Patient-Assisted Compression - Impact on Image Quality and Workflow

Patient-Assisted Compression - Impact on Image Quality and Workflow

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03196635
Enrollment
31
Registered
2017-06-23
Start date
2017-06-19
Completion date
2017-06-29
Last updated
2018-08-31

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Breast Cancer

Brief summary

This study is being conducted to compare the image quality of breast images obtained using standard (technologist-controlled \[TC\]) compression and patient-assisted (PA) compression and to evaluate the impact of PA compression on clinical workflow.

Detailed description

Patient-assisted (PA) compression allows the patient to participate in controlling the amount of compression force during mammography and is a personalized approach that has demonstrated successful reduction in discomfort experienced during mammography. GE Healthcare's Senographe Pristina, an innovative mammography platform that provides both two-dimensional (2D) and three-dimensional (3D) imaging capabilities, offers both standard and PA compression modes. This study is being conducted to compare the image quality of breast images obtained using TC compression and PA compression, and to evaluate the impact of PA compression on clinical workflow. The study population will consist of adult asymptomatic women presenting for screening 2D mammography.

Interventions

DEVICEPatient-Assisted (PA) Breast Compression

The technologist will properly position the breast and apply minimum compression. The subject will be instructed to apply compression as the technologist ensures the breast tissue is in appropriate position and tautness. The technologist will then guide the subject to achieve appropriate compression level, sufficient but not excessive, and the image will be acquired. This will be done for both standard views CC & MLO.

DEVICETechnologist-Controlled (TC) Breast Compression

TC compression will be conducted per standard of care practices at the site.

Sponsors

GE Healthcare
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
SCREENING
Masking
NONE

Eligibility

Sex/Gender
FEMALE
Age
40 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Are women aged 40 years or older; * Are asymptomatic and scheduled for FFDM screening mammography; * Have left and right breasts; * Have breast sizes compatible with the dimensions of a 24 x 31 cm image detector, without anatomical cut-off; * Are documented as non-pregnant based on the investigator's medical judgment and in consideration of local clinical practice standards for evidence of non-pregnancy; * Are able and willing to comply with study procedures; and * Are able and willing to provide written informed consent to participate.

Exclusion criteria

* Have been previously included in this study or are participating in another study expected to interfere with study procedures or outcomes; * Have undergone diagnostic or surgical intervention(s) or procedure(s) on either breast, including breast biopsy, lumpectomy, or reconstruction, within five (5) years (≤ 5 years) of the study exam date; * Are currently undergoing radiotherapy or chemotherapy, or have a history of prior radiotherapy treatment on either breast; * Are currently lactating; or * Have breast implants.

Design outcomes

Primary

MeasureTime frameDescription
Number of Subjects With Acceptable Overall Clinical Image Quality for Patient-assisted (PA) and Technologist-controlled (TC) Image SetsThrough study completion, on average 1 monthOne PA image set and one TC image set was acquired from each completed subject. The overall clinical image quality acceptability was collected and summarized on a per subject-basis using binary responses of either acceptable or unacceptable for unilateral, two-view PA and TC compression image sets. Two readers evaluated each of the 60 image sets (30 PA and 30 TC compression image sets from 30 completed participants). In cases of disagreement between Readers 1 and 2, a third reader provided adjudication.

Secondary

MeasureTime frameDescription
Repeat Image AcquisitionThrough study completion, on average 1 monthNumber of incidences per image set when the technologists or readers indicated a repeat acquisition for PA and TC compression modes. More than one incident (i.e. reason for repeat from technologist or view indicated for repeat by readers) could have been selected for a given image set.
Acceptability of Mammographic AttributesThrough study completion, on average 1 monthAcceptability of mammographic attributes using a binary response of either acceptable or unacceptable for unilateral, two-view PA and TC compression image sets were summarized.

Other

MeasureTime frameDescription
Breast Thickness for PA and TC CompressionsThrough study completion, on average 1 monthBreast thickness (millimeter \[mm\]) for each compression mode (TC and PA compression) were summarized for craniocaudal (CC) and mediolateral oblique (MLO) views. For each subject, both views were collected using both compression modes.
Comparison of Image Acquisition TimeThrough study completion, on average 1 monthThe length of time it takes for image acquisition using each compression mode (TC Compression and PA Compression) were compared.
Radiation Dose for PA and TC CompressionsThrough study completion, on average 1 monthRadiation dose (entrance skin air kerma \[ESAK\] in milligray \[mGy\]) for each compression mode (TC and PA compression) were summarized for craniocaudal (CC) and mediolateral oblique (MLO) views. For each subject, both views were collected using both compression modes.
Technologist Interventions in PA CompressionThrough study completion, on average 1 monthData will be collected regarding any interventions made by the technologist during PA compression and this data were summarized.
Compression Force for PA and TC CompressionsThrough study completion, on average 1 monthCompression forces (decanewton \[daN\]) for each compression mode (TC and PA compression) were summarized for craniocaudal (CC) and mediolateral oblique (MLO) views. For each subject, both views were collected using both compression modes.

Countries

United States

Participant flow

Participants by arm

ArmCount
All Study Participants
All subjects will undergo their regularly scheduled full-field digital mammogram, consisting of bilateral, 2-view (CC/MLO) image acquisition. In addition, a study-specific, unilateral 2-view image set will obtained, utilizing the patient-assisted breast compression mode. Patient-Assisted Breast Compression: The technologist will properly position the breast and apply minimum compression. The subject will be instructed to apply compression as the technologist ensures the breast tissue is in appropriate position and tautness. The technologist will then guide the subject to achieve appropriate compression level, sufficient but not excessive, and the image will be acquired. This will be done for both standard views CC/MLO.
31
Total31

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyWithdrawal by Subject1

Baseline characteristics

CharacteristicAll Study Participants
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
8 Participants
Age, Categorical
Between 18 and 65 years
23 Participants
Breast of Interest
First breast of interest
16 Participants
Breast of Interest
Second breast of interest
14 Participants
Breast of Interest
Withdrawn subject
1 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
1 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
29 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
1 Participants
Sex: Female, Male
Female
31 Participants
Sex: Female, Male
Male
0 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 30
other
Total, other adverse events
0 / 30
serious
Total, serious adverse events
0 / 30

Outcome results

Primary

Number of Subjects With Acceptable Overall Clinical Image Quality for Patient-assisted (PA) and Technologist-controlled (TC) Image Sets

One PA image set and one TC image set was acquired from each completed subject. The overall clinical image quality acceptability was collected and summarized on a per subject-basis using binary responses of either acceptable or unacceptable for unilateral, two-view PA and TC compression image sets. Two readers evaluated each of the 60 image sets (30 PA and 30 TC compression image sets from 30 completed participants). In cases of disagreement between Readers 1 and 2, a third reader provided adjudication.

Time frame: Through study completion, on average 1 month

Population: 60 image sets (30 PA and 30 TC compression image sets from 30 participants) were evaluated for acceptability of overall clinical image quality by two readers. The one (1) withdrawn subject did not undergo image acquisition study procedures and thus was not included in this analysis set.

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
All Study Participants, PA Compression Image SetsNumber of Subjects With Acceptable Overall Clinical Image Quality for Patient-assisted (PA) and Technologist-controlled (TC) Image Sets30 Participants
All Study Participants, TC Compression Image SetsNumber of Subjects With Acceptable Overall Clinical Image Quality for Patient-assisted (PA) and Technologist-controlled (TC) Image Sets30 Participants
Secondary

Acceptability of Mammographic Attributes

Acceptability of mammographic attributes using a binary response of either acceptable or unacceptable for unilateral, two-view PA and TC compression image sets were summarized.

Time frame: Through study completion, on average 1 month

Population: The PA and TC image sets for all 30 subjects were evaluated for acceptability of pre-defined set of mammography attributes by Readers 1 and 2. Adjudication was not required if there was disagreement between the readers. The one withdrawn subject did not undergo image acquisition study procedures and thus was not included in this analysis set.

ArmMeasureGroupValue (NUMBER)
All Study Participants, PA Compression Image SetsAcceptability of Mammographic AttributesAcceptable sharpness of the image30 image sets
All Study Participants, PA Compression Image SetsAcceptability of Mammographic AttributesAcceptable Breast Positioning30 image sets
All Study Participants, PA Compression Image SetsAcceptability of Mammographic AttributesAcceptable tissue visibility at the skin line30 image sets
All Study Participants, PA Compression Image SetsAcceptability of Mammographic AttributesAcceptable breast compression30 image sets
All Study Participants, PA Compression Image SetsAcceptability of Mammographic AttributesAcceptable noise level of the image30 image sets
All Study Participants, PA Compression Image SetsAcceptability of Mammographic AttributesAcceptable exposure and visualization of tissue30 image sets
All Study Participants, PA Compression Image SetsAcceptability of Mammographic AttributesAcceptable artifacts level30 image sets
All Study Participants, PA Compression Image SetsAcceptability of Mammographic AttributesAcceptable image contrast30 image sets
All Study Participants, TC Compression Image SetsAcceptability of Mammographic AttributesAcceptable artifacts level30 image sets
All Study Participants, TC Compression Image SetsAcceptability of Mammographic AttributesAcceptable Breast Positioning29 image sets
All Study Participants, TC Compression Image SetsAcceptability of Mammographic AttributesAcceptable exposure and visualization of tissue30 image sets
All Study Participants, TC Compression Image SetsAcceptability of Mammographic AttributesAcceptable breast compression30 image sets
All Study Participants, TC Compression Image SetsAcceptability of Mammographic AttributesAcceptable sharpness of the image30 image sets
All Study Participants, TC Compression Image SetsAcceptability of Mammographic AttributesAcceptable tissue visibility at the skin line30 image sets
All Study Participants, TC Compression Image SetsAcceptability of Mammographic AttributesAcceptable noise level of the image30 image sets
All Study Participants, TC Compression Image SetsAcceptability of Mammographic AttributesAcceptable image contrast30 image sets
Secondary

Repeat Image Acquisition

Number of incidences per image set when the technologists or readers indicated a repeat acquisition for PA and TC compression modes. More than one incident (i.e. reason for repeat from technologist or view indicated for repeat by readers) could have been selected for a given image set.

Time frame: Through study completion, on average 1 month

Population: Instances of repeat imaging for each subject's PA (N=30) and TC (N=30) image sets were indicated by technologists during image acquisitions and by readers during image evaluations. The one withdrawn subject did not undergo image acquisition study procedures and thus was not included in this analysis set.

ArmMeasureGroupValue (COUNT_OF_UNITS)
All Study Participants, PA Compression Image SetsRepeat Image AcquisitionReader-indicated Repeat4 Image sets
All Study Participants, PA Compression Image SetsRepeat Image AcquisitionTechnologist-indicated Repeat2 Image sets
All Study Participants, TC Compression Image SetsRepeat Image AcquisitionReader-indicated Repeat4 Image sets
All Study Participants, TC Compression Image SetsRepeat Image AcquisitionTechnologist-indicated Repeat7 Image sets
Other Pre-specified

Breast Thickness for PA and TC Compressions

Breast thickness (millimeter \[mm\]) for each compression mode (TC and PA compression) were summarized for craniocaudal (CC) and mediolateral oblique (MLO) views. For each subject, both views were collected using both compression modes.

Time frame: Through study completion, on average 1 month

Population: Breast thickness (mm) was captured for each view under each compression for all subjects. The one withdrawn subject did not undergo image acquisition study procedures and thus was not included in this analysis set.

ArmMeasureGroupValue (MEAN)Dispersion
All Study Participants, PA Compression Image SetsBreast Thickness for PA and TC CompressionsCC52.0 mmStandard Deviation 14.3
All Study Participants, PA Compression Image SetsBreast Thickness for PA and TC CompressionsMLO54.6 mmStandard Deviation 16.1
All Study Participants, TC Compression Image SetsBreast Thickness for PA and TC CompressionsCC53.2 mmStandard Deviation 14
All Study Participants, TC Compression Image SetsBreast Thickness for PA and TC CompressionsMLO53.7 mmStandard Deviation 15.1
Other Pre-specified

Comparison of Image Acquisition Time

The length of time it takes for image acquisition using each compression mode (TC Compression and PA Compression) were compared.

Time frame: Through study completion, on average 1 month

Population: Imaging time was collected for each image set. 4 image sets were excluded from TC analysis due to repeat imaging per standard of care; repeat imaging wasn't allowed for PA compression. 1 subject's PA \& TC image sets were excluded from analysis due to a protocol deviation. 1 withdrawn subject did not undergo imaging, so was not part of the analysis.

ArmMeasureValue (MEAN)Dispersion
All Study Participants, PA Compression Image SetsComparison of Image Acquisition Time3.9 minutesStandard Deviation 1
All Study Participants, TC Compression Image SetsComparison of Image Acquisition Time3.4 minutesStandard Deviation 1
Other Pre-specified

Compression Force for PA and TC Compressions

Compression forces (decanewton \[daN\]) for each compression mode (TC and PA compression) were summarized for craniocaudal (CC) and mediolateral oblique (MLO) views. For each subject, both views were collected using both compression modes.

Time frame: Through study completion, on average 1 month

Population: Compression force (daN) was captured for each view under each compression for all subjects. The one withdrawn subject did not undergo image acquisition study procedures and thus was not included in this analysis set.

ArmMeasureGroupValue (MEAN)Dispersion
All Study Participants, PA Compression Image SetsCompression Force for PA and TC CompressionsCC9.1 daNStandard Deviation 3
All Study Participants, PA Compression Image SetsCompression Force for PA and TC CompressionsMLO8.2 daNStandard Deviation 2.7
All Study Participants, TC Compression Image SetsCompression Force for PA and TC CompressionsCC8.3 daNStandard Deviation 2
All Study Participants, TC Compression Image SetsCompression Force for PA and TC CompressionsMLO8.7 daNStandard Deviation 2.2
Other Pre-specified

Radiation Dose for PA and TC Compressions

Radiation dose (entrance skin air kerma \[ESAK\] in milligray \[mGy\]) for each compression mode (TC and PA compression) were summarized for craniocaudal (CC) and mediolateral oblique (MLO) views. For each subject, both views were collected using both compression modes.

Time frame: Through study completion, on average 1 month

Population: ESAK was captured for each view under each compression for all subjects. The one withdrawn subject did not undergo image acquisition study procedures and thus was not included in this analysis set.

ArmMeasureGroupValue (MEAN)Dispersion
All Study Participants, PA Compression Image SetsRadiation Dose for PA and TC CompressionsCC4.5 mGyStandard Deviation 2.4
All Study Participants, PA Compression Image SetsRadiation Dose for PA and TC CompressionsMLO4.9 mGyStandard Deviation 2.6
All Study Participants, TC Compression Image SetsRadiation Dose for PA and TC CompressionsCC4.7 mGyStandard Deviation 2.4
All Study Participants, TC Compression Image SetsRadiation Dose for PA and TC CompressionsMLO4.8 mGyStandard Deviation 2.4
Other Pre-specified

Technologist Interventions in PA Compression

Data will be collected regarding any interventions made by the technologist during PA compression and this data were summarized.

Time frame: Through study completion, on average 1 month

Population: Subjects underwent PA compression on their assigned Breast of Interest. Technologists were asked to provide confirmation if any intervention was necessary while the subject was controlling the compression. The one withdrawn subject did not undergo image acquisition study procedures and thus was not included in this analysis set.

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
All Study Participants, PA Compression Image SetsTechnologist Interventions in PA CompressionTechnologist had to increase compression0 Participants
All Study Participants, PA Compression Image SetsTechnologist Interventions in PA CompressionTechnologist had to decrease compression1 Participants
All Study Participants, PA Compression Image SetsTechnologist Interventions in PA CompressionNo intervention required29 Participants

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026