Skip to content

Retrospective CI532 Hearing Performance

Retrospective Post-market Hearing Performance Outcome in a Cohort of CI532 Recipients

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT03039283
Enrollment
162
Registered
2017-02-01
Start date
2017-03-14
Completion date
2018-08-28
Last updated
2019-10-30

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Hearing Loss

Keywords

Moderate to profound sensorineural hearing loss

Brief summary

The aim of this retrospective study is to collect and assess hearing performance data that have been measured by five clinics in Germany as part of their clinical routine in recipients implanted with a commercial CI532.

Detailed description

Speech understanding data in quiet and in noise and aided and unaided thresholds, medical history, anomalies in terms of electrode placement are collected through an eCRF. Recipient's device characteristics are collected through cdx files. Surgeon handling and usability of CI532 are collected through paper questionnaire. The retrospective study is aimed at collection of data for CI532 recipients who have routine hearing measures at pre-implant and post-implant intervals in their medical records. Patients are approached for study participation by the caring clinician on a consecutive and voluntary basis and on condition of signing the patient informed consent.

Interventions

Retrospective study of the commercial CI532 cochlear implant

Sponsors

Cochlear
Lead SponsorINDUSTRY

Study design

Observational model
COHORT
Time perspective
RETROSPECTIVE

Eligibility

Sex/Gender
ALL
Healthy volunteers
No

Inclusion criteria

* Ability to conduct adult hearing performance test material * Good German language skills to assess clinical hearing performance * CI532 recipients assessed via routine clinical measurements at pre implant, and post implant intervals with available data records in hospital files. * Patients that have read, understood and signed the patient informed consent.

Exclusion criteria

* Recipients that have participated in the CLTD5446 study.

Design outcomes

Primary

MeasureTime frameDescription
Change From Pre-operative (Daily Listening Condition) Baseline Speech Understanding in Quiet at 6 Months Post-operative (Best Aided Conditions).pre-operatively and at 6 months post-operativelyIs tested using the centre's clinical routine speech tests. Participants are listening in their normal hearing configuration: often with acoustic hearing aids in both ears preoperatively, an implant in one ear and hearing aid in the opposite, or using two implants, one in each ear postoperatively. All centres used Freiburger German monosyllable lists. Recorded lists of everyday words are presented to participants at 65 dB SPL (loud conversational speech level) from loudspeakers and participants repeat back what they hear. Lists are scored as a percentage correct words, with two lists being used per condition to represent speech understanding in quiet.
Change From Pre-operative (Daily Listening Condition) Baseline Speech Understanding in Noise at 6 Months Post-operative (Best Aided Conditions).pre-operatively and at 6 months post-operativelyIs tested using the centre's clinical routine speech tests. Participants are listening in their normal hearing configuration: often with acoustic hearing aids in both ears preoperatively, an implant in one ear and hearing aid in the opposite, or using two implants, one in each ear postoperatively. Centres used different types of speech in noise testing. Lists of sentences were presented in competing background noise. An adaptive procedure is used such that after each sentence is presented the speech level is decreased if \>50% of the words are correctly repeated in the sentence, or increased if otherwise. The test result is the mean signal-to-noise ratio of the last eight presentations. An average of two lists is used as the result representing the signal-to-noise ratio that gives 50% sentence understanding.
Percentage of Participants Showing Post-operative Improvement in the Ipsilateral Ear in Quiet.pre-operatively and at 6 months post-operativelyIs tested using the centre's clinical routine speech. Participants are listening using only the ear which was treated using hearing aid preoperatively and the implant postoperatively. All centres used Freiburger German monosyllable lists. Recorded lists of everyday words are presented to participants at 65 dB SPL (loud conversational speech level) from loudspeakers and participants repeat back what they hear. Lists are scored as a percentage correct words, with two lists being used per condition to represent speech understanding in quiet.
Percentage of Participants Showing Post-operative Improvement in Best Aided Condition in QuietPre-operatively and at 6 months post-operativelyIs tested using the centre's clinical routine speech. Participants are listening in their normal hearing configuration: often with acoustic hearing aids in both ears preoperatively, an implant in one ear and hearing aid in the opposite, or using two implants, one in each ear postoperatively. All centres used Freiburger German monosyllable lists. Recorded lists of everyday words are presented to participants at 65 dB SPL (loud conversational speech level) from loudspeakers and participants repeat back what they hear. Lists are scored as a percentage correct words, with two lists being used per condition to represent speech understanding in quiet.
Percentage of Participants Showing Post-operative Improvement in the Ipsilateral Ear in Noise.Pre-operatively and at 6 months post-operativelyIs tested using the centre's clinical routine speech. Participants are listening using only the ear which was treated using hearing aid preoperatively and the implant postoperatively. Centres used different types of speech in noise testing. Lists of sentences were presented in competing background noise. In some cases the noise level was fixed at 10 dB signal-to-noise ratio, and the percentage of words correctly repeated was recorded (over two lists). Alternatively, an adaptive procedure is used such that after each sentence is presented the speech level is decreased if \>50% of the words are correctly repeated in the sentence, or increased if otherwise. The test result is the mean signal-to-noise ratio of the last eight presentations. An average of two lists is used as the result representing the signal-to-noise ratio that gives 50% sentence understanding.
Percentage of Participants Showing Post-operative Improvement in Best Aided Condition in Noise.Pre-operatively and at 6 months post-operativelyIs tested using the centre's clinical routine speech. Participants are listening in their normal hearing configuration: often with acoustic hearing aids in both ears preoperatively, an implant in one ear and hearing aid in the opposite, or using two implants, one in each ear postoperatively. Centres used different types of speech in noise testing. Lists of sentences were presented in competing background noise. In some cases the noise level was fixed at 10 dB signal-to-noise ratio, and the percentage of words correctly repeated was recorded (over two lists). Alternatively, an adaptive procedure is used such that after each sentence is presented the speech level is decreased if \>50% of the words are correctly repeated in the sentence, or increased if otherwise. The test result is the mean signal-to-noise ratio of the last eight presentations. An average of two lists is used as the result representing the signal-to-noise ratio that gives 50% sentence understanding.

Countries

Germany

Participant flow

Recruitment details

162 participants were recruited in five German centres from March 2017 to August 2018

Participants by arm

ArmCount
Nucleus CI532 Cochlear Implant
Nucleus CI532 cochlear implant: Retrospective study of the commercial CI532 cochlear implant
159
Total159

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyProtocol Violation2

Baseline characteristics

CharacteristicNucleus CI532 Cochlear Implant
Age, Continuous56.6 years
STANDARD_DEVIATION 18
Race and Ethnicity Not Collected— Participants
Region of Enrollment
Germany
159 Participants
Sex: Female, Male
Female
73 Participants
Sex: Female, Male
Male
86 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 0
other
Total, other adverse events
0 / 0
serious
Total, serious adverse events
0 / 0

Outcome results

Primary

Change From Pre-operative (Daily Listening Condition) Baseline Speech Understanding in Noise at 6 Months Post-operative (Best Aided Conditions).

Is tested using the centre's clinical routine speech tests. Participants are listening in their normal hearing configuration: often with acoustic hearing aids in both ears preoperatively, an implant in one ear and hearing aid in the opposite, or using two implants, one in each ear postoperatively. Centres used different types of speech in noise testing. Lists of sentences were presented in competing background noise. An adaptive procedure is used such that after each sentence is presented the speech level is decreased if \>50% of the words are correctly repeated in the sentence, or increased if otherwise. The test result is the mean signal-to-noise ratio of the last eight presentations. An average of two lists is used as the result representing the signal-to-noise ratio that gives 50% sentence understanding.

Time frame: pre-operatively and at 6 months post-operatively

Population: 15 participants with routine speech understanding data in noise available at pre-op and at 6 months were analyzed.

ArmMeasureValue (MEAN)Dispersion
Nucleus CI532 Cochlear ImplantChange From Pre-operative (Daily Listening Condition) Baseline Speech Understanding in Noise at 6 Months Post-operative (Best Aided Conditions).-5.5 Speech Recognition Threshold in decibelsStandard Deviation 11.2
Primary

Change From Pre-operative (Daily Listening Condition) Baseline Speech Understanding in Quiet at 6 Months Post-operative (Best Aided Conditions).

Is tested using the centre's clinical routine speech tests. Participants are listening in their normal hearing configuration: often with acoustic hearing aids in both ears preoperatively, an implant in one ear and hearing aid in the opposite, or using two implants, one in each ear postoperatively. All centres used Freiburger German monosyllable lists. Recorded lists of everyday words are presented to participants at 65 dB SPL (loud conversational speech level) from loudspeakers and participants repeat back what they hear. Lists are scored as a percentage correct words, with two lists being used per condition to represent speech understanding in quiet.

Time frame: pre-operatively and at 6 months post-operatively

Population: 49 participants with routine speech understanding data available at pre-op and 6 months were analysed.

ArmMeasureValue (MEAN)Dispersion
Nucleus CI532 Cochlear ImplantChange From Pre-operative (Daily Listening Condition) Baseline Speech Understanding in Quiet at 6 Months Post-operative (Best Aided Conditions).26.1 Percent correctStandard Deviation 26.7
Primary

Percentage of Participants Showing Post-operative Improvement in Best Aided Condition in Noise.

Is tested using the centre's clinical routine speech. Participants are listening in their normal hearing configuration: often with acoustic hearing aids in both ears preoperatively, an implant in one ear and hearing aid in the opposite, or using two implants, one in each ear postoperatively. Centres used different types of speech in noise testing. Lists of sentences were presented in competing background noise. In some cases the noise level was fixed at 10 dB signal-to-noise ratio, and the percentage of words correctly repeated was recorded (over two lists). Alternatively, an adaptive procedure is used such that after each sentence is presented the speech level is decreased if \>50% of the words are correctly repeated in the sentence, or increased if otherwise. The test result is the mean signal-to-noise ratio of the last eight presentations. An average of two lists is used as the result representing the signal-to-noise ratio that gives 50% sentence understanding.

Time frame: Pre-operatively and at 6 months post-operatively

Population: 4 of the 159 participants had pre-operative and 6-month post-operative data available in the best aided condition to compare.

ArmMeasureValue (NUMBER)
Nucleus CI532 Cochlear ImplantPercentage of Participants Showing Post-operative Improvement in Best Aided Condition in Noise.75 Percentage of participants
Primary

Percentage of Participants Showing Post-operative Improvement in Best Aided Condition in Quiet

Is tested using the centre's clinical routine speech. Participants are listening in their normal hearing configuration: often with acoustic hearing aids in both ears preoperatively, an implant in one ear and hearing aid in the opposite, or using two implants, one in each ear postoperatively. All centres used Freiburger German monosyllable lists. Recorded lists of everyday words are presented to participants at 65 dB SPL (loud conversational speech level) from loudspeakers and participants repeat back what they hear. Lists are scored as a percentage correct words, with two lists being used per condition to represent speech understanding in quiet.

Time frame: Pre-operatively and at 6 months post-operatively

Population: 60 of the 159 participants had pre-operative and 6-month post-operative data available in the best aided condition to compare.

ArmMeasureValue (NUMBER)
Nucleus CI532 Cochlear ImplantPercentage of Participants Showing Post-operative Improvement in Best Aided Condition in Quiet73.3 Percentage of participants
Primary

Percentage of Participants Showing Post-operative Improvement in the Ipsilateral Ear in Noise.

Is tested using the centre's clinical routine speech. Participants are listening using only the ear which was treated using hearing aid preoperatively and the implant postoperatively. Centres used different types of speech in noise testing. Lists of sentences were presented in competing background noise. In some cases the noise level was fixed at 10 dB signal-to-noise ratio, and the percentage of words correctly repeated was recorded (over two lists). Alternatively, an adaptive procedure is used such that after each sentence is presented the speech level is decreased if \>50% of the words are correctly repeated in the sentence, or increased if otherwise. The test result is the mean signal-to-noise ratio of the last eight presentations. An average of two lists is used as the result representing the signal-to-noise ratio that gives 50% sentence understanding.

Time frame: Pre-operatively and at 6 months post-operatively

Population: 36 of the 159 participants had pre-operative and 6-month post-operative data available in the ipsilateral condition to compare.

ArmMeasureValue (NUMBER)
Nucleus CI532 Cochlear ImplantPercentage of Participants Showing Post-operative Improvement in the Ipsilateral Ear in Noise.67 Percentage of participants
Primary

Percentage of Participants Showing Post-operative Improvement in the Ipsilateral Ear in Quiet.

Is tested using the centre's clinical routine speech. Participants are listening using only the ear which was treated using hearing aid preoperatively and the implant postoperatively. All centres used Freiburger German monosyllable lists. Recorded lists of everyday words are presented to participants at 65 dB SPL (loud conversational speech level) from loudspeakers and participants repeat back what they hear. Lists are scored as a percentage correct words, with two lists being used per condition to represent speech understanding in quiet.

Time frame: pre-operatively and at 6 months post-operatively

Population: 99 of the 159 participants had pre-operative and 6-month post-operative data available in the ipsilateral condition to compare.

ArmMeasureValue (NUMBER)
Nucleus CI532 Cochlear ImplantPercentage of Participants Showing Post-operative Improvement in the Ipsilateral Ear in Quiet.96 Percentage of participants

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026