Blood Pressure, Morbid Obesity, Parturient, Pre-Eclampsia
Conditions
Brief summary
The study will compare agreement of invasive blood pressure measurements with non-invasive blood pressure measurements measured with a conical blood pressure and large standard upper arm rectangular cuff in morbidly obese severely hypertensive (systolic blood pressure \> 160 mmHg) parturients.
Detailed description
A morbidly obese \[Body Mass Index (BMI) ≥ 40 kg/m2\] pregnant woman is at singular risk for all the complications of pregnancy, most notably preeclampsia (PE). Furthermore a woman who is already chronically hypertensive is likely to develop superimposed PE . Uncontrolled systolic hypertension in pregnancy prompts placental abruption, hemorrhagic stroke, and systolic or diastolic heart failure. Accurate blood pressure measurement is therefore a prerequisite to controlling dangerously high systolic blood pressure (SBP) to enable labor and delivery to be conducted safely in association with PE. Precise SBP measurement is also a precondition for clinically testing hypotheses concerning the existence of druggable targets that will allow prolongation of pregnancy in the face of severe PE and /or intra-uterine growth restriction (IUGR) remote from term. Oscillometric noninvasive blood pressure (NIBP) measurement is the customary standard blood pressure monitoring method most often used in the labor suite and obstetric operating room today. For this purpose a rectangular, cylindrical blood pressure (BP) cuff placed on the upper arm is connected to an oscillometric blood pressure device. Yet inaccuracies related to the fact that oscillometric NIBP devices under-read high SBP relative to IBP measurements have been recognized as a potential source of maternal morbidity for years. The study will compare IBP readings obtained from morbidly obese, severely hypertensive (SBP \> 160 mm Hg) parturients with NIBP consecutively measured with new innovative conical Ultracheck Curve BP cuffs and large standard upper arm rectangular cylindrical BP cuffs.
Interventions
Non-invasive SBP, DAP and MAP will be measured in both arms using conical Ultracheck Curve BP cuff
Non-invasive SBP, DAP and MAP will be measured in both arms using traditional cylindrical BP cuff
Direct invasive arterial pressure readings will be recorded simultaneously with non-invasive measurements
Sponsors
Study design
Eligibility
Inclusion criteria
* Written, signed and dated informed consent * ≥ 18 years of age * BMI \>40 kg/m2 * Has a clinically indicated radial artery line already in situ or who agree to placement of an arterial line * Systolic blood pressure \>160 mm Hg * Gestational age greater than or equal to 24 weeks * Parturients admitted for induction of labor
Exclusion criteria
* Parturients admitted in labor
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Comparison of Noninvasive Systolic Blood Pressure Readings to Invasive Blood Pressure Measurement | Single time point measurement (1 day) | Absolute difference between the mean of 2 non-invasive systolic blood pressure readings and mean of 2 invasive systolic blood pressure readings |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Comparison of Mean Arterial Pressure (MAP) Measured Withnon-invasive Cuffs to MAP Measured by Invasive Blood Pressure Monitoring | Single time point measurement (1 day) | Absolute difference in the mean of 2 non-invasive MAP readings and mean of 2 invasive MAP readings |
| Comparison of Noninvasive Diastolic Blood Pressure Readings to Invasive Blood Pressure Measurement | Single time point measurement (1 day) | Absolute difference between the mean of 2 non-invasive diastolic blood pressure readings and mean of 2 invasive diastolic blood pressure readings |
| Mean of 2 Systolic Blood Pressure Readings Taken With Cylindrical and Conical BP Cuffs | Single time point measurement (1 day) | mean of 2 systolic blood pressure readings taken with cylindrical and conical BP cuffs |
| Mean of 2 MAP Readings Taken With Cylindrical and Conical BP Cuffs | Single time point measurement (1 day) | Mean of 2 MAP readings taken with cylindrical and conical BP cuffs |
| Mean of 2 Diastolic Blood Pressure Non-invasive BP Measurements Taken With Cylindrical and Conical BP Cuffs | Single time point measurement (1 day) | Mean of 2 diastolic blood pressure non-invasive BP measurements taken with cylindrical and conical BP cuffs |
Countries
United States
Participant flow
Pre-assignment details
3 participants were randomized, but data was not analyzable on 2 participants. Randomization assignment was not recorded on these 2 participants.
Participants by arm
| Arm | Count |
|---|---|
| Arm A: Conical Then Cylindrical BP recording with noninvasive conical Ultracheck Curve BP cuff and direct invasive arterial pressure monitoring; then with traditional cylindrical BP cuff and direct invasive arterial pressure monitoring
Non-invasive BP recording with conical Ultracheck Curve BP cuff: Non-invasive SBP, DAP and MAP will be measured in both arms using conical Ultracheck Curve BP cuff
Non-invasive BP recording with traditional cylindrical BP cuff: Non-invasive SBP, DAP and MAP will be measured in both arms using traditional cylindrical BP cuff
Direct invasive arterial pressure: Direct invasive arterial pressure readings will be recorded simultaneously with non-invasive measurements | 0 |
| Arm B: Cylindrical Then Conical BP recording with noninvasive traditional cylindrical BP cuff and direct invasive arterial pressure monitoring; then with conical Ultracheck Curve BP cuff and direct invasive arterial pressure monitoring
Non-invasive BP recording with conical Ultracheck Curve BP cuff: Non-invasive SBP, DAP and MAP will be measured in both arms using conical Ultracheck Curve BP cuff
Non-invasive BP recording with traditional cylindrical BP cuff: Non-invasive SBP, DAP and MAP will be measured in both arms using traditional cylindrical BP cuff
Direct invasive arterial pressure: Direct invasive arterial pressure readings will be recorded simultaneously with non-invasive measurements | 1 |
| Total | 1 |
Baseline characteristics
| Characteristic | Arm B: Cylindrical Then Conical | Total | Arm A: Conical Then Cylindrical |
|---|---|---|---|
| Age, Categorical <=18 years | 0 Participants | 0 Participants | 0 Participants |
| Age, Categorical >=65 years | 0 Participants | 0 Participants | 0 Participants |
| Age, Categorical Between 18 and 65 years | 1 Participants | 1 Participants | 0 Participants |
| Race and Ethnicity Not Collected | — | 0 Participants | — |
| Region of Enrollment United States | 1 participants | 1 participants | — |
| Sex: Female, Male Female | 1 Participants | 1 Participants | 0 Participants |
| Sex: Female, Male Male | 0 Participants | 0 Participants | 0 Participants |
Adverse events
| Event type | EG000 affected / at risk | EG001 affected / at risk |
|---|---|---|
| deaths Total, all-cause mortality | 0 / 0 | 0 / 1 |
| other Total, other adverse events | 0 / 0 | 0 / 1 |
| serious Total, serious adverse events | 0 / 0 | 0 / 1 |
Outcome results
Comparison of Noninvasive Systolic Blood Pressure Readings to Invasive Blood Pressure Measurement
Absolute difference between the mean of 2 non-invasive systolic blood pressure readings and mean of 2 invasive systolic blood pressure readings
Time frame: Single time point measurement (1 day)
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Conical | Comparison of Noninvasive Systolic Blood Pressure Readings to Invasive Blood Pressure Measurement | -28 mm Hg |
| Cylindrical | Comparison of Noninvasive Systolic Blood Pressure Readings to Invasive Blood Pressure Measurement | -6.5 mm Hg |
Comparison of Mean Arterial Pressure (MAP) Measured Withnon-invasive Cuffs to MAP Measured by Invasive Blood Pressure Monitoring
Absolute difference in the mean of 2 non-invasive MAP readings and mean of 2 invasive MAP readings
Time frame: Single time point measurement (1 day)
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Conical | Comparison of Mean Arterial Pressure (MAP) Measured Withnon-invasive Cuffs to MAP Measured by Invasive Blood Pressure Monitoring | -25 mm Hg |
| Cylindrical | Comparison of Mean Arterial Pressure (MAP) Measured Withnon-invasive Cuffs to MAP Measured by Invasive Blood Pressure Monitoring | -13.5 mm Hg |
Comparison of Noninvasive Diastolic Blood Pressure Readings to Invasive Blood Pressure Measurement
Absolute difference between the mean of 2 non-invasive diastolic blood pressure readings and mean of 2 invasive diastolic blood pressure readings
Time frame: Single time point measurement (1 day)
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Conical | Comparison of Noninvasive Diastolic Blood Pressure Readings to Invasive Blood Pressure Measurement | -9.5 mm Hg |
| Cylindrical | Comparison of Noninvasive Diastolic Blood Pressure Readings to Invasive Blood Pressure Measurement | 0.5 mm Hg |
Mean of 2 Diastolic Blood Pressure Non-invasive BP Measurements Taken With Cylindrical and Conical BP Cuffs
Mean of 2 diastolic blood pressure non-invasive BP measurements taken with cylindrical and conical BP cuffs
Time frame: Single time point measurement (1 day)
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Conical | Mean of 2 Diastolic Blood Pressure Non-invasive BP Measurements Taken With Cylindrical and Conical BP Cuffs | 73.5 mm Hg | Standard Deviation 3.54 |
| Cylindrical | Mean of 2 Diastolic Blood Pressure Non-invasive BP Measurements Taken With Cylindrical and Conical BP Cuffs | 85 mm Hg | Standard Deviation 2.83 |
Mean of 2 MAP Readings Taken With Cylindrical and Conical BP Cuffs
Mean of 2 MAP readings taken with cylindrical and conical BP cuffs
Time frame: Single time point measurement (1 day)
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Conical | Mean of 2 MAP Readings Taken With Cylindrical and Conical BP Cuffs | 87.5 mm Hg | Standard Deviation 2.12 |
| Cylindrical | Mean of 2 MAP Readings Taken With Cylindrical and Conical BP Cuffs | 101.5 mm Hg | Standard Deviation 0.71 |
Mean of 2 Systolic Blood Pressure Readings Taken With Cylindrical and Conical BP Cuffs
mean of 2 systolic blood pressure readings taken with cylindrical and conical BP cuffs
Time frame: Single time point measurement (1 day)
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Conical | Mean of 2 Systolic Blood Pressure Readings Taken With Cylindrical and Conical BP Cuffs | 132.5 mm Hg | Standard Deviation 2.12 |
| Cylindrical | Mean of 2 Systolic Blood Pressure Readings Taken With Cylindrical and Conical BP Cuffs | 154 mm Hg | Standard Deviation 1.41 |