens was reported.

Time frame: Up to 2-Week Follow-up

Population: Subjects that were dispensed at least one study lens.

ArmMeasureValue (NUMBER)
TestNumber of Grade 3 or Higher Slit Lamp Findings0 Slit Lamp Finding
ControlNumber of Grade 3 or Higher Slit Lamp Findings1 Slit Lamp Finding
Comparison: It was calculated that 120 participants randomized in a 1:1 fashion between the two sequences would have at least 80% power to detect 5% difference between the Test and Control lens with respect to the proportion of eyes with Grade 3 or higher slit lamp findings across all study visits. Sample size was determined using simulations methods. Sample size for this study was primarily driven by the slit lamp findings.95% CI: [-0.0237, 0.0129]Bayesian beta-binomial model
Primary

Overall Comfort

Overall comfort was assessed using the Contact Lens User Experience™ (CLUE) questionnaire. CLUE is a validated patient-reported outcomes (PRO) questionnaire to assess patient-experience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65. Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120. The average comfort score from the 3 study periods was reported.

Time frame: 2-Week Follow-up

Population: All subjects that completed all study visits without a major protocol deviation.

ArmMeasureValue (MEAN)Dispersion
TestOverall Comfort67.509 Units on a ScaleStandard Deviation 23.8724
ControlOverall Comfort61.185 Units on a ScaleStandard Deviation 24.2024
Comparison: It was calculated that 60 participants randomized in a 1:1 fashion between the two sequences would have at least 80% power to detect a 5 points difference in mean overall comfort at the 2-week follow-up. Sample size was determined using Stroup's Method (alpha=0.05).95% CI: [1.16, 9.09]Bayesian multivariate normal random-effe
Secondary

Overall Handling

Overall Handling was assessed using the Contact Lens User Experience™ (CLUE) questionnaire. CLUE is a validated patient-reported outcomes (PRO) questionnaire to assess patient-experience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65. Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120. The average handling score from the 3 study periods was reported.

Time frame: 2-Week Follow-up

Population: All subjects that completed all study visits without a major protocol deviation.

ArmMeasureValue (MEAN)Dispersion
TestOverall Handling72.11 Units on a ScaleStandard Deviation 20.046
ControlOverall Handling66.79 Units on a ScaleStandard Deviation 20.007
Comparison: Sample size calculations were based on the primary endpoints.95% CI: [0.36, 6.29]Bayesian multivariate normal random-effe
Secondary

Overall Quality of Vision

Overall quality of vision was assessed using the Contact Lens User Experience™ (CLUE) questionnaire. CLUE is a validated patient-reported outcomes (PRO) questionnaire to assess patient-experience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65. Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120. The average vision score from the 3 study periods was reported.

Time frame: 2-Week Follow-up

Population: All subjects that completed all study visits without a major protocol deviation.

ArmMeasureValue (MEAN)Dispersion
TestOverall Quality of Vision69.16 Units on a ScaleStandard Deviation 20.362
ControlOverall Quality of Vision60.329 Units on a ScaleStandard Deviation 22.289
Comparison: Sample size calculations were based on the primary endpoints.95% CI: [3.69, 10.74]Bayesian multivariate normal random-effe