Comparative Study of ELLIPTA Dry Powder Inhaler (DPI) Versus DISKUS DPI Used With HandiHaler DPI in Subjects With Chronic Obstructive Pulmonary Disease (COPD)

A Randomized, Open-label, Cross-over, Placebo Inhaler Study to Evaluate the Correct Use of ELLIPTA™ Dry Powder Inhaler (DPI) Compared to DISKUS™ DPI Used in Combination With HandiHaler DPI in Participants With Chronic Obstructive Pulmonary Disease (COPD)

Status
Completed
Phases
Phase 4
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03227445
Enrollment
240
Registered
2017-07-24
Start date
2017-09-20
Completion date
2018-01-04
Last updated
2020-07-08

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Pulmonary Disease, Chronic Obstructive

Keywords

DISKUS, HandiHaler, COPD, Placebo, Cross-over, Dry powder inhaler, ELLIPTA

Brief summary

This randomized, cross over study aims to find out the benefits of delivering triple therapy using a single ELLIPTA® DPI (fixed-dose combination triple therapy) versus delivering triple therapy using two different types of inhalers (open triple therapy) including DISKUS® with HandiHaler® to subjects with COPD. Correct inhaler use, critical errors and performance attributes will also be assessed. Approximately 240 subjects with COPD will be randomized in the study. The study will be conducted in 3 visits and will be completed in approximately 56 days. At Visit 1 (Day 1) and Visit 2 (Day 28) subjects will be randomized to receive a placebo ELLIPTA inhaler once daily (QD) or a placebo DISKUS twice daily (BID) with placebo HandiHaler QD inhaler in 1:1 ratio in a cross-over manner for the study period (28 days for each period). At Visit 3 (Day 56), subjects will be asked to complete preference questionnaire 1 or 2. There will be no active treatment and subjects will continue to take their own prescribed COPD maintenance and rescue medication during the entire study period. ELLIPTA and DISKUS are the registered trademarks of GlaxoSmithKline group of companies. HandiHaler is the registered trademark of Boehringer Ingelheim group of companies.

Interventions

DEVICEELLIPTA placebo DPI

ELLIPTA is a dry powder inhaler used via oral route. It will be a placebo DPI with two strips with 30 blisters per strip. First strip will contain lactose monohydrate and the second strip will contain lactose monohydrate blended with magnesium stearate.

DEVICEDISKUS placebo DPI

DISKUS is a dry powder inhaler used via oral route. It will be a placebo DPI with one blister strip that will contain lactose monohydrate.

DEVICEHandiHaler placebo DPI

HandiHaler is a dry powder inhaler used via oral route. It will be a DPI with placebo capsules that will contain lactose monohydrate.

OTHERInhaler preference questionnaires

Preference questionnaires will be given to subjects to understand the inhaler preference. There will be 2 types of questionnaire, preference questionnaires 1 and 2, which will be randomized at visit 3 (Day 56).

Sponsors

GlaxoSmithKline
Lead SponsorINDUSTRY

Study design

Allocation
RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
TREATMENT
Masking
NONE

Masking description

This will an open-label study and blinding will not be performed.

Intervention model description

Subjects will be randomized to receive a placebo ELLIPTA inhaler QD or a placebo DISKUS BID with placebo HandiHaler QD inhaler at Visit 1 and Visit 2 in a cross-over manner followed by Preference Questionnaire 1 or 2

Eligibility

Sex/Gender
ALL
Age
40 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Subjects must be capable of giving signed informed consent. * Subjects must have a diagnosis of COPD with a documented history of COPD for at least 12 months, in accordance with the definition by the American Thoracic Society/European Respiratory Society. * Subjects must be 40 years of age inclusive, at the time of signing the informed consent. * Male or female subjects will be included. Females must not be pregnant or planning pregnancy during the study or not lactating. * Subjects must have a documented post albuterol forced expiratory volume in one second (FEV1)/ forced vital capacity (FVC) ratio \<0.70 and FEV1 \<=70% of predicted obtained within two years of Visit 1. * Current or former (defined as subjects who have quit smoking for at least 3 months prior to Screening/Visit 1) cigarette smokers with a \> 10 pack-year smoking history \[Number of pack-years = (number of cigarettes per day/20) x number of years smoked (example, 10 pack-years is equal to 20 cigarettes per day for 10 years, or 10 cigarettes per day for 20 years\]. * All subjects should be currently receiving maintenance treatment for COPD for at least 4 weeks prior to Randomization/Visit 1 and evaluated as unlikely to change COPD treatment within 4 weeks of Visit 1. * All subjects should be able to stay on their prescribed maintenance COPD inhaler (s) without change throughout the entire treatment period. * Subjects must be able to read, comprehend, and record information in English.

Exclusion criteria

* Subjects must not have a current diagnosis of asthma. * Subjects must not have used the ELLIPTA, DISKUS, or HandiHaler inhalers in the 12 months prior to Visit 1. * Subjects must not be receiving their current COPD medications with the ELLIPTA, DISKUS, or HandiHaler inhalers. * Subjects must not be receiving only inhaled short-acting beta-adrenergic agonists, i.e., albuterol as their daily COPD therapy (as needed or regularly scheduled). * Subjects must not have experienced more than 1 COPD exacerbation which required hospitalization in the 12 months prior to Visit 1. * Subjects must not have a known or suspected history of alcohol or drug abuse within the last 2 years. * Subjects must not have a history of hypersensitivity to any component of the study inhalers (example, lactose, magnesium stearate). In addition, subjects with a history of severe milk protein allergy that, in the opinion of the study physician, contraindicates participation will also be excluded. * Subjects with other respiratory disorders, including active tuberculosis, active lung cancer, sarcoidosis, lung fibrosis, pulmonary hypertension, or pulmonary disease (including but not confined to asthma, bronchiectasis with the need for treatment, cystic fibrosis, and bronchopulmonary dysplasia), interstitial lung diseases or other active pulmonary diseases. * Subjects with historical, or current evidence of clinically significant or rapidly progressing or unstable cardiovascular, neurological, renal, hepatic, immunological, endocrine (including uncontrolled diabetes or thyroid disease) or hematological abnormalities that are uncontrolled. Significant is defined as any disease that, in the opinion of the investigator, would put the safety of the participant at risk through participation, or which would affect the analysis if the disease/condition exacerbated during the study will be excluded. * Subjects with history of psychiatric disease, intellectual impairment, poor motivation or other conditions that will limit the validity of informed consent to participate in the study will be excluded. * Subjects at risk of non-compliance, or unable to comply with the study procedures, or unable to continue their current medications. * Subjects who have received an investigational drug and/or medical device/inhaler within 30 days of entry into this study (Screening/Visit 1), or within five drug half-lives of the investigational drug, whichever is longer. * Subjects will not be eligible for this study if he/she is an immediate family member of the participating investigator, sub investigator, study coordinator, or employee of the participating investigator.

Design outcomes

Primary

MeasureTime frameDescription
Percentage of Participants With Zero Errors After 28 Days of Inhaler Use in Each Treatment Phase (Primary Estimand: Hypothetical)Up to Day 56A checklist for correct use of each inhaler was developed in Patient Instruction Leaflets (PIL's). Participants were guided by trained health care provider (HCP) to demonstrate correct use of inhaler. Baseline assessment was conducted when participant initially dispensed the inhaler. Second assessment was conducted after each 28day dosing period. Correct Use Check list was completed by HCP at each visit. Primary hypothetical estimand is the estimate of treatment effect of all participants who stayed on their randomized study device sequence and did not change their standard COPD maintenance medication device to one delivered via ELLIPTA, DISKUS or HANDIHALER, throughout both 28 day treatment periods. Participants could attend visit without the device/s they were randomised to, in which case correct use cannot be assessed as described in the protocol. Participants who completed error assessments for both randomized treatment groups and experienced no intercurrent events were included
Percentage of Participants With Atleast One Error After 28 Days of Inhaler Use in Each Treatment Phase (Primary Estimand: Hypothetical)Up to Day 56Primary estimand is the treatment effect estimated in participants who stayed on their randomised study device sequence and did not change their standard COPD maintenance medication device to one delivered via ELLIPTA, DISKUS or HANDIHALER, throughout both 28 day treatment periods. The participant could attend the visit without the device/s they were randomised to, in which case correct use cannot be assessed as described in the protocol. For primary estimand, a sensitivity analysis using Cochran-Mantel-Haenszel test was performed on participants with discordant results. Only those participants who completed error assessments for both randomized treatment groups and experienced no intercurrent events were included
Percentage of Participants With Atleast One Error After 28 Days of Inhaler Use in Each Treatment Phase (Supplementary Estimand: Composite)Up to Day 56Supplementary estimand estimated the composite effect of initial randomized treatment. A sensitivity analysis using the Cochran-Mantel-Haenszel test was performed on participants with discordant results. Participants who withdrew during period 2 were included in the analysis using early withdrawal data where available or imputation otherwise. Participants who experienced an intercurrent event in Period 1 were excluded from the analysis.
Percentage of Participants With Zero Errors After 28 Days of Inhaler Use in Each Treatment Phase(Supplementary Estimand: Composite)Up to Day 56Supplementary estimand estimated the composite effect of initial randomized treatment. The analysis was performed using stratified exact logistic model. Participants were included in the model as fixed strata, treatment option was included in the exact statement and period included as fixed effects. Participants who withdrew during period 2 were included in the analysis using early withdrawal data where available or imputation otherwise. Participants who experienced an intercurrent event in Period 1 were excluded from the analysis.

Secondary

MeasureTime frameDescription
Change in Errors for Each Treatment Group in Participants With One or More Errors After 28 Days of UseDay 1 and Day 28 of each treatment groupAssessment of errors was conducted by health care professionals trained in the correct inhaler use of the three inhalers based on the checklist of errors. The median of overall errors made by each participant was assessed for ELLIPTA and DISKUS + HandiHaler both on Day 1 and Day 28. The difference was calculated by subtracting values of Day 28 from Day1 for each treatment regimen. Only those participants who completed error assessments for both randomized treatment groups and experienced no intercurrent events were included.
Number of Participants With Zero Critical Errors After 28 Days of Each Treatment Group Use (Primary Estimand: Hypothetical)Up to Day 56A checklist for correct use of each inhaler was developed based on the steps identified in the PIL. A critical error was defined as an error that was most likely to result in no, or a significantly reduced amount, of medication being inhaled by the participant. Primary estimand is the treatment effect estimated in participants who stayed on their randomised study device sequence and did not change their standard COPD maintenance medication device to one delivered via ELLIPTA, DISKUS or HANDIHALER, throughout both 28 day treatment periods. The participant could attend the visit without the device/s they were randomised to, in which case correct use cannot be assessed as described in the protocol. For primary estimand, a sensitivity analysis using Cochran-Mantel-Haenszel test was performed on participants with discordant results. Only those participants who completed error assessments for both randomized treatment groups and experienced no intercurrent events were included.
Number of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Up to Day 56The occurrence of each type of error for each inhaler (ELLIPTA, DISKUS or HANDIHALER) were evaluated based on the information collected in Correct Use Checklists. Primary estimand is the treatment effect estimated in participants who stayed on their randomised study device sequence & did not change their standard COPD maintenance medication device to one delivered via ELLIPTA, DISKUS or HANDIHALER, throughout both 28 day treatment periods. The participant could attend the visit without the device/s they were randomised to, in which case correct use cannot be assessed as described in the protocol. For primary estimand, a sensitivity analysis using Cochran-Mantel-Haenszel test was performed on participants with discordant results. Only those participants who completed error assessments for both randomized treatment groups & experienced no intercurrent events were included. The number of error is reported as NA for the type of error which was not applicable to the particular inhaler type
Number of Participants With Atleast One Critical Error After 28 Days of Each Treatment Group Use (Supplementary Estimand: Composite)Up to Day 56Supplementary estimand estimated the composite effect of initial randomized treatment. A sensitivity analysis using the Cochran-Mantel-Haenszel test was performed on participants with discordant results. Participants who withdrew during period 2 were included in the analysis using early withdrawal data where available or imputation otherwise. Participants who experienced an intercurrent event in Period 1 were excluded from the analysis.
Number of Participants With Atleast One Critical Error After 28 Days of Each Treatment Group Use (Primary Estimand: Hypothetical)Up to Day 56A critical error was defined as an error that was most likely to result in no, or a significantly reduced amount, of medication being inhaled by the participant. Primary estimand is the treatment effect estimated in participants who stayed on their randomised study device sequence and did not change their standard COPD maintenance medication device to one delivered via ELLIPTA, DISKUS or HANDIHALER, throughout both 28 day treatment periods. The participant could attend the visit without the device/s they were randomised to, in which case correct use cannot be assessed as described in the protocol. For primary estimand, a sensitivity analysis using Cochran-Mantel-Haenszel test was performed on participants with discordant results. Only those participants who completed error assessments for both randomized treatment groups and experienced no intercurrent events were included.
Number of Errors Per Participant for Each Treatment Group After 28 Days of Use (Primary Estimand: Hypothetical)Up to Day 56Participants were provided with PIL explaining correct use of inhaler. Overall error includes both critical and non-critical errors. Primary estimand is the treatment effect estimated in participants who stayed on their randomised study device sequence and did not change their standard COPD maintenance medication device to one delivered via ELLIPTA, DISKUS or HANDIHALER, throughout both 28 day treatment periods. The participant could attend the visit without the device/s they were randomised to, in which case correct use cannot be assessed as described in the protocol. For primary estimand, a sensitivity analysis using Cochran-Mantel-Haenszel test was performed on participants with discordant results. Only those participants who completed error assessments for both randomized treatment groups and experienced no intercurrent events were included.
Change in Errors Per Participant for Each Treatment Group After 28 Days of UseDay 1 and Day 28 of each treatment groupAssessment of errors was conducted by HCP trained in the correct inhaler use of the three inhalers based on the checklist of errors. The median number of overall errors made by per participant was assessed for ELLIPTA and DISKUS + HandiHaler both on Day 1 and Day 28 for each treatment group. The difference was calculated by subtracting values of Day 28 from Day1 for each treatment regimen. Only those participants who completed error assessments for both randomized treatment groups and experienced no intercurrent events were included.
Number of Errors for Each Treatment Group in Participants With One or More Errors After 28 Days of Use (Primary Estimand: Hypothetical)Up to Day 56Participants were provided with the PIL, explaining correct use of the inhaler. Overall error includes both critical and non-critical errors. Assessment of errors was conducted by HCP trained in the correct inhaler use of the three inhalers based on the checklist of errors. Primary estimand is the treatment effect estimated in participants who stayed on their randomised study device sequence and did not change their standard COPD maintenance medication device to one delivered via ELLIPTA, DISKUS or HANDIHALER, throughout both 28 day treatment periods. The participant could attend the visit without the device/s they were randomised to, in which case correct use cannot be assessed as described in the protocol. For primary estimand, a sensitivity analysis using Cochran-Mantel-Haenszel test was performed on participants with discordant results. Only those participants who completed error assessments for both randomized treatment groups and experienced no intercurrent events were included.

Countries

United States

Participant flow

Recruitment details

This is a randomized, multi-center, open-label study comparing placebo ELLIPTA with placebo DISKUS + HandiHaler to assess correct inhaler use in participants with Chronic Obstructive Pulmonary Disease (COPD). A total of 17 centers across United States participated in the study.

Pre-assignment details

Out of 240 randomized participants (par), 239 entered the study as 1 par was randomized in error. In order to minimize potential bias the order of placebo devices & order of responses, in preference questionnaire (PQ), was randomized in a crossover design. All par on study received placebo only & participant flow reflects this as the only treatment

Participants by arm

ArmCount
All Study Participants
Participants received placebo via ELLIPTA dry powder inhaler (DPI) or DISKUS DPI in combination with HandiHaler DPI based on the following treatment sequences: Treatment Sequence A (ELLIPTA in period 1 and DISKUS + HandiHaler in period 2, followed by PQ 1.); Treatment Sequence B (ELLIPTA in period 1 and DISKUS + HandiHaler in period 2, followed by PQ2); Treatment Sequence C (DISKUS + HandiHaler in period 1 and ELLIPTA in period 2, followed by PQ1); or Treatment Sequence D (DISKUS + HandiHaler in period 1 and ELLIPTA in period 2, followed by PQ2). Participants continued to take their own prescribed COPD medication and rescue medications for the entire 56-day study period.
239
Total239

Withdrawals & dropouts

PeriodReasonFG000FG001FG002FG003
Period 1 (28 Days)Adverse Event1201
Period 1 (28 Days)Protocol Violation1000
Period 1 (28 Days)Withdrawal by Subject0011
Period 2 (28 Days)Adverse Event1021
Period 2 (28 Days)Protocol Violation0001
Period 2 (28 Days)Withdrawal by Subject0101

Baseline characteristics

CharacteristicAll Study Participants
Age, Continuous65.9 Years
STANDARD_DEVIATION 8.25
Race/Ethnicity, Customized
African American/African Heritage
23 Count of Participants
Race/Ethnicity, Customized
White - Arabic/North African Heritage
3 Count of Participants
Race/Ethnicity, Customized
White - White/Caucasian/European Heritage
213 Count of Participants
Sex: Female, Male
Female
108 Participants
Sex: Female, Male
Male
131 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 239
other
Total, other adverse events
0 / 239
serious
Total, serious adverse events
10 / 239

Outcome results

Primary

Percentage of Participants With Atleast One Error After 28 Days of Inhaler Use in Each Treatment Phase (Primary Estimand: Hypothetical)

Primary estimand is the treatment effect estimated in participants who stayed on their randomised study device sequence and did not change their standard COPD maintenance medication device to one delivered via ELLIPTA, DISKUS or HANDIHALER, throughout both 28 day treatment periods. The participant could attend the visit without the device/s they were randomised to, in which case correct use cannot be assessed as described in the protocol. For primary estimand, a sensitivity analysis using Cochran-Mantel-Haenszel test was performed on participants with discordant results. Only those participants who completed error assessments for both randomized treatment groups and experienced no intercurrent events were included

Time frame: Up to Day 56

Population: ITT Population. Only those participants with data available at specified time point were analyzed.

ArmMeasureGroupValue (NUMBER)
ELLIPTAPercentage of Participants With Atleast One Error After 28 Days of Inhaler Use in Each Treatment Phase (Primary Estimand: Hypothetical)Atleast 1 error-ELLIPTA&0 errors-DISKUS+HandiHaler3 Percentage of participants
ELLIPTAPercentage of Participants With Atleast One Error After 28 Days of Inhaler Use in Each Treatment Phase (Primary Estimand: Hypothetical)Atleast 1error-DISKUS+HandiHaler&0errors-ELLIPTA23 Percentage of participants
p-value: <0.001Cochran-Mantel-Haenszel
Primary

Percentage of Participants With Atleast One Error After 28 Days of Inhaler Use in Each Treatment Phase (Supplementary Estimand: Composite)

Supplementary estimand estimated the composite effect of initial randomized treatment. A sensitivity analysis using the Cochran-Mantel-Haenszel test was performed on participants with discordant results. Participants who withdrew during period 2 were included in the analysis using early withdrawal data where available or imputation otherwise. Participants who experienced an intercurrent event in Period 1 were excluded from the analysis.

Time frame: Up to Day 56

ArmMeasureGroupValue (NUMBER)
ELLIPTAPercentage of Participants With Atleast One Error After 28 Days of Inhaler Use in Each Treatment Phase (Supplementary Estimand: Composite)Atleast 1 error-ELLIPTA&0 errors-DISKUS+HandiHaler4 Percentage of participants
ELLIPTAPercentage of Participants With Atleast One Error After 28 Days of Inhaler Use in Each Treatment Phase (Supplementary Estimand: Composite)Atleast 1error-DISKUS+HandiHaler&0errors-ELLIPTA27 Percentage of participants
p-value: <0.001Cochran-Mantel-Haenszel
Primary

Percentage of Participants With Zero Errors After 28 Days of Inhaler Use in Each Treatment Phase (Primary Estimand: Hypothetical)

A checklist for correct use of each inhaler was developed in Patient Instruction Leaflets (PIL's). Participants were guided by trained health care provider (HCP) to demonstrate correct use of inhaler. Baseline assessment was conducted when participant initially dispensed the inhaler. Second assessment was conducted after each 28day dosing period. Correct Use Check list was completed by HCP at each visit. Primary hypothetical estimand is the estimate of treatment effect of all participants who stayed on their randomized study device sequence and did not change their standard COPD maintenance medication device to one delivered via ELLIPTA, DISKUS or HANDIHALER, throughout both 28 day treatment periods. Participants could attend visit without the device/s they were randomised to, in which case correct use cannot be assessed as described in the protocol. Participants who completed error assessments for both randomized treatment groups and experienced no intercurrent events were included

Time frame: Up to Day 56

Population: Intent-to-treat (ITT) Population comprised of all randomized participants, excluding those who were randomized in error, and did not have at least one error assessment at Visit 1. Only those participants with data available at specified time point were analyzed.

ArmMeasureValue (NUMBER)
ELLIPTAPercentage of Participants With Zero Errors After 28 Days of Inhaler Use in Each Treatment Phase (Primary Estimand: Hypothetical)96 Percentage of participants
DISKUS + HANDIHALERPercentage of Participants With Zero Errors After 28 Days of Inhaler Use in Each Treatment Phase (Primary Estimand: Hypothetical)87 Percentage of participants
p-value: <0.00195% CI: [1.97, 54.28]stratified exact logistic model
Primary

Percentage of Participants With Zero Errors After 28 Days of Inhaler Use in Each Treatment Phase(Supplementary Estimand: Composite)

Supplementary estimand estimated the composite effect of initial randomized treatment. The analysis was performed using stratified exact logistic model. Participants were included in the model as fixed strata, treatment option was included in the exact statement and period included as fixed effects. Participants who withdrew during period 2 were included in the analysis using early withdrawal data where available or imputation otherwise. Participants who experienced an intercurrent event in Period 1 were excluded from the analysis.

Time frame: Up to Day 56

Population: ITT Population. Only those participants with data available at specified time point were analyzed.

ArmMeasureValue (NUMBER)
ELLIPTAPercentage of Participants With Zero Errors After 28 Days of Inhaler Use in Each Treatment Phase(Supplementary Estimand: Composite)95 Percentage of participants
DISKUS + HANDIHALERPercentage of Participants With Zero Errors After 28 Days of Inhaler Use in Each Treatment Phase(Supplementary Estimand: Composite)85 Percentage of participants
p-value: <0.00195% CI: [2.08, 24.55]stratified exact logistic model
Secondary

Change in Errors for Each Treatment Group in Participants With One or More Errors After 28 Days of Use

Assessment of errors was conducted by health care professionals trained in the correct inhaler use of the three inhalers based on the checklist of errors. The median of overall errors made by each participant was assessed for ELLIPTA and DISKUS + HandiHaler both on Day 1 and Day 28. The difference was calculated by subtracting values of Day 28 from Day1 for each treatment regimen. Only those participants who completed error assessments for both randomized treatment groups and experienced no intercurrent events were included.

Time frame: Day 1 and Day 28 of each treatment group

Population: ITT Population. Only those participants with data available at specified time point were analyzed.

ArmMeasureValue (MEDIAN)
ELLIPTAChange in Errors for Each Treatment Group in Participants With One or More Errors After 28 Days of Use0.7 Errors per participant
DISKUS + HANDIHALERChange in Errors for Each Treatment Group in Participants With One or More Errors After 28 Days of Use2.6 Errors per participant
Secondary

Change in Errors Per Participant for Each Treatment Group After 28 Days of Use

Assessment of errors was conducted by HCP trained in the correct inhaler use of the three inhalers based on the checklist of errors. The median number of overall errors made by per participant was assessed for ELLIPTA and DISKUS + HandiHaler both on Day 1 and Day 28 for each treatment group. The difference was calculated by subtracting values of Day 28 from Day1 for each treatment regimen. Only those participants who completed error assessments for both randomized treatment groups and experienced no intercurrent events were included.

Time frame: Day 1 and Day 28 of each treatment group

Population: ITT Population. Only those participants with data available at specified time point were analyzed.

ArmMeasureValue (MEDIAN)
ELLIPTAChange in Errors Per Participant for Each Treatment Group After 28 Days of Use0.0 Errors per participant
DISKUS + HANDIHALERChange in Errors Per Participant for Each Treatment Group After 28 Days of Use0.0 Errors per participant
Secondary

Number of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)

The occurrence of each type of error for each inhaler (ELLIPTA, DISKUS or HANDIHALER) were evaluated based on the information collected in Correct Use Checklists. Primary estimand is the treatment effect estimated in participants who stayed on their randomised study device sequence & did not change their standard COPD maintenance medication device to one delivered via ELLIPTA, DISKUS or HANDIHALER, throughout both 28 day treatment periods. The participant could attend the visit without the device/s they were randomised to, in which case correct use cannot be assessed as described in the protocol. For primary estimand, a sensitivity analysis using Cochran-Mantel-Haenszel test was performed on participants with discordant results. Only those participants who completed error assessments for both randomized treatment groups & experienced no intercurrent events were included. The number of error is reported as NA for the type of error which was not applicable to the particular inhaler type

Time frame: Up to Day 56

Population: ITT Population. Only those participants with data available at specified time point were analyzed.

ArmMeasureGroupValue (NUMBER)
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par closed the cover of the inhalerNA Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par takes long steady deep breath in through mouth2 Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par heard or felt the capsule vibrateNA Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par breathed out slowlyNA Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par breathes out slowly and gently1 Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par closed lips tightly around mouthpieceNA Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par removed Diskus from mouth,held breath for longNA Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par opened coverNA Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par placed capsule in center chamber of inhalerNA Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par took a quick deep breath through the DiskusNA Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par slid lever away from mouthpieceNA Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par held breath for few seconds & removed inhalerNA Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par didn't tilt/shake device post dose preparationNA Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par exhaled fullyNA Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par opened only 1 capsule from the blister cardNA Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par resumed normal breathingNA Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par breathed in deeplyNA Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par opened mouthpiece & center chamber is showingNA Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par didn't pierce use capsule for 2nd inhalationNA Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par didn't block air intake vents with fingersNA Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par did not exhale into the mouthpieceNA Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par exhales while holding inhaler away from mouth6 Errors
ELLIPTANumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par pressed green piercing button&released buttonNA Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par placed capsule in center chamber of inhalerNA Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par closed lips tightly around mouthpieceNA Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par didn't block air intake vents with fingersNA Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par breathed in deeplyNA Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par heard or felt the capsule vibrateNA Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par held breath for few seconds & removed inhalerNA Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par resumed normal breathingNA Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par didn't pierce use capsule for 2nd inhalationNA Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par exhales while holding inhaler away from mouth9 Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par takes long steady deep breath in through mouthNA Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par breathes out slowly and gentlyNA Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par opened cover1 Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par slid lever away from mouthpiece1 Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par didn't tilt/shake device post dose preparation2 Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par took a quick deep breath through the Diskus1 Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par removed Diskus from mouth,held breath for long2 Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par breathed out slowly1 Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par closed the cover of the inhaler2 Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par opened mouthpiece & center chamber is showingNA Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par opened only 1 capsule from the blister cardNA Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par did not exhale into the mouthpieceNA Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par pressed green piercing button&released buttonNA Errors
DISKUS + HANDIHALERNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par exhaled fullyNA Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par breathed in deeply10 Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par breathed out slowlyNA Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par exhales while holding inhaler away from mouth14 Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par exhaled fully7 Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par closed the cover of the inhalerNA Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par didn't pierce use capsule for 2nd inhalation1 Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par did not exhale into the mouthpiece10 Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par resumed normal breathing9 Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par pressed green piercing button&released button4 Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par opened mouthpiece & center chamber is showing1 Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par held breath for few seconds & removed inhaler10 Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par didn't block air intake vents with fingers9 Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par opened only 1 capsule from the blister card5 Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par slid lever away from mouthpieceNA Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par heard or felt the capsule vibrate14 Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par didn't tilt/shake device post dose preparation1 Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par opened coverNA Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par closed lips tightly around mouthpiece9 Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par took a quick deep breath through the DiskusNA Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par breathes out slowly and gentlyNA Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par placed capsule in center chamber of inhaler1 Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par removed Diskus from mouth,held breath for longNA Errors
HandiHalerNumber of Errors by Type for Each Inhaler After 28 Days of Use in Each Treatment Phase (Primary Estimand: Hypothetical)Par takes long steady deep breath in through mouthNA Errors
Secondary

Number of Errors for Each Treatment Group in Participants With One or More Errors After 28 Days of Use (Primary Estimand: Hypothetical)

Participants were provided with the PIL, explaining correct use of the inhaler. Overall error includes both critical and non-critical errors. Assessment of errors was conducted by HCP trained in the correct inhaler use of the three inhalers based on the checklist of errors. Primary estimand is the treatment effect estimated in participants who stayed on their randomised study device sequence and did not change their standard COPD maintenance medication device to one delivered via ELLIPTA, DISKUS or HANDIHALER, throughout both 28 day treatment periods. The participant could attend the visit without the device/s they were randomised to, in which case correct use cannot be assessed as described in the protocol. For primary estimand, a sensitivity analysis using Cochran-Mantel-Haenszel test was performed on participants with discordant results. Only those participants who completed error assessments for both randomized treatment groups and experienced no intercurrent events were included.

Time frame: Up to Day 56

Population: ITT Population. Only those participants with data available at specified time point were analyzed.

ArmMeasureValue (MEDIAN)
ELLIPTANumber of Errors for Each Treatment Group in Participants With One or More Errors After 28 Days of Use (Primary Estimand: Hypothetical)1.0 Errors per participant
DISKUS + HANDIHALERNumber of Errors for Each Treatment Group in Participants With One or More Errors After 28 Days of Use (Primary Estimand: Hypothetical)4.4 Errors per participant
Secondary

Number of Errors Per Participant for Each Treatment Group After 28 Days of Use (Primary Estimand: Hypothetical)

Participants were provided with PIL explaining correct use of inhaler. Overall error includes both critical and non-critical errors. Primary estimand is the treatment effect estimated in participants who stayed on their randomised study device sequence and did not change their standard COPD maintenance medication device to one delivered via ELLIPTA, DISKUS or HANDIHALER, throughout both 28 day treatment periods. The participant could attend the visit without the device/s they were randomised to, in which case correct use cannot be assessed as described in the protocol. For primary estimand, a sensitivity analysis using Cochran-Mantel-Haenszel test was performed on participants with discordant results. Only those participants who completed error assessments for both randomized treatment groups and experienced no intercurrent events were included.

Time frame: Up to Day 56

Population: ITT Population. Only those participants with data available at specified time point were analyzed.

ArmMeasureValue (MEDIAN)
ELLIPTANumber of Errors Per Participant for Each Treatment Group After 28 Days of Use (Primary Estimand: Hypothetical)0.0 Errors per participant
DISKUS + HANDIHALERNumber of Errors Per Participant for Each Treatment Group After 28 Days of Use (Primary Estimand: Hypothetical)0.6 Errors per participant
Secondary

Number of Participants With Atleast One Critical Error After 28 Days of Each Treatment Group Use (Primary Estimand: Hypothetical)

A critical error was defined as an error that was most likely to result in no, or a significantly reduced amount, of medication being inhaled by the participant. Primary estimand is the treatment effect estimated in participants who stayed on their randomised study device sequence and did not change their standard COPD maintenance medication device to one delivered via ELLIPTA, DISKUS or HANDIHALER, throughout both 28 day treatment periods. The participant could attend the visit without the device/s they were randomised to, in which case correct use cannot be assessed as described in the protocol. For primary estimand, a sensitivity analysis using Cochran-Mantel-Haenszel test was performed on participants with discordant results. Only those participants who completed error assessments for both randomized treatment groups and experienced no intercurrent events were included.

Time frame: Up to Day 56

Population: ITT Population. Only those participants with data available at specified time point were analyzed.

ArmMeasureGroupValue (NUMBER)
ELLIPTANumber of Participants With Atleast One Critical Error After 28 Days of Each Treatment Group Use (Primary Estimand: Hypothetical)Atleast 1 critical error with ELLIPTA1 Participants
ELLIPTANumber of Participants With Atleast One Critical Error After 28 Days of Each Treatment Group Use (Primary Estimand: Hypothetical)Atleast 1 critical error with DISKUS+HandiHaler23 Participants
p-value: <0.001Cochran-Mantel-Haenszel
Secondary

Number of Participants With Atleast One Critical Error After 28 Days of Each Treatment Group Use (Supplementary Estimand: Composite)

Supplementary estimand estimated the composite effect of initial randomized treatment. A sensitivity analysis using the Cochran-Mantel-Haenszel test was performed on participants with discordant results. Participants who withdrew during period 2 were included in the analysis using early withdrawal data where available or imputation otherwise. Participants who experienced an intercurrent event in Period 1 were excluded from the analysis.

Time frame: Up to Day 56

Population: ITT Population. Only those participants with data available at specified time point were analyzed.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
ELLIPTANumber of Participants With Atleast One Critical Error After 28 Days of Each Treatment Group Use (Supplementary Estimand: Composite)Atleast 1 error-ELLIPTA&0 errors-DISKUS+HandiHaler3 Participants
ELLIPTANumber of Participants With Atleast One Critical Error After 28 Days of Each Treatment Group Use (Supplementary Estimand: Composite)Atleast 1error-DISKUS+HandiHaler&0errors-ELLIPTA27 Participants
p-value: <0.001Cochran-Mantel-Haenszel
Secondary

Number of Participants With Zero Critical Errors After 28 Days of Each Treatment Group Use (Primary Estimand: Hypothetical)

A checklist for correct use of each inhaler was developed based on the steps identified in the PIL. A critical error was defined as an error that was most likely to result in no, or a significantly reduced amount, of medication being inhaled by the participant. Primary estimand is the treatment effect estimated in participants who stayed on their randomised study device sequence and did not change their standard COPD maintenance medication device to one delivered via ELLIPTA, DISKUS or HANDIHALER, throughout both 28 day treatment periods. The participant could attend the visit without the device/s they were randomised to, in which case correct use cannot be assessed as described in the protocol. For primary estimand, a sensitivity analysis using Cochran-Mantel-Haenszel test was performed on participants with discordant results. Only those participants who completed error assessments for both randomized treatment groups and experienced no intercurrent events were included.

Time frame: Up to Day 56

Population: ITT Population. Only those participants with data available at specified time point were analyzed.

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
ELLIPTANumber of Participants With Zero Critical Errors After 28 Days of Each Treatment Group Use (Primary Estimand: Hypothetical)215 Participants
DISKUS + HANDIHALERNumber of Participants With Zero Critical Errors After 28 Days of Each Treatment Group Use (Primary Estimand: Hypothetical)193 Participants
p-value: <0.001stratified exact logistic model