Brief summary

To evaluate the change in patient Oral Health-related Quality of Life (OHRQoL) from wearing a cobalt chrome (CoCr) removable partial denture (RPD) to wearing the Solvay Dental 360TM polymer Removable Partial Denture (RPD) for 8 weeks.

Detailed description

Cohort study conducted in the United States & United Kingdom to assess Oral Health Related Quality of Life (OHRQoL) in patients receiving the study polymer Removable Partial Denture (RPD) as compared to their baseline Cobalt Chrome (CoCr) Removable Partial Denture (RPD) Oral Health Quality of Life (OHRQoL).

Interventions

DEVICESolvay Dental 360™

polymer Removable Partial Denture (RPD)

Sponsors

Solvay Dental 360
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* with the absence of 3 or less teeth per saddle, excluding third molars * partial denture in one arch only * with previous denture wearing experience, and Cobalt Chrome (CoCr) Removable Partial Denture (RPD) fitted in the previous 24 months and the patient is currently and routinely wearing the denture * With a stable oral health in terms of absence of disease activity affecting the periodontium, dental hard tissues (caries), pulp and the structural/aesthetic integrity of restored teeth * With at least 1 posterior natural tooth occlusal stop (molar or premolar) * Class I, Class II and Class III (Kennedy's Classification) * Occlusal spacing (static and dynamic) around clasp assembly including the occlusal rest

Exclusion criteria

* Patients requiring extensive treatment prior to the provision of a removable partial denture (RPD) * Patients with active primary disease; caries, periodontal disease or symptoms of pulpal or apical pathology to the remaining teeth * Patients with irreversibly compromised structural/aesthetic integrity of restored teeth that cannot be restored as part of the provision of treatment. * Patients with an opposing Removable Partial Denture (RPD). Removable Partial Denture (RPD) to only be in one arch and must oppose a full denture or a dentate arch which provides a stable occlusal relationship * Canines are NOT to be replaced with the Removable Partial Denture's (RPDs). Patients must have their natural canines present

Design outcomes

Primary

MeasureTime frameDescription
Change in Patient Oral Health-related QualityBaseline for the Cobalt Chrome (CoCr) Removable Partial Denture (RPD) and 8 weeks for Polymer Removable Partial DentureSubjects conducted the Oral Health-related Quality of Life (OHRQoL) before and after wearing the Solvay Dental 360™ polymer Removable Partial Denture (RPD). OHRQoL is a multidimensional survey which can be defined as a person's assessment of how functional, psychological, social factors, pain, or discomfort affect his/her well-being in the context of oral health. The purpose of the survey is to evaluate the change in these factors from wearing a cobalt chrome (CoCr) removable partial denture (RPD) to wearing the Solvay Dental 360™ polymer Removable Partial Denture (RPD). This score is calculated by a summation of the score from each question; each having the following scoring: Response Score Never 0 Points Hardly Ever 1 Point Occasionally 2 Points Fairly Often 3 Points Very Often 4 Points It ranges from 0 to 56 points. The overall scores are from the baseline Cobalt Chrome (CoCr) RPD and 8 weeks for Polymer RPD. Lower scores are better outcomes.

Secondary

MeasureTime frameDescription
Operator AssessmentBaseline for the Cobalt Chrome (CoCr) Removable Partial Denture (RPD) and visits 5 (5 weeks), 6 (6 weeks), 7 (7 weeks) and 8 (8 weeks) for the Solvay Dental 360™: Polymer Removable Partial DentureOperator assessment of the adaptation of the polymer frame, adaption of the tissue base areas and their relationship to the frame adaptation, functional evaluation, aesthetic assessment and oral health assessment of abutment teeth and the periodontium. This assessment was done with a Likert scale at baseline for the Cobalt Chrome (CoCr) Removable Partial Denture (RPD) and visits 5, 6, 7 and 8 for the Solvay Dental 360™: Polymer Removable Partial Denture. This assessment of aesthetics was determined by the operator based on a five point scale rating from very dissatisfied through very satisfied. More satisfied is a better outcome.
Patient Frame PreferenceBaseline for the Cobalt Chrome (CoCr) Removable Partial Denture (RPD) and 8 weeks for the Solvay Dental 360™: Polymer Removable Partial DenturePatient-centered assessment of preference at the final visit for the study. Patients were asked to choose their preference of Removable Partial Denture based on comfort, aesthetics and chewing. Preference for a specific denture is a better outcome.
Review for Mechanical FailureApproximately 8 weeksThis is an evaluation of the Solvay Dental 360™ Polymer Removable Partial Denture (RPD) for chips, cracks, fractures, permanent deformation or other signs mechanical failure. The number in the reported data represents a cumulative number of Solvay Dental 360™Polymer Removable Partial Denture (RPD) devices with these mechanical failures. A device deficiency was defined as the inadequacy of a medical device with respect to its identity, quality, durability, reliability, safety or performance.
Comparison of Health of Abutment TeethBaseline for the Cobalt Chrome (CoCr) Removable Partial Denture (RPD) and 8 weeks for the Solvay Dental 360™: Polymer Removable Partial DentureTo investigate the difference in health of the abutment teeth following the use of the polymer RPD as compared to the CoCr RPD as assessed by the plaque score and gingival bleeding index. The plaque score is determined by assessing each tooth and the amount of plaque on the individual surfaces. The total number of plaque containing surfaces is divided by the total number of surfaces per arch of the mouth. Arch plaque is a subset of the whole plaque score calculated as the number of arch surfaces with plaque divided by the number of arch surfaces assessed. The gingival bleeding index is a designation of the degree of inflammation, as determined by a bleeding on probing assessment. Each tooth is probed at six sites and the bleeding is scored as present or absent. The number of sites where the bleeding is noted is divided by the total number of surfaces evaluated. All assessments are expressed as a percentage and an increased percentage indicates a poorer patient outcome.
Oral Health Parameters: Operator Assessment of Mobility:Baseline for the Cobalt Chrome (CoCr) Removable Partial Denture (RPD) and 8 weeks for the Solvay Dental 360™: Polymer Removable Partial DentureOral Health parameters associated with the Removable Partial Denture (RPD) abutment teeth and Removable Partial Denture (RPD): mobility (measured as a categorical value of either Class I, Class II or Class III). Mobility of the teeth abutting the missing tooth as well as adjacent to the abutment tooth are classified via the Miller Classification as the following: * Class I: Tooth can be moved less than 1mm in the buccolingual or mesiodistal direction; this included no mobility * Class II: Tooth can be moved 1mm or more in the buccolingual or mesiodistal direction. No mobility in the occlusoapical direction (vertical mobility) * Class III: Tooth can be moved 1mm or more in the buccolingual or mesiodistal direction. Mobility in the occlusoapical direction is also present The mobility score assessment is noted as proportion of teeth that moved from a lower classification to a higher classification. Class I is a better outcome.
Oral Health Parameters: Mucosal Bearing AreasBaseline for the Cobalt Chrome (CoCr) Removable Partial Denture (RPD) and 8 weeks for the Solvay Dental 360™: Polymer Removable Partial DentureOral Health parameters associated with the Solvay Dental 360™ Removable Partial Denture (RPD): mucosal bearing areas (measured as an evaluation of bleeding on probing indicated as a categorical value of bleeding present or bleeding absent, as compared to the Cobalt Chrome (CoCr) Removable Partial Denture (RPD). Operators were asked to rate the health of the mucosal bearing areas as poor, fair or good. The mucosal bearing area is the cushion between the denture base and the supporting bone. Good is a better outcome.
Operator Assessment of Pocket DepthBaseline for the Cobalt Chrome (CoCr) Removable Partial Denture (RPD) and 8 weeks for the Solvay Dental 360™: Polymer Removable Partial DentureOral Health parameters associated with the Removable Partial Denture (RPD) on pocket depths based on Operator assessment. Pocket depths are expressed in mm increments for mesial, mid, and distal locations on the lingual and buccal surfaces of the tooth. To measure pocket depth, a probe is placed beside the tooth beneath the gumline and indicates the groove between the gums and teeth. Visit 2 with the Cobalt Chrome (CoCr) Removable Partial Denture (RPD) is compared to visit 8 with the Solvay Dental 360™: Polymer Removable Partial Denture. In general, pocket depths of 1 to 3 mm indicate a healthy mouth and 4mm and greater may indicate periodontitis. The score is calculated as an average of all depths reported for all surfaces assessed. A lower score is a better outcome.

Countries

United Kingdom, United States

Participant flow

Participants by arm

ArmCount
Polymer Removable Partial Denture
Evaluate the change in patient Oral Health-related Quality of Life while wearing the Solvay Dental 360™ polymer Removable Partial Denture (RPD) Solvay Dental 360™: polymer Removable Partial Denture (RPD)
46
Total46

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyPhysician Decision2
Overall StudyProtocol Violation2
Overall StudyStudy Ended1
Overall StudyWithdrawal by Subject1

Baseline characteristics

CharacteristicPolymer Removable Partial Denture
Age, Continuous64.63 years
STANDARD_DEVIATION 12.57
Education
College
15 Participants
Education
Elementry School
4 Participants
Education
Graduate School
8 Participants
Education
High School
17 Participants
Education
Other
2 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
Race (NIH/OMB)
Asian
6 Participants
Race (NIH/OMB)
Black or African American
6 Participants
Race (NIH/OMB)
More than one race
7 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants
Race (NIH/OMB)
White
27 Participants
Region of Enrollment
United States
46 Participants
Sex: Female, Male
Female
19 Participants
Sex: Female, Male
Male
27 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 46
other
Total, other adverse events
0 / 46
serious
Total, serious adverse events
0 / 46

Outcome results

Primary

Change in Patient Oral Health-related Quality

Subjects conducted the Oral Health-related Quality of Life (OHRQoL) before and after wearing the Solvay Dental 360™ polymer Removable Partial Denture (RPD). OHRQoL is a multidimensional survey which can be defined as a person's assessment of how functional, psychological, social factors, pain, or discomfort affect his/her well-being in the context of oral health. The purpose of the survey is to evaluate the change in these factors from wearing a cobalt chrome (CoCr) removable partial denture (RPD) to wearing the Solvay Dental 360™ polymer Removable Partial Denture (RPD). This score is calculated by a summation of the score from each question; each having the following scoring: Response Score Never 0 Points Hardly Ever 1 Point Occasionally 2 Points Fairly Often 3 Points Very Often 4 Points It ranges from 0 to 56 points. The overall scores are from the baseline Cobalt Chrome (CoCr) RPD and 8 weeks for Polymer RPD. Lower scores are better outcomes.

Time frame: Baseline for the Cobalt Chrome (CoCr) Removable Partial Denture (RPD) and 8 weeks for Polymer Removable Partial Denture

ArmMeasureValue (MEAN)Dispersion
Solvay Dental 360™: Polymer Removable Partial DentureChange in Patient Oral Health-related Quality9.01 score on a scaleStandard Deviation 8.14
Cobalt Chrome (CoCr) Removable Partial Denture (RPD)Change in Patient Oral Health-related Quality14.41 score on a scaleStandard Deviation 9.48
Secondary

Comparison of Health of Abutment Teeth

To investigate the difference in health of the abutment teeth following the use of the polymer RPD as compared to the CoCr RPD as assessed by the plaque score and gingival bleeding index. The plaque score is determined by assessing each tooth and the amount of plaque on the individual surfaces. The total number of plaque containing surfaces is divided by the total number of surfaces per arch of the mouth. Arch plaque is a subset of the whole plaque score calculated as the number of arch surfaces with plaque divided by the number of arch surfaces assessed. The gingival bleeding index is a designation of the degree of inflammation, as determined by a bleeding on probing assessment. Each tooth is probed at six sites and the bleeding is scored as present or absent. The number of sites where the bleeding is noted is divided by the total number of surfaces evaluated. All assessments are expressed as a percentage and an increased percentage indicates a poorer patient outcome.

Time frame: Baseline for the Cobalt Chrome (CoCr) Removable Partial Denture (RPD) and 8 weeks for the Solvay Dental 360™: Polymer Removable Partial Denture

ArmMeasureGroupValue (MEAN)Dispersion
Solvay Dental 360™: Polymer Removable Partial DentureComparison of Health of Abutment TeethWhole Plaque Score36.48 Percentage of assessed surfacesStandard Deviation 19.95
Solvay Dental 360™: Polymer Removable Partial DentureComparison of Health of Abutment TeethArch Plaque Score37.39 Percentage of assessed surfacesStandard Deviation 20.94
Solvay Dental 360™: Polymer Removable Partial DentureComparison of Health of Abutment TeethBleeding on Probing16.33 Percentage of assessed surfacesStandard Deviation 15.03
Cobalt Chrome (CoCr) Removable Partial Denture (RPD)Comparison of Health of Abutment TeethArch Plaque Score39.87 Percentage of assessed surfacesStandard Deviation 25.1
Cobalt Chrome (CoCr) Removable Partial Denture (RPD)Comparison of Health of Abutment TeethBleeding on Probing20.40 Percentage of assessed surfacesStandard Deviation 21.24
Cobalt Chrome (CoCr) Removable Partial Denture (RPD)Comparison of Health of Abutment TeethWhole Plaque Score39.42 Percentage of assessed surfacesStandard Deviation 25.42
Secondary

Operator Assessment

Operator assessment of the adaptation of the polymer frame, adaption of the tissue base areas and their relationship to the frame adaptation, functional evaluation, aesthetic assessment and oral health assessment of abutment teeth and the periodontium. This assessment was done with a Likert scale at baseline for the Cobalt Chrome (CoCr) Removable Partial Denture (RPD) and visits 5, 6, 7 and 8 for the Solvay Dental 360™: Polymer Removable Partial Denture. This assessment of aesthetics was determined by the operator based on a five point scale rating from very dissatisfied through very satisfied. More satisfied is a better outcome.

Time frame: Baseline for the Cobalt Chrome (CoCr) Removable Partial Denture (RPD) and visits 5 (5 weeks), 6 (6 weeks), 7 (7 weeks) and 8 (8 weeks) for the Solvay Dental 360™: Polymer Removable Partial Denture

Population: The professional (operator) is surveyed about the outcomes for the fit, function, occlusion and aesthetics.

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentSatisified27 Participants
Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentVery Dissatisified1 Participants
Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentVery Satisified6 Participants
Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentDissatisified5 Participants
Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentNeither Satisfied or Dissatisified5 Participants
Cobalt Chrome (CoCr) Removable Partial Denture (RPD)Operator AssessmentSatisified26 Participants
Cobalt Chrome (CoCr) Removable Partial Denture (RPD)Operator AssessmentNeither Satisfied or Dissatisified9 Participants
Cobalt Chrome (CoCr) Removable Partial Denture (RPD)Operator AssessmentDissatisified1 Participants
Cobalt Chrome (CoCr) Removable Partial Denture (RPD)Operator AssessmentVery Satisified5 Participants
Cobalt Chrome (CoCr) Removable Partial Denture (RPD)Operator AssessmentVery Dissatisified1 Participants
Visit 6: Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentNeither Satisfied or Dissatisified8 Participants
Visit 6: Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentVery Dissatisified0 Participants
Visit 6: Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentDissatisified4 Participants
Visit 6: Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentSatisified22 Participants
Visit 6: Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentVery Satisified6 Participants
Visit 7: Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentVery Satisified7 Participants
Visit 7: Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentVery Dissatisified0 Participants
Visit 7: Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentSatisified19 Participants
Visit 7: Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentNeither Satisfied or Dissatisified11 Participants
Visit 7: Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentDissatisified1 Participants
Visit 8: Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentNeither Satisfied or Dissatisified8 Participants
Visit 8: Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentSatisified24 Participants
Visit 8: Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentVery Dissatisified0 Participants
Visit 8: Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentVery Satisified5 Participants
Visit 8: Solvay Dental 360™: Polymer Removable Partial DentureOperator AssessmentDissatisified3 Participants
Secondary

Operator Assessment of Pocket Depth

Oral Health parameters associated with the Removable Partial Denture (RPD) on pocket depths based on Operator assessment. Pocket depths are expressed in mm increments for mesial, mid, and distal locations on the lingual and buccal surfaces of the tooth. To measure pocket depth, a probe is placed beside the tooth beneath the gumline and indicates the groove between the gums and teeth. Visit 2 with the Cobalt Chrome (CoCr) Removable Partial Denture (RPD) is compared to visit 8 with the Solvay Dental 360™: Polymer Removable Partial Denture. In general, pocket depths of 1 to 3 mm indicate a healthy mouth and 4mm and greater may indicate periodontitis. The score is calculated as an average of all depths reported for all surfaces assessed. A lower score is a better outcome.

Time frame: Baseline for the Cobalt Chrome (CoCr) Removable Partial Denture (RPD) and 8 weeks for the Solvay Dental 360™: Polymer Removable Partial Denture

ArmMeasureValue (MEAN)Dispersion
Solvay Dental 360™: Polymer Removable Partial DentureOperator Assessment of Pocket Depth2.19 pocket depths expressed in mmStandard Deviation 0.63
Cobalt Chrome (CoCr) Removable Partial Denture (RPD)Operator Assessment of Pocket Depth2.23 pocket depths expressed in mmStandard Deviation 0.64
Secondary

Oral Health Parameters: Mucosal Bearing Areas

Oral Health parameters associated with the Solvay Dental 360™ Removable Partial Denture (RPD): mucosal bearing areas (measured as an evaluation of bleeding on probing indicated as a categorical value of bleeding present or bleeding absent, as compared to the Cobalt Chrome (CoCr) Removable Partial Denture (RPD). Operators were asked to rate the health of the mucosal bearing areas as poor, fair or good. The mucosal bearing area is the cushion between the denture base and the supporting bone. Good is a better outcome.

Time frame: Baseline for the Cobalt Chrome (CoCr) Removable Partial Denture (RPD) and 8 weeks for the Solvay Dental 360™: Polymer Removable Partial Denture

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Solvay Dental 360™: Polymer Removable Partial DentureOral Health Parameters: Mucosal Bearing AreasGood36 Participants
Solvay Dental 360™: Polymer Removable Partial DentureOral Health Parameters: Mucosal Bearing AreasFair4 Participants
Solvay Dental 360™: Polymer Removable Partial DentureOral Health Parameters: Mucosal Bearing AreasPoor0 Participants
Cobalt Chrome (CoCr) Removable Partial Denture (RPD)Oral Health Parameters: Mucosal Bearing AreasFair4 Participants
Cobalt Chrome (CoCr) Removable Partial Denture (RPD)Oral Health Parameters: Mucosal Bearing AreasGood40 Participants
Cobalt Chrome (CoCr) Removable Partial Denture (RPD)Oral Health Parameters: Mucosal Bearing AreasPoor0 Participants
Secondary

Oral Health Parameters: Operator Assessment of Mobility:

Oral Health parameters associated with the Removable Partial Denture (RPD) abutment teeth and Removable Partial Denture (RPD): mobility (measured as a categorical value of either Class I, Class II or Class III). Mobility of the teeth abutting the missing tooth as well as adjacent to the abutment tooth are classified via the Miller Classification as the following: * Class I: Tooth can be moved less than 1mm in the buccolingual or mesiodistal direction; this included no mobility * Class II: Tooth can be moved 1mm or more in the buccolingual or mesiodistal direction. No mobility in the occlusoapical direction (vertical mobility) * Class III: Tooth can be moved 1mm or more in the buccolingual or mesiodistal direction. Mobility in the occlusoapical direction is also present The mobility score assessment is noted as proportion of teeth that moved from a lower classification to a higher classification. Class I is a better outcome.

Time frame: Baseline for the Cobalt Chrome (CoCr) Removable Partial Denture (RPD) and 8 weeks for the Solvay Dental 360™: Polymer Removable Partial Denture

Population: Denominator is the total number of teeth assessed in the jaw that the RPD was provided for.

ArmMeasureGroupValue (COUNT_OF_UNITS)
Solvay Dental 360™: Polymer Removable Partial DentureOral Health Parameters: Operator Assessment of Mobility:Class I211 Teeth
Solvay Dental 360™: Polymer Removable Partial DentureOral Health Parameters: Operator Assessment of Mobility:Class II7 Teeth
Solvay Dental 360™: Polymer Removable Partial DentureOral Health Parameters: Operator Assessment of Mobility:Class III1 Teeth
Cobalt Chrome (CoCr) Removable Partial Denture (RPD)Oral Health Parameters: Operator Assessment of Mobility:Class I221 Teeth
Cobalt Chrome (CoCr) Removable Partial Denture (RPD)Oral Health Parameters: Operator Assessment of Mobility:Class II11 Teeth
Cobalt Chrome (CoCr) Removable Partial Denture (RPD)Oral Health Parameters: Operator Assessment of Mobility:Class III1 Teeth
Secondary

Patient Frame Preference

Patient-centered assessment of preference at the final visit for the study. Patients were asked to choose their preference of Removable Partial Denture based on comfort, aesthetics and chewing. Preference for a specific denture is a better outcome.

Time frame: Baseline for the Cobalt Chrome (CoCr) Removable Partial Denture (RPD) and 8 weeks for the Solvay Dental 360™: Polymer Removable Partial Denture

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
Solvay Dental 360™: Polymer Removable Partial DenturePatient Frame Preference32 Participants
Cobalt Chrome (CoCr) Removable Partial Denture (RPD)Patient Frame Preference8 Participants
Secondary

Review for Mechanical Failure

This is an evaluation of the Solvay Dental 360™ Polymer Removable Partial Denture (RPD) for chips, cracks, fractures, permanent deformation or other signs mechanical failure. The number in the reported data represents a cumulative number of Solvay Dental 360™Polymer Removable Partial Denture (RPD) devices with these mechanical failures. A device deficiency was defined as the inadequacy of a medical device with respect to its identity, quality, durability, reliability, safety or performance.

Time frame: Approximately 8 weeks

Population: Number of Solvay Dental 360™: Polymer Removable Partial Denture with mechanical failures

ArmMeasureGroupValue (NUMBER)
Solvay Dental 360™: Polymer Removable Partial DentureReview for Mechanical FailureLoss1 Number of Removable Partial Dentures
Solvay Dental 360™: Polymer Removable Partial DentureReview for Mechanical FailureChips2 Number of Removable Partial Dentures
Solvay Dental 360™: Polymer Removable Partial DentureReview for Mechanical FailureCrack1 Number of Removable Partial Dentures
Solvay Dental 360™: Polymer Removable Partial DentureReview for Mechanical FailureDetachment1 Number of Removable Partial Dentures
Solvay Dental 360™: Polymer Removable Partial DentureReview for Mechanical FailureFracture14 Number of Removable Partial Dentures
Solvay Dental 360™: Polymer Removable Partial DentureReview for Mechanical FailureMilling1 Number of Removable Partial Dentures