Non-invasive Versus Invasive Blood Pressure Measurement in the Morbidly Obese Parturient With Severe Preeclampsia

Non-invasive Versus Invasive Blood Pressure Measurement in the Morbidly Obese Parturient (BMI >= 40 kg/m2) With Severe Preeclampsia: A Comparison of Direct Arterial Blood Pressure Measurements With Readings Obtained Using Either Large Cylindrical or Novel Conical Bariatric Upper Arm Blood Pressure Cuffs

Status
Terminated
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03222414
Enrollment
3
Registered
2017-07-19
Start date
2015-05-01
Completion date
2016-01-18
Last updated
2019-12-11

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Blood Pressure, Morbid Obesity, Parturient, Pre-Eclampsia

Brief summary

The study will compare agreement of invasive blood pressure measurements with non-invasive blood pressure measurements measured with a conical blood pressure and large standard upper arm rectangular cuff in morbidly obese severely hypertensive (systolic blood pressure \> 160 mmHg) parturients.

Detailed description

A morbidly obese \[Body Mass Index (BMI) ≥ 40 kg/m2\] pregnant woman is at singular risk for all the complications of pregnancy, most notably preeclampsia (PE). Furthermore a woman who is already chronically hypertensive is likely to develop superimposed PE . Uncontrolled systolic hypertension in pregnancy prompts placental abruption, hemorrhagic stroke, and systolic or diastolic heart failure. Accurate blood pressure measurement is therefore a prerequisite to controlling dangerously high systolic blood pressure (SBP) to enable labor and delivery to be conducted safely in association with PE. Precise SBP measurement is also a precondition for clinically testing hypotheses concerning the existence of druggable targets that will allow prolongation of pregnancy in the face of severe PE and /or intra-uterine growth restriction (IUGR) remote from term. Oscillometric noninvasive blood pressure (NIBP) measurement is the customary standard blood pressure monitoring method most often used in the labor suite and obstetric operating room today. For this purpose a rectangular, cylindrical blood pressure (BP) cuff placed on the upper arm is connected to an oscillometric blood pressure device. Yet inaccuracies related to the fact that oscillometric NIBP devices under-read high SBP relative to IBP measurements have been recognized as a potential source of maternal morbidity for years. The study will compare IBP readings obtained from morbidly obese, severely hypertensive (SBP \> 160 mm Hg) parturients with NIBP consecutively measured with new innovative conical Ultracheck Curve BP cuffs and large standard upper arm rectangular cylindrical BP cuffs.

Interventions

DEVICENon-invasive BP recording with conical Ultracheck Curve BP cuff

Non-invasive SBP, DAP and MAP will be measured in both arms using conical Ultracheck Curve BP cuff

DEVICENon-invasive BP recording with traditional cylindrical BP cuff

Non-invasive SBP, DAP and MAP will be measured in both arms using traditional cylindrical BP cuff

OTHERDirect invasive arterial pressure

Direct invasive arterial pressure readings will be recorded simultaneously with non-invasive measurements

Sponsors

Vanderbilt University Medical Center
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
FEMALE
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Written, signed and dated informed consent * ≥ 18 years of age * BMI \>40 kg/m2 * Has a clinically indicated radial artery line already in situ or who agree to placement of an arterial line * Systolic blood pressure \>160 mm Hg * Gestational age greater than or equal to 24 weeks * Parturients admitted for induction of labor

Exclusion criteria

* Parturients admitted in labor

Design outcomes

Primary

MeasureTime frameDescription
Comparison of Noninvasive Systolic Blood Pressure Readings to Invasive Blood Pressure MeasurementSingle time point measurement (1 day)Absolute difference between the mean of 2 non-invasive systolic blood pressure readings and mean of 2 invasive systolic blood pressure readings

Secondary

MeasureTime frameDescription
Comparison of Mean Arterial Pressure (MAP) Measured Withnon-invasive Cuffs to MAP Measured by Invasive Blood Pressure MonitoringSingle time point measurement (1 day)Absolute difference in the mean of 2 non-invasive MAP readings and mean of 2 invasive MAP readings
Comparison of Noninvasive Diastolic Blood Pressure Readings to Invasive Blood Pressure MeasurementSingle time point measurement (1 day)Absolute difference between the mean of 2 non-invasive diastolic blood pressure readings and mean of 2 invasive diastolic blood pressure readings
Mean of 2 Systolic Blood Pressure Readings Taken With Cylindrical and Conical BP CuffsSingle time point measurement (1 day)mean of 2 systolic blood pressure readings taken with cylindrical and conical BP cuffs
Mean of 2 MAP Readings Taken With Cylindrical and Conical BP CuffsSingle time point measurement (1 day)Mean of 2 MAP readings taken with cylindrical and conical BP cuffs
Mean of 2 Diastolic Blood Pressure Non-invasive BP Measurements Taken With Cylindrical and Conical BP CuffsSingle time point measurement (1 day)Mean of 2 diastolic blood pressure non-invasive BP measurements taken with cylindrical and conical BP cuffs

Countries

United States

Participant flow

Pre-assignment details

3 participants were randomized, but data was not analyzable on 2 participants. Randomization assignment was not recorded on these 2 participants.

Participants by arm

ArmCount
Arm A: Conical Then Cylindrical
BP recording with noninvasive conical Ultracheck Curve BP cuff and direct invasive arterial pressure monitoring; then with traditional cylindrical BP cuff and direct invasive arterial pressure monitoring Non-invasive BP recording with conical Ultracheck Curve BP cuff: Non-invasive SBP, DAP and MAP will be measured in both arms using conical Ultracheck Curve BP cuff Non-invasive BP recording with traditional cylindrical BP cuff: Non-invasive SBP, DAP and MAP will be measured in both arms using traditional cylindrical BP cuff Direct invasive arterial pressure: Direct invasive arterial pressure readings will be recorded simultaneously with non-invasive measurements
0
Arm B: Cylindrical Then Conical
BP recording with noninvasive traditional cylindrical BP cuff and direct invasive arterial pressure monitoring; then with conical Ultracheck Curve BP cuff and direct invasive arterial pressure monitoring Non-invasive BP recording with conical Ultracheck Curve BP cuff: Non-invasive SBP, DAP and MAP will be measured in both arms using conical Ultracheck Curve BP cuff Non-invasive BP recording with traditional cylindrical BP cuff: Non-invasive SBP, DAP and MAP will be measured in both arms using traditional cylindrical BP cuff Direct invasive arterial pressure: Direct invasive arterial pressure readings will be recorded simultaneously with non-invasive measurements
1
Total1

Baseline characteristics

CharacteristicArm B: Cylindrical Then ConicalTotalArm A: Conical Then Cylindrical
Age, Categorical
<=18 years
0 Participants0 Participants0 Participants
Age, Categorical
>=65 years
0 Participants0 Participants0 Participants
Age, Categorical
Between 18 and 65 years
1 Participants1 Participants0 Participants
Race and Ethnicity Not Collected0 Participants
Region of Enrollment
United States
1 participants1 participants
Sex: Female, Male
Female
1 Participants1 Participants0 Participants
Sex: Female, Male
Male
0 Participants0 Participants0 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 00 / 1
other
Total, other adverse events
0 / 00 / 1
serious
Total, serious adverse events
0 / 00 / 1

Outcome results

Primary

Comparison of Noninvasive Systolic Blood Pressure Readings to Invasive Blood Pressure Measurement

Absolute difference between the mean of 2 non-invasive systolic blood pressure readings and mean of 2 invasive systolic blood pressure readings

Time frame: Single time point measurement (1 day)

ArmMeasureValue (NUMBER)
ConicalComparison of Noninvasive Systolic Blood Pressure Readings to Invasive Blood Pressure Measurement-28 mm Hg
CylindricalComparison of Noninvasive Systolic Blood Pressure Readings to Invasive Blood Pressure Measurement-6.5 mm Hg
Secondary

Comparison of Mean Arterial Pressure (MAP) Measured Withnon-invasive Cuffs to MAP Measured by Invasive Blood Pressure Monitoring

Absolute difference in the mean of 2 non-invasive MAP readings and mean of 2 invasive MAP readings

Time frame: Single time point measurement (1 day)

ArmMeasureValue (NUMBER)
ConicalComparison of Mean Arterial Pressure (MAP) Measured Withnon-invasive Cuffs to MAP Measured by Invasive Blood Pressure Monitoring-25 mm Hg
CylindricalComparison of Mean Arterial Pressure (MAP) Measured Withnon-invasive Cuffs to MAP Measured by Invasive Blood Pressure Monitoring-13.5 mm Hg
Secondary

Comparison of Noninvasive Diastolic Blood Pressure Readings to Invasive Blood Pressure Measurement

Absolute difference between the mean of 2 non-invasive diastolic blood pressure readings and mean of 2 invasive diastolic blood pressure readings

Time frame: Single time point measurement (1 day)

ArmMeasureValue (NUMBER)
ConicalComparison of Noninvasive Diastolic Blood Pressure Readings to Invasive Blood Pressure Measurement-9.5 mm Hg
CylindricalComparison of Noninvasive Diastolic Blood Pressure Readings to Invasive Blood Pressure Measurement0.5 mm Hg
Secondary

Mean of 2 Diastolic Blood Pressure Non-invasive BP Measurements Taken With Cylindrical and Conical BP Cuffs

Mean of 2 diastolic blood pressure non-invasive BP measurements taken with cylindrical and conical BP cuffs

Time frame: Single time point measurement (1 day)

ArmMeasureValue (MEAN)Dispersion
ConicalMean of 2 Diastolic Blood Pressure Non-invasive BP Measurements Taken With Cylindrical and Conical BP Cuffs73.5 mm HgStandard Deviation 3.54
CylindricalMean of 2 Diastolic Blood Pressure Non-invasive BP Measurements Taken With Cylindrical and Conical BP Cuffs85 mm HgStandard Deviation 2.83
Secondary

Mean of 2 MAP Readings Taken With Cylindrical and Conical BP Cuffs

Mean of 2 MAP readings taken with cylindrical and conical BP cuffs

Time frame: Single time point measurement (1 day)

ArmMeasureValue (MEAN)Dispersion
ConicalMean of 2 MAP Readings Taken With Cylindrical and Conical BP Cuffs87.5 mm HgStandard Deviation 2.12
CylindricalMean of 2 MAP Readings Taken With Cylindrical and Conical BP Cuffs101.5 mm HgStandard Deviation 0.71
Secondary

Mean of 2 Systolic Blood Pressure Readings Taken With Cylindrical and Conical BP Cuffs

mean of 2 systolic blood pressure readings taken with cylindrical and conical BP cuffs

Time frame: Single time point measurement (1 day)

ArmMeasureValue (MEAN)Dispersion
ConicalMean of 2 Systolic Blood Pressure Readings Taken With Cylindrical and Conical BP Cuffs132.5 mm HgStandard Deviation 2.12
CylindricalMean of 2 Systolic Blood Pressure Readings Taken With Cylindrical and Conical BP Cuffs154 mm HgStandard Deviation 1.41