Breast Cancer
Conditions
Brief summary
This study is being conducted to compare the image quality of breast images obtained using standard (technologist-controlled \[TC\]) compression and patient-assisted (PA) compression and to evaluate the impact of PA compression on clinical workflow.
Detailed description
Patient-assisted (PA) compression allows the patient to participate in controlling the amount of compression force during mammography and is a personalized approach that has demonstrated successful reduction in discomfort experienced during mammography. GE Healthcare's Senographe Pristina, an innovative mammography platform that provides both two-dimensional (2D) and three-dimensional (3D) imaging capabilities, offers both standard and PA compression modes. This study is being conducted to compare the image quality of breast images obtained using TC compression and PA compression, and to evaluate the impact of PA compression on clinical workflow. The study population will consist of adult asymptomatic women presenting for screening 2D mammography.
Interventions
The technologist will properly position the breast and apply minimum compression. The subject will be instructed to apply compression as the technologist ensures the breast tissue is in appropriate position and tautness. The technologist will then guide the subject to achieve appropriate compression level, sufficient but not excessive, and the image will be acquired. This will be done for both standard views CC & MLO.
TC compression will be conducted per standard of care practices at the site.
Sponsors
Study design
Eligibility
Inclusion criteria
* Are women aged 40 years or older; * Are asymptomatic and scheduled for FFDM screening mammography; * Have left and right breasts; * Have breast sizes compatible with the dimensions of a 24 x 31 cm image detector, without anatomical cut-off; * Are documented as non-pregnant based on the investigator's medical judgment and in consideration of local clinical practice standards for evidence of non-pregnancy; * Are able and willing to comply with study procedures; and * Are able and willing to provide written informed consent to participate.
Exclusion criteria
* Have been previously included in this study or are participating in another study expected to interfere with study procedures or outcomes; * Have undergone diagnostic or surgical intervention(s) or procedure(s) on either breast, including breast biopsy, lumpectomy, or reconstruction, within five (5) years (≤ 5 years) of the study exam date; * Are currently undergoing radiotherapy or chemotherapy, or have a history of prior radiotherapy treatment on either breast; * Are currently lactating; or * Have breast implants.
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Number of Subjects With Acceptable Overall Clinical Image Quality for Patient-assisted (PA) and Technologist-controlled (TC) Image Sets | Through study completion, on average 1 month | One PA image set and one TC image set was acquired from each completed subject. The overall clinical image quality acceptability was collected and summarized on a per subject-basis using binary responses of either acceptable or unacceptable for unilateral, two-view PA and TC compression image sets. Two readers evaluated each of the 60 image sets (30 PA and 30 TC compression image sets from 30 completed participants). In cases of disagreement between Readers 1 and 2, a third reader provided adjudication. |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Repeat Image Acquisition | Through study completion, on average 1 month | Number of incidences per image set when the technologists or readers indicated a repeat acquisition for PA and TC compression modes. More than one incident (i.e. reason for repeat from technologist or view indicated for repeat by readers) could have been selected for a given image set. |
| Acceptability of Mammographic Attributes | Through study completion, on average 1 month | Acceptability of mammographic attributes using a binary response of either acceptable or unacceptable for unilateral, two-view PA and TC compression image sets were summarized. |
Other
| Measure | Time frame | Description |
|---|---|---|
| Breast Thickness for PA and TC Compressions | Through study completion, on average 1 month | Breast thickness (millimeter \[mm\]) for each compression mode (TC and PA compression) were summarized for craniocaudal (CC) and mediolateral oblique (MLO) views. For each subject, both views were collected using both compression modes. |
| Comparison of Image Acquisition Time | Through study completion, on average 1 month | The length of time it takes for image acquisition using each compression mode (TC Compression and PA Compression) were compared. |
| Radiation Dose for PA and TC Compressions | Through study completion, on average 1 month | Radiation dose (entrance skin air kerma \[ESAK\] in milligray \[mGy\]) for each compression mode (TC and PA compression) were summarized for craniocaudal (CC) and mediolateral oblique (MLO) views. For each subject, both views were collected using both compression modes. |
| Technologist Interventions in PA Compression | Through study completion, on average 1 month | Data will be collected regarding any interventions made by the technologist during PA compression and this data were summarized. |
| Compression Force for PA and TC Compressions | Through study completion, on average 1 month | Compression forces (decanewton \[daN\]) for each compression mode (TC and PA compression) were summarized for craniocaudal (CC) and mediolateral oblique (MLO) views. For each subject, both views were collected using both compression modes. |
Countries
United States
Participant flow
Participants by arm
| Arm | Count |
|---|---|
| All Study Participants All subjects will undergo their regularly scheduled full-field digital mammogram, consisting of bilateral, 2-view (CC/MLO) image acquisition. In addition, a study-specific, unilateral 2-view image set will obtained, utilizing the patient-assisted breast compression mode.
Patient-Assisted Breast Compression: The technologist will properly position the breast and apply minimum compression. The subject will be instructed to apply compression as the technologist ensures the breast tissue is in appropriate position and tautness. The technologist will then guide the subject to achieve appropriate compression level, sufficient but not excessive, and the image will be acquired. This will be done for both standard views CC/MLO. | 31 |
| Total | 31 |
Withdrawals & dropouts
| Period | Reason | FG000 |
|---|---|---|
| Overall Study | Withdrawal by Subject | 1 |
Baseline characteristics
| Characteristic | All Study Participants |
|---|---|
| Age, Categorical <=18 years | 0 Participants |
| Age, Categorical >=65 years | 8 Participants |
| Age, Categorical Between 18 and 65 years | 23 Participants |
| Breast of Interest First breast of interest | 16 Participants |
| Breast of Interest Second breast of interest | 14 Participants |
| Breast of Interest Withdrawn subject | 1 Participants |
| Ethnicity (NIH/OMB) Hispanic or Latino | 1 Participants |
| Ethnicity (NIH/OMB) Not Hispanic or Latino | 29 Participants |
| Ethnicity (NIH/OMB) Unknown or Not Reported | 1 Participants |
| Sex: Female, Male Female | 31 Participants |
| Sex: Female, Male Male | 0 Participants |
Adverse events
| Event type | EG000 affected / at risk |
|---|---|
| deaths Total, all-cause mortality | 0 / 30 |
| other Total, other adverse events | 0 / 30 |
| serious Total, serious adverse events | 0 / 30 |
Outcome results
Number of Subjects With Acceptable Overall Clinical Image Quality for Patient-assisted (PA) and Technologist-controlled (TC) Image Sets
One PA image set and one TC image set was acquired from each completed subject. The overall clinical image quality acceptability was collected and summarized on a per subject-basis using binary responses of either acceptable or unacceptable for unilateral, two-view PA and TC compression image sets. Two readers evaluated each of the 60 image sets (30 PA and 30 TC compression image sets from 30 completed participants). In cases of disagreement between Readers 1 and 2, a third reader provided adjudication.
Time frame: Through study completion, on average 1 month
Population: 60 image sets (30 PA and 30 TC compression image sets from 30 participants) were evaluated for acceptability of overall clinical image quality by two readers. The one (1) withdrawn subject did not undergo image acquisition study procedures and thus was not included in this analysis set.
| Arm | Measure | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|
| All Study Participants, PA Compression Image Sets | Number of Subjects With Acceptable Overall Clinical Image Quality for Patient-assisted (PA) and Technologist-controlled (TC) Image Sets | 30 Participants |
| All Study Participants, TC Compression Image Sets | Number of Subjects With Acceptable Overall Clinical Image Quality for Patient-assisted (PA) and Technologist-controlled (TC) Image Sets | 30 Participants |
Acceptability of Mammographic Attributes
Acceptability of mammographic attributes using a binary response of either acceptable or unacceptable for unilateral, two-view PA and TC compression image sets were summarized.
Time frame: Through study completion, on average 1 month
Population: The PA and TC image sets for all 30 subjects were evaluated for acceptability of pre-defined set of mammography attributes by Readers 1 and 2. Adjudication was not required if there was disagreement between the readers. The one withdrawn subject did not undergo image acquisition study procedures and thus was not included in this analysis set.
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| All Study Participants, PA Compression Image Sets | Acceptability of Mammographic Attributes | Acceptable sharpness of the image | 30 image sets |
| All Study Participants, PA Compression Image Sets | Acceptability of Mammographic Attributes | Acceptable Breast Positioning | 30 image sets |
| All Study Participants, PA Compression Image Sets | Acceptability of Mammographic Attributes | Acceptable tissue visibility at the skin line | 30 image sets |
| All Study Participants, PA Compression Image Sets | Acceptability of Mammographic Attributes | Acceptable breast compression | 30 image sets |
| All Study Participants, PA Compression Image Sets | Acceptability of Mammographic Attributes | Acceptable noise level of the image | 30 image sets |
| All Study Participants, PA Compression Image Sets | Acceptability of Mammographic Attributes | Acceptable exposure and visualization of tissue | 30 image sets |
| All Study Participants, PA Compression Image Sets | Acceptability of Mammographic Attributes | Acceptable artifacts level | 30 image sets |
| All Study Participants, PA Compression Image Sets | Acceptability of Mammographic Attributes | Acceptable image contrast | 30 image sets |
| All Study Participants, TC Compression Image Sets | Acceptability of Mammographic Attributes | Acceptable artifacts level | 30 image sets |
| All Study Participants, TC Compression Image Sets | Acceptability of Mammographic Attributes | Acceptable Breast Positioning | 29 image sets |
| All Study Participants, TC Compression Image Sets | Acceptability of Mammographic Attributes | Acceptable exposure and visualization of tissue | 30 image sets |
| All Study Participants, TC Compression Image Sets | Acceptability of Mammographic Attributes | Acceptable breast compression | 30 image sets |
| All Study Participants, TC Compression Image Sets | Acceptability of Mammographic Attributes | Acceptable sharpness of the image | 30 image sets |
| All Study Participants, TC Compression Image Sets | Acceptability of Mammographic Attributes | Acceptable tissue visibility at the skin line | 30 image sets |
| All Study Participants, TC Compression Image Sets | Acceptability of Mammographic Attributes | Acceptable noise level of the image | 30 image sets |
| All Study Participants, TC Compression Image Sets | Acceptability of Mammographic Attributes | Acceptable image contrast | 30 image sets |
Repeat Image Acquisition
Number of incidences per image set when the technologists or readers indicated a repeat acquisition for PA and TC compression modes. More than one incident (i.e. reason for repeat from technologist or view indicated for repeat by readers) could have been selected for a given image set.
Time frame: Through study completion, on average 1 month
Population: Instances of repeat imaging for each subject's PA (N=30) and TC (N=30) image sets were indicated by technologists during image acquisitions and by readers during image evaluations. The one withdrawn subject did not undergo image acquisition study procedures and thus was not included in this analysis set.
| Arm | Measure | Group | Value (COUNT_OF_UNITS) |
|---|---|---|---|
| All Study Participants, PA Compression Image Sets | Repeat Image Acquisition | Reader-indicated Repeat | 4 Image sets |
| All Study Participants, PA Compression Image Sets | Repeat Image Acquisition | Technologist-indicated Repeat | 2 Image sets |
| All Study Participants, TC Compression Image Sets | Repeat Image Acquisition | Reader-indicated Repeat | 4 Image sets |
| All Study Participants, TC Compression Image Sets | Repeat Image Acquisition | Technologist-indicated Repeat | 7 Image sets |
Breast Thickness for PA and TC Compressions
Breast thickness (millimeter \[mm\]) for each compression mode (TC and PA compression) were summarized for craniocaudal (CC) and mediolateral oblique (MLO) views. For each subject, both views were collected using both compression modes.
Time frame: Through study completion, on average 1 month
Population: Breast thickness (mm) was captured for each view under each compression for all subjects. The one withdrawn subject did not undergo image acquisition study procedures and thus was not included in this analysis set.
| Arm | Measure | Group | Value (MEAN) | Dispersion |
|---|---|---|---|---|
| All Study Participants, PA Compression Image Sets | Breast Thickness for PA and TC Compressions | CC | 52.0 mm | Standard Deviation 14.3 |
| All Study Participants, PA Compression Image Sets | Breast Thickness for PA and TC Compressions | MLO | 54.6 mm | Standard Deviation 16.1 |
| All Study Participants, TC Compression Image Sets | Breast Thickness for PA and TC Compressions | CC | 53.2 mm | Standard Deviation 14 |
| All Study Participants, TC Compression Image Sets | Breast Thickness for PA and TC Compressions | MLO | 53.7 mm | Standard Deviation 15.1 |
Comparison of Image Acquisition Time
The length of time it takes for image acquisition using each compression mode (TC Compression and PA Compression) were compared.
Time frame: Through study completion, on average 1 month
Population: Imaging time was collected for each image set. 4 image sets were excluded from TC analysis due to repeat imaging per standard of care; repeat imaging wasn't allowed for PA compression. 1 subject's PA \& TC image sets were excluded from analysis due to a protocol deviation. 1 withdrawn subject did not undergo imaging, so was not part of the analysis.
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| All Study Participants, PA Compression Image Sets | Comparison of Image Acquisition Time | 3.9 minutes | Standard Deviation 1 |
| All Study Participants, TC Compression Image Sets | Comparison of Image Acquisition Time | 3.4 minutes | Standard Deviation 1 |
Compression Force for PA and TC Compressions
Compression forces (decanewton \[daN\]) for each compression mode (TC and PA compression) were summarized for craniocaudal (CC) and mediolateral oblique (MLO) views. For each subject, both views were collected using both compression modes.
Time frame: Through study completion, on average 1 month
Population: Compression force (daN) was captured for each view under each compression for all subjects. The one withdrawn subject did not undergo image acquisition study procedures and thus was not included in this analysis set.
| Arm | Measure | Group | Value (MEAN) | Dispersion |
|---|---|---|---|---|
| All Study Participants, PA Compression Image Sets | Compression Force for PA and TC Compressions | CC | 9.1 daN | Standard Deviation 3 |
| All Study Participants, PA Compression Image Sets | Compression Force for PA and TC Compressions | MLO | 8.2 daN | Standard Deviation 2.7 |
| All Study Participants, TC Compression Image Sets | Compression Force for PA and TC Compressions | CC | 8.3 daN | Standard Deviation 2 |
| All Study Participants, TC Compression Image Sets | Compression Force for PA and TC Compressions | MLO | 8.7 daN | Standard Deviation 2.2 |
Radiation Dose for PA and TC Compressions
Radiation dose (entrance skin air kerma \[ESAK\] in milligray \[mGy\]) for each compression mode (TC and PA compression) were summarized for craniocaudal (CC) and mediolateral oblique (MLO) views. For each subject, both views were collected using both compression modes.
Time frame: Through study completion, on average 1 month
Population: ESAK was captured for each view under each compression for all subjects. The one withdrawn subject did not undergo image acquisition study procedures and thus was not included in this analysis set.
| Arm | Measure | Group | Value (MEAN) | Dispersion |
|---|---|---|---|---|
| All Study Participants, PA Compression Image Sets | Radiation Dose for PA and TC Compressions | CC | 4.5 mGy | Standard Deviation 2.4 |
| All Study Participants, PA Compression Image Sets | Radiation Dose for PA and TC Compressions | MLO | 4.9 mGy | Standard Deviation 2.6 |
| All Study Participants, TC Compression Image Sets | Radiation Dose for PA and TC Compressions | CC | 4.7 mGy | Standard Deviation 2.4 |
| All Study Participants, TC Compression Image Sets | Radiation Dose for PA and TC Compressions | MLO | 4.8 mGy | Standard Deviation 2.4 |
Technologist Interventions in PA Compression
Data will be collected regarding any interventions made by the technologist during PA compression and this data were summarized.
Time frame: Through study completion, on average 1 month
Population: Subjects underwent PA compression on their assigned Breast of Interest. Technologists were asked to provide confirmation if any intervention was necessary while the subject was controlling the compression. The one withdrawn subject did not undergo image acquisition study procedures and thus was not included in this analysis set.
| Arm | Measure | Category | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|---|
| All Study Participants, PA Compression Image Sets | Technologist Interventions in PA Compression | Technologist had to increase compression | 0 Participants |
| All Study Participants, PA Compression Image Sets | Technologist Interventions in PA Compression | Technologist had to decrease compression | 1 Participants |
| All Study Participants, PA Compression Image Sets | Technologist Interventions in PA Compression | No intervention required | 29 Participants |