Measuring Thoracic Impedance in Hemodialysis Patients With the u-Cor Monitoring System

Measuring Thoracic Impedance in Hemodialysis Patients With the µ-Cor System A Pre-Market Validation Study (MaTcH)

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03072732
Acronym
MaTcH
Enrollment
40
Registered
2017-03-07
Start date
2016-11-03
Completion date
2017-02-01
Last updated
2020-11-04

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Fluid Retention Tissue

Keywords

Heart Failure, Hemodialysis

Brief summary

The purpose of this study is to measure the fluid changes in the body using a device called the µ-Cor System, an investigational device. The µ-Cor System will record the fluid changes in the body, as well as ECG (electrocardiogram or an electrical tracing of your heart rhythm) heart rate, breathing rate, posture and activity at regular intervals. The information collected by the µ-Cor System will then be compared to the actual fluid removed through dialysis and to the information collected by an FDA (Food and Drug Administration) cleared device called ZOE (NonInvasive Medical Technologies), a monitor which also measures the fluid changes in the body. The objectives of this study are to document any differences in measurement of thoracic impedance obtained from the µ-Cor System and the ZOE system. Thoracic impedance is a measure of the electrical activity in the chest that varies with changes in body size and composition, fluid volume, breathing status and other variables. Measurements of the amount of fluid removed during dialysis will also be compared between the µ-Cor System, the ZOE system and the actual fluid removed. The ability of the µ-Cor System to measure thoracic impedance will be tested by placing the device randomly at one of two locations: * Study Arm 1: side location- below left axilla * Study Arm 2: front location - upper left pectoral area Both the µ-Cor System and the ZOE System will be worn simultaneously during one dialysis session. The ZOE system device is placed on the chest- 2 points: 1 at the base of your neck and 1 in your mid chest.

Detailed description

The µ-Cor System is intended to record, store, transmit, and display the following physiological data to medical professionals: thoracic impedance, ECG, heart rate, respiration rate, activity and posture. The µ-Cor System is indicated for patients who are 21 years of age or older who: * have fluid-management problems, * are taking diuretic medication, * are living with heart failure, * are living with end-stage renal disease, * are recovering from a coronary artery disease-related event, and/or * are suffering from recurrent dehydration. Objectives: * This clinical trial is intended to provide evidence of substantial equivalence between the µ-Cor System and the ZOE System in the ability to measure thoracic impedance, by comparing the correlation between µ Cor measurement and ultrafiltration volume (UFV) as compared with the correlation between ZOE measurement and UFV. The actual UFV changes will be used to arbitrate any differences. * The ability of the µ-Cor System to measure thoracic impedance will be demonstrated at 2 body locations: side location (below left axilla, Study Arm 1), and front location (upper left pectoral area, Study Arm 2). Subject population: Forty patients undergoing hemodialysis (n = 40) in 2 study arms with at least 50% enrollment of patients diagnosed with congestive heart failure (CHF) in each arm: * Study Arm 1: 20 patients undergoing hemodialysis, with at least 10 of these patients having CHF, will have the µ-Cor device placed on the side location * Study Arm 2: 20 patients undergoing hemodialysis, with at least 10 of these patients having CHF, will have the µ-Cor device placed on the front location Patients will wear the µ-Cor device and will be connected to the ZOE monitor during one hemodialysis session in the clinic. Patients will be randomly allocated to Study Arm 1 or Study Arm 2, with randomization stratified by CHF status (CHF or non-CHF). During the hemodialysis session, the patient will wear 1 µ-Cor device and will be connected simultaneously to 1 ZOE monitor (via 2 ZOE electrodes) for comparative measurements and UFV correlation, from at least 15 minutes before the start of dialysis to at least 15 minutes after the end of the dialysis session. The µ-Cor will record measurements during the dialysis session, including at least the 15 minutes before and 15 minutes after the session. ZOE (Z0) values will be measured every 6 minutes (± 1 minute) during the dialysis session. ZOE (Z0) values also will be measured every 3 minutes (± 1 minute) before and after the session, for a minimum of 5 measurements. The UFV during the course of hemodialysis will be measured by automated readings provided by the dialysis machine every 6 minutes (± 1 minute). 10. Any IV infusions administered, oral fluid or solid intake, and urine output during the course of hemodialysis will be recorded. Weight, blood pressure, and heart rate will be recorded before and after the dialysis session.The patient's involvement in the study ends after the completion of all procedures planned for the hemodialysis session.

Interventions

DEVICEu-Cor System

Monitoring System

DEVICEZOE Fluid Status Monitor

Monitoring System

Sponsors

Zoll Medical Corporation
Lead SponsorINDUSTRY

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
DIAGNOSTIC
Masking
NONE

Intervention model description

2-arm study

Eligibility

Sex/Gender
ALL
Age
21 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Men and women at least 21 years of age. * Is currently scheduled to undergo hemodialysis 3 times per week in a clinic setting and has been on this regimen for at least 3 months. * For those patients with CHF: were diagnosed with CHF by a qualified provider and show symptomatic signs of New York Heart Association (NYHA) Class II to IV at enrollment. * Is prescribed a net fluid removal of at least 2.5 L during the hemodialysis session. * Is willing and able to sign informed consent in English.

Exclusion criteria

* Is a female patient with a known pregnancy or is unsure of pregnancy status. * Has known allergies or skin sensitivities to electrode hydrogel and/or acrylic based adhesive. * Has skin breakdown in areas where device and electrode placement is required. * Was hospitalized within the 2 weeks prior to enrollment. * Had intradialytic hypotension requiring administration of intravenous (IV) fluids of ≥250 mL or that resulted in a referral to urgent care, within 2 weeks prior to enrollment. * Had myocardial infarction, acute coronary syndrome, or stroke within 4 weeks prior to enrollment. * Has active nephrotic syndrome * Has severe malnutrition, as diagnosed per a qualified provider. * Is participating in another clinical trial. * Has an implanted device that might interfere with the µ-Cor.

Design outcomes

Primary

MeasureTime frameDescription
Difference in Correlation of Thoracic Impedance With Ultra-filtration Volume (UFV) Between uCor and ZOE15 minutes prior the start of one hemodialysis session to 15 minutes post completion of one hemodialysis session, approximately 2 to 6 hours.The primary endpoint measurement is the difference of 1) the Pearson correlation of thoracic impedance made by the u-Cor System to UFV and 2) the Pearson correlation of thoracic impedance made by the ZOE to UFV. Differences in correlation coefficients were calculated for each subject and each study arm.

Countries

United States

Participant flow

Recruitment details

Patients were recruited at the dialysis clinics of three sites, from November 2016 through February 2017.

Participants by arm

ArmCount
Study Arm 1 - Below Left Axilla
20 patients undergoing hemodialysis, with at least 10 of these patients having congestive heart failure, will have the u-Cor system placed below left axilla and the ZOE Fluid Status Monitor u-Cor System: Monitoring System ZOE Fluid Status Monitor: Monitoring System
20
Study Arm 2 - Upper Left Pectoral Area
20 patients undergoing hemodialysis, with at least 10 of these patients having congestive heart failure, will have the u-Cor system placed on the upper left pectoral area and the ZOE Fluid Status Monitor u-Cor System: Monitoring System ZOE Fluid Status Monitor: Monitoring System
20
Total40

Baseline characteristics

CharacteristicStudy Arm 1 - Below Left AxillaTotalStudy Arm 2 - Upper Left Pectoral Area
Age, Continuous58 Years
STANDARD_DEVIATION 9.5
57 Years
STANDARD_DEVIATION 9.4
56 Years
STANDARD_DEVIATION 9.5
Ethnicity (NIH/OMB)
Hispanic or Latino
2 Participants2 Participants0 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
18 Participants38 Participants20 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants0 Participants0 Participants
Race (NIH/OMB)
American Indian or Alaska Native
3 Participants4 Participants1 Participants
Race (NIH/OMB)
Asian
0 Participants0 Participants0 Participants
Race (NIH/OMB)
Black or African American
13 Participants28 Participants15 Participants
Race (NIH/OMB)
More than one race
0 Participants0 Participants0 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
1 Participants2 Participants1 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants0 Participants0 Participants
Race (NIH/OMB)
White
3 Participants6 Participants3 Participants
Region of Enrollment
United States
20 participants40 participants20 participants
Sex: Female, Male
Female
6 Participants10 Participants4 Participants
Sex: Female, Male
Male
14 Participants30 Participants16 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 200 / 20
other
Total, other adverse events
0 / 200 / 20
serious
Total, serious adverse events
0 / 200 / 20

Outcome results

Primary

Difference in Correlation of Thoracic Impedance With Ultra-filtration Volume (UFV) Between uCor and ZOE

The primary endpoint measurement is the difference of 1) the Pearson correlation of thoracic impedance made by the u-Cor System to UFV and 2) the Pearson correlation of thoracic impedance made by the ZOE to UFV. Differences in correlation coefficients were calculated for each subject and each study arm.

Time frame: 15 minutes prior the start of one hemodialysis session to 15 minutes post completion of one hemodialysis session, approximately 2 to 6 hours.

Population: Adult patients undergoing hemodialysis 3x week, with an expected net fluid removal of at least 2.5 L per dialysis session.

ArmMeasureValue (NUMBER)
Study Arm 1 - Below Left AxillaDifference in Correlation of Thoracic Impedance With Ultra-filtration Volume (UFV) Between uCor and ZOE0.66 difference in correlation coefficients
Study Arm 2 - Upper Left Pectoral AreaDifference in Correlation of Thoracic Impedance With Ultra-filtration Volume (UFV) Between uCor and ZOE0.23 difference in correlation coefficients
Comparison: The null hypothesis was that the correlation coefficient of the ZOE device would be greater than the correlation coefficient of the uCor device by at least 0.05.~For each subject, a correlation coefficient was calculated between study arm 1 uCor readings and UFV, as well as ZOE readings and UFV.~The difference between uCor correlation coefficient and ZOE correlation coefficient for each subject was used as the Primary Measurement.95% CI: [0.57, 0.75]
Comparison: The null hypothesis was that the correlation coefficient of the ZOE device would be greater than the correlation coefficient of the uCor device by at least 0.05.~For each subject, a correlation coefficient was calculated between study arm 2 uCor readings and UFV, as well as ZOE readings and UFV.~The difference between uCor correlation coefficient and ZOE correlation coefficient for each subject was used as the Primary Measurement.95% CI: [0.12, 0.35]

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026