Skip to content

Tear Film Innovations iLux Safety Study

Tear Film Innovations iLux Safety Study

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03055897
Enrollment
15
Registered
2017-02-16
Start date
2016-05-03
Completion date
2016-05-18
Last updated
2023-04-12

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Evaporative Dry Eye, Meibomian Gland Dysfunction

Brief summary

The objective of this study was to measure the temperature of the cornea and surrounding eye tissue temperature immediately after heating the eyelids with the iLux device using the protocol treatment profile. Eligible participants attended one treatment visit, with follow-up one day following the procedure.

Detailed description

Alcon Research, LLC, acquired TearFilm Innovations, Inc., in December 2018. This study was designed and conducted by TearFilm Innovations, Inc. The study results were collected, analyzed, and provided by TearFilm Innovations, Inc., to Alcon Research, LLC.

Interventions

Medical device that applies localized heat and pressure therapy to the eyelid in order to express melted meibum from obstructed glands

Sponsors

Tear Film Innovations, Inc.
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

Key Inclusion Criteria: \- Sign written informed consent prior to study participation Key

Exclusion criteria

* History of ocular surgery within 1 year * Active ocular infection * Ocular surface abnormality; lid surface abnormalities * Pregnant, nursing or not utilizing adequate birth control measures * Other protocol-specified

Design outcomes

Primary

MeasureTime frameDescription
Mean Overall Maximum Temperature of the Cornea and Surrounding Eye Tissue Immediately Post-treatmentDay 0 immediately post-treatmentPost-heating values were obtained by infrared camera for each participant. The cornea, the eyelid, and the surface tissue surrounding the eye were measured. Both eyes contributed to the analysis. No formal hypothesis testing was specified.

Countries

United States

Participant flow

Recruitment details

Participants were recruited from one investigative site located in the United States.

Pre-assignment details

This reporting group includes all enrolled participants.

Participants by arm

ArmCount
iLux 2020 System
Meibomian gland treatment (Day 0) according to instructions for use (IFU)/User Manual
15
Total15

Baseline characteristics

CharacteristiciLux 2020 System
Age, Continuous39.9 years
STANDARD_DEVIATION 15.1
Race/Ethnicity, Customized
Hispanic/Latino
1 participants
Race/Ethnicity, Customized
Not Reported
3 participants
Race/Ethnicity, Customized
White
11 participants
Region of Enrollment
United States
15 participants
Sex/Gender, Customized
Male
7 participants
Sex/Gender, Customized
Not reported
8 participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 15
other
Total, other adverse events
0 / 15
serious
Total, serious adverse events
0 / 15

Outcome results

Primary

Mean Overall Maximum Temperature of the Cornea and Surrounding Eye Tissue Immediately Post-treatment

Post-heating values were obtained by infrared camera for each participant. The cornea, the eyelid, and the surface tissue surrounding the eye were measured. Both eyes contributed to the analysis. No formal hypothesis testing was specified.

Time frame: Day 0 immediately post-treatment

Population: Safety Analysis Set: All subjects enrolled and treated

ArmMeasureGroupValue (MEAN)Dispersion
iLux 2020 SystemMean Overall Maximum Temperature of the Cornea and Surrounding Eye Tissue Immediately Post-treatmentCorneal temperature37.7 degrees CelsiusStandard Deviation 0.5
iLux 2020 SystemMean Overall Maximum Temperature of the Cornea and Surrounding Eye Tissue Immediately Post-treatmentEyelid temperature38.5 degrees CelsiusStandard Deviation 0.8
iLux 2020 SystemMean Overall Maximum Temperature of the Cornea and Surrounding Eye Tissue Immediately Post-treatmentSurface temperature38.2 degrees CelsiusStandard Deviation 0.7

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026