Assessment of the Capability of PulmoVista 500 to Continuously Monitor Changes of Ventilation Over Time

Assessment of the Capability of PulmoVista 500 to Continuously Monitor Changes of Ventilation Over Time

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT03076983
Enrollment
80
Registered
2017-03-10
Start date
2017-05-30
Completion date
2018-04-12
Last updated
2019-11-27

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

ALI - Acute Lung Injury, ARDS, Mechanical Ventilation, Single-lung Ventilation, Weaning From Ventilator

Keywords

Ventilation Monitoring, Electrical Impedance Tomography

Brief summary

The results of this study will help to define the capability and reliability of PulmoVista 500 to detect changes in both global and regional ventilation.

Interventions

DEVICELung Monitoring with EIT device (PulmoVista 500)

Patient's lungs will be monitored during clinical routine interventions with Dräger's EIT device (PulmoVista 500). No medical interventions to routine patient care required.

Sponsors

Technische Universität Dresden
CollaboratorOTHER
Drägerwerk AG & Co. KGaA
Lead SponsorINDUSTRY

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

\- The population included in the clinical study will be selected from a pool of patients undergoing respiratory support who are scheduled to have their ventilation settings changed. As a subgroup, patients who are scheduled for surgeries that need one-lung-ventilation (OLV) will be enrolled to the study. * Male and female patients at the age of 18 years or older * On respiratory support in ICU care or scheduled for such ( e.g. postoperatively) or scheduled for surgery with OLV * Patients of which the monitoring of ventilation distribution may be of clinical interest * Patients being ventilated via an artificial airway with a mechanical ventilator that is compatible with PulmoVista 500 * Patients scheduled for changes in ventilation settings that may cause relevant changes in the ventilation * Chest circumference between 70 and 150 cm * Written informed consent to participate in the study provided by either the patient or the legal representative of the patient.

Exclusion criteria

* Currently has a permanent or temporary pacemaker, implantable cardiac device (ICD) or other device emitting electrical energy * a BMI ≥ 50 * tidal volume (VT) ≤ 200 mL * Current uncontrolled body movements such as tics, tremors or seizures, * Current wound dressings or infections on the chest that might interfere with the PulmoVista 500 electrode belt placement * Women of child bearing potential whose pregnancy cannot be excluded based on a pregnancy test or other proven facts. * Allergic to materials used in the electrode belt * Participation of the patient in an interventional trial within the last four weeks before enrollment in this trial * Evidence suggesting that the patient or his legal representative is not likely to follow the trial protocol (e.g. lacking compliance) * Known infectious diseases that require isolation of patient (e.g. MRSA) * Concomitant use of an air anti-decubitus medical mattress with dynamic inflation that cannot be deactivated during EIT measurements

Design outcomes

Primary

MeasureTime frameDescription
Monitoring Capabilities of PulmoVista 500Day 1The cross-correlation between the volume curve of the ventilator and the global impedance curve of PulmoVista 500 was expressed by the cross-correlation coefficient R. The cross-correlation coefficient is a measure to quantify the relationship between the volume curve and the EIT curve. The correlation coefficient R can assume values between \[-1, 1\] to describe the relationship between the two curves, as is usual for a correlation. To calculate the correlation coefficient, a phase shift between the two curves was taken into account, so the cross-correlation between them was calculated. The phase shift or delay occurs when the signal from the fan is transmitted to the PulmoVista500. The cross-correlation coefficient is determined by calculating the correlation coefficient for the curves within a narrow range of Lags(-15 to 15 ) for each pair of curves and using the maximum value as the value of correlation coefficient (therefore it is called cross-correlation coefficient)

Secondary

MeasureTime frameDescription
Detection of Changes in Regional VentilationDay 1To assess the capability of PulmoVista 500 to detect changes in regional ventilation by evaluating the cross correlation between impedance waveforms derived from PulmoVista 500 and volume curves derived from ventilator during one-lung-ventilation.
Assess Changes of Tidal VolumesDay 1Assess that changes of tidal volumes induced by ventilator settings can be monitored by the Trends view To assess if the changes of tidal volumes induced by ventilator settings can be monitored by the Trends view, the physician was asked to answer following question of a questionnaire: Open the view End-insp. trend after the PEEP setting has been changed significantly (2 mbar or more). Set C1 at the lower and C2 at the higher PEEP level. Does the image Change: C2 minus C1 and/or the corresponding regional Tidal Variations show the expected redistribution of ventilation from ventral to dorsal? Possible answers were: No, Yes and N/A
Assess Changes of the End-expiratory Lung VolumesDay 1Assess that changes of the end-expiratory lung volumes (induced by e.g. PEEP changes, recruitment and suctioning maneuvers) can be monitored by dEELI trend view To assess if the changes of the end-expiratory lung volumes (induced by e.g. PEEP changes, recruitment and suctioning maneuvers) can be monitored by dEELI trend view, the physician was asked to answer following two question of a questionnaire: Question 1: Open the view dEELI trend after the PEEP setting has been changed. Set C1 at the lower and C2 at the higher PEEP level. Does the image dEELI: C2 minus C1 show the expected increase of end-exp. lung volume in turquoise (light-blue)? Question 2: Open the view dEELI trend after lung suction has been conducted. Set C1 at a phase before and C2 during or immediately after the lung suction. Does the image dEELI: C2 minus C1 show the expected decrease of lung volume in orange color? Possible answers for the both above mentioned questions were No,Yes or N/A2
Safety - Documentation of Any Safety EventsDay 1Documentation of any safety events related to the use of PulmoVista 500
Assessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionDay 1How easy is it to... Q1: ... identify within the Tidal Image, whether the right and the left lung are evenly ventilated? Q2: ... quantify the regional ventilation distribution as expressed by the regional Parameters TV ROI 1 - TV ROI 4? Q3: ... set the cursors to respective positions which allow you to compare ventilation distribution before and after the intervention? Q4: ... detect increases and/or decreases of regional ventilation? Q5: ... relate the regional ventilation distribution to and the corresponding parameters from the ventilator, which are displayed in the trend table to each other? Q7: ... set the cursors to respective positions which allows you assessing lung volume changes induced by the intervention? Q8: ... detect increases and/or decreases of regional lung volume? How useful is the clinical Information on... Q6: ... ventilation redistribution caused by this intervention? Q9: ... regional lung volume changes caused by this intervention?

Countries

Germany

Participant flow

Participants by arm

ArmCount
ICU Patients
Patients currently on or scheduled for ventilator care in the intensive-care-unit (ICU) Lung Monitoring with EIT device (PulmoVista 500): Patient's lungs will be monitored during clinical routine interventions with Dräger's EIT device (PulmoVista 500). No medical interventions to routine patient care required.
55
OLV Patients
Patients scheduled for elective surgery receiving one-lung-ventilation (OLV) Lung Monitoring with EIT device (PulmoVista 500): Patient's lungs will be monitored during clinical routine interventions with Dräger's EIT device (PulmoVista 500). No medical interventions to routine patient care required.
25
Total80

Baseline characteristics

CharacteristicICU PatientsOLV PatientsTotal
Age, Continuous60.9 years
STANDARD_DEVIATION 13.8
59.1 years
STANDARD_DEVIATION 12.5
60.4 years
STANDARD_DEVIATION 13.4
Body mass index26.87 kg/m^2
STANDARD_DEVIATION 6.33
27.62 kg/m^2
STANDARD_DEVIATION 4.89
27.11 kg/m^2
STANDARD_DEVIATION 5.9
Ethnicity (NIH/OMB)
Hispanic or Latino
0 Participants0 Participants0 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
0 Participants0 Participants0 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
55 Participants25 Participants80 Participants
Height172.02 Centimetre
STANDARD_DEVIATION 9.29
170.96 Centimetre
STANDARD_DEVIATION 10.69
171.69 Centimetre
STANDARD_DEVIATION 9.7
Region of Enrollment
Germany
55 participants25 participants80 participants
Sex: Female, Male
Female
18 Participants10 Participants28 Participants
Sex: Female, Male
Male
37 Participants15 Participants52 Participants
Weight79.75 kilogram
STANDARD_DEVIATION 21.25
80.84 kilogram
STANDARD_DEVIATION 15.68
80.09 kilogram
STANDARD_DEVIATION 19.59

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 550 / 25
other
Total, other adverse events
2 / 5512 / 25
serious
Total, serious adverse events
7 / 551 / 25

Outcome results

Primary

Monitoring Capabilities of PulmoVista 500

The cross-correlation between the volume curve of the ventilator and the global impedance curve of PulmoVista 500 was expressed by the cross-correlation coefficient R. The cross-correlation coefficient is a measure to quantify the relationship between the volume curve and the EIT curve. The correlation coefficient R can assume values between \[-1, 1\] to describe the relationship between the two curves, as is usual for a correlation. To calculate the correlation coefficient, a phase shift between the two curves was taken into account, so the cross-correlation between them was calculated. The phase shift or delay occurs when the signal from the fan is transmitted to the PulmoVista500. The cross-correlation coefficient is determined by calculating the correlation coefficient for the curves within a narrow range of Lags(-15 to 15 ) for each pair of curves and using the maximum value as the value of correlation coefficient (therefore it is called cross-correlation coefficient)

Time frame: Day 1

Population: A Intention-to-Treat (ITT) collective was used as analysis population. The intention-to-treat collective comprises all patients for which both, the impedance waveforms and parallel to them volume waveforms have been generated during ventilation,so that the primary endpoint can be evaluated.

ArmMeasureValue (MEAN)Dispersion
ICU PatientsMonitoring Capabilities of PulmoVista 5000.9692 correlation coefficientStandard Deviation 0.0627
OLV PatientsMonitoring Capabilities of PulmoVista 5000.9093 correlation coefficientStandard Deviation 0.1351
Secondary

Assess Changes of the End-expiratory Lung Volumes

Assess that changes of the end-expiratory lung volumes (induced by e.g. PEEP changes, recruitment and suctioning maneuvers) can be monitored by dEELI trend view To assess if the changes of the end-expiratory lung volumes (induced by e.g. PEEP changes, recruitment and suctioning maneuvers) can be monitored by dEELI trend view, the physician was asked to answer following two question of a questionnaire: Question 1: Open the view dEELI trend after the PEEP setting has been changed. Set C1 at the lower and C2 at the higher PEEP level. Does the image dEELI: C2 minus C1 show the expected increase of end-exp. lung volume in turquoise (light-blue)? Question 2: Open the view dEELI trend after lung suction has been conducted. Set C1 at a phase before and C2 during or immediately after the lung suction. Does the image dEELI: C2 minus C1 show the expected decrease of lung volume in orange color? Possible answers for the both above mentioned questions were No,Yes or N/A2

Time frame: Day 1

Population: These two questions are only applicable to the patient group of ICU Patients as no assessment of the changes of the end-expiratory lung volume was planned a priori within the OLV patient group.

ArmMeasureGroupCategoryValue (COUNT_OF_PARTICIPANTS)
ICU PatientsAssess Changes of the End-expiratory Lung VolumesQuestion 1Answer: No1 Participants
ICU PatientsAssess Changes of the End-expiratory Lung VolumesQuestion 1Answer: Yes51 Participants
ICU PatientsAssess Changes of the End-expiratory Lung VolumesQuestion 1Answer: N/A3 Participants
ICU PatientsAssess Changes of the End-expiratory Lung VolumesQuestion 2Answer: No0 Participants
ICU PatientsAssess Changes of the End-expiratory Lung VolumesQuestion 2Answer: Yes29 Participants
ICU PatientsAssess Changes of the End-expiratory Lung VolumesQuestion 2Answer: N/A26 Participants
Secondary

Assess Changes of Tidal Volumes

Assess that changes of tidal volumes induced by ventilator settings can be monitored by the Trends view To assess if the changes of tidal volumes induced by ventilator settings can be monitored by the Trends view, the physician was asked to answer following question of a questionnaire: Open the view End-insp. trend after the PEEP setting has been changed significantly (2 mbar or more). Set C1 at the lower and C2 at the higher PEEP level. Does the image Change: C2 minus C1 and/or the corresponding regional Tidal Variations show the expected redistribution of ventilation from ventral to dorsal? Possible answers were: No, Yes and N/A

Time frame: Day 1

Population: This question is only applicable to the patient group of ICU Patients as no assessment of the changes of tidal volumes induced by ventilator settings was planned a priori for the OLV patient group.

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
ICU PatientsAssess Changes of Tidal VolumesAnswer:No3 Participants
ICU PatientsAssess Changes of Tidal VolumesAnswer: Yes49 Participants
ICU PatientsAssess Changes of Tidal VolumesAnswer: N/A3 Participants
Secondary

Assessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure Description

How easy is it to... Q1: ... identify within the Tidal Image, whether the right and the left lung are evenly ventilated? Q2: ... quantify the regional ventilation distribution as expressed by the regional Parameters TV ROI 1 - TV ROI 4? Q3: ... set the cursors to respective positions which allow you to compare ventilation distribution before and after the intervention? Q4: ... detect increases and/or decreases of regional ventilation? Q5: ... relate the regional ventilation distribution to and the corresponding parameters from the ventilator, which are displayed in the trend table to each other? Q7: ... set the cursors to respective positions which allows you assessing lung volume changes induced by the intervention? Q8: ... detect increases and/or decreases of regional lung volume? How useful is the clinical Information on... Q6: ... ventilation redistribution caused by this intervention? Q9: ... regional lung volume changes caused by this intervention?

Time frame: Day 1

ArmMeasureGroupCategoryValue (COUNT_OF_PARTICIPANTS)
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 11 = very difficult / not useful at all0 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 120 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 132 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 141 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 15 = very easy / very useful52 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 21 = very difficult / not useful at all0 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 220 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 232 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 243 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 25 = very easy / very useful50 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 31 = very difficult / not useful at all0 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 320 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 330 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 340 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 35 = very easy / very useful55 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 41 = very difficult / not useful at all0 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 420 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 432 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 442 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 45 = very easy / very useful51 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 51 = very difficult / not useful at all0 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 520 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 530 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 542 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 55 = very easy / very useful53 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 61 = very difficult / not useful at all0 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 620 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 639 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 6411 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 65 = very easy / very useful35 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 71 = very difficult / not useful at all0 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 720 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 730 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 740 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 75 = very easy / very useful55 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 81 = very difficult / not useful at all0 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 820 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 832 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 843 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 85 = very easy / very useful50 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 91 = very difficult / not useful at all0 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 920 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 9310 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 9410 Participants
ICU PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 95 = very easy / very useful35 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 532 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 11 = very difficult / not useful at all0 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 85 = very easy / very useful16 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 120 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 549 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 130 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 75 = very easy / very useful17 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 140 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 55 = very easy / very useful12 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 15 = very easy / very useful25 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 9314 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 21 = very difficult / not useful at all0 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 61 = very difficult / not useful at all1 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 220 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 81 = very difficult / not useful at all1 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 230 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 621 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 243 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 91 = very difficult / not useful at all2 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 25 = very easy / very useful22 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 6310 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 31 = very difficult / not useful at all0 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 822 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 320 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 6411 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 332 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 95 = very easy / very useful1 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 342 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 65 = very easy / very useful2 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 35 = very easy / very useful21 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 831 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 41 = very difficult / not useful at all1 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 71 = very difficult / not useful at all0 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 421 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 922 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 430 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 721 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 443 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 845 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 45 = very easy / very useful20 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 732 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 51 = very difficult / not useful at all1 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 946 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 521 Participants
OLV PatientsAssessing the Use Specific Clinical Helpfulness of PulmoVista 500 by Particular Questionnaire. The Questions Which Were Asked Within the Questionaire Are Listed in the Outcome Measure DescriptionQuestion 745 Participants
Secondary

Detection of Changes in Regional Ventilation

To assess the capability of PulmoVista 500 to detect changes in regional ventilation by evaluating the cross correlation between impedance waveforms derived from PulmoVista 500 and volume curves derived from ventilator during one-lung-ventilation.

Time frame: Day 1

Population: For this secondary endpoint the cross-correlation between between impedance waveforms derived from PulmoVista 500 and volume curves derived from Ventilator was calculated only for patients which received one-lung-ventilation (OLV).~Therefore no patients from the ICU patients Group were included into the analysis.

ArmMeasureValue (MEAN)Dispersion
OLV PatientsDetection of Changes in Regional Ventilation0.9093 correlation coefficientStandard Deviation 0.1351
Secondary

Safety - Documentation of Any Safety Events

Documentation of any safety events related to the use of PulmoVista 500

Time frame: Day 1

ArmMeasureGroupCategoryValue (COUNT_OF_PARTICIPANTS)
ICU PatientsSafety - Documentation of Any Safety EventsSubjects with any AEyes12 Participants
ICU PatientsSafety - Documentation of Any Safety EventsSubjects with any ADEyes0 Participants
ICU PatientsSafety - Documentation of Any Safety EventsSubjects with any AEno43 Participants
ICU PatientsSafety - Documentation of Any Safety EventsSubjects with any SAEno48 Participants
ICU PatientsSafety - Documentation of Any Safety EventsSubjects with any ADEno55 Participants
ICU PatientsSafety - Documentation of Any Safety EventsSubjects with any SAEyes7 Participants
OLV PatientsSafety - Documentation of Any Safety EventsSubjects with any ADEno25 Participants
OLV PatientsSafety - Documentation of Any Safety EventsSubjects with any AEno13 Participants
OLV PatientsSafety - Documentation of Any Safety EventsSubjects with any AEyes12 Participants
OLV PatientsSafety - Documentation of Any Safety EventsSubjects with any SAEyes1 Participants
OLV PatientsSafety - Documentation of Any Safety EventsSubjects with any ADEyes0 Participants
OLV PatientsSafety - Documentation of Any Safety EventsSubjects with any SAEno24 Participants