User Performance Evaluation of Contour® Next One, Accu-Chek® Aviva Connect, FreeStyle Freedom Lite, OneTouch® Verio and GlucoMen® Areo Blood Glucose Monitoring Systems Following ISO 15197:2013

User Performance Evaluation of Contour® Next One, Accu-Chek® Aviva Connect, FreeStyle Freedom Lite, OneTouch® Verio and GlucoMen® Areo Blood Glucose Monitoring Systems Following ISO 15197:2013

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT03033849
Enrollment
189
Registered
2017-01-27
Start date
2017-01-31
Completion date
2017-03-31
Last updated
2017-09-12

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes Mellitus

Brief summary

The user performance evaluation shows whether people with diabetes are able to obtain accurate measurement results with a blood glucose monitoring system. In this study, user performance evaluation will be performed for Contour® Next One (Ascensia Diabetes Care GmbH), Accu-Chek® Aviva Connect (Roche Diabetes Care GmbH), FreeStyle Freedom Lite (Abbott Diabetes Care Inc.), OneTouch® Verio (LifeScan Europe) and GlucoMen® areo A. Menarini Diagnostics S.r.l. (based on ISO 15197:2013; EN ISO 15197:2015, clause 8. For each BGMS, measurement procedures for user performance evaluation will be performed with 1 test meter and 1 reagent system lot by the study subjects. The same meter and an additional test meter will be used for double measurements performed by study personnel (with the same reagent system lot used by subjects).

Interventions

DEVICEBlood glucose monitoring system for self-testing FreeStyle Freedom Lite
DEVICEBlood glucose monitoring system for self-testing OneTouch® Verio
DEVICEBlood glucose monitoring system for self-testing GlucoMen® areo
DEVICEBlood glucose monitoring system for self-testing Contour® Next One

This is a user performance study, that is intended to show whether people with diabetes are able to obtain accurate measurement results with a blood glucose monitoring system (BGMS). The subjects' measurement technique (e.g. applying blood onto the system reagent, reading the result) will be observed and documented by study personnel. Immediately after the measurement by the subject, the study personnel will perform a measurement with the same test meter. Measurement results obtained by the subjects with the BGMS will be compared to measurement results obtained with the comparison method. Sample collection for the comparison measurement will be performed by study personnel within 5 min after the subject's measurement with the test meter. The samples will be collected from the subjects' skin puncture (if possible), an additional skin puncture (if required) will be documented. An additional skin puncture will be performed by study personnel with single-use lancing devices.

DEVICEBlood glucose monitoring system for self-testing Accu-Chek® Aviva Connect

Sponsors

Ascensia Diabetes Care
CollaboratorINDUSTRY
Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm
Lead SponsorOTHER

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* Male or female, with type 1 diabetes, type 2 diabetes or subjects without diabetes * Signed informed consent form * Minimum age of 18 years * Subjects are legally competent and capable to understand character, meaning and consequences of the study. * If blood glucose values \< 80 mg/dl or \> 300 mg/dl shall be measured after short term alteration in insulin therapy: * Male or female with type 1 diabetes and intensified insulin therapy or insulin pump therapy. * Signature of subjects to document consent with these procedures on informed consent form.

Exclusion criteria

* Pregnancy or lactation period * Severe acute disease (at the study physician's discretion) * Severe chronic disease with potential risk during the test procedures (at the study physician's discretion) * Current constitution that compromises the subject's capability to participate in the study (at the study physician's discretion) * Being unable to give informed consent * \< 18 years * Legally incompetent * Being committed to an institution (e.g. psychiatric clinic) * Language barriers potentially compromising an adequate compliance with study procedures * Dependent on investigator or sponsor * If blood glucose values \< 80 mg/dl shall be measured after short term alteration in insulin therapy, subjects with type 1 diabetes, suffering from: * Coronary heart disease * Condition after myocardial infarction * Condition after cerebral events * Peripheral arterial occlusive disease * Hypoglycemia unawareness

Design outcomes

Primary

MeasureTime frameDescription
Acceptance criteria defined by ISO 15197:2013 (E) will be appliedFor each subject, the experimental phase has an expected duration of up to 3 hours95 % of the individual glucose measured values shall fall within ± 15 mg/dl (0.83 mmol/l) of the measured values of the comparison measurement procedure at glucose concentrations \< 100 mg/dl (5.55 mmol/l) and within ± 15 % at glucose concentrations ≥ 100 mg/dl (5.55 mmol/l).