Blepharoptosis, Facial Paralysis, Lagophthalmos
Conditions
Brief summary
Objective 1: Determine the safety and feasibility of externally mounted magnets for extended management of chronic eyelid movement disorders by measuring visual acuity and corneal and skin integrity and comfort over 8 weeks of wear. Objective 2: Collect preliminary data on the relative efficacy of external magnetic devices by comparing them to externally mounted lid weights and ptosis crutches using rating scales and video analysis of blink biomechanics.
Detailed description
Two major types of eye lid movement disorders include lagophthalmos (incomplete closure of the eyelids) and blepharoptosis (incomplete opening of the eyelids). Both of these conditions occur because of a disruption in the normal agonist-antagonist neuro-muscular complex balance. Generally paralysis of the eyelid or facial muscles is responsible for the abnormality (the eyelids can close but not open, or open but not close). An external device, if able to generate an appropriately balanced force, could restore eyelid movement by performing the paralyzed function; for example, a ptotic (droopy) eyelid could be opened, and the functioning eyelid closure muscle could overcome the device's force. In prior work the investigators established proof-of-concept data demonstrating safety and efficacy for temporary management of ptosis up to 2 hour per day for 2 weeks. Preliminary data suggest similar safety for lagophthalmos. This study will expand on this preliminary work to examine longer wear times, home use, and feasibility for a wider range of eye lid disorders.
Interventions
Small externally mounted magnetic lid arrays are attached to the eye lid (s). Glasses with a second magnet system may be combined with the arrays depending on the lid condition. The device intends to facilitate eye opening and closing.
Sponsors
Study design
Eligibility
Inclusion criteria
* Presence of an eyelid movement disorder for at least one eye, moderate cognitive function or better defined as greater than or equal to 18 out of 30 on a pre-screening of the Mini-Mental State Exam, age 5 or older.
Exclusion criteria
* Absence of an eyelid movement disorder or presence of a corneal ulcer. Those with a corneal ulcer are at risk for permanent loss of vision and should be managed with proven methods. Age less than 5, Severe Cognitive impairment defined as Mini-mental score \<18, behaviors consistent with delirium (combinations of disorientation, hallucinations, delusions, and incoherent speech), or lethargy. These individuals must be excluded since participation requires competent self-care, reliable responses and cooperation during fitting of the devices.
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Rating Scale of Skin Integrity | Baseline and up to 32 weeks | Ratings at the beginning and end of treatment will be analyzed by calculating the mean and standard deviation of the skin erythema rating across all available time points. The outcome measure is the FDA scale, Assessing the Irritation and Sensitization Potential of Transdermal and Topical Delivery Systems for ANDAs Guidance for Industry. Range is 0 to 7 with higher numbers being worse skin erythema. |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Count of Participants With Adverse Corneal Events Based on National Eye Institute Rating Scale | Baseline and up to 32 weeks | Using the National Eye Institute Rating Scale (0 to 3 points with higher being more staining), cases where staining increased more than 2 points or if there was any ulceration, an adverse event was tabulated and summed for each participant over 1 up to 32 weeks of use, depending on how long the participant was followed. |
| Change in Visual Acuity | Baseline and up to 32 weeks | Visual acuity at 20ft before treatment and after 1 week or more. Difference between weekly acuity measures to baseline was calculated and averaged for each subject, and then a group mean and standard deviation calculated. |
Countries
United States
Participant flow
Participants by arm
| Arm | Count |
|---|---|
| Magnetic Lid System All participants will trial a commercially available device and an experimental magnetic device.
Magnetic lid system: Small externally mounted magnetic lid arrays are attached to the eye lid (s). Glasses with a second magnet system may be combined with the arrays depending on the lid condition. The device intends to facilitate eye opening and closing. | 11 |
| Total | 11 |
Baseline characteristics
| Characteristic | Magnetic Lid System |
|---|---|
| Age, Continuous | 52 years STANDARD_DEVIATION 18 |
| Ethnicity (NIH/OMB) Hispanic or Latino | 0 Participants |
| Ethnicity (NIH/OMB) Not Hispanic or Latino | 11 Participants |
| Ethnicity (NIH/OMB) Unknown or Not Reported | 0 Participants |
| Race (NIH/OMB) American Indian or Alaska Native | 0 Participants |
| Race (NIH/OMB) Asian | 0 Participants |
| Race (NIH/OMB) Black or African American | 2 Participants |
| Race (NIH/OMB) More than one race | 0 Participants |
| Race (NIH/OMB) Native Hawaiian or Other Pacific Islander | 0 Participants |
| Race (NIH/OMB) Unknown or Not Reported | 0 Participants |
| Race (NIH/OMB) White | 9 Participants |
| Region of Enrollment United States | 11 participants |
| Sex: Female, Male Female | 6 Participants |
| Sex: Female, Male Male | 5 Participants |
Adverse events
| Event type | EG000 affected / at risk |
|---|---|
| deaths Total, all-cause mortality | 0 / 11 |
| other Total, other adverse events | 2 / 11 |
| serious Total, serious adverse events | 0 / 11 |
Outcome results
Rating Scale of Skin Integrity
Ratings at the beginning and end of treatment will be analyzed by calculating the mean and standard deviation of the skin erythema rating across all available time points. The outcome measure is the FDA scale, Assessing the Irritation and Sensitization Potential of Transdermal and Topical Delivery Systems for ANDAs Guidance for Industry. Range is 0 to 7 with higher numbers being worse skin erythema.
Time frame: Baseline and up to 32 weeks
Population: Erythema rating in participants who had worn the MLP for 1 to 32 weeks. All the ratings each subject had across all their visits were averaged, and then a group average and std dev calculated.
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Magnetic Lid System | Rating Scale of Skin Integrity | 0.3 units on a scale | Standard Deviation 0.5 |
Change in Visual Acuity
Visual acuity at 20ft before treatment and after 1 week or more. Difference between weekly acuity measures to baseline was calculated and averaged for each subject, and then a group mean and standard deviation calculated.
Time frame: Baseline and up to 32 weeks
Population: Participants who wore the MLP for 1 week or more
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Magnetic Lid System | Change in Visual Acuity | 0.03 LogMAR | Standard Deviation 0.08 |
Count of Participants With Adverse Corneal Events Based on National Eye Institute Rating Scale
Using the National Eye Institute Rating Scale (0 to 3 points with higher being more staining), cases where staining increased more than 2 points or if there was any ulceration, an adverse event was tabulated and summed for each participant over 1 up to 32 weeks of use, depending on how long the participant was followed.
Time frame: Baseline and up to 32 weeks
Population: Participants with ptosis wearing the MLP for 1 or more weeks
| Arm | Measure | Value (COUNT_OF_PARTICIPANTS) |
|---|---|---|
| Magnetic Lid System | Count of Participants With Adverse Corneal Events Based on National Eye Institute Rating Scale | 1 Participants |