Skip to content

Clinical Investigation of the eyeWatch Glaucoma Drainage Device

Clinical Investigation of the eyeWatch Glaucoma Drainage Device

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03210571
Enrollment
27
Registered
2017-07-07
Start date
2018-05-01
Completion date
2020-10-27
Last updated
2021-10-28

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Open Angle Glaucoma

Brief summary

This study aims to test a new medical device, called the eyeWatch, used in filtering surgery for the treatment of glaucoma. This medical device has the following property: it allows better control of intraocular pressure in the initial postoperative phase by adjusting aqueous flow going through the device. The target criteria for the study are defined as: keeping intraocular pressure in target range; reducing the number of anti-glaucoma drugs; evaluating the number and type of postoperative complications. So far, this device has been already tested on 20 patients in Switzerland.

Interventions

DEVICEeyeWatch system

The eyeWatch device is composed of an AGDD, a control unit and a draining plate

Sponsors

Rheon Medical SA
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Patient must be 18 years of age or older. * Patient shall suffer from primary open angle glaucoma, pseudo-exfoliative glaucoma, or refractory POAG or PEXG after previous failed filtering surgery, in the study eye. Diagnosis is based on glaucomatous optic neuropathy, Shaffer angle greater than 2 as seen on gonioscopy, and visual field defect attributed to glaucoma. * Eye to be treated shall be phakic or pseudophakic with no other active ocular disease or disorder except for incipiens cataract. Prior ophthalmic surgery in the study eye shall have been performed more than 3 months before enrollment in this study. * Patient condition is indicated for primary and secondary filtration surgery. * Presence of elevated intraocular pressure defined as an intraocular corrected pressure (IOP) \> 20 mmHg in the study eye, under maximally tolerated medications. The IOP level shall be obtained on 2 consecutive measurements (not taken on the same day) prior to surgery. * Optic neuropathy is exclusively attributed to glaucoma. * Patient agreed to sign the written inform consent prior to entering into the investigation. * Patient is able and willing to complete post-operative follow-up requirements.

Exclusion criteria

* Diagnosis of neovascular glaucoma, congenital glaucoma. * History of previous intraocular surgery in the study eye referring to but not limited to extraocular muscles (strabismus), corneal transplant, retinal surgery. * Proliferative or severe non-proliferative retinopathy in either eye. * Congenital anomaly of the anterior chamber angle in the study eye. * Optic neuropathy other than glaucoma in the study eye. * Patient with retinal vein occlusion in the study eye. * Patient with retinal artery occlusion in the study eye. * Patient with corneal opacifications, endotheliopathy, or irregularities that may interfere with the optic nerve evaluation or the IOP measurements in the study eye. * Patient with a history of severe eye trauma in the study eye. * Patient with ocular malformations such as microphthalmia in the study eye. * Patient with concurrent inflammatory/infective eye disorder in the study eye. * Patient with severe systemic disease or disabling conditions such as chronic renal failure, post organ transplants. * Patient participating in another clinical trial or having participated in another clinical trial less than 3 months prior to entering into the investigation. * Patient is pregnant, breast-feeding or unable to make the decision to participate in a clinical investigation (e.g. mentally ill person, mentally handicapped person)

Design outcomes

Primary

MeasureTime frameDescription
Number and type of serious adverse device eventsFollow-up: day 1, Week 1, week 2, month 1, month 2, month 3, month 6, month 12 compared to baselineDemonstration of safety. The number and type of serious adverse device event per patient will be measured. Results of biomicroscopy, gonioscopy, pachymetry visual field and endothelial count will be compared to baseline.
Effectiveness in reducing the intraocular pressureFollow-up: day 1, Week 1, week 2, month 1, month 2, month 3, month 6, month 12 compared to baselineDemonstration of the performance. The intraocular pressure (IOP in mmHg) will be measured at different time points to show the effective reduction compared to baseline. IOP reduced by ≥ 20% or IOP \< 21 mmHg, and no IOP \< 5 mmHg on two consecutive visits after 1 month (included)

Countries

United Kingdom

Outcome results

None listed